Abstract
The NMISA Humidity Laboratory maintains the South African national measurement standards for humidity and provides traceability for commercial calibration laboratories. The laboratory performs both relative humidity and dew-point calibrations. The laboratory has been accredited since January 2004 for both parameters by the South African National Accreditation System (SANAS). Industries such as the pharmaceutical industry, power stations, and the food industry make use of the calibration services provided by the laboratory. The facilities and the measurement capabilities of the laboratory are presented in this paper. Two calibration methods are used for the calibration of relative humidity measuring instruments (RH meters). Depending on the type of instrument, they are calibrated against aqueous salt solutions or by comparison with reference RH meters. Measurements are performed over the humidity range 5–95 %rh (Best Measurement Capabilities: 0.4–1.4 %rh). Two chilled-mirror hygrometers are used as reference standards for dew-point calibrations over the range from −75°Cfp to +17°Cdp (Best Measurement Capabilities: 1.5°Cdp to 0.4°Cdp). The laboratory has participated in two international interlaboratory comparisons, one for each accredited parameter. The NMISA Humidity Laboratory is working to reduce its calibration uncertainties for relative humidity instruments: since first accredited in 2004, the best measurement capabilities have been improved from 2.1 %rh to 0.4 %rh for some humidity ranges.
Similar content being viewed by others
References
C. Jones, Humidity measurement in South Africa – no longer wet behind the ears, in Papers and Abstracts from the 3rd International Symposium on Humidity and Moisture (National Physical Laboratory, Teddington, 1998)
D. Jonker, R. Mokhutshoane, H. Liedberg, Characterisation of relative humidity sensors, in Proceedings of 4th International Symposium on Humidity and Moisture (CMS-ITRI, Taipei, 2002), pp. 576–582
D. Jonker, R. Mnguni, H. Liedberg, Reduction of relative humidity calibration uncertainties at CSIR NML, in Proceedings of 5th International Symposium on Humidity and Moisture (INMETRO, Rio de Janeiro, 2006)
D. Jonker, R. Mokhutshoane, Comparison of the NML dewpoint and relative humidity standards, in Proceedings of NLA/MIG Test and Measurement Workshop and Conference (South Africa, 2001)
M.R. Mnguni, D. Jonker, V. Ramnath, H.G. Liedberg, in Proceedings of TEMPMEKO 2004, 9th International Symposium on Temperature and Thermal Measurements in Industry and Science, ed. by D. Zvizdić, L.G. Bermanec, T. Veliki, T. Stašić (FSB/LPM, Zagreb, Croatia, 2004), pp. 717–722
H.G. Liedberg, M.R. Mnguni, D. Jonker, in Proceedings of TEMPMEKO 2007, Int. J. Thermophys., doi:10.1007/s10765-008-0423-z
R. Benyon, Report on the EA Interlaboratory Comparison Th 8 “Relative Humidity” (NAC, Spain, 2000)
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Jonker, D., Mnguni, M.R. & Liedberg, H.G. The Humidity Calibration Facility of the National Metrology Institute of South Africa (NMISA). Int J Thermophys 29, 1644–1651 (2008). https://doi.org/10.1007/s10765-008-0426-9
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10765-008-0426-9