Abstract
Objective
The aim of our study was to evaluate the level of agreement between tricuspid annular plane systolic excursion (TAPSE) measured by transthoracic echocardiography (TTE) and TAPSE measured using transesophageal echocardiography (TEE) in anesthetized patients.
Materials and methods
Thirty patients scheduled for elective cardiac surgery were prospectively studied. Shortly after induction of anesthesia before the operation, TAPSE was measured by TTE using M-mode in apical 4chamber view (4CH) and by TEE in six different views: using 2D echocardiography in midesophageal (ME) 4CH view, using M-mode in deep transgastric right ventricle (dTG RV) view at 0° and dTG RV longaxis view (LAX) as well as using anatomical M-mode (AM-mode) in ME 4CH, dTG RV at 0° and dTG RV LAX views.
Results
Bland–Altman analysis showed a good agreement for TAPSE measured using M-mode in TTE and using AM-mode in TEE in the ME 4CH and dTG RV at 0° views (−2.5 ± 18 and −2.2 ± 14% respectively). The agreement between TAPSE measured in TTE and TEE using 2D in ME 4CH, using M-mode in dT GRV 0° and using M-mode and AM-mode in dTG RV LAX view showed a significant systematic underestimation of the measurements (−8.8 ± 21, −8.8 ± 24, −17.8 ± 28 and −6.4 ± 20%).
Conclusion
Our study showed that the right ventricular function can be accurately and precisely estimated using TAPSE measurement by TEE in the midesophageal four chamber and deep transgastric right ventricle view at 0° using anatomical M-mode.
Similar content being viewed by others
References
Denault AY, Haddad F, Jacobsohn E, Deschamps A (2013) Perioperative right ventricular dysfunction. Curr Opin Anaesthesiol 26(1):71–81. doi:10.1097/ACO.0b013e32835b8be2
Vandenheuvel MA, Bouchez S, Wouters PF, De Hert SG (2013) A pathophysiological approach towards right ventricular function and failure. Eur J Anaesthesiol 30(7):386–394. doi:10.1097/EJA.0b013e3283607a2d
Haddad F, Elmi-Sarabi M, Fadel E, Mercier O, Denault AY (2016) Pearls and pitfalls in managing right heart failure in cardiac surgery. Curr Opin Anaesthesiol 29(1):68–79. doi:10.1097/ACO.0000000000000284
Lella LK, Sales VL, Goldsmith Y, Chan J, Iskandir M, Gulkarov I, Tortolani A, Brener SJ, Sacchi TJ, Heitner JF (2015) Reduced right ventricular function predicts long-term cardiac re-hospitalization after cardiac surgery. PLoS ONE 10(7):e0132808. doi:10.1371/journal.pone.0132808
Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, Flachskampf FA, Foster E, Goldstein SA, Kuznetsova T, Lancellotti P, Muraru D, Picard MH, Rietzschel ER, Rudski L, Spencer KT, Tsang W, Voigt JU (2015) Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr 28 (1):1–39 e14. doi:10.1016/j.echo.2014.10.003
Rudski LG, Lai WW, Afilalo J, Hua L, Handschumacher MD, Chandrasekaran K, Solomon SD, Louie EK, Schiller NB (2010) Guidelines for the echocardiographic assessment of the right heart in adults: a report from the American Society of Echocardiography endorsed by the European Association of Echocardiography, a registered branch of the European Society of Cardiology, and the Canadian Society of Echocardiography. J Am Soc Echocardiogr 23 (7):685–713. doi:10.1016/j.echo.2010.05.010. quiz 786–688
Kaul S, Tei C, Hopkins JM, Shah PM (1984) Assessment of right ventricular function using two-dimensional echocardiography. Am Heart J 107(3):526–531
Ho SY, Nihoyannopoulos P (2006) Anatomy, echocardiography, and normal right ventricular dimensions. Heart 92(Suppl 1):i2–i13. doi:10.1136/hrt.2005.077875
Leather HA, Ama R, Missant C, Rex S, Rademakers FE, Wouters PF (2006) Longitudinal but not circumferential deformation reflects global contractile function in the right ventricle with open pericardium. Am J Physiol Heart Circ Physiol 290(6):H2369–H2375. doi:10.1152/ajpheart.01211.2004
Hashimoto I, Watanabe K (2016) Geometry-related right ventricular systolic function assessed by longitudinal and radial right ventricular contractions. Echocardiography 33(2):299–306. doi:10.1111/echo.13039
Wouters P (2010) The right ventricle. In: Feneck RO, Kneeshaw J, Ranucci M (eds) Core topics in transesophageal echocardiography. Cambridge University Press, Cambridge; New York, p 416
Carerj S, Micari A, Trono A, Giordano G, Cerrito M, Zito C, Luzza F, Coglitore S, Arrigo F, Oreto G (2003) Anatomical M-mode: an old-new technique. Echocardiography 20(4):357–361
Donal E, Coisne D, Pham B, Ragot S, Herpin D, Thomas JD (2004) Anatomic M-mode, a pertinent tool for the daily practice of transthoracic echocardiography. J Am Soc Echocardiogr 17(9):962–967. doi:10.1016/j.echo.2004.05.009
Mele D, Pedini I, Alboni P, Levine RA (1998) Anatomic M-mode: a new technique for quantitative assessment of left ventricular size and function. Am J Cardiol 81 (12A):82G-85G
Hahn RT, Abraham T, Adams MS, Bruce CJ, Glas KE, Lang RM, Reeves ST, Shanewise JS, Siu SC, Stewart W, Picard MH, American Society of Echocardiography, Society of Cardiovascular A (2014) Guidelines for performing a comprehensive transesophageal echocardiographic examination: recommendations from the American Society of Echocardiography and the Society of Cardiovascular Anesthesiologists. Anesth Analg 118 (1):21–68. doi:10.1213/ANE.0000000000000016
Kasper J, Bolliger D, Skarvan K, Buser P, Filipovic M, Seeberger MD (2012) Additional cross-sectional transesophageal echocardiography views improve perioperative right heart assessment. Anesthesiology 117(4):726–734. doi:10.1097/ALN.0b013e318269054b
Tan CO, Harley I (2014) Perioperative transesophageal echocardiographic assessment of the right heart and associated structures: a comprehensive update and technical report. J Cardiothorac Vasc Anesth 28(4):1100–1121. doi:10.1053/j.jvca.2013.05.031
Ling LF, Obuchowski NA, Rodriguez L, Popovic Z, Kwon D, Marwick TH (2012) Accuracy and interobserver concordance of echocardiographic assessment of right ventricular size and systolic function: a quality control exercise. J Am Soc Echocardiogr 25(7):709–713. doi:10.1016/j.echo.2012.03.018
Bilic-Zulle L (2011) Comparison of methods: Passing and Bablok regression. Biochem Med (Zagreb) 21(1):49–52
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310
Bland JM, Altman DG (1990) A note on the use of the intraclass correlation coefficient in the evaluation of agreement between two methods of measurement. Comput Biol Med 20(5):337–340
Chhapola V, Kanwal SK, Brar R (2015) Reporting standards for Bland–Altman agreement analysis in laboratory research: a cross-sectional survey of current practice. Ann Clin Biochem 52 (Pt 3):382–386. doi:10.1177/0004563214553438
Giavarina D (2015) Understanding Bland Altman analysis. Biochem Med (Zagreb) 25(2):141–151. doi:10.11613/BM.2015.015
Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J (2000) Comparing methods of clinical measurement: reporting standards for bland and altman analysis. Anesth Analg 90(3):593–602
Focardi M, Cameli M, Carbone SF, Massoni A, De Vito R, Lisi M, Mondillo S (2015) Traditional and innovative echocardiographic parameters for the analysis of right ventricular performance in comparison with cardiac magnetic resonance. Eur Heart J Cardiovasc Imaging 16(1):47–52. doi:10.1093/ehjci/jeu156
Galli E, Guirette Y, Feneon D, Daudin M, Fournet M, Leguerrier A, Flecher E, Mabo P, Donal E (2015) Prevalence and prognostic value of right ventricular dysfunction in severe aortic stenosis. Eur Heart J Cardiovasc Imaging 16(5):531–538. doi:10.1093/ehjci/jeu290
Nguyen T, Cao L, Movahed A (2014) Altered right ventricular contractile pattern after cardiac surgery: monitoring of septal function is essential. Echocardiography 31(9):1159–1165. doi:10.1111/echo.12657
Pleister A, Kahwash R, Haas G, Ghio S, Cittadini A, Baliga RR (2015) Echocardiography and heart failure: a glimpse of the right heart. Echocardiography 32(Suppl 1):S95–S107. doi:10.1111/echo.12678
Portnoy SG, Rudski LG (2015) Echocardiographic evaluation of the right ventricle: a 2014 perspective. Curr Cardiol Rep 17(4):21. doi:10.1007/s11886-015-0578-8
Tousignant C, Kim H, Papa F, Mazer CD (2012) Evaluation of TAPSE as a measure of right ventricular output. Can J Anaesth 59(4):376–383. doi:10.1007/s12630-011-9659-3
Morita Y, Nomoto K, Fischer GW (2016) Modified tricuspid annular plane systolic excursion using transesophageal echocardiography for assessment of right ventricular function. J Cardiothorac Vasc Anesth 30(1):122–126. doi:10.1053/j.jvca.2015.07.024
Fusini L, Tamborini G, Gripari P, Maffessanti F, Mazzanti V, Muratori M, Salvi L, Sisillo E, Caiani EG, Alamanni F, Fiorentini C, Pepi M (2011) Feasibility of intraoperative three-dimensional transesophageal echocardiography in the evaluation of right ventricular volumes and function in patients undergoing cardiac surgery. J Am Soc Echocardiogr 24(8):868–877. doi:10.1016/j.echo.2011.05.002
Qureshi MY, Eidem BW, Reece CL, O’Leary PW (2015) Two-dimensional measurement of tricuspid annular plane systolic excursion in children: can it substitute for an m-mode assessment? Echocardiography 32(3):528–534. doi:10.1111/echo.12687
David JS, Tousignant CP, Bowry R (2006) Tricuspid annular velocity in patients undergoing cardiac operation using transesophageal echocardiography. J Am Soc Echocardiogr 19(3):329–334. doi:10.1016/j.echo.2005.09.013
Atsumi A, Ishizu T, Kameda Y, Yamamoto M, Harimura Y, Machino-Ohtsuka T, Kawamura R, Enomoto M, Seo Y, Aonuma K (2013) Application of 3-dimensional speckle tracking imaging to the assessment of right ventricular regional deformation. Circ J 77(7):1760–1768
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Rights and permissions
About this article
Cite this article
Flo Forner, A., Hasheminejad, E., Sabate, S. et al. Agreement of tricuspid annular systolic excursion measurement between transthoracic and transesophageal echocardiography in the perioperative setting. Int J Cardiovasc Imaging 33, 1385–1394 (2017). https://doi.org/10.1007/s10554-017-1128-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10554-017-1128-9