Abstract
Purpose
Retinal microsurgery requires extremely delicate manipulation of retinal tissue where tool-to-tissue interaction forces are usually below the threshold of human perception. Creating a force-sensing surgical instrument that measures the forces directly at the tool tip poses great challenges due to the interactions between the tool shaft and the sclerotomy opening.
Methods
We present the design and analysis of a force measurement device that senses distal forces interior to the sclera using 1-cm long, 160 μm diameter Fiber Bragg Grating (FBG) strain sensors embedded in a 0.5 mm diameter tool shaft. Additionally, we provide an algorithm developed to cancel the influence of environmental temperature fluctuations.
Results
The force-sensing prototype measures forces with a resolution of 0.25 mN in 2 DOF while being insensitive to temperature.
Conclusion
Sub-millinewton resolution force sensors integrated into microsurgical instruments are feasible and have potential applications in both robotic and freehand microsurgery.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Patkin M (1977) Ergonomics applied to the practice of microsurgery. Aust NZ J Surg 47: 320–329. doi:10.1111/j.1445-2197.1977.tb04297.x
Gupta PK, Jensen PS, de Juan E Jr (1999) Surgical forces and tactile perception during retinal microsurgery, MICCAI’99. LNCS 1679: 1218–1225
Iordachita I et al (2006) Steady-hand manipulator for retinal surgery. In: Proceedings of the MICCAI medical robotics workshop
Mitchell B et al (2007) Development and application of a New steady-hand manipulator for retinal surgery. In: Proceedings of the IEEE international conference robotics & automation. Rome, pp 623–629
Okamura AM (2004) Methods for haptic feedback in teleoperated robot-assisted surgery. Ind Robot 31: 499–508. doi:10.1108/01439910410566362
Zhou Y, Nelson BJ, Vikramaditya B (1998) Fusing force and vision feedback for micromanipulation. In: Proceedings of the IEEE international conference in robotics & automation, Leuven, Belgium, pp 1220–1225
Kim K, Sun Y, Voyles R, Nelson B (2007) Calibration of multi-axis MEMS force sensors using the shape-from-motion method. IEEE Sens J 7(3):. doi:10.1109/JSEN.2006.890141
Menciassi A, Eisinberg A, Scalari G, Anticoli C, Carrozza MC, Dario P (2001) Force feedback-based microinstrument for measuring tissue properties and pulse in microsurgery. In: Proceedings of the IEEE international conference robotics & automation, Seoul, Korea (South), pp 626–631
Berkelman PJ, Whitcomb LL, Taylor RH, Jensen P (2003) A miniature microsurgical instrument tip force sensor for enhanced force feedback during robot-assisted manipulation. IEEE Trans Robot Autom 19(5): 917–921. doi:10.1109/TRA.2003.817526
Berkelman PJ, Whitcomb LL, Taylor RH, Jensen P (2000) A miniature Instrument Tip Force Sensor for Robot/Human Cooperative Microsurgical Manipulation with Enhanced Force Feedback. In: Medical Image Computing and Computer-Assisted Interventions, Pittsburgh, pp 897–906
Jagtap D, Riviere CN (2004) Applied force during vitreoretinal microsurgery with handheld instruments. In: Proc 26th IEEE engineering in medicine and biology conference (EMBS), San Francisco, pp 2771–2773
Hill KO, Meltz G (1997) Fiber Bragg grating technology fundamentals and overview. J Lightwave Technol 15(8): 1263–1276. doi:10.1109/50.618320
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License ( https://creativecommons.org/licenses/by-nc/2.0 ), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
About this article
Cite this article
Iordachita, I., Sun, Z., Balicki, M. et al. A sub-millimetric, 0.25 mN resolution fully integrated fiber-optic force-sensing tool for retinal microsurgery. Int J CARS 4, 383–390 (2009). https://doi.org/10.1007/s11548-009-0301-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11548-009-0301-6