Abstract
Objective
The identification of the interhemispheric fissure (IF) is important in clinical applications for brain landmark identification, registration, symmetry assessment, and pathology detection. The IF is usually approximated by the midsagittal plane (MSP) separating the brain into two hemispheres. We present a fast accurate, automatic, and robust algorithm for finding the MSP for CT scans acquired in emergency room (ER) with a large slice thickness, high partial volume effect, and substantial head tilt.
Materials and methods
An earlier algorithm for MSP identification from MRI using the Kullback–Leibler’s measure was extended for CT by estimating patient’s head orientation using model fitting, image processing, and atlas-based techniques. The new algorithm was validated on 208 clinical scans acquired mainly in the ER with slice thickness ranging from 1.5 to 6 mm and severe head tilt.
Results
The algorithm worked robustly for all 208 cases. An angular discrepancy (°) and maximum distance (mm) between the calculated MSP and ground truth have the mean value (SD) 0.0258° (0.9541°) and 0.1472 (0.7373) mm, respectively. In average, the algorithm takes 10 s to process of a typical CT case.
Conclusion
The proposed algorithm is robust to head rotation, and correctly identifies the MSP for a standard clinical CT scan with a large slice thickness. It has been applied in our several CT stroke CAD systems.
Similar content being viewed by others
References
Bhanuprakash KN, Hu Q, Aziz A, Nowinski WL (2006) Rapid and automatic localization of the anterior and posterior commissure point landmarks in MR volumetric neuroimages. Acad Radiol 13(1): 36–54. doi:10.1016/j.acra.2005.08.023
Verard L, Allain P, Travere JM, Baron JC, Bloyet D (1997) Fully automatic identification of AC and PC landmarks on brain MRI using scene analysis. IEEE Trans Med Imaging 16(5): 610–616. doi:10.1109/42.640751
Talairach J, Tournoux P (1988) Co-planar stereotaxic atlas of the human brain. Thieme, Stuttgart, pp 5–8
Schaltenbrand G, Wahren W (1977) Atlas of stereotaxy of the human brain. Thieme, Stuttgart, pp 35–39
Nowinski WL, Qian G, Bhanuprakash KN, Hu Q, Aziz A (2006) Fast talairach transformation for magnetic resonance neuroimages. J Comput Assist Tomogr 30(4): 629–641. doi:10.1097/00004728-200607000-00013
Joshi S, Lorenzen P, Gerig G, Bullitt E (2003) Structural and radiometric asymmetry in brain images. Med Image Anal 7: 155–170. doi:10.1016/S1361-8415(03)00002-1
Liu Y, Collins RT, Rothfus WE (2001) Robust midsagittal plane extraction from normal and pathological 3-D neuroradiology image. IEEE Trans Med Imaging 20: 173–192
Volkau I, Bhanuprakash KN, Ananthasubramaniam A, Gupta V, Aziz A, Nowinski WL (2006) Quantitative analysis of brain symmetry by using the divergence measure: normal-pathological brain discrimination. Acad Radiol 13: 752–758. doi:10.1016/j.acra.2006.01.043
Crow TJ (1990) Temporal lobe asymmetry as the key to the etiology of schizophrenia. Schizophr Bull 16: 433–443
Shirakawa O, Kitamura N, Lin XH, Hashimoto T, Maeda K (2001) Abnormal neurochemical asymmetry in the temporal lobe of schizophrenia. Prog Neuropsychopharmacol Biol Psychiatry 25(4): 867–877. doi:10.1016/S0278-5846(01)00149-X
Onitsuka T, McCarley RW, Kuroki N, Dickey CC, Kubicki M, Demeo SS, Frumin M, Kikinis R, Jolesz FA, Shenton ME (2007) Occipital lobe gray matter volume in male patients with chronic schizophrenia: a quantitative MRI study. Schizophr Res 92: 197–206. doi:10.1016/j.schres.2007.01.027
Gupta V, Bhanuprakash KN, Nowinski WL (2008) Automatic and rapid identification of infarct slices and hemisphere in DWI scans. Acad Radiol 15: 24–39. doi:10.1016/j.acra.2007.07.024
Chun-Chih L, I-Jen C, Furen X, Jau-Min W (2006) Tracing the deformed midline on brain CT. Biomed Eng Appl Basis Comm 18: 305–311. doi:10.4015/S1016237206000452
Liu X, Imielińska C, Connolly ES, D’Ambrosio A (2006) Automatic correction of the 3D orientation of the brain imagery. ISSPIT, 27–30 August 2006. Vancouver, Canada
Tuzikov A, Colliot O, Bloch I (2002) Brain symmetry plane computation in MR images using inertia axes and optimization. In: Proceedings of 16th international conference on pattern recognition (ICPR), vol 1, Quebec City, Canada, pp 1051–1054
Ardekani BA, Kershaw J, Braun M, Kanno I (1997) Automatic detection of the mid-sagittal plane in 3D brain images. IEEE Trans Med Imaging 16: 947–952
Junck L, Moen JG, Hutchins GD, Brown MB, Kuhl DE (1990) Correlation methods for the centering, rotation, and alignment of functional brain images. J Nucl Med 31: 1220–1226
Prima S, Ourselin S, Ayache N (2002) Computation of the mid- sagittal plane in 3D brain images. IEEE Trans Med Imaging 21: 122–138. doi:10.1109/42.993131
Guillemaud R, Marais P, Zisserman A, McDonald B, Crow TJ, Brady M (1996) A three dimensional mid sagittal plane for brain asymmetry measurement. Schizophr Res 18(2): 183–184. doi:10.1016/0920-9964(96)85575-7
Hu Q, Nowinski WL (2003) A rapid algorithm for robust and automatic extraction of the midsagittal plane of the human cerebrum from neuroimages based on local symmetry and outlier removal. Neuroimage 20: 2153–2165. doi:10.1016/j.neuroimage.2003.08.009
Volkau I, Bhanuprakash KN, Ananthasubramaniam A, Aziz A, Nowinski WL (2006) Extraction of the midsagittal plane from morphological neuroimages using the Kullback–Leibler’s measure. Med Image Anal 10: 863–874. doi:10.1016/j.media.2006.07.005
Nowinski WL, Bhanuprakash KN, Volkau I, Ananthasubramaniam A, Beauchamp N Jr (2006) Rapid and automatic calculation of the midsagittal plane in magnetic resonance diffusion and perfusion images. Acad Radiol 13(5): 652–663. doi:10.1016/j.acra.2006.01.051
Nowinski WL, Qian G, Bhanu Prakash KN, Volkau I, Thirunavuukarasuu A, Hu Q, Ananthasubramaniam A, Liu J, Gupta V, Ng TT, Leong WK, Beauchamp NJ (2007) A CAD system for acute ischemic stroke image processing. Int J CARS 2(Suppl 1): 220–222. doi:10.1007/s11548-007-0132-2
Darius G, Mecislovas M (2007) Automatic symmetry plane extraction from falx cerebri areas in CT slices. Bildverarbeitung Med 2007: 267–271
Bergo Felipe PG, Falcão Alexandre X, Yasuda Clarissa L, Ruppert Guilherme CS (2009) Fast, accurate and precise mid-sagittal plane location in 3D MR images of the brain. Biomedical engineering systems and technologies. Commun Comput Inf Sci 25: 278–290. doi:10.1007/978-3-540-92219-3_21
Jackson SA, Thomas RM (2004) Cross-sectional imaging made easy. Churchill Livingstone, Edinburgh
Fitzgibbon A, Pilu M, Fisher R (1999) Direct least-square fitting of ellipses. IEEE Trans Pattern Anal Mach Intell 21: 476–480. doi:10.1109/34.765658
Nowinski WL, Qian G, Leong WK, Liu J, Kazmierski R, Urbanik A et al (2007) A stroke CAD in the ER. 93 Radiological Society of North America Scientific Assembly and Annual Meeting Program 2007, Chicago, USA, 25–30 November 2007, p 837
Nowinski WL, Qian G, Banuprakash KN, Volkau I, Leong WK, Huang S, Ananthasubramaniam A, Liu J, Ng TT, Gupta V (2008) Stroke suite: CAD systems for acute ischemic stroke, hemorrhagic stroke, and stroke in ER. Med Imaging Inform 4987: 377–386. doi:10.1007/978-3-540-79490-5_44
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Puspitasari, F., Volkau, I., Ambrosius, W. et al. Robust calculation of the midsagittal plane in CT scans using the Kullback–Leibler’s measure. Int J CARS 4, 535–547 (2009). https://doi.org/10.1007/s11548-009-0366-2
Received:
Accepted:
Published:
Issue Date: