Introduction

The Boulby Underground Laboratory of the Institute of Underground Science (IUS), University of Sheffield, is located deeper underground than any other underground research center in Europe [1]. The Laboratory is located 1,100 m below ground in the Boulby Mine (NE England). The total area of the Lab is over 1,000 m2, including associated facilities. The scientific programs at IUS comprise dark matter detection and astroparticle physics [2]. Because the experiments represent investigations of extremely rare events, it is necessary to determine precisely the natural radioactivity background levels. After cosmic rays, the most important sources of background radiation comprise the decay of primordial radionuclides, including 40K, 232Th, and 238U in rocks, building materials, and materials used in constructing detectors [3]. The main sources of background radiation are the neutrons that originate from (α, n) reactions and the spontaneous fission of U and Th, which can imitate some of the signals expected from the experiments. In this study, we performed in situ gamma-ray measurements inside the IUS, and we measured samples of the parent rocks that surrounded the Lab, which consisted mainly of halite (NaCl) with minor impurities of sylvite (KCl) and mudstone (ca 92 wt% SiO2).

Materials and methods

The measurements were made in 2006. We measured the background gamma radiation in the IUS in situ with a portable gamma-ray spectrometry (Fig. 1). The GX3020 system consisted of a coaxial HPGe detector (32 % efficiency, 59 mm crystal length, 56.6 mm diameter) with a cryostat mounted on a tripod or a special table, a 50 × 180 mm collimator (diameter 80 mm), a multichannel buffer (InSpector 2000 DSP), and a laptop computer. The detector had a bias voltage of 4,000 V and energy resolutions of 0.8 keV at 122 keV and 1.7 keV at 1.33 MeV. For efficiency calibrations and determination of radionuclides, we used In Situ Object Counting Software (ISOCS), Laboratory Sourceless Calibration Software (LabSOCS), and Genie 2000 v.3. software packages. The ISOCS software was designed to generate efficiency calibrations for many types of large objects, walls, and floors based on mathematical modeling instead of calibration sources. The total duration of a single measurement varied from about 1 to 67 h.

Fig. 1
figure 1

Locations and configurations of in situ measurements

Samples of halite and mudstone collected from the longwalls outside the Lab were crushed. Several months after collection, they were placed in a 450 cm3 Marinelli beaker, and measured with the same GX3020 HPGe detector in a lead and copper shield (60 mm). The energy calibration of the spectrometer was carried out with homogeneously dispersed 241Am, 109Cd, 139Ce, 57Co, 60Co, 137Cs, 113Sn, 85Sr, 88Y, and 203Hg radioisotopes embedded in silicone resin (certificate source type Marinelli Beaker Standard Source [MBSS], supplied by the Czech Metrological Institute). These measurements were performed at the Laboratory of Natural Radioactivity (Faculty of Earth Sciences, University of Silesia).

Locations of in situ measurements

Location point 1 was in the transport hall (Fig. 1.1). The end cap of the detector was mounted 90 cm above the drift level, 10 m from the transport entrance, and 2 m from the left longwall (facing the ground).

Location point 2 was in the HPGe detector room, with the detector facing the floor. The detector was placed 90 cm above the floor and 2 m from the side walls (Fig. 1.2).

Location point 3 was in a drift, 29 m to the left side of the main Lab entrance (Fig. 1.3). The detector was set on a table, oriented horizontally, 10 cm from the longwall, and 90 cm above the drift level.

Location point 4 was near the DRIFT detector. The spectrometer was placed on a tripod at 1.6 and 3.6 m from the side walls, facing the floor, at 90 cm above the floor plane (Fig. 1.4).

Location point 5 was in the transport hall (Fig. 1.5). The detector was mounted horizontally on a table, 10 cm from the longwall, and 90 cm above the drift level.

Location point 6 was in the Main Lab, 6.5 m from the service entrance, and 2.5 m from the side walls. The detector was set horizontally on a table, 90 cm above the floor and directed to the inside of the Lab. This geometry ensured close to a 4π field of view of the entire room (Fig. 1.6).

Location point 7 was in an area called the “H area”. The spectrometer was set on a tripod, facing down, 90 cm above the drift level, and 3 m from the side walls (Fig. 1.7).

Location point 8 was in the transport hall. The detector was equipped with a collimator; it was mounted on a table, set horizontally at 5 cm from the longwall, 90 cm above the drift level, and 8.5 m from the Lab exit (Fig. 1.8).

Location point 9 was the same as point 6, except that the detector was facing down and placed on a tripod at 90 cm above the floor plane (Fig. 1.9).

Location point 10 was at the border between the Lab area and the transport hall (Fig. 1.10). The detector was on a table, set horizontally, pointing into the transport hall, placed at 130 cm above the ground level.

The gamma-ray spectra, measured at locations inside and outside the Lab, are shown in Figs. 2 and 3, respectively.

Fig. 2
figure 2

In situ gamma-ray spectra at specified locations inside the Lab. The characteristic gamma-ray emitters are indicated above the corresponding peaks

Fig. 3
figure 3

In situ gamma-ray spectra at specified locations outside the Lab. The characteristic gamma-ray emitters are indicated above the corresponding peaks

Results and discussion

The count rates (γ s−1) at all location points are listed in Table 1. The gamma-ray fluxes (γ cm−2 s−1) from location 8 are given in Table 2. The count rates in the main gamma peaks and the gamma fluxes from these peaks at location 8 are presented in Tables 3 and 4, respectively. Table 5 summarizes the results of the activity measurements in halite and mudstone.

Table 1 Count rates (γ s−1) in particular energy ranges
Table 2 Gamma flux in γ cm−2 s−1 in the specified energy ranges measured at location 8
Table 3 Individual peak sizes in counts per seconds (CPS) for all 10 measurement locations
Table 4 Gamma flux in γ cm−2 s−1 from the main gamma peaks at location 8
Table 5 Measured 40K, 232Th, and 238U activity concentrations in halite and mudstone

In situ measurements

Inside the Lab (locations 2, 4, 6, 9, and 10), in the energy range of 7.4–2734.2 keV, the total count rates varied from 18 γ s−1 (location 4) to 28 γ s−1 (locations 6 and 9), with an arithmetic mean of 24(4) γ s−1 (Table 1; Fig. 4a). Outside the Lab (locations 1, 3, 5, 7, and 8), in the same energy range, the total count rates varied from 6.5 (location 8) to ~23 γ s−1 (location 1). The arithmetic mean for measurements taken without a collimator (locations 1, 3, 5, and 7) was 20(3) γ s−1 (Table 1; Fig. 4b). The average count rate inside the Lab was slightly higher than that outside the Lab. This indicated that building materials, detectors, and calibration sources contributed an additional gamma background. As expected, the highest count rates (between 7.4 and 249.8 keV) contributed, on average, 66 % of the total counts (0.66 fractional contribution; Table 5) at all measurement locations. In the higher energy ranges, the count rates noticeably decreased as the energy range increased. The average fractional contributions to the total counts were 0.16, 0.11, 0.07, 5 × 10−3, and 4 × 10−3 for ranges 250–500, 501–1005, 1006–1556, 1556–2056, and 2056–2734 keV, respectively (Table 1; Fig. 5). For measurements taken with a collimator (location 8), the total gamma flux was 0.128 γ cm−2 s−1 (Table 2). Similar to the count rates, the fractional contributions of the gamma fluxes rapidly decreased as the energy ranges increased; they were on the order of 10−2 in the 250–1,556 keV range, and 10−4 in the 1,556–2,734 keV range.

Fig. 4
figure 4

a Count rates at specified locations inside the Lab. Thick solid line—average count rate. b Count rates at specified locations outside the Lab. Thick solid line—average count rate from the measurements taken without a collimator

Fig. 5
figure 5

Average fractional contributions of count rates in the different energy ranges at all locations

The highest integral areas from the main gamma transitions were noted under the peak at 1,460 keV (40K), and under the peaks at 351.9 keV (214Pb) and 609.3 keV (214Bi) (Table 3). Compared to those, much smaller integrals were noted for the two most intense gamma transitions from the thorium series, i.e., at 911.6 keV (228Ac) and 2614.5 keV (208Tl). The gamma transitions at 2204.2 keV showed the lowest count rates, ranging from 1.22 × 10−3 (location 8) to 6.6 × 10−3 γ s−1 (location 10). Despite the low yield (1.28 %) of the 2204.2 keV transition [4], its contribution may be important to geo-neutrino and similar experiments, because it may overlap with the deuteron binding energy of 2.2 MeV. This energy is released in gamma rays as a result of inverse beta-decay in a liquid scintillator [5]. Gamma fluxes at location 8 (measured with a collimator, Table 4) showed peaks on the order of 10−3 to 10−4 γ cm−2 s−1 at most energy levels, and a peak on the order of 10−5 at 2204.2 keV. The gamma background radiation in the Boulby Underground Laboratory is lower than those observed in the Gran Sasso National Laboratory and the Laboratoire Souterrain de Modane [6, 7].

Laboratory measurements

The activity measurements for halite and mudstone (Fig. 6) are given in Table 5. The activity concentrations of 232Th and 238U were very low in both halite and mudstone, but they were significantly higher in mudstone than in halite. The average mudstone to halite ratios were 11, 6, and 18 for 40K, 232Th and 238U, respectively (Table 5). The activity concentrations calculated directly during in situ measurements (location 8) were 112(2), 1.88(6), and 1.63(6) Bq kg−1 for 40K, 232Th, and 238U, respectively. These results indicated that the activity around the Lab resulted from a mixture of mudstone and halite activities.

Fig. 6
figure 6

a A photograph of a halite sample with its corresponding gamma-ray spectrum. b A photograph of a mudstone sample with its corresponding gamma-ray spectrum

Conclusions

The gamma background was very low inside the Boulby Underground Laboratory. It was characterized by average count rates of 24 γ s−1 in the energy range of 7–2,734 keV. The gamma flux outside the Lab was 0.128 γ cm−2 s−1. The construction materials and calibration sources inside the Lab did not contribute significantly to the total gamma background.