Introduction

Precision requirements regarding power calculation for intraocular lens (IOL) are growing. With new multifocal IOLs on board and refractive or presbyopic lens exchange becoming more popular, there is no room for inaccurate measurements. Addressing posterior corneal astigmatism has become an important issue when dealing with toric monofocal or toric multifocal IOLs [1,2,3,4]. The posterior corneal curvature can be assessed with optical coherence tomography (OCT) systems (Casia SS-1000, Tomey Corp.) and Scheimpflug camera-based devices such as Pentacam HR® (Oculus Optikgerate GmbH), Galilei® (Ziemer Ophthalmic Systems AG) and Sirius® (Schwind). Some of these devices, used primarily for 3-dimensional anterior segment analysis, have been recently redesigned to feature an optical biometry function. The new Galilei G6® and Pentacam AXL® can now also be used also as biometers for IOL power calculations [5, 6].

There are few studies comparing these new devices to the “gold standard” optical biometer, the IOL Master 500 [7,8,9]. Recent study retrospectively compared the Galilei G6 with the IOL Master 500. The devices provided similar mean IOL power, axial length (AL), keratometric and anterior chamber depth (ACD) measurements, but the authors suggested they should not be used interchangeably for biometric measurements and IOL power calculations [7]. Studies comparing the Pentacam AXL to other optical biometers found inconsistent results. Shajari compared the Pentacam AXL with the IOL Master 500 and the IOL Master 700 and found no statistically significant differences in AL, keratometry and ACD measurements performed with the three devices [8]. Sel comparing the Pentacam AXL with the IOL Master 700 found significant differences for keratometry values, AL and ACD measurements [9].

The Pentacam AXL is a single rotating Scheimpflug camera device combined with optical biometry based on partial coherence interferometry (PCI) [5]. The repeatability of measurements obtained with the Pentacam HR and the Pentacam AXL was confirmed [8, 10, 11]. However, the question arises, if the Pentacam AXL can be used as a solo optical biometer in an ophthalmic setting.

The aim of our study was to compare the measurement results for AL, keratometry, ACD and white-to-white (WTW) obtained with the Pentacam AXL and the IOL Master 500.

Patients and methods

This cross-sectional study enrolled 97 eyes of 97 consecutive patients who came for preoperative measurements before cataract surgery or refractive lens exchange in a private hospital (2500 cases/year). The study was conducted in compliance with the principles of the Declaration of Helsinki and approved by local ethics committee of Lower Silesia Medical Council. Written informed consent was obtained from all study participants.

Patients with corneal or retinal diseases, including age-related macular degeneration, and previous ocular trauma were excluded. Only patients with good fixation were eligible. For each subject, we assessed only one eye, either the one with the more advanced cataract, planned to be operated first, or the phakic eye if the other was already pseudophakic. The subjects were examined prospectively on two devices: IOL Master 500 (software version 5.02.0590) and Pentacam AXL (software version 1.20r111). IOL Master 500 was considered a reference biometer. The measurements for individual subjects on the two devices were taken no more than 15 min apart to avoid the effects of diurnal variations in corneal indices. The IOL Master exam was always performed as first. Exams were included only if positively evaluated by automatic quality checks on both devices.

The Pentacam AXL is a rotating Scheimpflug camera combined with optical biometry based on PCI. Its light source is a blue light-emitting diode (LED) with a wavelength of 475 nm. The rotational measuring procedure generates Scheimpflug images in three dimensions, with the dot matrix fine-meshed in the centre due to the rotation. It takes a maximum of 2 s to generate a complete image of the anterior eye segment. Any eye movement is detected by a second camera and corrected for in the process. The entire cornea is analysed in multiple ways. The simulated K (SimK) is derived from images taken exclusively from the anterior corneal surface, over a 3-mm ring [5].

IOL Master 500 is a PCI optical biometer, with a 780-nm laser diode infrared light. It measures the ACD, which corresponds to the distance between the corneal epithelium and the anterior lens surface, through a lateral slit illumination. For keratometry, it uses a hexagonal pattern of six points in a 2.3-mm-diameter zone.

Both devices use a keratometric index of 1.3375 to convert anterior corneal curvature measurements in millimetres to corneal power in dioptres (D). Pentacam AXL calculates two ACDs: internal (from corneal endothelium to the anterior lens surface) and external (from corneal epithelium to the anterior lens surface). For the purpose of this study, only external ACD was included.

The data from the two devices taken for analysis included: axial length (AL), corneal curvature in steep meridian (Ks), corneal curvature in flat meridian (Kf), mean corneal curvature (Km), anterior chamber depth (ACD) and white-to-white distance (WTW).

For all eyes, IOL power (SN60WF, Alcon®) targeting emmetropia was calculated with Haigis and SRK/T formulas using the IOL Master measurements with the User Group for Laser Interference Biometry (ULIB) constants for IOL Master 500 [12,13,14]. The ULIB constants for Pentacam AXL have been derived from constants for IOL Master. The IOL power indicating an absolute value of target refraction closest to zero was used for analysis.

Statistical analysis

Statistical analyses were performed using MATLAB R2009b software (MathWorks, Inc.). We calculated the sample size needed to detect an AL difference of 0.02 mm between two devices. Fifty-seven eyes were required for a significance level (α) of 0.05 and a test power of 0.80. Normality of data was assessed by the Kolmogorov–Smirnov test. All measurement data exhibited normal distribution. A paired sample t test was performed to evaluate statistical significance of differences between readings from the two devices. A difference was deemed statistically significant if its P value was less than 0.05. Intraclass correlation coefficient (ICC) was calculated to assess agreement between measurements [15]. ICC values higher than 0.900 were considered as having a high degree of agreement. Additionally, Bland–Altman plots were used to evaluate the agreement between devices. The mean difference and 95% limits of agreement (LoA) were calculated. The plots were created with Origin 8.0 software (OriginLab, Co.).

Results

Out of 97 enrolled eyes, 10 eyes of 10 patients were excluded because of persistent errors in repeated AL measurements with the Pentacam AXL, due to dense crystalline lens opacity (seven eyes) or high axial myopia (three eyes), although the respective AL measurement presented no difficulties with the IOL Master 500. The final analysis included 87 eyes of 87 subjects. Subject mean age was 58 ± 15 (SD) (range 27–82 years), 51 were female. The mean axial length was 23.29 ± 1.82 mm (range 19.34–28.49 mm), as measured with IOL Master 500.

Table 1 shows mean values and differences with 95% confidence interval (CI), LoA, ICC and P values for all measured parameters, comparing between IOL Master and Pentacam AXL. The paired sample t test showed statistically significant differences between IOL Master 500 and Pentacam AXL measurements for AL, Kf, Ks, Km and WTW (P < 0.05), while no significant difference was found for ACD measurements (P = 0.36). The ICC values between the two devices were high and exceeding 0.900 for all analysed parameters, except WTW (ICC = 0.431). The Bland–Altman plots for AL, Kf, Ks, Km and WTW are shown in Figs. 1, 2, 3, 4, 5 and 6. K values are reported in mm.

Table 1 Comparison of measurements obtained from IOL Master 500 and Pentacam AXL
Fig. 1
figure 1

Bland–Altman plots showing the agreement between IOL Master 500 and Pentacam AXL in AL measurements. The bold lines represent the mean difference between two devices. The dashed lines represent 95% limits of agreement (LoA). The percentage of points within 95% LoA is 93

Fig. 2
figure 2

Bland–Altman plots showing the agreement between IOL Master 500 and Pentacam AXL in Kf measurements. The bold lines represent the mean difference between two devices. The dashed lines represent 95% limits of agreement (LoA). The percentage of points within 95% LoA is 96

Fig. 3
figure 3

Bland–Altman plots showing the agreement between IOL Master 500 and Pentacam AXL in Ks measurements. The bold lines represent the mean difference between two devices. The dashed lines represent 95% limits of agreement (LoA). The percentage of points within 95% LoA is 94

Fig. 4
figure 4

Bland–Altman plots showing the agreement between IOL Master 500 and Pentacam AXL in Km measurements. The bold lines represent the mean difference between two devices. The dashed lines represent 95% limits of agreement (LoA). The percentage of points within 95% LoA is 96

Fig. 5
figure 5

Bland–Altman plots showing the agreement between IOL Master 500 and Pentacam AXL in ACD measurements. The bold lines represent the mean difference between two devices. The dashed lines represent 95% limits of agreement (LoA). The percentage of points within 95% LoA is 93

Fig. 6
figure 6

Bland–Altman plots showing the agreement between IOL Master 500 and Pentacam AXL in WTW measurements. The bold lines represent the mean difference between two devices. The dashed lines represent 95% limits of agreement (LoA). The percentage of points within 95% LoA is 95

The Pentacam AXL yielded lower values for AL and WTW as well as flatter corneal curvatures in flat and step meridians. The mean difference between devices in AL measurements was 0.01 mm (95% CI 0.01–0.02). The difference between mean curvature radii reached significance (P < 0.05) with Pentacam AXL results being higher by a mean difference of 0.05 mm and 0.06 mm for the flat and the steep meridian, respectively. WTW distance measured with Pentacam AXL was significantly shorter then with IOL Master 500 (11.5 ± 0.4 mm and 11.9 ± 0.5 mm, respectively). ACD was the only parameter that showed no statistically significant difference between devices, with mean value of − 0.01 ± 0.13 (P = 0.36) and good agreement (ICC = 0.945).

Table 2 shows the mean IOL powers (SN60WF, Alcon®) calculated on both devices, aiming for “zero” postoperative refraction. Ten eyes were excluded from the analysis because their calculated IOL power exceeded the commercially available range. There was a statistically significant difference between IOL power selected by the two devices, with Pentacam AXL showing higher IOL power in 75% of eyes with the Haigis formula and in 62% of eyes with the SRK/T formula (Fig. 7). The differences were within ± 0.5 D in 71.6 and 86.4% of eyes with Haigis and SRK/T formula, respectively (Table 2).

Table 2 Comparison of Haigis and SRK/T IOL power calculations (SN60WF, Alcon®) with IOL Master 500 and Pentacam AXL data
Fig. 7
figure 7

Distribution of IOL power differences between IOL Master 500 and Pentacam AXL for two calculation formulas (Haigis and SRK/T) to reach emmetropia

The distribution of IOL powers differences between devices, as calculated with two formulas (Haigis and SRK/T) to achieve emmetropia, is shown in Fig. 7.

Discussion

We planned the present study to establish if we can rely on the Pentacam AXL as a single optical biometer in our setting (private hospital). It seems economically questionable as well as time-consuming to use two optical biometers to examine every patient, while the option of measuring the posterior corneal curvature is a promising feature of an optical biometer. We therefore compared the new Pentacam AXL with the existing IOL Master 500, performing examinations on consecutive patients presenting for cataract surgery or refractive lens exchange.

We found that axial length measurements obtained with the IOL Master 500 were slightly but significantly higher (0.01 mm) than those obtained with the Pentacam AXL. This is close to the results of two recent studies, which report that the Pentacam AXL returns readings that, respectively, differ by a mean of 0.05 mm relative to the IOL Master 700 [9] and 0.026 mm relative to the IOL Master 500 [8]. Such a small difference should not influence IOL power calculation.

Both devices use the same proven PCI biometry. Nevertheless, we were not able to measure AL with the Pentacam AXL for 10 eyes of 10 patients out of the total 97, even though these eyes were successfully measured with the with IOL Master 500, with a wide range of AL (19.34–28.49 mm). The eyes in question featured advanced cataract or high axial myopia. New swept source OCT biometers are expected to have higher success rates in measuring AL in cases of dense and posterior subcapsular cataracts [16].

The agreement between Pentacam AXL and IOL Master 500 in ACD measurements was very good, with a statistically nonsignificant difference of − 0.01 mm (95% CI − 0.04–0.02). An older study (2009) reported that average IOL Master ACD measurement results were 0.11 mm smaller than those obtained with the Pentacam. Still, it was not enough to produce clinically noticeable differences in refractive outcomes of cataract surgery [17]. It was also shown that for ACD values, the Pentacam had the smallest standard deviation when compared to IOL Master and A-scan [18].

In our study, the mean difference in corneal curvature measurements between the IOL Master 500 and the Pentacam AXL was largest for steep meridian (− 0.06 mm, equal to 0.33 D for a standardised keratometric refraction index of 1.3375). Another group recently found a difference of − 0.033 mm (equal to 0.19 D), also regarding the steep meridian [8]. Studies based on the former version of Pentacam or Pentacam HR also indicated differences in keratometric measurements [19,20,21,22,23,24]. Karunaratne found statistically significant differences between the mean keratometric values of the IOL Master and Pentacam equivalent keratometry readings at 2-, 3- and 4.5-mm zone measurements. The authors concluded that after constant optimisation, keratometry from the IOL Master and the Pentacam equivalent 4.5-mm zone keratometry reading may be similarly effective when used in IOL power calculation formulas [22].

The Pentacam AXL measures posterior curvature of the cornea and corneal thickness, yielding total corneal power and true net power information. We provide no comparisons of the devices in this respect, as it is well known, that the cited parameters and anterior curvature readings are not comparable [21, 25] and cannot be used for most of the present IOL calculation formulas [26].

For calculation of IOL toricity, the Pentacam AXL offers meridional analysis method, which uses total corneal power in a 3-mm zone [27, 28]. For the spherical equivalent power, anterior corneal curvature values are applied. We did not analyse calculations for toric IOLs in our subjects. It is worth mentioning that the Pentacam’s repeatability of corneal curvature readings was found to be better than that of the IOL Master 500 or IOL Master 700 [8].

It was interesting to find that WTW measurements were significantly lower on the Pentacam AXL in comparison to the IOL Master 500 with poor agreement (ICC = 0.470) between the devices. A statistically significantly lower WTW distance compared to the IOL Master was also found with the Lenstar® (Haag Streit, USA) in a meta-analysis published in 2017 [29]. Another study reported a lower average WTW value on a Scheimpflug/Placido topography system as compared to the IOL Master [30]. Moreover, the IOL Master constantly yielded values exceeding measurements based on direct eye visualisation [31]. These differences must be considered in IOL formulas incorporating WTW data and in calculating phakic IOLs.

Since the ACD measurements were not statistically different and differences in AL were clinically insignificant, the tendency of the Pentacam AXL to provide higher calculated IOL power in our cohort can likely be explained by its flatter keratometric measurements. Compared to the IOL Master, the Pentacam AXL indicated higher IOL power, according to Haigis and SRK/T, for 75 and 62% of eyes, respectively. The ULIB SN60WF constants for the Pentacam are the same as for IOL Master 500 (A SRKT = 119; a0 = − 0.769; a1 = 0.234; a2 = 0.217). Our results showed that these constants are obviously not optimal for the Pentacam AXL, and all users should start with a period of “sham” measurements, which will form a basis for the needed corrections. Our cohort was not suitable for constants optimisation, because in reality we implanted our subjects with different types of IOLs. Nevertheless, our theoretical comparison indicates that optimisation of constants for the Pentacam AXL is a must. This is a time-consuming process, and new Pentacam AXL users should be aware of this burden.

Conclusions

The measurements results required for IOL power calculation performed on the compared devices were found to be significantly different. Pentacam AXL and IOL Master 500 should not be used interchangeably. Due to its constantly flatter corneal curvature measurements, the Pentacam AXL requires optimisation of constants used in formulas.