Abstract
A new multi-sphere spectrometer based on water moderator equipped with a high efficiency 6LiI(Eu) crystal scintillation detector was suggested to establish a reliable neutron spectrometer. The energy response matrix of these spectrometers was calculated from 0.01 to 20 MeV by applying the MCNPX Monte Carlo code. Simulations were done for six spheres with a diameter range of 2–18 in. These simulations included a detailed description of the geometry of the system. The obtained responses were parameterized as a function of sphere diameter by a log-normal function, and the parameters were calculated for 30 discrete neutron energies. The newly calculated energy responses were compared to polyethylene sphere responses. In other computations, the response of the water and polyethylene spheres exposed to 252Cf, 239Pu–Be and Po–Be neutron sources were calculated by Monte Carlo modeling.
Similar content being viewed by others
References
ICRP Publication 103 (2007) The 2007 Recommendations of the international commission on radiological protection. Elsevier Health Sciences, Pergamon Press, Oxford
Wrixon AD (2008) New ICRP recommendations. J Radiol Prot 28:161–168
Nakamura T (2003) Recent development of advanced neutron detection technology. J Nucl Radiochem Sci 4:15–24
Thomas DJ, Alevra AV (2002) Bonner sphere spectrometers a critical review. Nucl Instrum Methods A 476:12–20
Khabaz R, Miri H (2011) Development of a Bonner sphere spectrometer with emphasis on decreasing the contribution of scattering by using a new designed shadow cone. J Radioanal Nucl Chem 289:789–794
Thomas PM, Harrison KG, Scott MC (1984) A multisphere neutron spectrometer using a central 3He detector. Nucl Instrum Methods A 224:225–232
Jacobs GJH, Van den Bosch RLP (1980) Calibration measurements with the multisphere method and neutron spectrum analyses using the SAND-II program. Nucl Instrum Methods 175:483–489
Vega-Carrillo HR (2002) TLD pairs, as thermal neutron detectors in neutron multisphere spectrometry. Radiat Meas 35(251):254
Vega-Carrillo HR, Ortiz-Hernandez A, Hernandez-Davila VM, Hernandez-Almaraz B, Montalvo TR (2010) H*(10) and neutron spectra around linacs. J Radioanal Nucl Chem 283:537–540
Vega-Carrillo HR, Hernandez-Almaraz B, Hernandez-Davila VM, Montalvo TR, Ortiz-Hernandez A (2010) Neutron spectrum and doses in a 18 MV LINAC. J Radioanal Nucl Chem 283:261–265
Tripathy SP et al (2009) Measurement of 241Am–Be spectra (bare and Pb-covered) using TLD pairs in multi-spheres: spectrum unfolding by different methods. Nucl Instrum Methods A 598:556–560
Bramblett RL, Ewing RI, Bonner TW (1960) A new type of neutron spectrometer. Nucl Instrum Methods 9:1–12
Vega-Carrillo HR, Gallego E, Lorente A (2009) Response matrix calculation of a Bonner sphere spectrometer based on a 6LiI(Eu) scintillator. Nucl Tech 168:359–363
Khabaz R, Miri H (2011) Measurement of neutron spectrum with multi-sphere using BF3 and evaluation of scattering effect on spectrum. Nucl Technol Radiat Prot 26:140–146
Mares V, Schraube H (1995) Improved response matrices of Bonner sphere spectrometers with 6LiI scintillation detector and 3He proportional counter between 15 and 100 MeV. Nucl Instrum Methods A 366:203–206
Khabaz R, Miri H (2011) Determination of 241Am–Be spectra using Bonner sphere spectrometer by applying shadow cone technique in calibration”. J Appl Sci 11:2849–2854
Vega-Carrillo HR et al (2007) Monte Carlo study to measure the neutron flux using the prompt gamma-rays in hydrogen. Rev Mex Fis S 53:8–11
Sanna R. (1973) Thirty one group response matrices for the multisphere neutron spectrometer over the energy range thermal to 400 MeV. USAEC, HASL-267
Chadwick MB, Oblozinsky P, Herman M et al (2006) ENDF/BVII.0: next generation evaluated nuclear data library for nuclear science and technology. Nucl Data Sheets 107:2931–3060
Waters LS (2002) MCNPX-A general Monte-Carlo N-particle transport code, Version 2.4.0, LANL report LA-CP-02-408, Los Alamos
Hakim M, Zahir I, Tassadit S, Malika A (2010) MCNP5 evaluation of a response matrix of a Bonner sphere spectrometer with a high efficiency 6LiI (Eu) detector from 0.01 to 20 MeV neutron. J Radioanal Nucl Chem 284:253–263
Anderson ME, Neff RA (1972) Neutron energy spectra of different size 239Pu–Be(α, n) sources. Nucl Instrum Methods 99:231–235
Notarrigo S, Parisi R, Ricamo R, Rubbino A (1962) Neutrons from Po–Be sources. Nucl Phys 29:507–514
Zaborowski HL (1981) Proceedings of the 4th symposium on neutron dosimetry, EUR 7448:575
ISO 8529-2 (2000) Reference neutron radiations-part 2: calibration fundamentals of radiation protection devices related to the basic quantities characterizing the radiation field. International Organization for Standardization, Geneva
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Khabaz, R. Study of a new multi-sphere spectrometer based on water moderator with a high efficiency 6LiI(Eu) detector. J Radioanal Nucl Chem 293, 383–389 (2012). https://doi.org/10.1007/s10967-012-1748-4
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-012-1748-4