Comparison Between a 7 French 6 MM Tip Cryothermal Catheter and a 9 French 8 MM Tip Cryothermal Catheter for Cryoablation Treatment of Common Atrial Flutter

  • Published:
Journal of Interventional Cardiac Electrophysiology Aims and scope Submit manuscript

Abstract

Background: Larger tipped cryothermal catheters may deliver efficacy and procedure benefits in the treatment of patients with atrial flutter.

Objective: To compare 7 French 6 mm and 9 French 8 mm tip cryothermal catheters (Freezor®, Xtra or Freezor®, MAX, CryoCath Technologies Inc., Kirkland, Canada) in terms of acute and chronic efficacy, and procedure characteristics in the treatment of atrial flutter.

Methods: This non-randomized clinical investigation determined bi-directional isthmus block at intervention, procedure characteristics, and symptom and conduction recurrence rates post procedure in consecutive patients with symptomatic atrial flutter ablated either with the 7 French 6 mm tip cryothermal catheter (n = 43) at −75C for 4 minutes or the 9 French 8 mm tip cryothermal catheter (n = 51) at −75C for 8 minutes.

Results: Clinical data showed a higher acute success rate for the larger tipped catheter (100% vs. 88%). Symptom recurrence rates were 0% for both catheters at 3, 6, and 9 month follow-up. Conduction recurrence rates were similar for both catheters on repeat electrophysiological study at 3 months post procedure (35% vs. 32%). Procedure benefits were fewer cryotests (20 ± 17 vs. 26 ± 21) and ablations (4 ± 4 vs. 12 ± 18), and shorter procedure (80 ± 61 min vs. 87 ± 48 min), fluoroscopy (14 ± 9 min vs. 24 ± 10 min), and cryoapplication times (37 ± 18 min vs. 44 ± 23 min) with the larger tipped catheter. There were no adverse events reported.

Conclusions: Clinical results showed differences in catheter performance that favoured the larger tipped catheter. However, increased acute success rate did not translate into reduced conduction recurrence rate post ablation, although clinical recurrence was completely absent long-term.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Langberg JJ, Lee MA, Chin MC, Rosenqvist M. Radiofrequency catheter ablation: The effect of electrode size on lesion volume in vivo. Pacing Clin Electrophysiol 1990;13(10):1242–1248.

    CAS  PubMed  Google Scholar 

  2. Chan RC, Johnson SB, Seward JB, Packer DL. The effect of ablation electrode length and catheter tip to endocardial orientation on radiofrequency lesion size in the canine right atrium. Pacing Clin Electrophysiol 2002;25(1):4–13.

    Article  Google Scholar 

  3. Feld G, Wharton M, Plumb V, Daoud E, Friehling T, Epstein L. Radiofrequency catheter ablation of type 1 atrial flutter using large-tip 8- or 10-mm electrode catheters and a high-output radiofrequency energy generator: Results of a multicenter safety and efficacy study. J Am Coll Cardiol 2004;43(8):1466–1472.

    Article  PubMed  Google Scholar 

  4. Marrouche NF, Schweikert R, Saliba W, Pavia SV, Martin DO, Dresing T, Cole C, Balaban K, Saad E, Perez-Lugones A, Bash D, Tchou P, Natale A. Use of different catheter ablation technologies for treatment of typical atrial flutter: Acute results and long-term follow-up. Pacing Clin Electrophysiol 2003;26(3):743–746.

    Article  PubMed  Google Scholar 

  5. Ventura R, Willems S, Weiss C, Flecke J, Risius T, Rostock T, Hoffmann M, Meinertz T. Large tip electrodes for successful elimination of atrial flutter resistant to conventional catheter ablation. J Interv Card Electrophysiol 2003;8(2):149–154.

    Article  PubMed  Google Scholar 

  6. Montenero AS, Bruno N, Antonelli A, et al. Safety and efficacy of catheter cryoablation for the treatment of atrial flutter: Acute and long term results. Ital Heart J 2004;5(Suppl. 1):131S–137S

    Google Scholar 

  7. Gaita F, Haissaguerre M, Giustetto C, Grossi S, Caruzzo E, Bianchi F, Richiardi E, Riccardi R, Hocini M, Jais P. Safety and efficacy of cryoablation of accessory pathways adjacent to the normal conduction system. J Cardiovasc Electrophysiol 2003;14(8):825–829.

    Article  PubMed  Google Scholar 

  8. Manusama R, Timmermans C, Limon F, Philippens S, Crijns HJGM, Rodriguez L-M. Catheter-based cryoablation permanently cures patients with common atrial flutter. Circulation 2004;109(13):1636–1639.

    Article  PubMed  Google Scholar 

  9. Riccardi R, Gaita F, Caponi D, Grossi S, Scaglione M, Caruzzo E, Di Donna P, Pistis G, Richiardi E, Giustetto C, Bocchiardo M. Percutaneous catheter cryothermal ablation of atrioventricular nodal reentrant tachycardia: Efficacy and safety of a new ablation technique. Ital Heart J 2003;4(1):35–43.

    PubMed  Google Scholar 

  10. Skanes AC, Dubuc M, Klein GJ, Thibault B, Krahn AD, Yee R, Roy D, Guerra P, Talajic M. Cryothermal ablation of the slow pathway for the elimination of atrioventricular nodal reentrant tachycardia. Circulation 2000;102(23):2856–2860.

    CAS  PubMed  Google Scholar 

  11. Lanzotti ME, De Ponti R, Tritto M, Spadacini G, Salerno-Uriarte JA. Successful treatment of anteroseptal accessory pathways by transvenous cryomapping and cryoablation. Ital Heart J 2002;3(2):128–132.

    PubMed  Google Scholar 

  12. Dubuc M, Talajic M, Roy D, Thibault B, Leung TK, Friedman PL. Feasibility of cardiac cryoablation using a transvenous steerable electrode catheter. J Interv Card Electrophysiol 1998;2(3):285–292.

    Article  CAS  PubMed  Google Scholar 

  13. Montenero AS, Bruno N, Zumbo F, et al. Safety and efficacy of a 7FR, 6 mm tip cryoablation catheter for the treatment of atrial flutter in 98 consecutive patients. Europace 2003;4(Suppl A):A4.

    Google Scholar 

  14. Timmermans C, Ayers GM, Crijns HJ, Rodriguez LM. Randomized study comparing radiofrequency ablation with cryoablation for the treatment of atrial flutter with emphasis on pain perception. Circulation 2003;107(9):1250–1252.

    Article  PubMed  Google Scholar 

  15. Timmermans C, Rodriguez LM, Van Suylen RJ, et al. Catheter-based cryoablation produces permanent bidirectional cavotricuspid isthmus conduction block in dogs. J Interv Card Electrophysiol 2002;7(2):149–155.

    Article  PubMed  Google Scholar 

  16. Skanes AC, Yee R, Krahn AD, Klein GJ. Cryoablation of atrial arrhythmias. Card Electrophysiol Rev 2002;6(4):383–388.

    Article  PubMed  Google Scholar 

  17. Cauchemez B, Haissaguerre M, Fischer B, Thomas O, Clementy J, Coumel P. Electrophysiological effects of catheter ablation of inferior vena cava-tricuspid annulus isthmus in common atrial flutter. Circulation 1996;93(2):284–294.

    CAS  PubMed  Google Scholar 

  18. Lowe MD, Meara M, Mason J, Grace AA, Murgatroyd FD. Catheter cryoablation of supraventricular arrhythmias: A painless alternative to radiofrequency energy. Pacing Clin Electrophysiol 2003;26(1 Pt 2):500–503.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Annibale S. Montenero MD.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Montenero, A.S., Bruno, N., Antonelli, A. et al. Comparison Between a 7 French 6 MM Tip Cryothermal Catheter and a 9 French 8 MM Tip Cryothermal Catheter for Cryoablation Treatment of Common Atrial Flutter. J Interv Card Electrophysiol 13, 59–69 (2005). https://doi.org/10.1007/s10840-005-0353-7

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10840-005-0353-7

Navigation