Abstract
Introduction
Catheter ablation (CA) is an established therapy for atrial fibrillation (AF). The SmartTouch catheter (STc) provides information about catheter tip to tissue contact force (CF). The Surround Flow catheter (SFc) provides a uniform cooling of the tip during ablation. We sought to analyze the impact of STc and SFc on CA of paroxysmal AF in terms of feasibility and acute efficacy.
Methods and results
Sixty-three patients (mean age 57.6 ± 9.8 years, 53 males) with paroxysmal AF underwent pulmonary veins (PVs) antral isolation, by using standard ThermoCool catheter (TCc) in 21, STc in 21, and SFc in 21. Total procedural, fluoroscopy, and radiofrequency (RF) delivery times; percentage of persistently deconnected PVs after 30 min; and percentage of isolated PVs at the end of the procedure were measured.
The use of both STc and SFc obtained a reduction of fluoroscopy time (TCc 34 ± 18 min, STc 20 ± 10 min, p < 0.001; SFc 21 ± 13 min, p = 0.02 vs TCc) and RF time (TCc 41 ± 13 min, STc 30 ± 14 min, p = 0.013; SFc 30 ± 9 min, p < 0.01 vs TCc). The use of STc resulted in a reduction of procedural time (TCc 181 ± 53 min, STc 140 ± 53 min, p < 0.001; SFc 170 ± 51 min, p = NS vs TCc). The percentage of isolated PVs was comparable between groups (TCc 96 % vs STc 98 % vs SFc 96 %; p = NS). The percentage of deconnected PVs at 30 min was lower in TCc (89 %) than in STc (95 %) and in SFc (95 %) group (p < 0.05).
Conclusions
Both STc and SFc allowed a simplification of CA of paroxysmal AF. In addition, they reduced early PVs reconnection.
Condensed abstract
Sixty-three patients with paroxysmal AF underwent ablation by standard ThermoCool, SmartTouch, or Surround Flow catheter. Both the SmartTouch and the Surround Flow significantly reduced radiofrequency and fluoroscopy times, as well as pulmonary veins reconnection rate at 30 min. Moreover, the SmartTouch reduced overall duration of the procedure.
Similar content being viewed by others
References
Calkins, H., Kuck, K. H., Cappato, R., Brugada, J., Camm, A. J., Chen, S. A., Crijns, H. J. G., Damiano, R. J., Davies, D. W., DiMarco, J., Edgerton, J., Ellenbogen, K., Ezekowitz, M. D., Haines, D. E., Haissaguerre, M., Hindricks, G., Iesaka, Y., Jackman, W., Jalife, J., Jais, P., Kalman, J., Keane, D., Kim, Y. H., Kirchhof, P., Klein, G., Kottkamp, H., Kumagai, K., Lindsay, B. D., Mansour, M., Marchlinski, F. E., McCarthy, P. M., Mont, J. L., Morady, F., Nademanee, K., Nakagawa, H., Natale, A., Nattel, S., Packer, D., Pappone, C., Prystowsky, E., Raviele, A., Reddy, V., Ruskin, J. N., Shemin, R. J., Tsao, H. M., & Wilber, D. (2012). 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Europace, 14, 528–606.
Haissaguerre, M., Jais, P., Shah, D. C., Takahashi, A., Hocini, M., Quiniou, G., Garrigue, S., Le Mouroux, A., Le Metayer, P., & Clementy, J. (1998). Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med, 339, 659–666.
Haissaguerre, M., Jais, P., Shah, D. C., Garrigue, S., Takahashi, A., Lavergne, T., Hocini, M., Peng, J. T., Roudaut, R., & Clementy, J. (2000). Electrophysiological end point for catheter ablation of atrial fibrillation initiated from multiple pulmonary venous foci. Circulation, 101, 1409–1417.
Pappone, C., Rosanio, S., Oreto, G., Tocchi, M., Gugliotta, F., Vicedomini, G., Salvati, A., Dicandia, C., Mazzone, P., Santinelli, V., Gulletta, S., & Chierchia, S. (2000). Circumferential radiofrequency ablation of pulmonary vein ostia: a new anatomic approach for curing atrial fibrillation. Circulation, 102, 2619–2628.
Oral, H., Scharf, C., Chugh, A., Hall, B., Cheung, P., Good, E., Veerareddy, S., Pelosi, F., Jr., & Morady, F. (2003). Catheter ablation for paroxysmal atrial fibrillation: segmental pulmonary vein ostial ablation versus left atrial ablation. Circulation, 108, 2355–2360.
Cappato, R., Negroni, S., Pecora, D., Bentivegna, S., Lupo, P. P., Carolei, A., Esposito, C., Furlanello, F., & De Ambroggi, L. (2003). Prospective assessment of late conduction recurrence across radiofrequency lesions producing electrical disconnection at the pulmonary vein ostium in patients with atrial fibrillation. Circulation, 108, 1599–1604.
Nanthakumar, K., Plumb, V. J., Epstein, A. E., Veenhuyzen, G. D., Link, D., & Kay, G. N. (2004). Resumption of electrical conduction in previously isolated pulmonary veins: rationale for a different strategy? Circulation, 109, 1226–1229.
Ouyang, F., Antz, M., Ernst, S., Hachiya, H., Mavrakis, H., Deger, F. T., Schaumann, A., Chun, J., Falk, P., Hennig, D., Liu, X., Bansch, D., & Kuck, K. H. (2005). Recovered pulmonary vein conduction as a dominant factor for recurrent atrial tachyarrhythmias after complete circular isolation of the pulmonary veins: lessons from double Lasso technique. Circulation, 111, 127–135.
Verma, A., Kilicaslan, F., Pisano, E., Marrouche, N. F., Fanelli, R., Brachmann, J., Geunther, J., Potenza, D., Martin, D. O., Cummings, J., Burkhardt, J. D., Saliba, W., Schweikert, R. A., & Natale, A. (2005). Response of atrial fibrillation to pulmonary vein antrum isolation is directly related to resumption and delay of pulmonary vein conduction. Circulation, 112, 627–635.
Thiagalingam, A., D’Avila, A., McPherson, C., Malchano, Z., Ruskin, J., & Reddy, V. Y. (2007). Impedance and temperature monitoring improve the safety of closed-loop irrigated-tip radiofrequency ablation. J Cardiovasc Electrophysiol, 18, 318–325.
Yokoyama, K., Nakagawa, H., Shah, D. C., Lambert, H., Leo, G., Aeby, N., Ikeda, A., Pitha, J. V., Sharma, T., Lazzara, R., & Jackman, W. M. (2008). Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythmia Electrophysiol, 1, 354–362.
Nakagawa, H., Ikeda, A., Govari, A., Ephrath, Y., Ariel, G., Pitha, J. V., Sharma, T., Lazzara, R., & Jackman, W. M. (2009). Electrogram amplitude and impedance are poor predictors of electrode-tissue contact force for radiofrequency ablation (abstract). Heart Rhythm, 6, S12.
Reddy, V. Y., Shah, D., Kautzner, J., Schmidt, B., Saoudi, N., Herrera, C., Jais, P., Hindricks, G., Peichl, P., Yulzari, A., Lambert, H., Neuzil, P., Natale, A., & Kuck, K. H. (2012). The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm, 9, 1789–1795.
Cheema, A., Dong, J., Dalal, D., Marine, J. E., Henrikson, C. A., Spragg, D., Cheng, A., Nazarian, S., Bilchick, K., Sinha, S., Scherr, D., Almasry, I., Halperin, H., Berger, R., & Calkins, H. (2007). Incidence and time course of early recovery of pulmonary vein conduction after catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol, 18, 387–391.
Stabile, G., Scaglione, M., Del Greco, M., De Ponti, R., Bongiorni, M. G., Zoppo, F., Soldati, E., Marazzi, R., Marini, M., Gaita, F., Iuliano, A., & Bertaglia, E. (2012). Reduced fluoroscopy exposure during ablation of atrial fibrillation using a novel electroanatomical navigation system: a multicentre experience. Europace, 14, 60–65.
Nakagawa, H., Yamanashi, W. S., Pitha, J. V., Arruda, M., Wang, X., Ohtomo, K., Beckman, K. J., McClelland, J. H., Lazzara, R., & Jackman, W. M. (1995). Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation, 91, 2264–2273.
Martinek, M., Lemes, C., Sigmund, E., Derndorfer, M., Aichinger, J., Winter, S., Nesser, H. J., & Puererfellner, H. (2012). Clinical impact of a new open-irrigated radiofrequency catheter with direct force measurement on atrial fibrillation ablation. Pacing Clin Electrophysiol, 35, 1312–1318.
Haines, D. E. (1991). Determinants of lesion size during radiofrequency catheter ablation: the role of electrode-tissue contact force and duration of energy delivery. J Cardiovasc Electrophysiol, 2, 509–515.
Perna, F., Heist, E. K., Danik, S. B., Barrett, C. D., Ruskin, J. N., & Mansour, M. (2011). Assessment of catheter tip contact force resulting in cardiac perforation in swine atria using force sensing technology. Circ Arrhythmia Electrophysiol, 4, 218–224.
Di Biase, L., Natale, A., Barrett, C., Tan, C., Elayi, C. S., Ching, C. K., Wang, P., Al-Ahmad, A., Arruda, M., Burkhardt, J. D., Wisnoskey, B. J., Chowdhury, P., De Marco, S., Armaganijan, L., Litwak, K. N., Schweikert, R. A., & Cummings, J. E. (2009). Relationship between catheter forces, lesion characteristics, “popping”, and char formation: experience with robotic navigation system. J Cardiovasc Electrophysiol, 20, 436–440.
Okumura, Y., Johnson, S. B., Bunch, T. J., Henz, B. D., O’Brien, C. J., & Packer, D. L. (2008). A systematical analysis of in vivo contact forces on virtual catheter tip/tissue surface contact during cardiac mapping and intervention. J Cardiovasc Electrophysiol, 19, 632–640.
Shah, D. C., Schmidt, B., Arentz, T., Kuck, K.-H., Neuzil, P., Latcu, G., Hindricks, G., Kautzner, J., Aeby, N., & Lambert, H. (2009). Catheter contact force during human right and left atrial mapping in humans (abstract). Heart Rhythm, 6, S274.
Kuck, K. H., Reddy, V. Y., Schmidt, B., Natale, A., Neuzil, P., Saoudi, N., Kautzner, J., Herrera, C., Hindricks, G., Jais, P., Nakagawa, H., Lambert, H., & Shah, D. (2012). A novel radiofrequency ablation catheter using contact force sensing: Toccata study. Heart Rhythm, 9, 18–23.
Scaglione, M., Blandino, A., Raimondo, C., Caponi, D., Di Donna, P., Toso, E., Ebrille, E., Cesarani, F., Ferrarese, E., & Gaita, F. (2012). Impact of ablation catheter irrigation design on silent cerebral embolism after radiofrequency catheter ablation of atrial fibrillation: results from a pilot study. J Cardiovasc Electrophysiol, 23, 801–805.
Macle, L., Weerasooriya, R., Jais, P., Scavee, C., Raybaud, F., Choi, K.-J., Hocini, M., Clementy, J., & Haissaguerre, M. (2003). Radiation exposure during radiofrequency catheter ablation for atrial fibrillation. Pacing Clin Electrophysiol 26, 26(Pt. II), 288–291.
Scaglione, M., Biasco, L., Caponi, D., Anselmino, M., Negro, A., Di Donna, P., Corleto, A., Montefusco, A., & Gaita, F. (2011). Visualization of multiple catheters with electroanatomical mapping reduces X-ray exposure during atrial fibrillation ablation. Europace, 13, 955–962.
Kerst, G., Weig, H. J., Weretka, S., Seizer, P., Hofbeck, M., Gawaz, M., & Schreieck, J. (2012). Contact force-controlled zero-fluoroscopy catheter ablation of right-sided and left atrial arrhythmia substrates. Heart Rhythm, 9, 709–714.
Acknowledgements
The authors thank Dr. Samuele Guerzoni for his assistance in data collection.
Conflict of Interest
The authors declared no conflicts of interest.
A.N. and S.D. are employees of Biosense Webster Italy
Author information
Authors and Affiliations
Corresponding author
Additional information
Luigi Sciarra and Paolo Golia equally contributed to the manuscript.
Rights and permissions
About this article
Cite this article
Sciarra, L., Golia, P., Natalizia, A. et al. Which is the best catheter to perform atrial fibrillation ablation? A comparison between standard ThermoCool, SmartTouch, and Surround Flow catheters. J Interv Card Electrophysiol 39, 193–200 (2014). https://doi.org/10.1007/s10840-014-9874-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10840-014-9874-2