Skip to main content
Log in

Dimensional analysis of the distal phalanx with consideration of distal interphalangeal joint arthrodesis using a headless compression screw

  • Surgery Articles
  • Published:
HAND

Abstract

Background

Headless compression screws have been used for arthrodesis of the distal interphalangeal joint with good clinical results. However, some distal phalanges are too small, thus precluding their use, or increasing complication rates.

Methods

In order to validate our digital measurements, radiographs of five cadaveric hands were obtained. The distal phalanges of these fingers were measured using our institution’s digital radiograph software. The cadavers were then dissected, and the actual phalanges were measured to confirm the accuracy of the digital measurements; 200 consecutive hand radiographs were then measured to obtain average width and heights of the narrowest measurements of the distal phalanx of each finger. Subgroup analysis of each finger for age and gender was completed.

Results and Conclusions

Females were routinely smaller than their male counterparts. No age-related differences were seen. A substantial number of fingers were found to be too small to allow for use of headless compression screws; 42 % of small finger distal phalanges are too small to accommodate 2.8-mm threads, and that number increases to 81 % with 3.2-mm threads and 97 % with 3.5-mm threads. Few distal phalanges, regardless of which finger, are large enough to accommodate 3.5-mm-diameter threads; 22% of female ring finger distal phalanges are smaller than 2.8 mm, and 66 % of ring fingers are smaller than 3.2 mm. Based on our measurements, a surprising number of distal phalanges are smaller than the trailing thread diameter of commonly commercially available headless screws. Care must be taken in selecting implants for distal interphalangeal (DIP) fusion.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Arata J, Ishikawa K, Soadea H, Kitayama T. Arthrodesis of the distal interphalangeal joint using a bioabsorbable rod as an intramedullary nail. Scand J Plast Reconstr Surg Hand Surg. 2003;37:228–31.

    Article  PubMed  Google Scholar 

  2. Brutus JP, Palmer AK, Mosher JF, Harley BJ, Loftus JB. Use of a headless compressive screw for distal interphalangeal joint arthrodesis in digits: clinical outcome and review of complications. J Hand Surg [Am]. 2006;31:85–9.

    Article  Google Scholar 

  3. Cobb T. Arthroscopic distal interphalangeal joint arthrodesis. Tech Hand Upper Extrem Surg. 2008;12:266–9.

    Article  Google Scholar 

  4. Henry M. Variable pitch headless compression screw treatment of distal phalangeal nonunions. Tech Hand Upper Extrem Surg. 2010;14:230–3.

    Article  Google Scholar 

  5. Kocak E, Carruthers K, Kobus R. Distal interphalangeal joint arthrodesis with the Herbert headless compression screw: outcomes and complications in 64 consecutively treated joints. Hand. 2011;6:56–9.

    Article  PubMed Central  PubMed  Google Scholar 

  6. Lamas-Gomez C, Proubasta I, Escriba I, Itarte J, Caceres E. Distal interphalangeal joint arthrodesis: treatment with Herbert screw. J South Orthop Assoc. 2003;12:154–9.

    CAS  PubMed  Google Scholar 

  7. Leibovic S. Internal fixation for small joint arthrodesis in the hand. The interphalangeal joints. Hand Clin. 1997;13:601–13.

    CAS  PubMed  Google Scholar 

  8. Leibovic S. Instructional course lecture. Arthrodesis of the interphalangeal joints with headless compression screws. J Hand Surg [Am]. 2007;32:1113–9.

    Article  Google Scholar 

  9. Olivier L, Gensigk F, Board T, Kendoff, Krehmeier U, Wolfhard U. Arthrodesis of the distal interphalangeal joint: description of a new technique and clinical follow-up at 2 years. Arch Orthop Trauma Surg. 2008;128:307–11.

    Article  CAS  PubMed  Google Scholar 

  10. Sabbagh W, Grobbelaar A, Clarke C, Smith P, Harrison D. Long-term results of digital arthrodesis with the Harrison-Nicolle peg. J Hand Surg (Br). 2001;26:568–71.

    Article  CAS  Google Scholar 

  11. Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psych Bull. 1979;86:420–8.

    Article  CAS  Google Scholar 

  12. Tomaino MM. Distal interphalangeal joint arthrodesis with screw fixation: why and how. Hand Clin. 2006;22:207–10.

    Article  PubMed  Google Scholar 

  13. Wyrsch B, Dawson J, Aufranc S, Weikert D, Milek M. Distal interphalangeal joint arthrodesis comparing tension-band wire and Herbert screw: a biomechanical and dimensional analysis. J Hand Surg [Am]. 1996;21:438–43.

    Article  CAS  Google Scholar 

  14. Zavitsanos G, Watkins F, Britton E, Somia N, Gupta A, Kleinbert G. Distal interphalangeal joint arthrodesis using intramedullary and interosseus fixation. Hand Surg. 1999;4:51–5.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

This work was supported by an Orthopaedic Research Initiation Grant from the Department of Orthopaedics and Rehabilitation of the Penn State Milton S. Hershey Medical Center.

Conflict of Interest

Rodney Brenneman has no conflicts of interest to disclose.

Justin Bigger has no conflicts of interest to disclose.

Michael Darowish has no conflicts of interest to disclose.

Statement of Human and Animal Rights

All procedures followed were in accordance with the ethical standards of the responsible committee on human experimentation (institutional and national) and with the Helsinki Declaration of 1975, as revised in 2008.

Statement of Informed Consent

No informed consent was obtained; 200 consecutive deidentified hand radiographs were used. The study was approved by the institutional IRB.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Michael Darowish.

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Darowish, M., Brenneman, R. & Bigger, J. Dimensional analysis of the distal phalanx with consideration of distal interphalangeal joint arthrodesis using a headless compression screw. HAND 10, 100–104 (2015). https://doi.org/10.1007/s11552-014-9679-x

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11552-014-9679-x

Keywords

Navigation