INTRODUCTION

Over the past few decades, therapeutic proteins have become established as a major drug class for the pharmaceutical industry. A major challenge in commercializing proteins as drug candidates is their inherent physical and chemical instability. Of the numerous degradation pathways, protein aggregation is a particular cause for concern (1,2). It is believed that protein aggregates may have the propensity in vivo to trigger an antibody response against the monomeric form of that protein, also known as unwanted immunogenicity (1,3). The presence of antibodies against a therapeutic protein can have serious clinical consequences, such as loss of therapeutic efficacy or even the neutralization of the equivalent endogenous protein (4). Therefore, determination of the amount and type of aggregates necessary to allegedly trigger such immune reactions is of major importance both for pharmaceutical companies and regulatory agencies. With regard to their size, protein-containing samples can be very heterogeneous, and individual aggregates can range in diameter from a few nanometers to several micrometers or larger (2,5,6).

A number of characterization tools are available for determining the size of protein aggregates, e.g. size exclusion chromatography and asymmetrical flow field-flow fractionation combined with UV and light scattering detectors, dynamic light scattering (DLS) and nanoparticle tracking analysis (NTA), analytical ultracentrifugation, light obscuration particle counting and micro flow imaging (711). A major limitation of most of these techniques is that it is very difficult, if not impossible, to use them for analyzing therapeutic proteins or their aggregates in biological media. In fact, very little is known about the fate of protein pharmaceuticals following administration to a patient. To address this problem, we set out to identify a new strategy and analytical tool that would allow us to characterize aggregates of a specific protein in the presence of serum or plasma.

Very recently, fluorescence Single Particle Tracking (fSPT) was suggested as a powerful method to size submicron matter in undiluted biological fluids (12). Using fSPT, it was shown that the aggregation of fluorescently labeled liposomal drug carriers can be followed in real time in undiluted whole blood. In the present work, we have evaluated the potential of the fSPT technique to detect and determine the size of submicron protein aggregates in serum and plasma. To this end, we have covalently labeled the proteins of interest with a fluorescent probe. This strategy has previously been applied for the characterization of a specific protein interaction in human serum using analytical ultracentrifugation in an instrument modified to incorporate a fluorescence detection system (13). The authors demonstrated that the fluorescence probe did not affect either the characteristics or the potency of the protein. The same labeling strategy may also be used to characterize the aggregates of therapeutic proteins in complex formulations containing, for example, high amounts of stabilizing proteins, such as human serum albumin (HSA).

In this work, a monoclonal antibody (MAb) was fluorescently labeled without affecting its aggregation profile. Submicron MAb aggregates were generated by applying stress and were successfully analyzed in biological fluids and in a formulation containing a large amount of HSA.

MATERIAL AND METHODS

Diluents

The formulation buffer used to prepare the polystyrene beads and proteins contained 10 mM sodium citrate (Merck, Darmstadt, Germany), 5% (w/v) sucrose (Sigma-Aldrich, Buchs, Switzerland), pH 6.0. The formulation buffer was filtered using a 0.22-μm PES low binding syringe-driven filter unit (Millex™ GP, Millipore, Ireland).

Human serum and plasma were collected from three healthy volunteers free of medications. Serum was collected in Vacutainer SST tubes, and plasma was collected in Vacutainer heparin tubes (Becton Dickinson, Franklin Lakes, USA). The plasma samples were centrifuged for 2 min at 2,000 rpm with a Sigma 1-15 centrifuge (Osterode, Germany) mainly to spin down the red blood cells. The serum samples were centrifuged for 15 min at 3,000 rpm in a Beckman Coulter Alegra X-12 centrifuge (Brea, USA) to remove all the blood cells and clotting factors. The serum and plasma samples were stored at 4°C for a maximum period of 48 h before being used for measurements. The viscosities of the buffer, serum and plasma were measured in an AR-G2 rheometer from TA Instruments (New Castle, USA) at 37°C. The average values used for sizing calculations are 0.80 cP, 1.29 cP and 1.30 cP for buffer, serum and plasma, respectively.

Preparation of Fluorescent Polystyrene Beads

Polystyrene FluoSpheres carboxylate-modified microspheres with a nominal size of 0.1 μm, yellow-green fluorescent (excitation 505 nm/emission 515 nm), were purchased from Invitrogen (Merelbeke, Belgium) and covalently linked with methoxy-polyethylene glycol-amine (m-PEG-amine) (5000 Da, Creative PEGWorks, Winston Salem, U.S.A.) polymer chains using a method adapted from Suh et al. in order to minimize interactions with serum components (14). The fluorescent PEGylated beads were separately diluted with formulation buffer, serum and plasma until the concentration was acceptable for either NTA or fSPT measurements, i.e. 1:5,000 volume-based dilution for NTA measurements and 1:1,000 volume-based dilution for fSPT measurements.

Preparation of Alexa Fluor-Labeled IgG

Alexa Fluor 488, Alexa Fluor 546 and Alexa Fluor 594 carboxylic acid, N-hydroxysuccinimide esters were obtained from Invitrogen (Merelbeke, Belgium). IgG and HSA labeling was performed according to the manufacturer’s instructions. In brief, a recombinant human MAb of the IgG1 subclass (11) was prepared at a concentration of 10 mg/ml in 100 mM sodium phosphate (Fluka, Sigma-Aldrich, Buchs, Switzerland) pH 7.4. To this was added a solution of dye in DMSO (10 mg/ml) to give a molar ratio of 4:1 (dye:protein). The same procedure was followed for HSA. A pH of 7.4 was chosen in order to achieve selective labeling of the amine termini. The samples were mixed and left for 1 h at room temperature. The conjugates were purified by passing the samples through disposable PD10 desalting columns (GE Healthcare, Uppsala, Sweden) packed with Sephadex G25 medium that had been equilibrated with phosphate-buffered saline (PBS). Finally, the samples were dialyzed using a Float-A-Lyzer® G2 (Spectrum, Rancho Dominguez, USA) with a 100 kDa molecular weight cut-off membrane to remove excess dye and to exchange the buffer from PBS back to the original formulation buffer (see Diluents subsection). The final labeling ratio achieved was about four Alexa Fluor 546 labels per IgG, about two Alexa Fluor 594 labels per IgG and about two Alexa Fluor 488 labels per HSA, according to the formula suggested by Invitrogen:

$$ \frac{{{n_{{Dye}}}}}{{{n_{{Protein}}}}} = \frac{{{A_{{\max }}}}}{{{\varepsilon_{{Dye}}} \times {c_{{\Pr otein}}} \times l}} $$

where n Dye is the moles of dye, n Protein is the moles of protein, A max is the measured absorbance at the dye’s absorption maximum, εdye is the molar extinction coefficient of the dye at its absorption maximum, c Protein is the molar protein concentration (determined by absorbance at 280 and corrected for the presence of dye, according to the correction formulas given by the dye supplier) and l is the path length.

Preparation of Fluorescently Labeled Protein Aggregates

Covalent aggregates of Alexa Fluor 546-labeled IgG (A546 IgG) were prepared by using glutaraldehyde (Sigma-Aldrich, Steinheim, Germany) as a cross-linker. The aggregates were obtained by incubating 0.5 mg/ml of A546 IgG with 0.2% (v/v) of glutaraldehyde for 2 min at room temperature. The cross-linking reaction was terminated by adding sodium borohydride (Sigma-Aldrich, Steinheim, Germany) at a final concentration of 0.008% (w/v). This reaction was then allowed to proceed for 24 h at room temperature. The sample was diluted 250-fold with the formulation buffer, serum or plasma before the fSPT analysis and 5-fold before the NTA measurements.

For the preparation of fluorescently labeled protein aggregates in a complex formulation, Alexa Fluor 594-labeled IgG (A594 IgG) was prepared at a concentration of 0.5 mg/ml in formulation buffer (see Diluents subsection) containing Alexa Fluor 488-labeled HSA (A488 HSA) at a concentration of 3.2 mg/ml. The sample was placed in a 1.5-ml reaction tube (Eppendorf, Hamburg, Germany) and incubated for 12 min at 74°C in a VWR Digital Heat Block (VWR, West Chester, USA). The sample was diluted 1,000-fold with the formulation buffer before the fSPT and 50-fold before the NTA measurements.

Nanoparticle Tracking Analysis (NTA)

NTA measurements were performed essentially as described before with a NanoSight LM20 (NanoSight, Amesbury, United Kingdom), equipped with a sample chamber with a 640-nm laser and a Viton fluoroelastomer O-ring (11). The samples were injected in the sample chamber with sterile syringes (BD Discardit II, New Jersey, USA), and all measurements were performed at 37°C. The software used for capturing and analyzing the data was the NTA 2.0 Build 127. The samples were measured for 40 s with manual shutter and gain adjustments. Three measurements of the same sample were performed for the polystyrene beads and six measurements for the protein aggregates. The standard deviations of the mean size values were calculated from the values obtained in each measurement.

Light Obscuration (LO)

LO measurements were performed on a PAMAS SVSS system (PAMAS GmbH, Rutesheim, Germany) equipped with a HCB-LD-25/25 sensor and a 1 ml syringe. Each sample was measured three times, with each measurement consisting of a pre-run volume of 0.3 ml followed by three runs of 0.2 ml at a flow rate of 10 ml/min. The measurement settings differ significantly from the United States Pharmacopeia (USP) method, but they were chosen in order to reduce sample volume, since it has been shown that small sampling volumes are appropriate for quantifying size and levels of subvisible particles at the amounts typically present in protein therapeutics (15). The samples were diluted 100-fold with the formulation buffer before the measurements in order to keep the particle counts below the upper count limit of the device.

Fluorescence Single Particle Tracking (fSPT)

The fSPT technique was recently described in detail elsewhere (12). In short, widefield laser illumination was provided to an inverted epi-fluorescence microscope (Nikon TE2000E, NIKON BELUX, Brussels, Belgium) for excitation of the fluorescently labeled sample. The fSPT setup was equipped with three solid state lasers: a 100 mW Calypso 491 nm (Cobolt, Solna, Sweden), a 75 mW Jive 561 nm (Cobolt) and a IQ1C 30 mW 636 nm (Power Technology, Little Rock, AR). Using a fast and sensitive electron-multiplying CCD camera (Cascade II:512; Roper Scientific, Tucson, AZ), videos were acquired (Nikon Elements R imaging software) of the fluorescently labeled protein aggregates diffusing in the medium (formulation buffer, serum, or plasma). A Nikon Plan Apochromat 100× NA1.4 oil immersion objective lens was used for imaging.

Using custom-built software, diffusional motion trajectories were calculated for individual protein aggregates from the fSPT videos. Since the diffusion coefficient for each trajectory could be calculated, a distribution of diffusion coefficients was obtained by analyzing a large number of trajectories. A maximum entropy deconvolution algorithm was used to reduce statistical broadening and sampling noise in the distribution of diffusion coefficients. Finally, this distribution of diffusion coefficients was converted to a size distribution using the Stokes-Einstein equation for spherical particles:

$$ d = \frac{{kT}}{{3D\pi \eta }} $$

where d is the particle diameter, k the Boltzmann constant, T the absolute temperature, η the dynamic viscosity of the solution and D the diffusion coefficient.

Each measurement consisted of the recording and analysis of ten videos of the same sample for the polystyrene beads and 20 videos for the protein aggregates. Each video was recorded for only 5 s in order to minimize photobleaching. The standard deviations of the mean size values were calculated from the values obtained in each video.

The upper and lower size and concentration limits of fSPT are hard to specify, since they are dependent on several complex analytical factors. The lower size limit for detecting a particle depends on the ratio between the background intensity and the number of photons that can be collected from that particle, which in turn is dependent on the brightness of the fluorescent label, laser intensity and illumination time. Nevertheless, in our experiments, particles as small as 50 nm could be successfully measured. The upper size limit is mostly related to the particle density and fluid viscosity in such a way that particle sedimentation becomes the limiting factor. In a general way, an upper size limit of about 1000 nm can be set for fSTP measurements.

With this technique, the upper limit of particle concentration is dictated by the optical resolution and the tracking algorithm, resulting in a maximum useful particle concentration that is typically in the nanomolar range (12). Theoretically, there is no lower limit of particle concentration, since the size is calculated for individual particles. However, it is clear that there should be a sufficient number of particles in the system to obtain good statistics within a reasonable time period.

Colocalization Analysis

In order to verify whether IgG and HSA can form mixed aggregates, A594 IgG (red fluorescent) and A488 HSA (green fluorescent) were combined, and the mixture was subsequently subjected to heat stress (12 min at 74°C). The resulting aggregate-containing solution was then analyzed by dual color fSPT.

Dual color is an fSPT mode in which videos can be recorded simultaneously in two different spectral ranges (green and red in this case), enabling separate video analysis and the detection of particle colocalization between both videos. Since aggregates are seen in fSPT videos as separate objects (spots of light), an object-based algorithm was employed for quantification. For each frame of the movie, all green and red objects were identified by image processing (for details, see Braeckmans et al.) (16). For each object, the contour and centre position were determined. A green object was considered colocalized with a red object if its center position fell within the contour of the red object. Vice versa, a red object was considered colocalized with a green object if its center position fell within the contour of the green object. By performing this analysis for all frames of a movie, the percentage of mixed aggregates relative to the pure IgG or HSA aggregates could be estimated.

RESULTS AND DISCUSSION

Fluorescent Polystyrene Beads in Biological Fluids

In line with previous studies, we first verified that fSPT can accurately detect and size submicron particles in biological fluids (12). PEGylated standard fluorescent polystyrene beads of 0.1 μm nominal size were separately diluted in buffer, serum and plasma, and the size distributions were then measured by fSPT (Fig. 1). The samples in buffer were also analyzed by NTA as a control, since both sizing techniques are based on the same principle of tracking individual particles and calculating the particle hydrodynamic size from the measured diffusion coefficient. However, the standard NTA instrument, as described by Filipe et al. (11), relies on light scattered by the particles and, therefore, cannot be used for biological media like serum or plasma, which contain a variety of light scattering components.

The size distributions obtained by fSPT and NTA for the PEGylated beads in buffer are comparable, confirming the accuracy of fSPT to size nanoparticles. The mean size given by NTA was 129 ± 7 nm, while the mean sizes given by fSPT were 139 ± 10 nm in buffer, 143 ± 7 nm in serum and 142 ± 8 nm in plasma. The source of the small difference between the means given by the two techniques in buffer is unknown, but may have to do with the contribution of small impurities (between 50 and 120 nm) associated with the PEGylation process. Such impurities would decrease the NTA mean but cannot be detected by fSPT, since they would not be fluorescent. In any case, the mean values are within the standard deviations obtained for both techniques. These standard deviations of the means are small and very similar between the two techniques, confirming the precision of fSPT.

The fSPT size distributions are also very similar in each of the media, indicating that the beads do not aggregate in biological fluids and that these fluids do not interfere with the fluorescence properties of the beads. Moreover, it can be concluded that the clotting factors, which are present in plasma and not in serum, have no major effect on the size distribution of the beads. These results demonstrate the potential of this technique to measure the size distribution of nanoparticles in biological fluids.

Effect of Labeling on Aggregation Profile

With the aim of evaluating the effect of the fluorescent label on the aggregation profile of this specific IgG, a comparative aggregation study between A546-labeled and unlabeled IgG was performed. The samples were stressed by glutaraldehyde cross-linking under the same conditions, and the aggregates were analyzed by NTA and LO (Fig. 2). Combined, these two techniques focus on the size range covered by fSPT (submicron range) and on the size range that can interfere with fSPT measurements (micron range).

Fig. 1
figure 1

Size distribution from fSPT and NTA measurements of PEGylated fluorescent polystyrene beads in buffer, serum and plasma.

Fig. 2
figure 2

Size distribution from NTA and LO measurements in buffer of A546 IgG and unlabeled IgG aggregates induced by glutaraldehyde cross-linking.

The NTA size distribution of the labeled IgG samples is very similar to that of the unlabeled one. The height difference between the two peaks is within the variations inherent to the stress method itself and therefore should not be taken into account. The LO measurements also show a similar size distribution and particle count per size bin between the two samples. These results lead to the conclusion that this fluorescent label does not have a major impact on the aggregation profile in the subvisible and submicron range of this specific IgG with this particular stress method.

These findings should not be generalized for other proteins, other fluorescent labels or other stress methods. During optimization studies, we verified that some fluorescent dyes may significantly change the aggregation profile of some proteins in certain given conditions (data not shown). Therefore, a case-by-case approach should be followed to evaluate the effect of a fluorescent label on a certain protein’s stability.

Covalent A546 IgG Aggregates in Biological Fluids

In order to evaluate the analytical performance of fSPT for protein aggregates in biological fluids, covalent submicron A546 IgG aggregates were analyzed in buffer, serum and plasma (Fig. 3). Glutaraldehyde cross-linked covalent aggregates were used, since they cannot be dissociated and can be considered to be stable aggregates. Unstressed A546 IgG was also analyzed by fSPT in each of the media and was shown not to contain aggregates. As an additional control, the sample diluted in buffer was also measured by NTA.

Fig. 3
figure 3

Size distribution from fSPT and NTA measurements of glutaraldehyde cross-linked A546 IgG in buffer, serum and plasma.

The size average of the A546 IgG aggregates in buffer measured with NTA was 215 ± 38 nm and with fSPT 198 ± 36 nm. The difference observed in peak width between the NTA and fSPT size distributions in buffer can be attributed to a difference in the size analysis algorithm. In the fSPT sizing method used here, the raw size distribution as measured directly from the particle trajectories is refined by a maximum entropy deconvolution step that filters out statistical broadening and sampling noise so as to not over-interpret features in the distribution that are not statistically warranted by the data (12). Although a similar function is available in the NTA software (model curve fitting), we chose to use the log-normal fit of the data for this sample, given that according to the manufacturer, this function should only be used for either very monodisperse or clearly bimodal samples. Another factor that can contribute to the difference in peak width is the difference in optimal dilutions (50-fold) required for measurements with each technique, since higher dilutions may affect the aggregate size distribution (7).

These glutaraldehyde cross-linked aggregates showed an increase in their mean size in serum and plasma compared to buffer (198 ± 36 nm, 265 ± 38 nm and 245 ± 34 nm in buffer, serum and plasma, respectively). This size increase could be due to self-association of the aggregates or to interactions between the aggregates and other serum and plasma components, a phenomenon normally described as opsonization (17). In general, opsonization is the adsorption of plasma components (i.e. opsonins) onto the surface of a foreign particle to render it more susceptible to phagocytosis. Opsonins include mainly immunoglobulins, blood clotting factors and components of the complement system (18,19). Given that the IgG used in this work is humanized and that IgG is one of the most abundant classes of proteins circulating in blood, one would not expect it to be opsonized. However, protein cross-linking with glutaraldehyde leads to a high amount of intermolecular and intramolecular cross-links, which normally lead to conformational changes (2022). Consequently, since opsonins bind to nanoparticles mainly via hydrophobic interactions, it is possible that these unnatural aggregates of conformationally changed IgG may lead to some opsonization (23).

Given that the size distribution of these aggregates is relatively the same in serum and in plasma, it is possible to conclude that the presence of clotting factors has little impact on the hypothetical opsonization, since they do not have a major effect on the size average of these aggregates.

A594 IgG Aggregates Formulated with A488 HSA

Over the last few decades, the need to stabilize therapeutic proteins progressively led to rather complex formulations containing sugars, surfactants, amino acids and even other proteins. However, such intricate environments often make the selective characterization of the therapeutic protein very difficult. For example, several therapeutic protein formulations contain high concentrations of HSA, which is known to act as a general stabilizer for liquid and lyophilized protein formulations by effectively inhibiting protein surface adsorption (24,25). Unfortunately, with most protein characterization tools, it is virtually impossible to distinguish between the excipient HSA and the therapeutic protein without prior separation. When considering protein aggregates, the situation becomes even more complex, as both the protein drug and HSA may form separate or mixed aggregates.

This section aims to show the potential of fSPT for the selective characterization of protein aggregates in formulations containing high amounts of HSA. For this purpose, A594 IgG was mixed with A488 HSA at an IgG:HSA concentration ratio of 1:6 (w/w); the mixture was heated in order to stimulate the formation of aggregates with diameters of the order of a hundred nanometers (monitored by fSPT). The sample was then analyzed by fSPT using dual color imaging (Fig. 4), whereby Alexa Fluor 488 and Alexa Fluor 594-labeled aggregates could be tracked simultaneously to examine whether HSA and IgG form mixed aggregates or not. The overall aggregate size was also measured by NTA, although it cannot distinguish between IgG aggregates, HSA aggregates or mixed aggregates (Fig. 4).

Fig. 4
figure 4

Size distribution from fSPT and NTA measurements of a heat-stressed mixture of A594 IgG and A488 HSA in buffer.

The average size of the aggregates determined by NTA was 152 nm. By fSPT, the average size of the IgG-containing and HSA-containing aggregates was 161 nm and 122 nm, respectively (Table I). These results suggest that either the aggregates of both proteins have coincidently the same size range or that these aggregates are composed of both proteins. In order to clarify this uncertainty, the videos from both IgG (red fluorescence) and HSA (green fluorescence) channels were aligned, and an object-based colocalization was calculated. A representative fSPT video frame is shown on Fig. 5, wherein colocalization is clearly visible (see also Supplementary Material for complete overlaid video). The results show that the IgG-containing aggregates colocalize significantly with the HSA-containing aggregates. About 70 ± 9% of the aggregates tracked in the IgG channel were found to be colocalized with HSA-containing aggregates, while only about 36 ± 4% of the aggregates tracked in the HSA channel colocalized with IgG-containing aggregates. The amount of aggregates tracked in the HSA channel was approximately twice the amount of aggregates tracked in the IgG channel. This indicates that there were more HSA-containing aggregates than IgG-containing aggregates, and this explains the low colocalization observed for the aggregates in the HSA channel. The fact that there were more HSA aggregates is not surprising considering the IgG:HSA concentration ratio of 1:6 (w/w). An example of this occurrence can be seen in Fig. 5, where two out of five HSA aggregates do not colocalize with IgG aggregates.

Table I Mean Size from NTA and fSPT Measurements of a Heat-Stressed Mixture of A594 IgG and A488 HSA in Buffer
Fig. 5
figure 5

A video frame containing aggregates captured in the IgG and HSA channels using the dual color mode of fSPT, with the corresponding overlay image.

Previous studies have already suggested the interaction between therapeutic proteins and HSA to form mixed aggregates in solution, mainly by enzyme-linked immunosorbent assays (26,27). In the present study, using dual color fSPT, we were for the first time able to visualize and get a clear glimpse at the aggregation profile of therapeutic proteins in formulations containing HSA. Here, we have clearly demonstrated that under conditions of heat stress, HSA can form mixed aggregates with the therapeutic protein: in this case, IgG. Given that heat stress induces protein denaturation and consequently exposure of hydrophobic regions, it is likely that HSA starts to bind to the partially denatured intermediates of the therapeutic protein and ends up involved in the overall aggregation process (25,28). In fact, the same heat stress was applied to the A594 IgG in a formulation without A488 HSA, and it resulted in a much higher amount of IgG aggregates with a higher size average, confirming the stabilizing properties of HSA in solution (data not shown).

It should be noted that control experiments in buffer with only one of the two proteins labeled indicated that the labeling of HSA or IgG did not change the size distribution. However, labeling of HSA (but not IgG) substantially reduced the amount of submicron aggregates (as measured by NTA), even though the amount of micron-sized aggregates (as measured by LO) remained approximately the same (data not shown). This may be due to the hydrophobic nature of the label and/or the (random) HSA labeling procedure. Further studies with other labels or different, site-specific labeling strategies should be done in order to understand and avoid this problem.

CONCLUDING REMARKS

In this work, we evaluated the potential of fSPT for the characterization of protein aggregates in biological fluids and HSA-containing formulations. fSPT gave comparable size distributions to NTA for aggregates and particles in buffer and has the added advantage that it can also be applied to look at samples in serum and plasma. It seems that some serum components adsorb to the beads and protein aggregates once these come into contact with biological fluids. fSPT has also been shown to be suitable for selectively analyzing protein aggregates in complex formulations, such as those containing HSA. Using dual color fSPT, it was possible to verify that most of the IgG aggregates created by heat stress in an HSA-containing formulation were in fact composed of a mixture of IgG and HSA.

The requirement for prior labeling of the protein of interest with a fluorescent probe may be seen as a drawback of this technique. However, if it can be proven that the stability and aggregation profile of the proteins remain unaltered after the labeling, the potential of this technique to reveal the fate of protein aggregates in biological fluids is evident. This paves the way for several research areas, such as formulation and immunogenicity studies, where it is important to know if a certain therapeutic protein aggregates once injected into the bloodstream or if certain types of aggregates disappear or change their size in biological fluids.