Abstract
Venous blood is a preferred matrix for the determination of total mercury (Hg) in human biomonitoring but has some drawbacks such as the requirement for an uninterrupted cold chain for transport and storage and the need of medical personnel for sample collection. Therefore, we tested and implemented a simpler and less expensive method for measuring Hg in human blood using dried blood spots (DBS). For method development, we investigated the influence of different storage conditions (temperature, storage vessel, time) on DBS samples. For method validation, we compared DBS and venous blood and investigated whether DBS sampling is suitable for measuring Hg in the general population in countries with low Hg exposure such as Germany. Based on our results, we found that pre-cleaned glass tubes were most suitable for storage of DBS samples, as this allowed the samples to remain stable for at least 4 weeks even at high temperatures (40 °C). When comparing venous blood and DBS, a very good correlation (r = 0.95, p < 0.01, Spearman-Rho) and high precision of DBS (mean relative standard deviation 8.2% vs. 7.2% in venous blood samples) were observed. Comparing the recoveries of both matrices in different concentration ranges, the variation of the recoveries decreases with increasing Hg concentration. The mean recoveries also decreased with increasing Hg concentration. Overall, we found comparable results for DBS and venous blood using direct Hg analysis. Furthermore, we demonstrated that DBS are suitable for Hg biomonitoring in the general population in Germany and improved the storage conditions for the DBS.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Mercury (Hg) is a toxic metal and a hazard for humans and the environment (UNEP, 2013). Even low exposure to Hg has negative effects on the health of humans, e.g., on cardiovascular, reproductive, renal, and central nervous systems, and, in particular, during pregnancy and infanthood, cognitive development can be disturbed at low levels of exposure (Afrifa et al., 2019; Bose-O’Reilly et al., 2010; Genchi et al., 2017). To be able to study these associations in detail, simple methods to sample Hg in those populations, e.g., newborns, are needed. Hg can occur in the elemental, inorganic (e.g., HgCl2), or organic form (e.g., methylmercury), with each species having different toxicodynamic properties (Bernhoft, 2012; Bose-O’Reilly et al., 2010). The major anthropogenic sources of Hg emissions are the burning of fossil fuels and artisanal and small-scale gold mining (ASGM) activities (UNEP, 2019). In ASGM areas, the use of elemental Hg is a major source of Hg exposure (Bose-O’Reilly et al., 2017; Nyanza et al., 2019; Sundseth et al., 2017). In high-income countries, the main source of Hg exposure is the consumption of fish, e.g., tuna, which may contain a high amount of organic Hg due to the accumulation in the food chain (Bose-O’Reilly et al., 2010; Sundseth et al., 2017).
To assess Hg exposure in human blood, venous blood sampling represents the current gold standard. Challenges in using venous blood as a sample material include the necessary availability of trained medical personnel and the maintenance of an uninterrupted cooling chain prior to laboratory analysis. This may lead to relatively high logistical efforts and costs. Furthermore, venipuncture is a more invasive technique compared to other methods of specimen collection and is ethically concerning for certain populations such as infants and children (Basu et al., 2017, 2018; Funk et al., 2015; Nyanza et al., 2019; Ostler et al., 2014; Santa-Rios et al., 2020). As an alternative to venous blood sampling, dried blood spot (DBS) sampling can be used for Hg biomonitoring (Basu et al., 2017; Chaudhuri et al., 2009; Funk et al., 2013, 2015; Nelson et al., 2016; Nyanza et al., 2019; Perkins & Basu, 2018; Santa-Rios et al., 2020). Here, capillary blood, e.g., from the finger, is collected on special filter cards. In contrast to venous blood sampling, only a few drops of blood (approximately 60 µl per circle) are needed. DBS sampling usually do not require cooling, extensive storage space, or even medically trained personnel, therefore increasing the feasibility of this method for studies of large populations or in remote areas with limited laboratory infrastructure (Basu et al., 2017; Chaudhuri et al., 2009; Lehmann et al., 2013; McDade et al., 2007; Nelson et al., 2016; Ostler et al., 2014; Santa-Rios et al., 2020).
So far, only three studies compared DBS with venous blood sampling for human biomonitoring of Hg (Funk et al., 2015; Nyanza et al., 2019; Santa-Rios et al., 2020). One study found a relationship between venous blood and DBS, but it was confounded by the high background contamination of DBS cards, which is why the authors considered the retrospective use of DBS without pre-cleaning of the cards inappropriate (Funk et al., 2015). A second study showed that the Hg levels in DBS samples from pregnant women, who live in an ASGM area, were comparable to venous blood levels (Nyanza et al., 2019). Furthermore, it was demonstrated that DBS are comparable to associated venous blood levels with samples from low exposed persons, too, when methylmercury and inorganic mercury were analyzed using gas chromatography-cold vapor atomic fluorescence (GC-CVAFS) (Santa-Rios et al., 2020). However, DBS sampling has so far not been used for biomonitoring of total Hg in blood in the general population in countries with an expected low exposure, e.g., Germany, in combination with direct Hg analysis. The implementation of DBS sampling for Hg biomonitoring comes with multiple challenges such as the background contamination of DBS cards as well as potential contamination or loss of Hg during sampling, transport, and storage (Basu et al., 2017; Chaudhuri et al., 2009; Funk et al., 2013, 2015; Lehmann et al., 2013; Nyanza et al., 2019; Santa-Rios et al., 2020).
Only few studies dealt with the storage stability of Hg in DBS samples (Chaudhuri et al., 2009; Perkins & Basu, 2018; Santa-Rios et al., 2020). In these studies, Hg in DBS samples was found to generally stable under the investigated conditions. However, no studies have been carried out on the influence of different storage vessels for DBS, although its storage is an important factor for possible contamination of DBS samples.
The aim of this study was the development and validation of a biomonitoring method for total Hg in capillary blood using DBS sampling in combination with direct Hg analysis. Therefore, DBS and venous blood samples were collected from German adults to evaluate if both sampling methods will provide comparable results. Additionally, the influence of different storage vessels on the stability of Hg in DBS samples was investigated.
Material and methods
Used materials
Hg ICP standard (998 ± 8 mg/l in 10% nitric acid), hydrochloric acid (30%) for trace metal analysis, and nitric acid (65%) for trace metal analysis were obtained from Merck (Darmstadt, Germany). Whatman® 903 protein saver cards for DBS sampling were obtained from Sigma-Aldrich (St. Louis, USA). Disposable lancets (Solofix®) from B. Braun (Melsungen, Germany) were used for fingerpricks. LDPE plastic zip lock bags (22 × 16 cm) for DBS storage were obtained from Buerkle (Bad Bellingen, Germany). Borosilicate glass tubes with plastic screw caps (98 × 16 mm) for DBS storage were obtained from Schuett-biotec (Goettingen, Germany). Certified reference materials for blood (ClinChek®, RECIPE, Munich, Germany) from different batches were used for stability experiments (Hg concentration: 2.5 µg/l and 8.4 µg/l) and quality control during the analysis of the participants’ DBS samples (Hg concentration: 2.9 µg/l).
Storage stability of Hg in DBS
In order to investigate the influence of different storage conditions on Hg levels in DBS, fresh venous blood from one volunteer was used for the following experiments. In detail, three spots of one DBS card were spiked with 55 µl blood in order to be within the circled line and dried for 2 h at room temperature. To test the influence of the storage vessel, the prepared DBS cards were individually stored in plastic bags, untreated glass tubes, and glass tubes that had been pre-cleaned with an aqueous mixture of hydrochloric and nitric acid (each 5%, v/v), respectively (Figure S1). In detail, the glass tubes were filled with 5 ml acid mixture and put on a roll mixer for 2 h. Afterwards, the tubes were rinsed twice with ultrapure water and dried at 100 °C in an oven. In order to store DBS samples in the glass tubes, the DBS spots were cut out approximately 5 mm below the circled line using an acid-washed stainless steel scissor. To test the influence of the storage temperature, each storage vessel was stored at − 20 °C, room temperature, and 40 °C for 1, 2, and 4 weeks, respectively. Every single experiment was conducted in duplicate. As a reference, Hg levels were additionally analyzed in DBS cards prepared in the same way as described above immediately after drying. Furthermore, the stability of Hg in DBS was also tested using certified reference material for blood Hg concentration: 2.5 µg/l and 8.4 µg/l, storing the prepared DBS cards in plastic bags at 40 °C for up to 4 weeks.
Application of DBS for human biomonitoring of Hg
Study design
This proof-of-principle study was conducted at the Institute and Clinic for Occupational, Social and Environmental Medicine, LMU University Hospital Munich. The study was carried out in accordance with the Code of Ethics of the Declaration of Helsinki for experiments involving human subjects and reviewed and approved by the ethics committee of the Ludwig Maximilians University of Munich (20–091). Participants had to be at least 18 years old and no individual restrictions for venous and capillary blood sampling. In total, 53 participants were randomly recruited at the Institute and Clinic of Occupational, Social and Environmental medicine at the LMU University Hospital Munich in June and July 2020. Prior to the sampling, each participant was informed about the study and signed an informed consent form. Each participant was asked to fill out a questionnaire about potential Hg exposure (e.g., fish consumption and dental amalgam).
Sample collection
Venous blood samples from all participants were collected into 7 ml Lithium-Heparin-coated tubes for trace metal analyses (Sarstedt®) and stored at − 20 °C until analysis. For DBS sampling, the same participant was asked to wash his hands thoroughly to prevent contamination during the sampling. Afterwards, one finger was disinfected and pricked with a sterile disposable lancet. The first drop of blood was discarded before filling three spots of one DBS card with capillary blood. If the blood flow stopped before the circled area of three spots was completely filled, another finger was punctured with the consent of the participant. The DBS samples were dried for 2 h at room temperature. To test the influence of storage on real samples, some DBS samples (n = 18) were analyzed immediately after drying. The remaining samples (n = 32) were stored for 1 week at room temperature in a pre-cleaned glass tube as described above.
Sample analysis
All samples were analyzed by direct Hg analysis using a DMA80-evo® instrument (MLS Mikrowellen-Labor-Systeme GmbH, Leutkirch, Germany). Hg was detected by atomic absorption at 253.5 nm. The quantification was based on an external calibration. Before every analysis series, the sample boats were preconditioned to avoid interference by residual Hg. For venous blood, 100 µl blood were directly pipetted into the sample boats and analyzed. Each venous blood sample was at least analyzed in triplicate. For DBS, three completely filled circles were punched out using a pre-cleaned 0.5-inch stainless steel paper puncher. The punched circles, which contained approximately 60 µl (specification by the manufacturer, 55 µl for stability experiments) blood, were individually placed directly into the sample boat of the instrument and analyzed to yield three individual values per participant. One blank per participant was prepared in the same manner using an empty circle from the same DBS card. The limits of detection (LOD) were 0.02 µg/l for venous blood and 0.14 µg/l for DBS samples, respectively (for detailed information, see Table S1). The limits of quantitation (LOQ) were 0.04 µg/l for venous blood and 0.28 µg/l for DBS samples, respectively. For quality assurance, an aqueous Hg standard solution (10 µg/l) and certified reference material for blood (Hg concentration: 2.9 µg/l) were analyzed daily prior to the analysis of the samples. Day-to-day variation of the Hg standard and certified reference material was less than 10%.
Statistical analysis
For data processing, Excel 2016 was used. Statistical analysis was performed with IBM SPSS® Statistics, Version 25. Samples below the LOD or the LOQ were excluded from the statistical analysis. For evaluation of the accuracy of DBS, the recovery was calculated from the Hg levels in DBS and venous blood (HgDBS/HgVB*100%). Differences in recoveries due to the storage conditions were analyzed using single factor ANOVA and post-hoc tests (Sheffé, Bonferoni). For information about the Hg levels in venous blood and DBS samples of the study population, median, geometric mean, minimum, and maximum were used due to the non-normal distribution of Hg levels in blood. Correlation between Hg levels in DBS and venous blood was analyzed by Spearman-Rho test. Jonkheere-Terpstra tests were performed to evaluate the differences in the recoveries between individual concentration ranges (Hg levels in venous blood: < 0.5 µg/l, 0.5–1.0 µg/l, 1.0–1.5 µg/l, > 1.5 µg/l). Data was graphically displayed using bar charts, scatter plots, and Bland–Altman plot.
Results and discussion
Storage stability of Hg in DBS
To evaluate the effect of different storage conditions, DBS samples were stored in different vessels at different temperatures and for different times. The mean Hg level in DBS samples when analyzed directly after drying was 0.81 µg/l (100%). In Fig. 1, the recoveries of Hg in DBS cards that have been stored under different conditions are shown. For pre-cleaned glass tubes (a), recoveries varied from 78 to 99% and the differences between the groups were not significant. For plastic bags, recoveries varied between 70 and 240%. For storage at − 20 °C and room temperature, no significant differences between the groups were found. In contrast, recoveries significantly increased after 2 and 4 weeks of storage when the samples were kept at 40 °C (p < 0.01). Hg recoveries in the uncleaned glass tubes showed high variability that could not be explained by the storage conditions, although mean recoveries were good for the majority of the tested conditions (Figure S3).
Based on these results, we decided to use pre-cleaned glass tubes for the storage of DBS samples in the further course of this study. Consequently, we believe that the samples will be stable for at least 4 weeks, even when stored at 40 °C. Uncleaned glass showed good results, too. However, the results suggest that individual glass tubes may be contaminated with residual amounts of Hg, e.g., during manufacturing, and therefore require cleaning before they can be used as containers for DBS samples. The majority of previously published studies that use DBS samples for Hg biomonitoring use plastic bags for storage (Chaudhuri et al., 2009; Funk et al., 2015; Nelson et al., 2016; Perkins & Basu, 2018; Santa-Rios et al., 2020). Although the storage of DBS samples in plastic bags showed satisfying recoveries at − 20 °C and room temperature in our experiments, Hg levels were significantly elevated when the samples were stored at 40 °C. The elevated Hg levels may be explained by residual Hg in the bags or ambient Hg that could penetrate the plastic bag and bind to the dried blood. Other studies recommend the use of metal-free plastic bags or cleaning procedures (Funk et al., 2015; Nyanza et al., 2019). However, we preferred glass tubes to plastic bags, as the handling for cleaning and drying of the tubes was relatively simple.
Interestingly, accumulation of Hg in DBS samples has not been observed when certified reference material for blood, which also has been used in several other studies, was used instead of fresh blood for stability experiments (Figure S4) (Basu et al., 2017; Chaudhuri et al., 2009; Funk et al., 2015; Nelson et al., 2016; Nyanza et al., 2019; Santa-Rios et al., 2020). As blank values were also not affected by any storage condition (supplementary material), we assume that the accumulation of Hg during storage only affects DBS samples made from freshly drawn blood. Consequently, freshly drawn blood, e.g., venous blood, should be preferably used for method development.
Correlation of Hg in venous blood and DBS samples
In total, 53 participants were recruited. One participant was excluded because only one circle on the DBS card was completely filled. Two participants were excluded because of possible sample contamination. In contrast, six participants with only two DBS spots used for Hg analysis were included in the study. Consequently, the data from 50 participants was used for further analysis. The results for venous blood and DBS samples are shown in Table 1. The Hg levels in one venous blood sample were below the LOD, in another below the LOQ. For DBS, six samples were below the LOQ and therefore excluded from further analysis. Stratification of the Hg levels in venous blood by gender, age, fish consumption, and dental amalgam fillings can be found in the Supporting Information (Table S2). The individual mean Hg values in venous blood and DBS samples can be found in the Supporting Information (Table S3).
Regarding the precision of Hg analysis, venous blood sampling was slightly but not significantly superior to DBS sampling. Although more than 50% samples of both venous blood and DBS showed a relative standard deviation (RSD) of less than 10%, mean RSD of venous blood samples was lower compared to DBS (7.2% vs. 8.2%). Furthermore, 94% of the venous blood samples were below an RSD of 20% compared to 93% of the DBS samples. This may be explained by residual Hg in the filter paper. As other studies report, residual Hg is likely not homogenously distributed in the filter paper, limiting the correction for blank samples from the same DBS card (Funk et al., 2015). Nevertheless, precision of DBS sampling was relatively comparable to venous blood sampling.
The correlation of the Hg levels in venous blood and DBS samples is shown in Fig. 2. We found a very strong linear correlation between Hg levels in venous blood and DBS samples (r = 0.95, p < 0.001, Spearman-Rho). This is comparable to what has been found by Nyanza et al. (2019) and Santa-Rios et al. (2020). The majority of the samples were within the desired recovery range between 70 and 130% (75%). Overall, recoveries ranged between 62 and 210%. However, recoveries of more than 150% were only observed for samples with venous blood Hg levels of less than 0.5 µg/l. The associated Bland–Altman plot (Fig. 3) further confirms that DBS sampling and venous blood sampling are comparable methods, although the results for DBS samples were on average 0.09 µg/l lower than what has been found for venous blood samples (bias). Additionally, the recovery of Hg levels in DBS samples depended on the concentration of Hg in venous blood. In fact, there was statistically significant negative correlation (p < 0.01) between the mean Hg level of both methods and the difference between the Hg levels in DBS and venous blood samples.
In Fig. 4, the recoveries of DBS samples stratified by concentration ranges are shown. Median recoveries significantly decreased with increasing Hg levels in venous blood (p = 0.001, Jonkheere-Terpstra test). Additionally, the variation of the recoveries was lowest when the Hg levels in venous blood were above 1.5 µg/l. This may be explained by the fact that residual Hg on the DBS cards has a higher impact on the results when the Hg levels in blood are relatively low. Furthermore, it appears that results from the DBS samples tend to be elevated in samples with venous blood Hg levels below 0.5 µg/l. Above 0.5 µg/l, mean recoveries were found to be between 84 and 99%. Because the impact of residual Hg decreases with increasing Hg levels in venous blood, lowest scattering of recoveries was found in samples above Hg levels of 1.5 µg/l. The fact that the recoveries for DBS samples with Hg levels above 1.0 µg/l were below 100% could be due to a lower blood volume in the punched DBS than the estimated 60 µl.
Other studies report the influence of hematocrit on the volume of blood in a specific area of DBS card and recommend the weighing of punched spots to adjust for individual hematocrit values (Mei et al., 2001; Nyanza et al., 2019; O’Broin, 1993). However, we deliberately did not weigh the punched DBS spots because the hematocrit levels in humans are in sufficiently narrow physiological ranges for meaningful results (McDade et al., 2007). Finally, no significant difference for Hg recoveries was found between direct analysis of the DBS cards after drying (n = 17; median recovery: 103%) and analysis after storage for 1 week (n= 27; median recovery: 88%). Consequently, the storage of DBS samples in pre-cleaned glass tubes successfully prevented contamination.
Strengths and limitations of the study
The application of DBS sampling in combination with direct Hg analysis for Hg biomonitoring is the key strength of our study. Furthermore, our study sheds more light on the stability of Hg in DBS samples under different storage conditions, especially with regard to storage vessels and temperatures. Our study also emphasizes the importance of the use of fresh blood samples for method development instead of certified reference materials.
A limitation of our study is the fact that the exact blood volume of each DBS was not adjusted for hematocrit levels as suggested in other studies (Basu et al., 2017; Funk et al., 2015; Nyanza et al., 2019). Nevertheless, Funk et al. state that estimated blood volume sufficient for valid results (Funk et al., 2013). Additionally, Hg levels in capillary blood may be different from venous blood, e.g., due to the dilution by tissue fluid. Furthermore, the LOQ was too low for about 10% of the DBS samples. This is due to the relatively low blood volume of 60 µl on one hand and varying residual Hg amounts on the DBS cards that interfere greatly with low Hg amounts from the sampled blood on the other hand. Nonetheless, the LOQ of our method (0.24 µg/l) is, in our opinion, more than sensitive enough to identify people with medium to elevated Hg levels in blood. However, a better sensitivity may be achieved by using pre-cleaned DBS cards and ICP-MS for analysis as it has been described in other studies (Chaudhuri et al., 2009; Funk et al., 2013, 2015; Nelson et al., 2016; Nyanza et al., 2019). A further limitation of this method is that DBS spots cannot be used for the determination of other trace elements after direct mercury analysis. However, direct mercury analysis of DBS samples is a strength of this study.
Conclusion
Since DBS samples in studies and human biomonitoring, especially in areas with poor infrastructure, usually cannot be measured immediately after sample collection, the stability of Hg is an important prerequisite for its applicability. Therefore, we investigated the stability of Hg in DBS and the optimal storage conditions for DBS samples. To the best of our knowledge, this study is the first to investigate the stability of Hg in DBS samples in different storage vessels. We also demonstrated that it is advisable to use human blood for method development instead of certified reference materials to study the stability of Hg in DBS samples during storage. By comparing DBS with venous blood sampled under field conditions, we could show that the Hg levels in DBS samples have comparable precision and good correlation to venous blood samples. In conclusion, DBS sampling is a suitable sampling method in combination with direct Hg analysis for Hg biomonitoring. DBS are therefore an alternative to the common venous blood sampling. As DBS sampling only requires 60 µl capillary blood per spot, compared to several milliliters for venous blood sampling, this method may be particularly suitable for the Hg biomonitoring of infants and newborns. The second advantage is that DBS samples do not require a cooling chain and could therefore be used in remote areas, such as ASGM areas, to screen the Hg levels of miners. In summary, microsampling-assisted Hg biomonitoring may be a useful addition to assess and reduce Hg exposure in line with the Minamata convention. However, more research is required to further improve the sensitivity and sample stability of DBS.
Availability of data and materials
Individual anonymized Hg levels of the participants can be found in the supplementary material. Further anonymized data may be made available upon reasonable request.
References
Afrifa, J., Opoku, Y.K., Gyamerah, E.O., Ashiagbor, G., & Sorkpor, R.D. (2019). The clinical importance of the mercury problem in artisanal small-scale gold mining. Frontiers in Public Health, 7, 131–131. https://doi.org/10.3389/fpubh.2019.00131
Basu, N., Eng, J.W.L., Perkins, M., Santa-Rios, A., Martincevic, G., Carlson, K., et al. (2017). Development and application of a novel method to characterize methylmercury exposure in newborns using dried blood spots. Environmental Research, 159, 276–282. https://doi.org/10.1016/j.envres.2017.08.021
Basu, N., Horvat, M., Evers, D.C., Zastenskaya, I., Weihe, P., & Tempowski, J. (2018). A state-of-the-science review of mercury biomarkers in human populations worldwide between 2000 and 2018. Environmental Health Perspectives, 126(10), 106001. https://doi.org/10.1289/ehp3904
Bernhoft, R.A. (2012). Mercury toxicity and treatment: A review of the literature. Journal of Environmental and Public Health, 2012, 460508–460508. https://doi.org/10.1155/2012/460508
Bose-O’Reilly, S., Bernaudat, L., Siebert, U., Roider, G., Nowak, D., & Drasch, G. (2017). Signs and symptoms of mercury-exposed gold miners. International Journal of Occupational Medicine and Environmental Health, 30(2), 249–269. https://doi.org/10.13075/ijomeh.1896.00715
Bose-O’Reilly, S., McCarty, K.M., Steckling, N., & Lettmeier, B. (2010). Mercury exposure and children’s health. Current Problems in Pediatric and Adolescent Health Care, 40(8), 186–215. https://doi.org/10.1016/j.cppeds.2010.07.002
Chaudhuri, S.N., Butala, S.J., Ball, R.W., & Braniff, C. T. (2009). Pilot study for utilization of dried blood spots for screening of lead, mercury and cadmium in newborns. Journal of Exposure Science & Environmental Epidemiology, 19(3), 298–316. https://doi.org/10.1038/jes.2008.19
Funk, W.E., McGee, J.K., Olshan, A.F., & Ghio, A.J. (2013). Quantification of arsenic, lead, mercury and cadmium in newborn dried blood spots. Biomarkers, 18(2), 174–177. https://doi.org/10.3109/1354750x.2012.750379
Funk, W.E., Pleil, J.D., Sauter, D.J., McDade, T., & Holl, J.L. (2015). Use of dried blood spots for estimating children’s exposures to heavy metals in epidemiological research. Journal of Environmental & Analytical Toxicology, S7(2).
Genchi, G., Sinicropi, M.S., Carocci, A., Lauria, G., & Catalano, A. (2017). Mercury exposure and heart diseases. International Journal of Environmental Research and Public Health, 14(1). https://doi.org/10.3390/ijerph14010074
Lehmann, S., Delaby, C., Vialaret, J., Ducos, J., & Hirtz, C. (2013). Current and future use of “dried blood spot” analyses in clinical chemistry. Clinical Chemistry and Laboratory Medicine, 51(10), 1897–1909. https://doi.org/10.1515/cclm-2013-0228
McDade, T.W., Williams, S., & Snodgrass, J.J. (2007). What a drop can do: Dried blood spots as a minimally invasive method for integrating biomarkers into population-based research. Demography, 44(4), 899–925. https://doi.org/10.1353/dem.2007.0038
Mei, J.V., Alexander, J.R., Adam, B.W., & Hannon, W.H. (2001). Use of filter paper for the collection and analysis of human whole blood specimens. Journal of Nutrition, 131(5), 1631s–1636s. https://doi.org/10.1093/jn/131.5.1631S
Nelson, J.W., Edhlund, B.L., Johnson, J., Rosebush, C.E., Holmquist, Z.S., Swan, S.H., et al. (2016). Assessing a new method for measuring fetal exposure to mercury: Newborn bloodspots. International Journal of Environmental Research and Public Health, 13(7), 692. https://doi.org/10.3390/ijerph13070692
Nyanza, E.C., Bernier, F.P., Manyama, M., Hatfield, J., Martin, J.W., & Dewey, D. (2019). Maternal exposure to arsenic and mercury in small-scale gold mining areas of Northern Tanzania. Environmental Research, 173, 432–442. https://doi.org/10.1016/j.envres.2019.03.031
O’Broin, S. (1993). Influence of hematocrit on quantitative analysis of “blood spots” on filter paper. Clinical Chemistry, 39(6), 1354–1355.
Ostler, M.W., Porter, J.H., & Buxton, O.M. (2014). Dried blood spot collection of health biomarkers to maximize participation in population studies. Journal of Visualized Experiments, (83), e50973. https://doi.org/10.3791/50973
Perkins, M., & Basu, N. (2018). Dried blood spots for estimating mercury exposure in birds. Environmental Pollution, 236, 236–246. https://doi.org/10.1016/j.envpol.2018.01.036
Santa-Rios, A., Barst, B.D., & Basu, N. (2020). Mercury speciation in whole blood and dried blood spots from capillary and venous sources. Analytical Chemistry, 92(5), 3605–3612. https://doi.org/10.1021/acs.analchem.9b04407
Sundseth, K., Pacyna, J.M., Pacyna, E.G., Pirrone, N., & Thorne, R.J. (2017). Global sources and pathways of mercury in the context of human health. International Journal of Environmental Research and Public Health, 14(1), 105. https://doi.org/10.3390/ijerph14010105
United Nations Environment Programme UNEP (2013). Mercury: Time to act
United Nations Environment Programme UNEP (2019). Global Mercury Assessment 2019. Sources, Emissions, Releases and Environmental Transport.
Funding
Open Access funding enabled and organized by Projekt DEAL. This study was funded by the Friedrich-Baur foundation (Reg.-Nr. 27/18).
Author information
Authors and Affiliations
Contributions
Ann-Kathrin Schweizer: sampling, sample analysis, data analysis, visualization, preparation of the original draft, writing-review & editing. Michael Kabesch: supervision, conceptualization, writing-review & editing. Caroline Quartucci: sampling, supervision, writing-review & editing. Stephan Bose-O’Reilly: supervision, conceptualization, methodology, data analysis, writing-review & editing. Stefan Rakete: conceptualization, methodology, funding acquisition, data analysis, writing-review & editing, preparation of the final manuscript. All authors have given approval to the final version of the manuscript.
Corresponding author
Ethics declarations
Ethics approval
The study was carried out in accordance with the Code of Ethics of the Declaration of Helsinki for experiments involving human subjects and reviewed and approved by the ethics committee of the Ludwig Maximilians University of Munich (20–091).
Consent to participate
Informed consent was obtained from all individual participants included in the study.
Consent for publication
The participants signed an informed consent for publishing aggregated or anonymized data.
Conflict of interest
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Schweizer, AK., Kabesch, M., Quartucci, C. et al. Implementation of mercury biomonitoring in German adults using dried blood spot sampling in combination with direct mercury analysis. Environ Monit Assess 193, 488 (2021). https://doi.org/10.1007/s10661-021-09254-0
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s10661-021-09254-0