Wear predictions for UHMWPE material with various surface properties used on the femoral component in total knee arthroplasty: a computational simulation study

  • Clinical Applications of Biomaterials
  • Original Research
  • Published:
Journal of Materials Science: Materials in Medicine Aims and scope Submit manuscript

Abstract

The wear of ultrahigh-molecular weight polyethylene (UHMWPE) tibial inserts in total knee arthroplasty (TKA) remains a major limitation that hinders the longevity of clinically successful devices. Surface properties significantly affect the overall performance of TKA, and surface modification with mechanically and chemically stable materials is an effective method for overcoming the wear of TKA. However, wear tests are not cost-efficient or time-efficient; thus, the effects of geometric, loading, and alignment perturbations are often evaluated via parametric studies. Computational wear prediction using a finite element (FE) model followed by validation through comparison with experimental data is effective for assessing new prosthetic designs or surface change methods prior to functional testing and surgical implementation. The aim of this study was to evaluate the weight loss, wear depth, and kinematics for different surface properties, including nanostructured diamond (NSD), diamond-like carbon (DLC), titanium-nitride (TiN), and oxidized zirconium (OxZr) on femoral components in TKA using FE analysis under gait-cycle loading conditions. Weight loss and wear depth were lowest with OxZr followed by TiN, NSD, and DLC. However, the DLC femoral component did not show any improvement in wear rate compared to an uncoated cobalt–chromium (Co–Cr) femoral component. Not all surface changes applied in this study did lead to improvement in wear performance. However, this study demonstrates the potential of OxZr and TiN for reducing UHMWPE wear and offers new insights into the effects of wear on TKA.

Graphical Abstract

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

References

  1. Kilgus DJ, Moreland JR, Finerman GA, Funahashi TT, Tipton JS. Catastrophic wear of tibial polyethylene inserts. Clin Orthop Relat Res. 1991;273:223–31.

    Google Scholar 

  2. Amstutz HC, Campbell P, Kossovsky N, Clarke IC. Mechanism and clinical significance of wear debris-induced osteolysis. Clin Orthop Relat Res. 1992;276:7–18.

    Google Scholar 

  3. Kamali A, Farrar R, Hatto P, Stone M, Fisher J. Wear of ultrahigh-molecular-weight polyethylene against titanium-nitride-coated counterfaces. Proc Inst Mech Eng, Part J. 2005;219(1):41–7.

    Article  Google Scholar 

  4. Bowsher J, Shelton J. A hip simulator study of the influence of patient activity level on the wear of crosslinked polyethylene under smooth and roughened femoral conditions. Wear. 2001;250(1):167–79.

    Article  Google Scholar 

  5. Dowson D, Taheri S, Wallbridge N. The role of counterface imperfections in the wear of polyethylene. Wear. 1987;119(3):277–93.

    Article  Google Scholar 

  6. Elfick AP, Smith SL, Green SM, Unsworth A. The quantitative assessment of UHMWPE wear debris produced in hip simulator testing: the influence of head material and roughness, motion and loading. Wear. 2001;249(5):517–27.

    Article  Google Scholar 

  7. Benson LC, DesJardins JD, Harman MK, LaBerge M. Effect of stair descent loading on ultra-high molecular weight polyethylene wear in a force-controlled knee simulator. Proc Inst Mech Eng H. 2002;216(6):409–18.

    Article  Google Scholar 

  8. Benson LC, DesJardins JD, LaBerge M. Effects of in vitro wear of machined and molded UHMWPE tibial inserts on TKR kinematics. J Biomed Mater Res. 2001;58(5):496–504.

    Article  Google Scholar 

  9. Nagashima S, Moon MW, Lee KR. Diamond-like carbon coatings for joint arthroplasty. In: Sonntag R, Kretzer JP, editors. Materials for total joint arthroplasty: Biotribology of potential bearings. London: Imperial college press; 2015. p. 395–412.

  10. Cho H, Wei W, Kao H, Cheng C. Wear behavior of UHMWPE sliding on artificial hip arthroplasty materials. Mater Chem Phys. 2004;88(1):9–16.

    Article  Google Scholar 

  11. Hill MR, Catledge SA, Konovalov V, Clem WC, Chowdhury SA, Etheridge BS, et al. Preliminary tribological evaluation of nanostructured diamond coatings against ultra-high molecular weight polyethylene. J Biomed Mater Res B Appl Biomater. 2008;85(1):140–8.

    Article  Google Scholar 

  12. Mallik AK, Dandapat N, Ghosh P, Ganguly U, Jana S, Das S, et al. Deposition and characterization of diamond-like nanocomposite coatings grown by plasma enhanced chemical vapour deposition over different substrate materials. Bull Mater Sci. 2013;36(2):193–202.

    Article  Google Scholar 

  13. Saikko V, Ahlroos T, Calonius O, Keranen J. Wear simulation of total hip prostheses with polyethylene against CoCr, alumina and diamond-like carbon. Biomaterials. 2001;22(12):1507–14.

    Article  Google Scholar 

  14. Fisher J, McEwen HM, Tipper JL, Galvin AL, Ingram J, Kamali A, et al. Wear, debris, and biologic activity of cross-linked polyethylene in the knee: benefits and potential concerns. Clin Orthop Relat Res. 2004;428:114–9.

    Article  Google Scholar 

  15. Zhao D, Sakoda H, Sawyer WG, Banks SA, Fregly BJ. Predicting knee replacement damage in a simulator machine using a computational model with a consistent wear factor. J Biomech Eng. 2008;130(1):011004

    Article  Google Scholar 

  16. Netter J, Hermida J, Flores-Hernandez C, Steklov N, Kester M, D’Lima DD. Prediction of wear in crosslinked polyethylene unicompartmental knee arthroplasty. Lubricants. 2015;3(2):381–93.

    Article  Google Scholar 

  17. Knight LA, Pal S, Coleman JC, Bronson F, Haider H, Levine DL, et al. Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading. J Biomech. 2007;40(7):1550–8.

    Article  Google Scholar 

  18. Fregly BJ, Sawyer WG, Harman MK, Banks SA. Computational wear prediction of a total knee replacement from in vivo kinematics. J Biomech. 2005;38(2):305–14.

    Article  Google Scholar 

  19. Walker PS, Blunn GW, Broome DR, Perry J, Watkins A, Sathasivam S, et al. A knee simulating machine for performance evaluation of total knee replacements. J Biomech. 1997;30(1):83–9.

    Article  Google Scholar 

  20. ISO 14,243-1 Implants for surgery. Wear of total knee joint prostheses. Loading and displacement parameters for wear-testing machines with load control and corresponding environmental conditions for test. 2002.

  21. Halloran JP, Easley SK, Petrella AJ, Rullkoetter PJ. Comparison of deformable and elastic foundation finite element simulations for predicting knee replacement mechanics. J Biomech Eng. 2005;127(5):813–8.

    Article  Google Scholar 

  22. Kang KT, Park JH, Lee KI, Shim YB, Jang JW, Chun HJ. Gait cycle comparions of cruciate sacrifice for total knee design.-explicit finite element. Int J Precis Eng Man. 2012;13(11):2043–9.

    Article  Google Scholar 

  23. Kwon OR, Kang KT, Son J, Kwon SK, Jo SB, Suh DS, et al. Biomechanical comparison of fixed- and mobile-bearing for unicomparmental knee arthroplasty using finite element analysis. J Orthop Res. 2014;32(2):338–45.

    Article  Google Scholar 

  24. Kwon OR, Kang KT, Son J, Suh DS, Baek C, Koh YG. Importance of joint line preservation in unicompartmental knee arthroplasty: finite element analysis. J Orthop Res. 2017;35(2):347–52.

  25. Godest AC, Beaugonin M, Haug E, Taylor M, Gregson PJ. Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. J Biomech. 2002;35(2):267–75.

    Article  Google Scholar 

  26. Archard JF. Contact and rubbing of flat surfaces. J Appl Phys. 1953;24(8):981–8.

    Article  Google Scholar 

  27. McGloughlin TM, Murphy DM, Kavanagh AG. A machine for the preliminary investigation of design features influencing the wear behaviour of knee prostheses. Proc Inst Mech Eng H. 2004;218(1):51–62.

    Article  Google Scholar 

  28. Pal S, Haider H, Laz PJ, Knight LA, Rullkoetter PJ. Probabilistic computational modeling of total knee replacement wear. Wear. 2008;264(7):701–7.

    Article  Google Scholar 

  29. Sarkar AD. Friction and wear. London: Academic Press; 1980.

  30. Abdelgaied A, Liu F, Brockett C, Jennings L, Fisher J, Jin Z. Computational wear prediction of artificial knee joints based on a new wear law and formulation. J Biomech. 2011;44(6):1108–16.

    Article  Google Scholar 

  31. DesJardins JD, Burnikel B, LaBerge M. UHMWPE wear against roughened oxidized zirconium and CoCr femoral knee components during force-controlled simulation. Wear. 2008;264(3):245–56.

    Article  Google Scholar 

  32. Morrison ML, Jani S, Parikh A. Design of an advanced bearing system for total knee arthroplasty. Lubricants. 2015;3(2):475–92.

    Article  Google Scholar 

  33. Ezzet KA, Hermida JC, Colwell CW Jr., D’Lima DD. Oxidized zirconium femoral components reduce polyethylene wear in a knee wear simulator. Clin Orthop Relat Res. 2004;428:120–4.

    Article  Google Scholar 

  34. McKellop HA, D’Lima D. How have wear testing and joint simulator studies helped to discriminate among materials and designs? J Am Acad Orthop Surg. 2008;16(Suppl 1):S111–9.

    Article  Google Scholar 

  35. Barnett PI, McEwen HM, Auger DD, Stone MH, Ingham E, Fisher J. Investigation of wear of knee prostheses in a new displacement/force-controlled simulator. Proc Inst Mech Eng H. 2002;216(1):51–61.

    Article  Google Scholar 

  36. Johnson T, Laurent M, Yao J, Gilbertson L. The effect of displacement control input parameters on tibiofemoral prosthetic knee wear. Wear. 2001;250(1):222–6.

    Article  Google Scholar 

  37. McEwen HM, Barnett PI, Bell CJ, Farrar R, Auger DD, Stone MH, et al. The influence of design, materials and kinematics on the in vitro wear of total knee replacements. J Biomech. 2005;38(2):357–65.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Yong-Gon Koh.

Ethics declarations

Conflict of interest

The authors declare that they have no competing interests.

Additional information

Kyoung-Tak Kang and Juhyun Son contributed equally to this work.

Oh-Ryong Kwon and Yong-Gon Koh contributed equally to this work.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kang, KT., Son, J., Kim, HJ. et al. Wear predictions for UHMWPE material with various surface properties used on the femoral component in total knee arthroplasty: a computational simulation study. J Mater Sci: Mater Med 28, 105 (2017). https://doi.org/10.1007/s10856-017-5916-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1007/s10856-017-5916-5

Navigation