Abstract
A EUROMET collaborative project has been set up between Instituto Nacional de Técnica Aeroespacial (INTA) and E+E ELEKTRONIK Ges.m.b.H, the two designated laboratories of the Spanish and Austrian National Metrology Institutes, Centro Español de Metrología (CEM) and Bundesamt für Eich- und Vermessungswesen (BEV), respectively. The objective of the project is to provide INTA with a new standard that covers the dew-point temperature range from − 27°C to +90°C with a gas flow up to 5 L· min−1 in the “two-pressure” mode, extended to 95°C when operated as a continuous flow “single-pressure” generator, and investigate the importance of the enhancement factors in the uncertainty estimations used in support of the participants’ calibration and measurement capabilities (CMC) (The CIPM Mutual Recognition Arrangement, http://www.bipm.fr/en/cipm-mra/). The equivalence of the Spanish and Austrian national standards is also to be evaluated, further supporting the outcomes of the Key Comparisons, in which both have already participated. The preliminary results obtained to date are reported and discussed in the context of the project and the consistency of the declared CMC’s.
Similar content being viewed by others
References
The CIPM Mutual Recognition Arrangement, http://www.bipm.fr/en/cipm-mra. Accessed 10 Dec 2007
H. Mitter, The BEV/E+E Elektronik standard humidity generator, in 5th International Symposium on Humidity and Moisture, Brazil (2006)
G. Scholz, A standard calibrator for air hygrometers, Bulletin OIML Nr. 97 (Dez. 1984), pp. 18–27
P. Mackrodt, R. Benyon, G. Scholz, State-of-the-art calibration of high-range chilled-mirrror hygrometers and their use in the intercomparison of humidity standard generators, in Papers and Abstracts from the Third International Symposium on Humidity and Moisture, (NPL, 1998), pp. 159–166
R. Benyon, P. Huang, A comparison of INTA and NIST humidity standard generators, in Papers and Abstracts from the Third International Symposium on Humidity and Moisture (NPL, 1998), pp. 28–36
Sonntag D. (1990). Z. Meteorol. 70: 340
Greenspan L. (1976). J. Res. Natl. Bur. Stds. – A Phys. Chem. 80A: 41
B. Hardy, ITS-90 formulations for vapour pressure, frostpoint temperature, dewpoint temperature and enhancement factors in the range − 100 to 100°C, in Third International Symposium on Humidity and Moisture, Papers and Abstracts of the Conference (1998), (National Physical Laboratory, UK), ISBN 0 946754 241 pp. 214–221
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Benyon, R., Mitter, H. The New INTA High-Range Standard Humidity Generator and Its Comparison with the Austrian National Humidity Standard Maintained at BEV/E+E. Int J Thermophys 29, 1623–1631 (2008). https://doi.org/10.1007/s10765-007-0310-z
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10765-007-0310-z