Abstract
In this paper, it is made a study of the accuracy of an isothermal titration calorimeter in the operating mode of ‘continuous injection’. The experimental equipment has been a TAM2277-201/2250 by Thermometric AB and the liquid mixtures used in the calibration have been the mixture cyclohexane+benzene and the mixture water+ethanol. The calibration contemplates different effects that affect the uncertainty in the determination of the sensitivity, the effect of the liquid injection, the treatment of the calorimetric signal, the variation of the experimental baseline and the different noises included in the calorimetric signal.
Similar content being viewed by others
References
A. Ababou and J. E. Ladbury, J. Mol. Recognit., 19 (2006) 79.
H. Jelesarov and H. R. Bosshard, J. Mol. Recognit., 12 (1999) 3.
S. Tai and K. C. Tam, Colloids. Surf. A: Physicochem. Eng. Aspects, 229 (2003) 157.
H. Tachoire and V. Torra, Thermochim. Acta, 110 (1987) 171.
F. Socorro, M. Rodríguez de Rivera, J. P. Dubès, H. Tachoire and V. Torra, Meas. Sci. Technol., 1 (1990) 1285.
V. Torra and H. Tachoire. J. Therm. Anal. Cal., 52 (1998) 665.
I. Wadsö and L. Wadsö. J. Therm. Anal. Cal., 82 (2005) 553.
R. Sabbah, A. Xu-Wu, J. S. Chickos, M. L. P. Leitão, M. V. Roux and L. A. Torres, Thermochim. Acta, 331 (1999) 93.
M. J. Costigan, L. J. Hodges, K. N. Marsh, R. H. Stokes and C. W. Tuxford, Aust. J. Chem., 33 (1980) 2103.
R. H. Stokes, K. N. Marsh and R. P. Tomlins, J. Chem. Thermodyn., 1 (1969) 211.
A. Isalgue, J. Ortin, V. Torra and J. Viñals, An. Fis., 76 (1980) 192.
R. Kirchner, M. Rodríguez de Rivera, J. Seidel and V. Torra, J. Therm. Anal. Cal., 82 (2005) 179.
F. Marco, M. Rodríguez de Rivera, J. Ortin, T. Serra and V. Torra, Thermochim. Acta, 102 (1986) 173.
M. Rodríguez de Rivera and F. Socorro, J. Therm. Anal. Cal., 80 (2005) 769.
M. Rodríguez de Rivera and F. Socorro, J. Therm. Anal. Cal., 85 (2006) 477.
Guide to the Expression of Uncertainty in Measurement, International Organization for Standardization (ISO), Geneva, Switzerland 1995.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Rodríguez de Rivera, M., Socorro, F. Signal processing and uncertainty in an isothermal titration calorimeter. J Therm Anal Calorim 88, 745–750 (2007). https://doi.org/10.1007/s10973-006-8104-2
Published:
Issue Date: