Accuracy and rate of coronary artery segment visualization with CT angiography for the non-invasive detection of coronary artery stenoses

  • Original Paper
  • Published:
The International Journal of Cardiovascular Imaging Aims and scope Submit manuscript

Abstract

Objectives

To evaluate CT coronary angiography (CTA) when compared with catheter coronary angiography (CCA), for the detection of coronary artery stenoses and rate of optimal coronary artery segment visualization.

Method

Retrospective, two-center study enrolling 26 patients who underwent CCA and ECG-gated 16-detector CTA (slice thickness 0.6 mm; rotation 500 ms).

Results and conclusion

283 segments were available for postprocessing. Sensitivity, specificity, and positive predictive value were, respectively, 80, 100, and 100%, for detecting more than 50% luminal stenoses, when optimally visualized segments were considered, in comparison to CCA. Negative predictive value was excellent (98%). Rate of non-optimally visualized coronary segments was 26%. Most clinical benefits of coronary CT angiography should probably be obtained when it is performed to exclude significant stenoses on selected populations of patients with a low pre-test probability of severe coronary artery disease, and under optimal conditions of controlled heart rate and minimal presence of calcium.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Nieman K, Cademartiri F, Lemos PA et al (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054

    Article  PubMed  Google Scholar 

  2. Ropers D, Baum U, Pohle K et al (2003) Detection of coronary artery stenoses with thin-slice multid-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666

    Article  PubMed  Google Scholar 

  3. Mollet NM, Cademartiri F, Nieman K et al (2004) Multislice spiral computed tomography coronary angiography in patients with stable angina pectoris. J Am Coll Cardiol 43:2265–2270

    Article  PubMed  Google Scholar 

  4. Martuscelli E, Romagnoli A, D’Eliseo A et al (2004) Accuracy of thin-slice computed tomography in the detection of coronary stenoses. Eur Heart J 25:1043–1048

    Article  PubMed  Google Scholar 

  5. Hoffmann U, Moselewski F, Cury RC et al (2004) Predictive value of 16-slice multidetector spiral computed tomography to detect significant obstructive coronary artery disease in patients at high risk for coronary artery disease: patient- versus segment-based analysis. Circulation 110:2638–2643

    Article  PubMed  Google Scholar 

  6. Morgan-Hughes GJ, Roobottom CA, Owens PE, Marshall AJ (2005) Highly accurate coronary angiography with submillimetre, 16 slice computed tomography. Heart 91:308–313

    Article  PubMed  CAS  Google Scholar 

  7. Mollet NR, Cademartiri F, Krestin GP et al (2005) Improved diagnostic accuracy with 16-row multi-slice computed tomography coronary angiography. J Am Coll Cardiol 45:128–132

    Article  PubMed  Google Scholar 

  8. Hoffmann MHK, Shi H, Schmitz BL et al (2005) Noninvasive coronary angiography with multislice computed tomography. JAMA 293:2471–2478

    Article  PubMed  CAS  Google Scholar 

  9. Kuettner A, Beck T, Drosch T et al (2005) Diagnostic accuracy of noninvasive coronary imaging using 16-detector slice spiral computed tomography with 188 ms temporal resolution. J Am Coll Cardiol 45:123–127

    Article  PubMed  Google Scholar 

  10. Schuijf JD, Bax JJ, Salm LP, Jukema JW, Lamb HJ, van der Wall EE, de Roos A (2005) Noninvasive coronary imaging and assessment of left ventricular function using 16-slice computed tomography. Am J Cardiol 95:571–574

    Article  PubMed  Google Scholar 

  11. Achenbach S, Ropers D, Pohle FK et al (2005) Detection of coronary artery stenoses using multi-detector CT with 16 x 0.75 collimation and 375 ms rotation. Eur Heart J 26:1978–1986

    Article  PubMed  Google Scholar 

  12. Garcia MJ, Lessick J, Hoffmann MHK (2006) CATSCAN Study Investigators. Accuracy of 16-row multidetector computed tomography for the assessment of coronary artery stenosis. JAMA 296:403–411

    Article  PubMed  CAS  Google Scholar 

  13. Raff GL, Gallagher MJ, O’Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552–557

    Article  PubMed  Google Scholar 

  14. Leschka S, Alkadhi H, Plass A et al (2005) Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J 26:1482–1487

    Article  PubMed  Google Scholar 

  15. Leber AW, Knez A, von Ziegler F et al (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154

    Article  PubMed  Google Scholar 

  16. Pugliese F, Mollet NRA, Runza G, Van Mieghem C, Meijboom WB, Malagutti P, Baks T, Krestin GP, DeFeyter PJ, Cademartiri F (2006) Diagnostic accuracy of non-invasive 64-slice CT coronary angiography in patients with stable angina pectoris. Eur Radiol 16:575–582

    Article  PubMed  Google Scholar 

  17. Mollet NR, Cademartiri F, Van Mieghem CAG, Runza G, McFadden EP, Baks T, Serruys PW, Krestin GP, De Feyter PJ (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:2318–2323

    Article  PubMed  Google Scholar 

  18. Ropers D, Rixe J, Anders K, Küttner A, Baum U, Bautz W, Daniel WG, Achenbach S (2006) Usefulness of multidetector row spiral computed tomography with 64- x 0.6-mm collimation and 330-ms rotation for the noninvasive detection of significant coronary artery stenoses. Am J Cardiol 97:343–348

    Article  PubMed  Google Scholar 

  19. Fine JJ, Hopkins CB, Ruff N, Newton C (2006) Comparison of accuracy of 64-slice cardiovascular computed tomography with coronary angiography in patients with suspected coronary artery disease. Am J Cardiol 97:173–174

    Article  PubMed  Google Scholar 

  20. Austen WH, Edwards JE, Frye RL, Gensini GG, Gott VL, Griffith LSC, McGoon DC, Murphy M, Roe BB (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51:5–40

    Google Scholar 

  21. Hamon M, Biondi-Zoccai GGL, Malagutti P, Agostoni P, Morello R, Valgimigli M, Hamon M (2006) Diagnostic performance of multislice spiral computed tomography of coronary arteries as compared with conventional invasive coronary angiography: a meta-analysis. J Am Coll Cardiol 48:1896–1910

    Article  PubMed  Google Scholar 

  22. Agatston AS, Janowitz WR, Hildner FJ et al (1990) Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 15:827–832

    Article  PubMed  CAS  Google Scholar 

  23. Nikolaou K, Flohr T, Knez A, Carsten Rist C, Wintersperger B, Johnson T, Reiser MF, Becker CR (2004) Advances in cardiac CT imaging: 64-slice scanner. Int J Cardiovasc Imaging 20:535–540

    Article  PubMed  Google Scholar 

  24. Chartrand-Lefebvre C, Cadrin-Chênevert A, Bordeleau E, Ugolini P, Ouellet R, Sablayrolles J-L, Prenovault J Coronary CT angiography: overview of technical aspects, current concepts and perspective. Can Assoc Radiol J (in press)

  25. Giesler T, Baum U, Ropers D et al (2002) Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT: influence of heart rate on image quality and stenosis detection. AJR Am J Roentgenol 179:911–916

    PubMed  Google Scholar 

  26. Hong C, Becker CR, Huber A et al (2001) ECG-gated reconstructed multidetector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220:712–717

    Article  PubMed  CAS  Google Scholar 

  27. Nieman K, Rensing BJ, van Geuns RJM et al (2002) Non-invasive coronary angiography with multislice spiral computed tomography: impact of heart rate. Heart 88:470–474

    Article  PubMed  CAS  Google Scholar 

  28. Mollet NR, Cademartiri F, de Feyter PJ (2005) Non-invasive multislice CT coronary imaging. Heart 91:401–407

    Article  PubMed  Google Scholar 

  29. Heuschmid M, Kuettner A, Schroeder S et al (2005) ECG-gated 16-CT of the coronary arteries: assessment of image quality and accuracy in detecting stenoses. AJR Am J Roentgenol 184:1413–1419

    PubMed  Google Scholar 

  30. Cademartiri F, Mollet NR, Lemos PA et al (2005) Impact of coronary calcium score on diagnostic accuracy for the detection of significant coronary stenosis with multislice computed tomography angiography. Am J Cardiol 95:1225–1227

    Article  PubMed  CAS  Google Scholar 

  31. Cordeiro MAS, Lardo AC, Brito MSV, Neto MAR, Siqueira MHA, Parga JR, Avila LF, Ramires JAF, Lima JAC, Rochitte CE (2006) CT angiography in highly calcified arteries: 2D manual vs. modified automated 3D approach to identify coronary stenoses. Int J Cardiovasc Imaging 22:507–516

    Google Scholar 

  32. Achenbach S, Ropers D, Kuettner K, Flohr T, Ohnesorge B, Bruder H, Theessen H, Karakaya M, Daniel WG, Bautz W, Kalender WA, Anders K (2006) Contrast-enhanced coronary artery visualization by dual-source computed tomography—Initial experience. Eur J Radiol 57:331–335

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Carl Chartrand-Lefebvre.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Bordeleau, E., Lamonde, A., Prenovault, J. et al. Accuracy and rate of coronary artery segment visualization with CT angiography for the non-invasive detection of coronary artery stenoses. Int J Cardiovasc Imaging 23, 771–780 (2007). https://doi.org/10.1007/s10554-006-9198-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10554-006-9198-0

Keywords

Navigation