Scaling Marker Position Determines the Accuracy of Digital Templating for Total Hip Arthroplasty

  • Original Article
  • Published:
HSS Journal ®

Abstract

Background

Digital templating systems foster patient-specific measurements for preoperative planning.

Questions/Purposes

We aim (1) to verify the accuracy of a templating system, (2) to describe the effects of scaling marker position on the accuracy of digital templating of the hip, and (3) to provide a practical guide for scaling marker position using patient body mass index (BMI).

Methods

A scaling sphere was placed in five positions along the anterior-posterior axis of an acetabular implant and pelvis phantom, and x-rays were obtained. Each radiograph was templated for the acetabular component and recorded. A retrospective review identified CT scans of preoperative hip arthroplasty cases. The center of the greater trochanter was calculated from these CT scans as the percent distance from the anterior thigh and recorded with the patient’s BMI.

Results

By centering the scaling sphere on the acetabular component, an accurate cup size was achieved. A difference of 3.5 cm in sphere placement resulted in a full cup size magnification error. Positioning the scaling sphere at the level of the pubic symphysis resulted in a difference of four cup sizes. This patient population had an average BMI of 28.72 kg/m2 (standard deviation 6.26 kg/m2) and an average position of the center of the greater trochanter of 51% (standard deviation of 6%) from the anterior surface of thigh.

Conclusions

Digital templating relies on scaling marker position to accurately estimate implant size. Based on the findings in this study, scaling markers for hip imaging should be placed laterally, mid-thigh in the anterior-posterior direction for patients with a BMI between 25 and 40 kg/m2. If abnormal hip anatomy or extremes of BMI are discovered, then scaling sphere positioning should be optimized on a case-by-case basis. Digital templating systems for total hip arthroplasty must use precisely placed scaling markers at the level of the hip joint to allow for accurate implant size estimation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Boese CK, Lechler P, Rose L, et al. Calibration Markers for Digital Templating in Total Hip Arthroplasty. PLoS One. 2015; 10:e0128529.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Brew CJ, Simpson PM, Whitehouse SL, et al. Scaling digital radiographs for templating in total hip arthroplasty using conventional acetate templates independent of calibration markers. J Arthroplasty. 2012; 27:643–647.

    Article  PubMed  Google Scholar 

  3. Hsu AR, Gross CE, Bhatia S, et al. Template-directed instrumentation in total knee arthroplasty: cost savings analysis. Orthopedics. 2012; 35:e1596–1600.

    Article  PubMed  Google Scholar 

  4. Iorio R, Siegel J, Specht LM, et al. A comparison of acetate vs digital templating for preoperative planning of total hip arthroplasty: is digital templating accurate and safe? J Arthroplasty. 2009; 24:175–179.

    Article  PubMed  Google Scholar 

  5. Issa K, Pivec R, Boyd B, et al. Comparing the accuracy of radiographic preoperative digital templating for a second- versus a first-generation THA stem. Orthopedics. 2012; 35:1028–1034.

    Article  PubMed  Google Scholar 

  6. Ramme AJ, Iorio R, Smiaronksi J, et al. Organizational and Technical Considerations for the Implementation of a Digital Orthopaedic Templating System. Bull Hosp Jt Dis (2013); 74:254–261.

    Google Scholar 

  7. Sinclair VF, Wilson J, Jain NP, et al. Assessment of accuracy of marker ball placement in pre-operative templating for total hip arthroplasty. J Arthroplasty. 2014; 29:1658–1660.

    Article  PubMed  Google Scholar 

  8. White SP and Shardlow DL. Effect of introduction of digital radiographic techniques on pre-operative templating in orthopaedic practice. Ann R Coll Surg Engl. 2005; 87:53–54.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Wimsey S, Pickard R and Shaw G. Accurate scaling of digital radiographs of the pelvis. A prospective trial of two methods. J Bone Joint Surg Br. 2006; 88:1508–1512.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to thank George Rodriguez, CRT, for his assistance in image acquisition.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jonathan M. Vigdorchik MD.

Ethics declarations

Conflict of Interest

Austin J. Ramme MD, PhD; Nina D. Fisher, BS; Jonathan Egol; Gregory Chang, MD; Jonathan M. Vigdorchik, MD have declared that they have no conflict of interest.

Human/Animal Rights

All procedures followed were in accordance with the ethical standards of the responsible committee on human experimentation (institutional and national) and with the Helsinki Declaration of 1975, as revised in 2008 (5).

Informed Consent

Informed consent was waived from all patients for being included in the study.

Required Author Forms

Disclosure forms provided by the authors are available with the online version of this article.

Additional information

Work performed at the NYU Hospital for Joint Diseases.

Electronic supplementary material

ESM 1

(PDF 1224 kb)

ESM 2

(PDF 1224 kb)

ESM 3

(PDF 1224 kb)

ESM 4

(PDF 1224 kb)

ESM 5

(PDF 1224 kb)

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ramme, A.J., Fisher, N.D., Egol, J. et al. Scaling Marker Position Determines the Accuracy of Digital Templating for Total Hip Arthroplasty. HSS Jrnl 14, 55–59 (2018). https://doi.org/10.1007/s11420-017-9578-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11420-017-9578-0

Keywords

Navigation