Diffusion-weighted MRI in evaluating liver fibrosis: a feasibility study in cirrhotic patients

RM pesata in diffusione nella valutazione della fibrosi epatica: uno studio in pazienti cirrotici

  • ABDOMINAL RADIOLOGY
  • Published:
La radiologia medica Aims and scope Submit manuscript

Abstract

Purpose.

This study was designed to establish whether the measurement of apparent diffusion coefficients (ADCs) is clinically accurate in diagnosing liver fibrosis in a selected series of cirrhotic patients.

Materials and methods.

Twenty-eight cirrhotic patients (mean age 58.1 years) with histologically proven liver fibrosis and 29 healthy controls (mean age 43.8 yeas) underwent liver diffusion-weighted magnetic resonance (MR) using a 1.5-Tesla (T) magnet equipped with a phased-array coil. Diffusion studies with parallel imaging [generalized autocalibrating partially parallel acquisitions (GRAPPA)] were performed within a single breath-hold using a single-shot spin-echo echo-planar sequence (TE 74 ms) using four b values: b=0, 150, 250 and 400 s/mm2. A unidirectional diffusion gradient was applied. ADCs were measured on ADC maps.

Results.

Mean ADC was significantly lower in cirrhotic livers than in controls (1.11±0.16 vs. 1.54±0.12.10-3mm2/s) (p<0.0001). Receiver operating characteristic (ROC) analysis showed an area under the curve (AUC) of 0.96 [confidence interval (CI) 95%:(0.87; 0.94)], demonstrating higher sensitivity and specificity (92.9% and 100%, respectively) for an ADC cutoff of 1.31.10-3mm2/s. Positive predictive value (PPV), negative predictive value (NPV) and overall accuracy were 100%, 99.9% and 96.4%, respectively.

Conclusions.

Diffusion-weighted MRI is an accurate tool in evaluating advanced liver fibrosis if an optimised single-shot spinecho echo-planar sequence with maximum intermediate b value is used. The ADC threshold for liver fibrosis was 1.31.10-3mm2/s.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to R. Girometti.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Girometti, R., Furlan, A., Bazzocchi, M. et al. Diffusion-weighted MRI in evaluating liver fibrosis: a feasibility study in cirrhotic patients. Radiol med 112, 394–408 (2007). https://doi.org/10.1007/s11547-007-0149-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11547-007-0149-1

Key words

Navigation