Abstract
High-resolution gamma-spectrometry provides fast, accurate, non-destructive isotopic analysis of natural radionuclides in environmental samples. Low energy HPGe detectors (30%) with resolution less than 2.0 keV at 1332 keV of 60Co and with linked analytical software, provide good peakto-background performance and accurate unfolding of multiplet regions. A procedure regarding high-resolution gamma-spectrometry measurements of naturally occurring radioactive samples is presented here including calibration, background interference, self-absorption, practical considerations and validation with respect to IAEA reference materials. Recommended peaks for quantitative measurements, deriving from the analysis of observed spectral interferences as well as from radioactive equilibrium considerations, are reported.
Similar content being viewed by others
References
M. Eisenbud, Environmental Radioactivity, Academic Press, Orlando, Florida, USA, 1987.
R. D. Evans, The Atomic Nucleus, Mc Graw Hill book Co., New York, USA, 1955.
J. N. Rosholt Jr., Anal. Chem., 29 (1954) 1307.
K. H. Lieser, Nuclear and Radiochemistry: Fundamentals and Applications, VCH, Weinheim, Federal Republic of Germany, 1997, p. 41.
Preparation of Gamma-ray Spectrometry Reference Materials RGU-1, RGTh-1 and RGK-1 Report-IAEA/RL/148, Vienna, 1987.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Oddone, M., Giordani, L., Giacobbo, F. et al. Practical considerations regarding high resolution gamma-spectrometry measurements of naturally occurring radioactive samples. J Radioanal Nucl Chem 277, 579–585 (2008). https://doi.org/10.1007/s10967-007-7113-3
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-007-7113-3