1 Introduction

Black coatings featuring a high emissivity corresponding to a low reflectivity from the visible to the far-infrared spectral range are widely used in optical instruments to reduce stray light, in control panels or automotive instrumentation to enhance contrast and readability, in thermal applications to control and maximize radiative cooling or heating, in spectroscopy as reference surfaces and radiance standards. For these applications, reliable data of their optical properties are mandatory. However, emissivity data of black coatings published in the literature or in standardization documents often do not provide angularly and/or spectrally resolved information and/or lack uncertainties. The Physikalisch-Technische Bundesanstalt (PTB) determined the directional spectral and directional total emissivities under several angles of observation and also the hemispherical total emissivities of several widely used diffuse black coatings: Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black.

2 Measurements Scheme

The measurements were taken under vacuum (at a pressure of \(10^{-6}\) hPa) at the reduced background calibration facility (RBCF) [1, 2] of PTB to achieve low uncertainties and avoid atmospheric interferences. Furthermore, some of the measurements are compared with measurements obtained with the emissivity measurement in air facility (EMAF) [3], which is routinely operated at PTB for several years. The EMAF successfully took part in an international emissivity comparison organized by the consultative committee for thermometry (CCT) [4].

The measurement scheme for the determination of the emissivity is a comparison of the spectral radiance of the sample which is mounted on a heater and placed inside a temperature-stabilized spherical enclosure against the spectral radiances of two reference blackbodies at different temperatures [1,2,3]. The comparison is performed with an infrared Fourier transform spectrometer (IR-FTS). By using two reference blackbodies at different temperatures, the residual background radiation of the instrumentation can be corrected. The “main” reference blackbody is usually operated at a temperature close to the radiance temperature of the sample to achieve a similar signal level. The second reference blackbody is operated at a temperature significantly lower than the temperature of the main blackbody. At the RBCF, this is an liquid nitrogen-cooled (\(\hbox {LN}_{2}\)) blackbody, which is operated at the temperature of boiling nitrogen. At the EMAF, the second reference blackbody is a liquid-cooled blackbody operated at \(18~{^{\circ }}\)C.

3 Measurement Uncertainty

The calculation of the emissivity from the measured detector signals is based on the complete radiation budget, considering every radiation contribution in the optical path. The uncertainty of the directional spectral emissivity is calculated via a Monte–Carlo method [5, 6] based on the final determination equation of the directional spectral emissivity and is spectrally dependent. Furthermore, the uncertainty budget is calculated for each individual measurement, since the relevance of the individual uncertainty contributions depends significantly on the conditions of the measurement and the sample investigated. A detailed description of the data evaluation procedure and the uncertainty calculation is given in [6, 7].

4 Experimental Results on Different Black Coatings

The emissivity measurements of the black coatings were taken at two setups at PTB—under vacuum at the RBCF and in air at the EMAF—by applying the measurement schemes and data, and uncertainty evaluation described shortly above and described in detail in [1,2,3, 6, 7]. Two sets of detectors and beamsplitters were used with the IR-FTS according to the wavelength and temperature range of the measurements. In the range from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\), a deuterated l-alanine-doped triglycine sulfate (DLaTGS) or liquid nitrogen-cooled mercury cadmium telluride (MCT) detector in combination with a potassium bromide (KBr) beamsplitter were used. Due to its higher responsivity, the MCT was used for the measurements at lower sample temperatures. In the range from \(14~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\), the emissivity was determined by using a far-infrared deuterated triglycine sulfate (FDTGS) detector in combination with a \(6~\upmu \hbox {m}\) Multilayer Mylar beamsplitter.

Each of the investigated black coatings—Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black—was applied to copper substrates. The disk-shaped substrates have a diameter of 90 mm and a thickness of 5 mm and feature a hole of 2 mm diameter and 50 mm depth from the circumference to accommodate a platinum resistance thermometer (PRT) temperature sensor for monitoring the substrate temperature. The surface of the copper substrates was plane and untreated.

In the following, the individual emissivity results for the four different coatings are presented.

4.1 Nextel 811-21

Nextel 811-21 was in the past provided by 3M and is now available from “Mankiewicz Gebr. & Co” [8] where we obtained it. The substrates were coated according to the manufactures instructions, i.e., spray coating of the primer and spray coating of the Nextel 811-21 coating and subsequent heating at \(120~{^{\circ }}\)C. Some of the emittance properties of Nextel 811-21 are already provided in the literature [9, 10]. Here, the spectral emissivity of the sample Nextel 811-21 was determined under vacuum in the wavelength range from \(5~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at a temperature of \(120~{^{\circ }}\)C under different angles of observation. The resulting spectral curves are plotted in Fig. 1. Some of the measured angles are omitted here for clarity of the figure. The spectrally integrated directional total emissivities are shown in the inset of Fig. 1 with their corresponding standard uncertainties. A theoretical model based on Fresnel equations is fitted to these values and plotted as a solid line. Its integration over the angular range from \(0{^{\circ }}\) to \(90{^{\circ }}\) yields the hemispherical total emissivity. A decrease in the directional spectral emissivities and accordingly of the directional total emissivities toward larger angles is clearly visible, which is typical for high-emitting samples. The numerical values of the directional total and hemispherical total emissivities of Nextel 811-21 in the wavelength range from \(5~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at temperature of \(25~{^{\circ }}\)C and from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at temperature of \(120~{^{\circ }}\)C are provided in Table 1. Within the uncertainty of the measurements, no significant spectral change in emissivity can be observed between \(25~{^{\circ }}\)C and \(120~{^{\circ }}\)C.

Fig. 1
figure 1

Directional spectral emissivity of Nextel 811-21 measured at a temperature of \(120~{^{\circ }}\)C for different angles of observation. The spectrally integrated values of the directional total emissivity with their standard uncertainties as well as the hemispherical total emissivity are shown in the inset

Table 1 Directional total and hemispherical total emissivities of Nextel 811-21 in the wavelength range from \(5~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(25~{^{\circ }}\)C and from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(120~{^{\circ }}\)C with their corresponding standard uncertainties

The directional spectral emissivity of Nextel 811-21 from \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) at a temperature of \(120~{^{\circ }}\)C observed under an angle of \(10{^{\circ }}\) is plotted in Fig. 2. The shaded area around the curve is the range of the standard uncertainty of the measurement. The blue curve in the lower part of the graph and the corresponding right-hand ordinate axis show the spectral distribution of the uncertainty. The numerical values of the directional total and hemispherical total emissivities of Nextel 811-21 in the wavelength range from \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) are provided in Table 2.

Fig. 2
figure 2

Directional spectral emissivity of Nextel 811-21 up to a wavelength of \(100~\upmu \hbox {m}\) measured at a temperature of \(120~{^{\circ }}\)C and under an angle of observation of \(10{^{\circ }}\) with respect to the surface normal. The spectral distribution of the standard uncertainty is shown in the lower part of the plot with the respective scale on the right-hand ordinate axis

Table 2 Directional total and hemispherical total emissivities of Nextel 811-21 in the wavelength range from \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) with their corresponding standard uncertainties
Fig. 3
figure 3

Directional spectral emissivity of Herberts 1534 in the wavelength range from \(5.7~\upmu \hbox {m}\) to \(18.2~\upmu \hbox {m}\) measured at a temperature of \(120~{^{\circ }}\)C in air and at a temperature of \(25~{^{\circ }}\)C under vacuum. Both measurements were taken under an angle of observation of \(10{^{\circ }}\) with respect to the surface normal. The spectral distribution of the standard uncertainty of measurement at \(25~{^{\circ }}\)C under vacuum is shown in the lower half of the plot with the respective scale on the right-hand ordinate axis

4.2 Herberts 1534

Herberts 1534 is no longer in production but was used in the past for the far-infrared wavelength range (FIR) and space applications. There are still a number of instrumentations and blackbodies in use which are coated with this paint. Here, the spectral emissivity of the sample Herberts 1534 was measured under vacuum at a temperature of \(25~{^{\circ }}\)C and in air at a temperature of \(120~{^{\circ }}\)C and is plotted in Fig. 3. The brown curve, with its range of standard uncertainty given as a shaded area, shows the directional spectral emissivity under vacuum. Additionally, the uncertainty is also shown separately with the blue curve in the lower part of Fig. 3 and the corresponding right-hand ordinate axis. The green curve in Fig. 3 shows the directional spectral emissivity measured in air at a temperature of \(120~{^{\circ }}\)C. The directional spectral emissivity of Herberts 1534 measured in the wavelength range from \(14.5~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) and, for comparison, the results obtained in the mid-infrared wavelength (MIR) range are given in Fig. 4. The results of both measurements in the overlapping spectral range from \(14.5~\upmu \hbox {m}\) to \(18~\upmu \hbox {m}\) are consistent within the range of their standard uncertainties. The numerical values of the directional total and hemispherical total emissivities are presented in Tables 3 and 4.

Fig. 4
figure 4

Directional spectral emissivity of Herberts 1534 measured under vacuum up to a wavelength of \(100~\upmu \hbox {m}\) and compared in the overlapping wavelength range with measurements obtained under vacuum and in air in the MIR range

Table 3 Directional total and hemispherical total emissivities of Herberts 1534 in the wavelength range from \(5~\upmu \hbox {m}\) to \(18.2~\upmu \hbox {m}\) at \(25~{^{\circ }}\)C and from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(120~{^{\circ }}\)C with their corresponding standard uncertainties
Table 4 Directional total and hemispherical total emissivities of Herberts 1534 in the wavelength range from \(14.1~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) with their corresponding standard uncertainties
Fig. 5
figure 5

Directional spectral emissivities of three Aeroglaze Z306 samples with thicknesses of \(44~\upmu \hbox {m}\), \(99~\upmu \hbox {m}\) and \(236~\upmu \hbox {m}\) on polished copper substrates under an angle of observation of \(10{^{\circ }}\)

Fig. 6
figure 6

Directional spectral emissivities of Aeroglaze Z306 sample with thickness of \(99~\upmu \hbox {m}\) under an angle of observation of \(10{^{\circ }}\) at a temperatures of \(25 {^{\circ }}\)C and \(150 {^{\circ }}\)C

Table 5 Directional total and hemispherical total emissivities of Aeroglaze Z306 in the wavelength range from \(5~\upmu \hbox {m}\) to \(16.6~\upmu \hbox {m}\) at \(25~{^{\circ }}\)C and from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(150~{^{\circ }}\)C with their corresponding standard uncertainties
Table 6 Directional total and hemispherical total emissivities of Aeroglaze Z306 at three different thicknesses. All measurements were taken at a temperature of \(150~{^{\circ }}\)C and in the wavelength range from \(14.7~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\)
Fig. 7
figure 7

Directional spectral emissivity of Acktar Fractal Black from \(5~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) measured at a temperature of \(120~{^{\circ }}\)C and under an angle of observation of \(10{^{\circ }}\) with respect to the surface normal. The spectral distribution of the standard uncertainty is shown in the lower half of the plot with the respective scale on the right-hand ordinate axis

Fig. 8
figure 8

Directional spectral emissivity of Acktar Fractal Black under different angles of observation with respect to the surface normal

4.3 Aeroglaze Z306

The Aeroglaze Z306 is an absorptive polyurethane coating which is often used in aerospace applications and which is well suited for vacuum conditions. Some of its emittance properties can be found in [11, 12]. In preparation of the measurements of Aeroglaze Z306, three copper substrates were spray coated with Aeroglaze according to the instructions given in the European Cooperation for Space Standardization document ECSS-Q-70-25A [13] with thicknesses of \(44~\upmu \hbox {m}\), \(99~\upmu \hbox {m}\) and \(236~\upmu \hbox {m}\). The measurements were taken at a temperature of \(150~{^{\circ }}\)C and are shown in Fig. 5 for an angle of observation of \(10{^{\circ }}\). A significant decrease in emissivity above \(22~\upmu \hbox {m}\) can be seen for the sample with a coating thickness of \(44~\upmu \hbox {m}\). The sample with the coating thickness of \(236~\upmu \hbox {m}\) shows a nearly constant (spectrally flat) emissivity. The clearly visible modulation in the emissivity of all three samples for longer wavelengths is explained by an onset of transparency of the coating toward longer wavelengths and multiple beam interference within the coating due to the highly reflective copper substrate. The observed modulation period is inversely proportional to the optical thickness of the coating.

The Aeroglaze Z306 coating of all three thicknesses is fully opaque in the MIR range, but the directional spectral emissivity shows a slight temperature dependence as shown in Fig. 6. The directional spectral emissivity at \(25~{^{\circ }}\)C is shown as the brown curve with the corresponding uncertainty range plotted as the shaded area around the curve. The green curve in Fig. 6 illustrates the directional spectral emissivity at \(150~{^{\circ }}\)C.

Tables 5 and 6 provide the numerical values of the directional total and hemispherical total emissivities of Aeroglaze Z306.

Table 7 Directional total and hemispherical total emissivities of Acktar Fractal Black in the wavelength range from \(5.7~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(25~{^{\circ }}\)C and from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(120~{^{\circ }}\)C with their corresponding standard uncertainties
Table 8 Directional total and hemispherical total emissivities of Acktar Fractal Black with their corresponding standard uncertainties. The measurements were taken at a temperature of \(120~{^{\circ }}\)C and in the wavelength range from \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\)

4.4 Acktar Fractal Black

Acktar Fractal Black is a highly emitting coating of the company “Acktar” [14]. Measurements of its reflectivity and other characteristics of this coating can be found in [14, 15], but direct emissivity measurements are missing. Here, the directional spectral emissivity of Acktar Fractal Black was measured under vacuum from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(120~{^{\circ }}\)C. The measurement under an angle of observation of \(10{^{\circ }}\) is plotted in Fig. 7. The spectral distribution of the uncertainty is also shown in the lower half of Fig. 7 with the corresponding ordinate axis on the right-hand side. The directional spectral emissivity was also measured under different angles of observation from \(10{^{\circ }}\) to \(70{^{\circ }}\) and is plotted in Fig. 8. Tables 7 and 8 provide the numerical values of the directional total and hemispherical total emissivities of Acktar Fractal Black.

4.5 Total and Hemispherical Emissivities

The integrated quantities, total and hemispherical emissivity, of the four black coatings are presented here for an overview, depending on temperature and the integrated wavelength range (MIR: \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\), FIR: \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\)). The directional total emissivities with their standard uncertainties are plotted in Figs. 9 and 10 and the hemispherical total emissivities in Fig. 11.

Fig. 9
figure 9

Directional total emissivities of the investigated coatings in the wavelength range from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at the temperatures of \(120~{^{\circ }}\)C, respectively, \(150~{^{\circ }}\)C with their corresponding standard uncertainties

Fig. 10
figure 10

Directional total emissivities of the investigated coatings in the wavelength range from \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) at the temperatures of \(120~{^{\circ }}\)C, respectively, \(150~{^{\circ }}\)C with their corresponding standard uncertainties

Fig. 11
figure 11

Hemispherical total emissivities of the investigated coatings with their corresponding standard uncertainties. The results are grouped according to the integrated spectral ranges and measurement temperatures

5 Conclusion

Highly accurate measurements of directional spectral, directional total and hemispherical total emissivities for four widely used black coatings—Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black—were taken by PTB in the wavelengths range from \(4~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) at different temperatures in the range from \(25~{^{\circ }}\)C to \(150~{^{\circ }}\)C.

The achieved absolute standard uncertainty (\(k=1\)) is calculated according to [5,6,7]. It is 0.005 or better for measurements at high temperatures (\(T=120~{^{\circ }}\)C and \(T=150~{^{\circ }}\)C) and it is better than 0.022 (for some measurements better than 0.011) for measurements at \(25~{^{\circ }}\)C. It should be noted that the achieved uncertainty significantly depends on the experimental conditions and the investigated coating.

The results achieved in this work are traceable to the International Temperature Scale of 1990 (ITS-90) due to the use of two radiance standards in terms of two reference blackbodies which are directly linked to the ITS-90. Furthermore, some of the measurements at the RBCF were validated by comparing them with measurements performed at EMAF. EMAF has successfully participated in CCT-S1, the supplementary comparison of CCT for spectral emissivity [4]. All emissivity results obtained at the RBCF and at the EMAF are consistent within the uncertainties of the measurements.

6 Note

The spectral emissivity data of the investigated coatings is made available as an electronic supplement to this article on the journals website.