Abstract
The critical clinical and scientific insights achieved through knowledge of in vivo musculoskeletal soft tissue strains has motivated the development of relevant measurement techniques. This review provides a comprehensive summary of the key findings, limitations, and clinical impacts of these techniques to quantify musculoskeletal soft tissue strains during dynamic movements. Current technologies generally leverage three techniques to quantify in vivo strain patterns, including implantable strain sensors, virtual fibre elongation, and ultrasound. (1) Implantable strain sensors enable direct measurements of tissue strains with high accuracy and minimal artefact, but are highly invasive and current designs are not clinically viable. (2) The virtual fibre elongation method tracks the relative displacement of tissue attachments to measure strains in both deep and superficial tissues. However, the associated imaging techniques often require exposure to radiation, limit the activities that can be performed, and only quantify bone-to-bone tissue strains. (3) Ultrasound methods enable safe and non-invasive imaging of soft tissue deformation. However, ultrasound can only image superficial tissues, and measurements are confounded by out-of-plane tissue motion. Finally, all in vivo strain measurement methods are limited in their ability to establish the slack length of musculoskeletal soft tissue structures. Despite the many challenges and limitations of these measurement techniques, knowledge of in vivo soft tissue strain has led to improved clinical treatments for many musculoskeletal pathologies including anterior cruciate ligament reconstruction, Achilles tendon repair, and total knee replacement. This review provides a comprehensive understanding of these measurement techniques and identifies the key features of in vivo strain measurement that can facilitate innovative personalized sports medicine treatment.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
In addition to considerable pain and loss of function experienced by patients,4 musculoskeletal soft tissue injuries present a considerable financial burden to health care systems worldwide. Some 200’000 anterior cruciate ligament (ACL) injuries76,109 occur annually in the Unites States alone, and nearly half of them require reconstructive surgery.37 The socioeconomic costs of the resulting medical treatment have been estimated to be over one billion dollars annually.58,123 In Sweden, the average cost of surgical management of acute Achilles tendon (AT) ruptures in 2009-2010 was around $10,000.178 Moreover, revision rates following initial treatment for AT rupture (2%-8%)89,130 and ACL revisions (3-16%)122,176 suggest that more effective treatment offer considerable potential exists for not only reducing costs, but also improving patients’ functional outcome.
Transforming sports medicine therapies to address current treatment shortcomings requires a comprehensive understanding regarding the functional demands withstood by musculoskeletal soft tissues during activities of daily living in healthy, injury, and pathologic conditions. A detailed understanding of in vivo ligament and tendon strains is not only important for identifying injury mechanisms, informing surgical reconstructions and optimizing rehabilitation protocols, but also essential for defining physiologic loading conditions for in vitro experiments and tissue engineered constructs. Therefore, the measurement of in vivo musculoskeletal soft tissue strain patterns during functional movements is critical for both facilitating scientific discoveries and driving innovative clinical treatments.
The importance of soft tissue strain to biomechanical tissue function and adaptation has been well established. In 1847, Wertheim demonstrated that the stress-strain relationship in animal tissue does not follow the linear relationship dictated by Hooke’s law.177 In the 1960s, the introduction of Fung’s Law provided the contemporary understanding of soft tissue viscoelasticity, building a more complete concept of the tissue deformation response to loading.53 Shortly afterwards, it was discovered that both soft tissue strain magnitude and rate had a direct influence on the failure properties of ligaments124 and tendons.61 In 1992, it was discovered that the strain magnitude, not stress, experienced by muscle fibres was directly related to damage severity.51 However, despite the importance of soft tissue strain being recognized for decades, clinical translation of this knowledge has been hindered by the limited in vivo measurements of tissue strain magnitudes and rates during dynamic movements.
In clinical settings, improved knowledge of in vivo ligament strains67 has facilitated key innovations in reconstruction techniques and rehabilitation for ligament injuries.92 For example, in the 1990s, it was thought that the bone tunnel during ACL reconstruction surgeries should be positioned to achieve graft isometry throughout flexion.6,190 However, research into the role of loading on graft tissue health25,29 and the contribution of ACL strain to knee stability179 led to an evolution of the surgical technique such that current approaches aim to position the tunnels in the centre of the native ACL attachment footprints to better replicate the strain patterns experienced by the native tissue.141 With respect to rehabilitation, sensor measurements of in vivo ACL strain showed a significant increase when the knee moved to extended postures,20,62 which has informed activity selection for post-operative rehabilitation. However, innovation of clinical therapies for many other pathologies is hindered by the limited in vivo strain data available for many other ligaments and tendons, especially during functional and rehabilitative activities.
In vivo strain measurements also provide critical insights into muscle-tendon function by enabling the elongation of the muscle fibres, tendon, and aponeurosis to be independently quantified.33 Measurements of muscle fibre lengths and velocities during movement enable investigation into how individuals leverage optimal force-length and force-velocity muscle contraction conditions. Ultrasound studies indicate that the tendon can act as a buffer to overall muscle-tendon unit stretch, resulting in strains within the individual tissues that do not reflect the overall length change of the muscle-tendon unit.33 For example, during early stance in walking, the overall length of the triceps surae muscle-tendon unit increases substantially, but the lateral gastrocnemius fibres lengthen only to a small degree,115 while the medial gastrocnemius and soleus fibres even remain isometric.3 At push-off, all of the triceps surae muscles shorten, but at a much slower rate than the tendon and muscle-tendon unit as a whole. This demonstrates the importance of in vivo strain measurements to reveal the complex interactions within muscle-tendon units to conserve energy during muscle contraction, amplify power output, and absorb shocks during impact.142
An improved understanding of in vivo soft tissue strains will also provide important insights into how humans modulate their neuromuscular coordination to generate movement. When coupled with traditional motion analysis techniques, in vivo strain measurements could enable an improved understanding of how the body leverages elastic energy storage in tendons.191 Furthermore, dynamic strain measurements of all the soft tissues crossing a joint may provide key insights towards resolving muscle redundancy.63 This problem seeks to determine how the externally measurable joint torques during movement are distributed to the redundant musculoskeletal system, resulting in one of the most famous unsolved problems in biomechanics.137
The critical need for in vivo quantification of musculoskeletal tissue strain for both clinical and basic science applications has inspired the development of three main measurement techniques: (1) implantable strain sensors, (2) virtual fibre elongation, and (3) ultrasound. Implantable strain sensors have largely been applied to measure in vivo ACL strains,17,44,140 but such technologies have mostly been implemented during controlled movements and in sterile surgical environments. The virtual fibre elongation method uses bone position from image-based methods including fluoroscopy,87,166,181 ultrasound,134 open magnetic resonance imaging (open-MRI),64,91 or computed tomography (CT)60,161 to derive in vivo ligament and tendon strains. Ultrasound technique enables dynamic non-invasive planar imaging of soft tissue structures, and has mostly been applied to measure muscle33,105 and tendon126,159 deformation. Each technique presents different advantages and limitations regarding accuracy, invasiveness, safety, and activities that can be measured. In addition, previous studies have chosen different approaches to define the fibre reference length.19,69,71,75 As a result, the reported strain data may not represent the real soft tissue strain, and thus understanding the true loading condition within the tissue remains challenging. Therefore, the abilities of these measurement techniques to determine the slack length of soft tissue should be fully reviewed and discussed. The technologies involved in in vivo strain measurement have been previously reviewed,16,44,140,144 however there has not yet been a comprehensive summary of the key findings and clinical impacts of each methodology. The purpose of this study is therefore to provide an overview of in vivo tendon and ligament strain measurement methodologies and their key findings, clinical implications, and limitations, as well as to identify technological developments that may lead to clinical and basic science advances in sports medicine.
Materials and Methods
Literature Search and Selection
Articles were searched using the keywords “sensor*”, “gauge*”, and “transduc*”, “imag*”, “ultrasound*”, “fluoroscop*” as well as “tendon*”, “ligament*, “muscle*”, “length change*”, “strain*” and “elongat*” in the Pubmed and EMBASE databases. A total of 9842 articles were found. The titles and abstracts of all articles were firstly screened by one author to exclude clearly irrelevant articles. Two authors then reviewed the full texts of the remaining articles to identify articles discussed in this review. As this is a narrative review, the two authors also searched the reference lists of these articles to identify any other relevant articles that had not been found through the above search. One experienced author discussed these candidate articles with the two authors to make a final decision on their inclusion. Finally, 151 articles were includedor reviewing.
Strain Calculation
Strain (ε) is traditionally defined using Eq. (1) in the biomechanics literature:
where \(l\) is the tissue length and \(l_{0}\) is the slack length or reference length. Slack length is the length at which a tissue begins exhibiting force when it is stretched. The ability of each measurement technique to quantify tissue slack length is detailed in the following sections. In many studies, the true slack length was not measured and instead a reference length, the length of the tissue in a defined body posture is used to calculate strain.71,108 Thus, comparing strains between studies requires careful consideration of the \(l_{0}\). definition.
Implantable Strain Sensors
Implantable sensors have been employed to directly measure in vivo ligament strain patterns in humans. Similar sensors have been applied to measure human tendon forces. Typically, these sensors leverage similar working principles, where mechanical loading or deformation of the tissue induces changes in their electric signals (voltage, resistance, or capacitance). Buckle transducers, liquid metal strain gauges, fibre optic transducers, and sonomicrometry crystals have all been used for in vivo force and strain measurements in animal studies to assess muscle, tendon, and ligament function during dynamic activities.16,44 For example, liquid metal strain gauges have been applied to measure strain patterns in rabbit ATs and in horse suspensory ligaments.26,84 In addition, buckle transducers and fibre optic transducers have been applied to measure AT force in humans during walking and jumping.43,52
While many aforementioned sensor designs have been demonstrated for force measurements in humans and strain measurements in animals, only Hall Effect Strain Transducers (HESTs) and Differential Variable Reluctance Transducers (DVRTs) have been used to directly measure in vivo strains in human musculoskeletal tissue.44 Therefore, the following discussion focuses on the properties, applications, and limitations of these sensors.
Measurement Technology and Experimental Methods
To detect displacement, HEST sensors measure the change in voltage resulting from relative movement between a magnet and a Hall-effect magnetic sensor, whereas DVRTs measure the movement of a magnetic core within two coil windings by measuring the resulting change in magnetic reluctance.17 The parameters of these two sensors are displayed in Fig. 1a.17,44 Beynnon and co-workers arthroscopically implanted a HEST sensor onto the ACL of subjects under local anaesthesia,21 mounting the sensor by pressing on two fixation barbs (Fig. 1b). ACL strains were then quantified during limited functional movements before removal of the sensor. The measurement procedures of DVRT sensors are similar to the HEST sensors (Fig. 1c).46
Slack Length
To assess in vivo ACL slack length, Beynnon and co-workers recorded the displacement of a HEST sensor fixed on the ACL while applying antero-posterior shear loads to the tibia.19,45 They found an “inflection point” on the plot of the sensor length against shear load (Fig. 1d), corresponding to the length where the ACL engages. The corresponding sensor length at this point was considered to be the reference length used to calculate strain. Hence, any subsequent relative change of the sensor displacement from this length was considered to directly reflect loading within the ACL.19,72 This method was later validated by Fleming and co-workers through concurrently implementing a force probe to measure ACL load on cadaveric knees.47
Key findings
In vivo strain sensor measurements have provided critical insights into rehabilitation protocols for ACL reconstruction by providing direct evidence for activity selection that ensure tissue strains are under damaging magnitudes,62 but high enough to facilitate healing.136 Specifically, squatting was long thought to be a safe activity for early rehabilitation of ACL reconstruction. However, in vivo ACL strain measurements using DVRT sensors have revealed that peak strains induced by squatting (3.6%)20 are higher than lunges (1.8-2.0%), sit-to-stand (2.8%), step-up/down (2.5-2.6%),62 and cycling activities (1.7%).46 These findings indicate that squatting exposes the ACL to greater loads than open-chain movements17 and should therefore be used cautiously during rehabilitation. In vivo sensor measurements have also revealed key insights into the role of muscle activation in both loading and protecting the ACL. The measurements indicate that isometric quadriceps contractions increase ACL strain at flexion angles of less than 50°,18,19,72 isometric gastrocnemius contractions increase ACL strain over all tested flexion angles, and that hamstrings contraction reduce ACL strain at flexion angles greater than 5°.17,48 These measurements have provided fundamental knowledge that can inform targeted muscle strengthening and neuromuscular coordination training to protect the ACL.
Advantages and Limitations
There are several unique advantages to the HEST and DVRT implantable sensor approaches for strain measurement. Firstly, it provides a direct strain measurement and thus the data is highly representative of the local tissue strain environments. Secondly, both HESTs and DVRTs are highly sensitive and able to detect minute strains in the musculoskeletal tissues.17 Finally, sensors provide high sampling frequencies for strain and strain rate measurements during dynamic activities.
Despite the insightful findings, however, the HEST and DVRT sensors have several major limitations:
-
(1)
Invasive nature: Sensor implantation is a highly invasive procedure, resulting in ethical difficulties to justify their use in healthy subjects. For ACL strain measurements, implantations have therefore typically been undertaken together with a pre-planned arthroscopic surgery to perform minor repairs of neighbouring tissues.19 Consequently, such patients may have deformities or injuries that may alter the tibiofemoral contact mechanics99 and potentially alter the measured soft tissue strain patterns.28,193
-
(2)
Tissue impingement: Current sensors can only be implanted in locations where there is sufficient room for the sensor and straight-forward surgical assess to the tissue. For example, the strain sensors have always been attached to the anteromedial bundle of the ACL because it is more accessible from the anterolateral surgical portal than the posterolateral bundle of the ACL. While HEST and DVRT sensors are small in size, their solid constitution can result in impingement with surrounding musculoskeletal tissues, and therefore impede certain postures during testing. Importantly, sensor disturbance with the femoral notch is unavoidable at full extension of the knee during ACL strain measurements,44 which hinders the assessment of ACL strain patterns throughout activities involving full extension or even hyperextension of the knee.44
-
(3)
Limited implantation duration: Previous experiments have been limited to a few hours because both HEST and DVRT sensors require a cable that crosses the skin to transmit the measurement signal, and are therefore not long-term biocompatible. These limitations have prevented subjects from performing many functional activities of daily living.
-
(4)
Sensor-body interference: Sensor cable migration induced by subject body movement can cause significant artefacts in the sensor signal. Sensor cables should be stabilized and well protected during experiments.
-
(5)
Sensor implantation alignment: Measurements are sensitive to implantation orientation of the sensor relative to the tissue. To assess tensile ligament strains, the measurement axis of the sensor must be aligned with the ligament fibres during implantation in order to record uniaxial strain.
Future Work
Recent advances in stretchable electronics may enable clinical translation of strain sensors towards intraoperative guidance and long-term assessment of healing. To facilitate these clinical applications and minimize risk to research subjects, these next generation strain sensors should incorporate three critical features: flexibility, wireless data transmission, and long-term biocompatibility. Thin and soft membrane-like strain sensors can be fabricated by embedding conductive materials in stretchable materials to reduce the risk of tissue impingement.7 Initial measurements include skin surface assessments of joint rotation 186 and muscle movement of the trachea,77 as well as ex vivo strains in the anterolateral ligament of a cadaveric knee (Fig. 2).192 Flexible electronic sensors can be fabricated using biocompatible materials (PDMS, gold, titanium, etc.), and several have demonstrated good biocompatibility through histologic examination.106,174 Wireless signal transmission with a passive sensor (i.e. no battery or chips) can be achieved using radio-frequency identification technology,73 enabling a resonant strain sensor containing a stretchable capacitor and a coil inductor whose resonance frequency can be wirelessly measured by a readout system to be developed (Fig. 2).158 However, improvement of sensor fixation,19,143 as well as long-term stability and quality of flexible sensor signals are still necessary for clinical application. The development of biodegradable strain sensors that are resorbed after a given duration is critical to avoid a second surgery for sensor removal.24 Such sensors would be ideal for clinical translation into soft tissue repair surgeries, especially for long-term assessment of tissue healing and data driven physical-therapy.
Virtual Fibre Elongation
Initial imaging assessments of ligament strain were achieved using open-MRI9,64 and CT scans55,85,138,188 to measure ligament length on the 2D slice images that best displayed the structure. However, substantial inaccuracies were clearly present in such 2D measurements due to out-of-plane errors. The virtual fibre elongation method overcomes this limitation by leveraging imaging data to reconstruct the relative 3D poses of the bones to determine the relative displacements of soft tissue attachment sites to non-invasively quantify tendon and ligament length change patterns.82,83,183 For investigating dynamic activities, fluoroscopy is the state-of-art imaging modality, as CT and MRI are limited by their long-capture times and small field of view (FOV). Thus the following discussion will primarily review fluoroscopic imaging studies.67
Measurement Technology and Experimental Methods
Typically, the joint of interest is statically imaged using MR171 or CT,161 then segmentation is performed to construct 3D surface models of the bones. The attachment footprints of the ligament bundles are identified on the bone models from the images (MR) or relative to landmarks (CT). Subsequently, the subject performs a functional movement while the joint is imaged using single or dual-plane fluoroscopes152,161 to acquire a time series of radiographic images of the bones. The 3D bone poses are calculated to match to the images using a manual or (semi-)automated 2D-3D registration software.27,57,167 This process is repeated for each radiographic image to quantify the joint kinematics throughout the movement (Fig. 3).
The length change of a ligament during a movement is assessed by tracking the relative displacements of the attachment points. For ligaments with small attachment footprints, a virtual fibre connecting the centroids of ligament insertion sites is defined as the ligament “longitudinal axis”.35 For larger attachments, multiple virtual fibres can be defined.131,173 The fibre path can either be defined as the straight line connection between attachment points,131 or wrapping surfaces can be defined to represent the real curvilinear path of the ligament and prevent penetration of the bone geometries.71 The length of the virtual fibre at each joint pose is compared to quantify the relative length change throughout the activity. As the virtual fibre is a geometric path, the orientation of the ligament force can also be quantified as the angle between the projection of the virtual fibre in each plane and the relevant axes in the predefined coordinate system of the joint.181,187 This enables insights into ligament function such as the restraint it can provide against external loads and graft bending in ligament reconstruction.162,163
Slack Length
The virtual fibre elongation method does not provide any inherent means to determine the slack length of a tissue. Accordingly, most studies using this method define the reference length as the length of a tissue in a reference joint posture, to normalize length change measurements and provide an estimate of the elongation patterns. For the knee ligaments, the length at full knee extension71,166 or heel strike of walking108 is commonly used as the reference length.
Key Findings
The primary advantage of the virtual fibre elongation method is that it enables the non-invasive assessment of multiple ligaments in a joint during dynamic movements. The resulting comprehensive understanding of ligament contributions to functional joint mechanics has informed many surgical and rehabilitative sports medicine procedures. Using the virtual fibre elongation method, ACL length was shown to decrease by 28% from full knee extension to 135° of knee flexion during single leg lunge,87 while PCL length increases by 35%.131 Ankle plantarflexion and supination were found to increase the length of the anterior talofibular ligament but decrease the length of the calcaneofibular ligament.11 Studying different ligaments in the same joint can also reveal how the ligaments interact during an activity. A comparison of intact and ACL-deficient knees confirmed the importance of the ACL to knee stability by showing that ACL injury caused significant MCL extension and LCL shortening during lunges.172
The virtual fibre elongation method also provides insights into the strain distribution within tissues by using multiple fibres. This techniques has revealed that the anterior, middle, and posterior bundles of the superficial MCL exhibit different length change patterns (19%, 0%, and -16%) from full knee extension to 145° knee flexion during lunges.71 Similarly, the length of the anterior and posterior bundles of the LCL exhibited a 6-7% increase and a 10-16% decrease, respectively.71,132 Running induced significantly greater elongation of the ACL anteromedial bundle compared to the posterolateral bundle,121 whereas, walking showed more uniform elongation within the ACL.181 This approach also revealed that reconstructing the medial patellofemoral ligament (MPFL) to a location posterior and proximal to its anatomic femoral attachment results in the most isometric graft behaviour during lunging.90 Similarly, at the shoulder, virtual fibre elongation techniques have allowed an insight into otherwise difficult to access dynamic musculoskeletal interactions. Maximal external rotation at 90° abduction significantly elongates the anterior bundle of the inferior glenohumeral ligaments while only slightly elongating the posterior bundle. Conversely, maximum internal shoulder rotation at 90° abduction significantly elongates the posterior bundle while only slightly elongating the anterior bundle.113 Thus, the virtual fibre elongation method can reveal different function within regions of a single ligament, and consequently inform surgical reconstruction techniques and selective ligament release in joint revision surgery.
The virtual fibre elongation method can also evaluate the outcomes of surgical procedures via the length change patterns of ligaments or grafts during functional movements. A possible mechanism for poor outcomes in PCL-retaining total knee arthroplasty (TKA) patients was identified because PCL elongation showed a significant increase compared to healthy knees at flexion angles of over 75° during lunges.189 Cruciate-retaining TKA was also found to have a significant effect (mostly increasing) on the elongation patterns of the knee collateral ligaments during lunging by comparing the results of cruciate sacrificing TKA and healthy knees.71,133 A new study investigated the effect of knee TKA with ultra-congruent implants on the length change patterns of MCL and LCL during a variety of daily functional movements such as walking and squatting.68 In the TKA knees, while the LCL consistently slackened with increased knee flexion in all movements, the MCL showed regional variation in the length changes. The anterior, middle, posterior MCL fibres lengthened, remained isometric, and shortened during knee flexions, respectively. This confirmed the pervious observed distinct regional behaviour of the MCL71,108 and indicated that the location of partial MCL releases performed intraoperatively to balance the knee will have important consequences on post-operative stability. For ACL reconstruction, anatomic ACL graft placement better restores the length and length change patterns of the native ACL compared to an anteroproximal graft placement.1 A similar study found that compared to non-anatomic ACL reconstructions, anatomic ACL reconstruction results in PCL length change patterns (both bundles) that are closer to native patterns.161 Therefore, the virtual elongation method provides a unique opportunity to non-invasively evaluate native and graft soft tissue strains, hence enabling quantitative evaluation of surgical techniques.
Advantages and Limitations
The virtual fibre elongation method leverages state-of-the-art dynamic imaging technologies to non-invasively evaluate soft tissue strains throughout functional activities at high frequencies (> 200 Hz). It can quantify regional tissue strains by using multiple fibres to represent the same ligament (Fig. 4). Furthermore, as the fibre attachment locations are manually selected, uncertainty quantification and sensitivity studies can be readily performed.68 Additionally, the measurement accuracy is not affected by whether the tissue is deep or superficial. Finally, the non-invasive nature of the method allows comparisons of injured or reconstructed tissue function against the healthy contralateral limb.
Despite the many advantages of the virtual elongation method, there are also several limitations:
-
(1)
Accuracy: Representing complex 3D tissue geometries with virtual fibre paths introduces several sources of error. Significant uncertainty in reported ligament lengths originates from both the identification of the tissue attachment footprints and virtual fibre insertion sites within the footprint.90 The determination of ligament insertion sites on MR images can show large variations especially in some ligaments like the sMCL, which has large attachment areas (nearly 400 mm2 on its tibial proximal attachment).98 However, for other ligaments such as the ACL, footprints can be accurately identified from MRI images.10 The simplification of ligament pathways to straight virtual fibre paths also affects the accuracy, as it neglects any variations in pathway length due to wrapping around skeletal structures. Furthermore, virtual fibre ligament elongations are also sensitive to the measurement accuracy of the bone kinematics. With longer fluoroscope shuttering times (fluoroscopic exposure time per frame), the X-ray image becomes blurred, leading to errors in the accuracy of the 2D–3D registration. As an example, Ellingson and co-workers reported that the errors in knee joint kinematics could be as large as 2.0° and 1.6 mm with an exposure time of 16 ms,36 which would propagate into errors in the assessment of virtual fibre elongation patterns. Finally, virtual fibres only provide bone-to-bone elongation measurements, which makes them well suited for ligaments, but they cannot assess muscle fibre lengths and tendon lengths independently.
-
(2)
Difficult to determine slack length: The resting length of ligaments cannot be measured using fluoroscopy. Therefore, while relative elongation patterns can be reported, true tissue strain patterns or loading conditions cannot be quantified. To circumvent this issue, many researchers define the reference length of a structure to be its length at e.g. full knee extension,35 or at an instant of the gait cycle e.g. heel strike.70 However, the lack of a consistent choice of reference length limits inter-study comparisons, and an appropriate solution should be agreed across the field.
-
(3)
Measurement frequency and radiation exposure: Fluoroscopes can generate either continuous 57 or pulsed 107 X-ray imaging. Continuous radiation modes enable higher measurement frequencies (speeds are then limited by the camera and fluoroscopic imaging formation), but also leads to greater radiation exposure to the subject, which constrains the measurement time and consequently the number of repetitions. Pulsed X-ray reduces the radiation dosage and therefore extends the measurement time. However, fluoroscopes with pulsed X-ray have limited measurement frequencies (often 25–30 Hz), which is insufficient to track highly dynamic movements such as impact situations.
-
(4)
Limited FOV: Traditional fluoroscopy systems are stationary with fields of view of 20-30 cm diameter. Thus, for many joints it is impossible to image complete activity cycles such as walking, running, or stair ascent/descent. Therefore, movements such as single legged lunge are often studied as the knee remains within the FOV.90,173 To track walking or running gait activities, the subject has to walk on a treadmill with only portions of the gait cycle such as stance phase being imaged.108,181
Future Work
The virtual fibre elongation method has demonstrated great potential for assessing the function of ligaments in healthy74,166,181 and surgically reconstructed states133,161,165 in dynamic activities such as running161 and landing.168 However, the space constraints and stationary field of view of the imaging systems limit the movements that can be studied. Recently, mobile fluoroscopy systems have been developed to track the horizontal and vertical displacements of the joint during full body locomotion. Two of these systems have been developed to image lower limb joints throughout complete activity cycles.59,107 However, such systems may affect the subject’s gait performance and introduce additional errors due to vibrations.66
Currently, the virtual fibre elongation method is too time consuming for clinical translation due to the segmentation of the joint anatomy, definition of ligament attachment points, and 2D/3D images registration steps. Introducing artificial intelligence techniques to automate these processes could eliminate the time-consuming nature of this method, and therefore make it clinically viable. For example, a segmentation method combining image registration and machine learning has been developed to automatically segment the region of interest on ultrasound 185 and MRI images.184 It could be envisaged that such technology will help improve the automation of the virtual fibre elongation method. Moreover, instead of using virtual fibres to model the ligament, future study could develop the finite element model of the ligament and apply the joint kinematic data as boundary conditions to simulate the 3D strain-stress enviroment within the ligament.117 Finally, roentgen stereophotogrammetric analysis (RSA) can extend the virtual fibre elongation method to enable tendon strain measurement, especially to evaluate the recovery of the mechanical properties and function of reconstructed tissues.2,86,147,148 RSA requires that titanium beads are implanted in bones or soft tissues through open surgery12 or injection,149 and then uses dynamic imaging to capture the distance between beads throughout an activity to quantify tissue length changes. Tashman and co-workers have developed a dynamic RSA system by combining X-ray radiographs and high-speed digital imaging to achieve high measurement frequency (250 Hz) and accuracy (± 0.1 mm),164 which could be used in measuring dynamic activities. However, the invasive nature of bead implantation limits its application on human subjects.
Ultrasound
Ultrasound is a non-invasive, radiation-free, and cost-effective technique to visualise the internal architecture of musculoskeletal soft tissues. Clinically, ultrasound is commonly used to identify pathological tendon abnormalities.114 In research settings, ultrasound has been leveraged to quantify muscle, tendon, and ligament deformation during dynamic activities. Numerous ultrasound studies quantified strains of large flat tendons such as the AT during walking 50,88 and running,39,102 however, very few ultrasound studies focused on ligaments.155 Ultrasound also enables muscle fibre lengths and pennation angle to be dynamically assessed.33,97
Typical ultrasound systems consist of a hand-held transducer connected by flexible wire to a cart that houses the electronics for signal generation and processing, as well as a screen for visualization. The working principle behind ultrasound imaging relies on propagating ultrasonic waves (sound waves with frequencies > 20,000 Hz) into the tissue of interest and recording their reflection.139,146 The ultrasonic waves reflect off the tissue according to the local properties, and the echoes are recorded and used to generate an image. Traditional ultrasound systems generate planar 2D images and can reach tissues up to 17 cm beneath the skin.38
Two primary techniques have been developed to measure soft tissue strain during dynamic movements using ultrasound, myotendinous junction (MJ) tracking (Fig. 5) and speckle tracking (Fig. 5). Freehand 3D ultrasound has also been applied to measure Achilles tendon strain, but only by imaging multiple static postures, and thus this review focuses on MJ and speckle tracking techniques.
Measurement Technology and Experimental Methods
For both strain measurement techniques, the ultrasound transducers must be fixed to the body during dynamic movement to image a tissue. Typically, the ultrasound transducer is housed in a foam orthotic and strapped to the body using athletic tape.50 In many experiments, a treadmill is used so that the ultrasound cart can remain stationary. Commonly, markers are attached to the transducer and the ultrasound system is synchronized with a motion analysis system so that the transformation between the image field of view and other anatomic landmarks can be calculated (Fig. 5).
The MJ tracking technique enables muscle and tendon lengths to be quantified by tracking anatomical landmarks in each ultrasound image of a series collected throughout an activity. For example, the elongation of the free AT can be measured by tracking the displacement between the calcaneus attachment and the soleus-AT myotendinous junction (sATMJ).94 An ultrasound transducer with a wide FOV (100 mm) can image the entire free AT from the calcaneus to sATMJ94 (Fig. 5). For many applications and ultrasound systems, however, the tissue length exceeds the transducer’s FOV and thus only one attachment site can be tracked. Synchronizing the ultrasound system with an optic tracking system provides another means to track the length change of a long tendon during movement.40,104,135 To measure the AT, a reflective marker is placed on the calcaneus attachment to track its position. An ultrasound transducer equipped with reflective markers is fixed on the shank to image the position of the sATMJ, and then transform it into the global coordinate system. The free AT length is then defined as the distance between the positions of the calcaneus and the sATMJ.
Speckle tracking provides an alternative method to quantify regional tissue strain using ultrasound. A speckle is an interference pattern with bright and dark spots in B-mode or radiofrequency (RF) ultrasound images.156 The speckle pattern remains constant when the tissue structure is static and thus is thought to be representative of the tissue structure rather than measurement noise. Therefore, when the tissue microstructure is deformed, the speckle pattern changes accordingly.22 After the B-mode or RF ultrasound images of the tendon are obtained throughout an activity, a region of interest (ROI) containing the tendon structure is manually selected on one image.95 Then, automated speckle tracking algorithms are used to calculate the frame-to-frame speckle displacement within the ROI, which is then converted into tendon elongations.50,134
Slack Length
Ultrasound provides two opportunities to assess the slack length of a tissue. Firstly, a tissue can be imaged at multiple joint postures until buckling can be identified in the B-mode images. This has been demonstrated on the patellar tendon.151,154 Secondly, a novel ultrasound imaging technique called shear wave elastography, which quantifies the speed of shear wave propagation in a tissue, has been applied to assess tissue slack length. Shear wave speed has been shown to be related to tissue stress,23,111 and thus can be applied to establish when the tissue goes slack if a corresponding joint is manipulated through its range of motion. This technique has been applied to assess the slack length of the AT75 and the ankle angle when the individual triceps surae muscles are slack.65
Key Findings
Ultrasound based strain measurements have provided many important insights into the dynamic function of the AT. For example, ultrasound was used to quantify in vivo AT strains during dynamic activities such as running (3.5-5.8%), walking (4.6%), and one-legged hopping (8-8.3%).40,41,104,105 These functional AT strain data provide some important implications for AT injury. Firstly, the in vivo AT strains during one-legged hopping are surprisingly close to the AT failure strains found in in vitro tendon tensile tests.15,104 As the AT remained healthy after hopping tests, this finding demonstrates the difficulty of comparing in vivo and in vitro strain measurements and perhaps inconsistency among the methods. Secondly, the in vivo strain data from ultrasound measurements can serve as a variable to assess other parameters associated with AT injury, such as the mechanical “safe factor” 0 and “core temperature” 9 of the AT. Franz and co-workers measured the regional deformations of AT during walking,50 and showed that the aggregate AT exhibited nearly twice as much elongation as the free AT. Moreover, the superficial AT exhibited larger elongations than the deep AT and only the superficial AT elongation was found to increase at higher walking speeds, indicating non-uniform regional strain patterns in the AT. Cronin and co-workers compared the length changes of Achilles tendons during walking between healthy and diabetic subjects.34 The length changes of Achilles tendons were smaller in diabetic subjects than healthy patients but the length changes of muscle-tendon units were similar. These findings indicate that diabetes may decrease the elastic energy storage of Achilles tendons and increase muscle work, resulting in decreased walking efficiency. Finally, ultrasound has provided insights to AT healing and protection. For example, AT strain during running with a 18 mm heel lift was significantly smaller than barefoot, suggesting that such a heel lift could help to reduce the AT strains during AT rehabilitation training.39
Ultrasound evaluations have also contributed towards understanding the function of human triceps surae muscles during walking and running.29,30,31,32,33,101 In these studies, either a single ultrasound transducer was positioned over the medial gastrocnemius (MG) so that the soleus was also visible,32 or two transducers were placed over the MG and soleus.80 The MG and soleus fascicle lengths and pennation angles were then measured using an automated tracking algorithm, and motion-capture derived knee and ankle joint kinematics were further used to estimate MG and soleus muscle-tendon unit lengths.32,102,116 These studies found different length change patterns and magnitudes between the MG and soleus fascicles during the stance phase of gait, indicating the different functional roles of the two muscles.80 Relevant studies have also determined the differences in triceps surae length changes and behavior between walking and running,81,102 between forefoot and rearfoot running,160 and at different running speeds.79 Investigating the length change patterns in both muscle fascicle and muscle-tendon unit also showed a decreased energy storing role of the AT in race-walking compared to running.32 The effect of age on gastrocnemius muscle-tendon behavior during walking has also been studied by ultrasound,116 and it was found that tendon lengthening is greater and muscle fascicle lengthening is smaller in older subjects compared to young subjects. In addition, ultrasound has been used to characterize the effect body mass on gastrocnemius medialis fascicle behavior during stair ascent.157 By calculating the difference between the muscle tendon unit length change and the fascicle length change, it was found that additional body mass could cause substantially greater stretching on the tendon. This finding may be meaningful in evaluating effect of obesity on pathological loading in triceps surae muscle tendon unit.
The application of ultrasound to quantify ligament strain is limited to a few studies investigating superficial ligaments during static movements, such as the dorsal lisfranc ligament,56,110,145 pisohamate ligament and pisometacarpal ligament.119 Using speckle-tracking, 5-7% strains were observed in the coracoacromial ligament during dynamic shoulder rotations such as forward flexion, horizontal abduction, and internal rotation at 90° abduction.134 The strain patterns and displacements of the coracoacromial ligament indicated contact with the rotator cuff in these frequently used shoulder movements. This indicates a possible mechanism for rotator cuff pathologies, and provides guidelines for rehabilitation protocols as certain movements such as forward flexion, horizontal abduction, and internal rotation at 90° abduction were found to induce excessive tissue strains.
Advantages and Limitations
Ultrasound provides several key advantages over other methods for assessing musculoskeletal soft tissue strain. Firstly, it is non-invasive, relatively inexpensive, and does not expose the subjects to radiation. Secondly, ultrasound can dynamically image soft tissues, which enables regional deformation within the tissue to be quantified. Finally, it enables the strains of muscle fibres and tendon to be independently measured as it can differentiate muscle-tendon interfaces.
Despite these advantages, ultrasound methods also suffer from several limitations:
-
(1)
Controversial reliability: Establishing good reliability of ultrasound tissue length measurement is fundamental to highly accurate measurements of tissue strain.114,169 However, the reliability of ultrasound measurement of static tissue geometry has been controversially reported,5,14,13,54,78,120,150 and these measurements are easier to perform than strain measurement during dynamic activities. Furthermore, transducer orientation and position have a significant effect on ultrasound measurements,93,96 and rigid fixation of the ultrasound transducer over the tissue of interest is difficult to achieve throughout dynamic activities. Some confidence can be gained from the good intra-class correlation coefficient (ICC) of ultrasound in measuring AT strains (0.72–0.86) and coefficient of multiple correlation of measuring PT strains (0.77–0.82),22,150 but it should be noted that these studies only measured tendon strain during static movements.
-
(2)
Challenging to measure ligament strains: Ultrasound is mostly used clinically to diagnose ligament injuries,100,153,175,180 but rarely to measure ligament strains. This is largely due to the difficulty of obtaining clear images of complete ligament structures as they are normally small and located beneath neighboring tissues. In addition, ligaments commonly are directly adjacent to bone and other hypoechoic tissues, which can lead to artefacts and reduced ligament visibility in ultrasound images. Furthermore, in traditional ultrasound images, the ligament appears hypoechoic at rest because of anisotropy artefacts and hyperechoic under loading due to tightening of the ligament microstructure.30,119 This change in appearance between loading conditions can help differentiate the ligament from other tissues but causes difficulties when trying to quantify ligament strain patterns throughout an entire movement.
-
(3)
Confounded by out-of-plane motion: The complexity of measuring the deformation of 3D soft tissues in planar 2D ultrasound images is another key limitation of ultrasound. MJ tracking and speckle tracking both require tracking landmarks or pixels in the 2D ultrasound image. Out-of-plane strains resulting from tissues bulging, rotating, or twisting may be missed or misinterpreted as in plane translation. Freehand 3D ultrasound addresses this limitation, but cannot be applied to dynamic movements because the transducer must be swept over the length of the tissue.42,128
-
(4)
Transducer fixation over tissue: For ultrasound measurement of dynamic activities, the transducer must be fixed over the tissue of interest as any motion of the transducer relative to the body can cause measurement error. This requires that the transducer is strapped to the body which may alter natural movement patterns. Finally, dynamic measurements are typically limited to a treadmill to avoid moving the ultrasound cart over-ground with the subject.
Future work
Ultrasound methods are currently the most clinically viable for assessing musculoskeletal soft tissue strains due to their non-invasive, non-radiative, and relatively low-cost nature. However, their 2D nature limits the accuracy and repeatability of this method. Freehand 3D ultrasound scanning has been developed to quantify the 3D AT deformation in vivo,42,125 including changes in AT length, AT width, thickness, cross-sectional area, and volume.125,127 This technique relies on a conventional ultrasound machine coupled with a transducer instrumented with reflective markers and an optical tracking system. The operator performs transverse ultrasound scans from the calcaneus along the AT to the gastrocnemius muscle,103,128 during which the spatial position and orientation of the ultrasound transducer are recorded. Then, the 2D ultrasound images (between-frame interval of 0.1-0.5 mm) are transformed into the global coordinate system to create a 3D AT reconstruction.129 Highly precise 2D-to-3D transformation of the image (error under 1 mm) can be achieved after temporal and spatial calibration of the ultrasound transducer.128,129 Measurements of 3D AT deformations then require a series of digital processing procedures such as segmenting and rendering AT cross-sections, reconstructing 3D AT volumes, and defining measurement sites.128 The applicability of 3D freehand ultrasound to dynamic movements, however, remains to be established (Fig. 5).
Summary and Perspectives
In vivo strain measurements have provided important insights into the mechanical function of musculoskeletal soft tissues, resulting in improved rehabilitative and surgical treatments in sports medicine. This review has presented three main state-of-the-art methodologies that enable quantification of soft tissue strain patterns during dynamic movements: implantable strain sensors, virtual fibre elongation, and ultrasound. Each methodology has provided key measurements that altered treatments and rehabilitation for musculoskeletal soft tissues. However, further innovations in the measurement technologies and their clinical implementation will be necessary to facilitate a data driven revolution to personalize sports medicine treatments.
Each strain measurement technique has its own virtues and limitations in terms of safety, application, and accuracy (Table 1). Implantable sensors enable strains in deep tissues to be measured with high accuracy and frequency. However, the sensor designs that have been implanted in humans are highly invasive, can impinge on neighbouring tissues, limit the movements that can be performed, and require a data transmission wire to cross the skin. The virtual fibre elongation method enables elongation measurements of multiple deep and superficial tissues during highly dynamic movements. However, it can expose the subjects to radiation depending on the imaging method, and can only quantify bone-to-bone tissue length changes and thus cannot differentiate between muscle and tendon strain. Ultrasound methods provide unique advantages because they are non-invasive, do not expose subjects to radiation, and provide direct imaging of the soft tissue structures. Ultrasound also provides the unique possibility to measure regional tissue strains compared to implantable sensors and the virtual fibre elongation methods which can only assess point-to-point strains. However, dynamic ultrasound has limited accuracy due to out-of-plane motion, and is largely only applicable to measure superficial muscles and tendons. Thus, researchers and clinicians must carefully consider the advantages and limitations of each measurement technique when planning new studies and interpreting results.
Assessment of soft tissue slack length remains significantly challenging for each strain measurement technique. The inflection point method has been well validated for determining ACL slack length with implantable strain sensors, but it is uncertain whether similar experiments can be replicated on other soft tissues. The virtual fibre method currently has no reported method for measuring slack length, but potentially could be coupled with ultrasound. Ultrasound can provide measurements of slack length through visualizing tissue buckling or using shear wave propagation,112 but these may not be applicable to all tissues. Future studies should at the very least report the method used to determine the reference length and provide quantitative values of the tissue lengths.
A key limitation of this study is that it only focused on measurement techniques that have been applied to measure in vivo human musculoskeletal soft tissue strains during functional movements. However, several techniques have been used to measure in vivo strains in animal tissue (sonomicrometry,118 liquid metal strain gauges26), in vivo strains in human bone,49 and in vivo loading in human tendons (fibre optic sensors,43 buckle transducers,52 shear wave tensiometers111). Furthermore, novel techniques such as near infrared spectroscopy170 show promise to quantify musculoskeletal soft tissue mechanical properties. The working principles behind any of these techniques may lead to the needed breakthrough that overcomes the limitations of current strain measurement methods.
In conclusion, we have found in vivo strain measurements have improved clinical treatments for many musculoskeletal pathologies including ACL reconstruction, AT repair, and TKA. However, a new generation of clinically viable technologies are needed to facilitate a data driven progression of personalized surgical and rehabilitative sports medicine treatments.
Abbreviations
- ACL:
-
Anterior cruciate ligament
- AT:
-
Achilles tendon
- MRI:
-
Magnetic resonance imaging
- CT:
-
Computed tomography
- HEST:
-
Hall Effect Strain Transducer
- DVRT:
-
Differential Variable Reluctance Transducer
- FOV:
-
Field of view
- sMCL:
-
Superficial medial collateral ligament
- MPFL:
-
Medial patellofemoral ligament
- TKA:
-
Total knee arthroplasty
- RSA:
-
Roentgen stereophotogrammetric analysis
- MJ:
-
Myotendinous junction
- sATMJ:
-
Soleus-AT myotendinous junction
- RF:
-
Radiofrequency
- ROI:
-
Region of interest
- MG:
-
Medial gastrocnemius
- ICC:
-
Intra-class correlation coefficient
References
Abebe, E. S., J. P. Kim, G. M. Utturkar, D. C. Taylor, C. E. Spritzer, C. T. Moorman, 3rd, W. E. Garrett, and L. E. DeFrate. The effect of femoral tunnel placement on ACL graft orientation and length during in vivo knee flexion. J. Biomech. 44:1914–1920, 2011.
Adam, F., D. Pape, D. Kohn, and R. Seil. Length of the patellar tendon after anterior cruciate ligament reconstruction with patellar tendon autograft: a prospective clinical study using Roentgen stereometric analysis. Arthroscopy 18:859–864, 2002.
af Klint, R., N. J. Cronin, M. Ishikawa, T. Sinkjaer, and M. J. Grey. Afferent contribution to locomotor muscle activity during unconstrained overground human walking: an analysis of triceps surae muscle fascicles. J Neurophysiol 103:1262–1274, 2010.
Agres, A. N., G. N. Duda, T. J. Gehlen, A. Arampatzis, W. R. Taylor, and S. Manegold. Increased unilateral tendon stiffness and its effect on gait 2-6 years after Achilles tendon rupture. Scand. J. Med. Sci. Sports 25:860–867, 2015.
Ahmad, A., M. A. M. Bandpei, S. A. Gilani, A. Munawar, I. Ahmed, and F. Tanveer. Reliability of musculoskeletal ultrasound imaging to measure supraspinatus tendon thickness in healthy subjects. J. Phys. Ther. Sci. 29:1394–1398, 2017.
Amis, A. A., and T. D. Zavras. Isometricity and graft placement during anterior cruciate ligament reconstruction. Knee 2:5–17, 1995.
Amjadi, M., K. U. Kyung, I. Park, and M. Sitti. Stretchable, skin-mountable, and wearable strain sensors and their potential applications: a review. Adv. Func. Mater. 26:1678–1698, 2016.
Amjadi M., A. Pichitpajongkit, S. Lee, S. Ryu, and I. Park. Highly stretchable and sensitive strain sensor based on silver nanowire-elastomer nanocomposite. ACS Nano 8:5154–5163, 2014.
Arai, Y., S. Nakagawa, T. Higuchi, A. Inoue, K. Honjo, H. Inoue, K. Ikoma, K. Ueshima, T. Ikeda, H. Fujiwara, and T. Kubo. Comparative analysis of medial patellofemoral ligament length change pattern in patients with patellar dislocation using open-MRI. Knee Surg. Sports Traumatol. Arthrosc. 25:2330–2336, 2017.
Araki, D., E. Thorhauer, and S. Tashman. Three-dimensional isotropic magnetic resonance imaging can provide a reliable estimate of the native anterior cruciate ligament insertion site anatomy. Knee Surg. Sports Traumatol. Arthrosc. 26:1311–1318, 2018.
de Asla, R. J., M. Kozanek, L. Wan, H. E. Rubash, and G. Li. Function of anterior talofibular and calcaneofibular ligaments during in-vivo motion of the ankle joint complex. J. Orthop. Surg. Res. 4:7, 2009.
Aspenberg, P., and T. Schepull. Substantial creep in healing human Achilles tendons. A pilot study. Muscles Ligaments Tendons J. 5:151–155, 2015.
Del Bano-Aledo, M. E., J. J. Martinez-Paya, J. Rios-Diaz, S. Mejias-Suarez, S. Serrano-Carmona, and A. de Groot-Ferrando. Ultrasound measures of tendon thickness: intra-rater, Inter-rater and Inter-machine reliability. Muscles Ligaments Tendons J. 7:192–199, 2017.
Barfod K. W., A. F. Riecke, A. Boesen, P. Hansen, J. F. Maier, S. Doessing and A. Troelsen. Validity and reliability of an ultrasound measurement of the free length of the Achilles tendon. Dan Med. J. 65: 2018.
Bennett, M. B., R. F. Ker, N. J. Dimery, and R. M. Alexander. Mechanical-properties of various mammalian tendons. J. Zool. 209:537–548, 1986.
Bey, M. J., and K. A. Derwin. Measurement of in vivo tendon function. J. Shoulder Elbow Surg. 21:149–157, 2012.
Beynnon, B. D., and B. C. Fleming. Anterior cruciate ligament strain in-vivo: a review of previous work. J. Biomech. 31:519–525, 1998.
Beynnon, B. D., B. C. Fleming, R. J. Johnson, C. E. Nichols, P. A. Renstrom, and M. H. Pope. Anterior cruciate ligament strain behavior during rehabilitation exercises in vivo. Am. J. Sports Med. 23:24–34, 1995.
Beynnon, B., J. G. Howe, M. H. Pope, R. J. Johnson, and B. C. Fleming. The measurement of anterior cruciate ligament strain in vivo. Int. Orthop. 16:1–12, 1992.
Beynnon, B. D., R. J. Johnson, B. C. Fleming, C. J. Stankewich, P. A. Renstrom, and C. E. Nichols. The strain behavior of the anterior cruciate ligament during squatting and active flexion-extension. A comparison of an open and a closed kinetic chain exercise. Am. J. Sports Med. 25:823–829, 1997.
Beynnon, B. D., M. H. Pope, C. M. Wertheimer, R. J. Johnson, B. C. Fleming, C. E. Nichols, and J. G. Howe. The effect of functional knee-braces on strain on the anterior cruciate ligament in vivo. J. Bone Jt. Surg. Am. 74:1298–1312, 1992.
Bogaerts, S., C. De Brito Carvalho, L. Scheys, K. Desloovere, J. D’Hooge, F. Maes, P. Suetens, and K. Peers. Evaluation of tissue displacement and regional strain in the Achilles tendon using quantitative high-frequency ultrasound. PLoS ONE 12:e0181364, 2017.
Bouillard, K., A. Nordez, and F. Hug. Estimation of individual muscle force using elastography. PLoS ONE 6:e29261, 2011.
Boutry, C. M., Y. Kaizawa, B. C. Schroeder, A. Chortos, A. Legrand, Z. Wang, J. Chang, P. Fox, and Z. N. Bao. A stretchable and biodegradable strain and pressure sensor for orthopaedic application. Nat. Electron. 1:314–321, 2018.
Brophy, R. H., D. Kovacevic, C. W. Imhauser, M. Stasiak, A. Bedi, A. J. Fox, X. H. Deng, and S. A. Rodeo. Effect of short-duration low-magnitude cyclic loading versus immobilization on tendon-bone healing after ACL reconstruction in a rat model. J. Bone Jt. Surg. Am. 93:381–393, 2011.
Brown, T. D., L. Sigal, G. O. Njus, N. M. Njus, R. J. Singerman, and R. A. Brand. Dynamic performance characteristics of the liquid metal strain gage. J. Biomech. 19:165–173, 1986.
Burckhardt, K., G. Szekely, H. Notzli, J. Hodler, and C. Gerber. Submillimeter measurement of cup migration in clinical standard radiographs. IEEE Trans. Med. Imaging 24:676–688, 2005.
Burks, R. T., M. H. Metcalf, and R. W. Metcalf. Fifteen-year follow-up of arthroscopic partial meniscectomy. Arthroscopy 13:673–679, 1997.
Camp, C. L., A. Lebaschi, G. T. Cong, Z. Album, C. Carballo, X. H. Deng, and S. A. Rodeo. Timing of postoperative mechanical loading affects healing following anterior cruciate ligament reconstruction: analysis in a murine model. J. Bone Jt. Surg. Am. 99:1382–1391, 2017.
Chen, P. T., C. H. Wu, C. W. Yu, J. H. Wang, T. F. Shih, T. G. Wang, and W. S. Chen. Sonography of the normal anterior cruciate ligament: a preliminary report. J. Med. Ultrasound 21:16–20, 2013.
Cronin, N. J., and T. Finni. Treadmill versus overground and barefoot versus shod comparisons of triceps surae fascicle behaviour in human walking and running. Gait Posture 38:528–533, 2013.
Cronin, N. J., B. Hanley, and A. Bissas. Mechanical and neural function of triceps surae in elite racewalking. J. Appl. Physiol. 121:101–105, 2016.
Cronin, N. J., and G. Lichtwark. The use of ultrasound to study muscle-tendon function in human posture and locomotion. Gait Posture 37:305–312, 2013.
Cronin, N. J., J. Peltonen, M. Ishikawa, P. V. Komi, J. Avela, T. Sinkjaer, and M. Voigt. Achilles tendon length changes during walking in long-term diabetes patients. Clin. Biomech. 25:476–482, 2010.
DeFrate, L. E., T. J. Gill, and G. Li. In vivo function of the posterior cruciate ligament during weightbearing knee flexion. Am. J. Sports Med. 32:1923–1928, 2004.
Ellingson, A. M., J. D. Mozingo, D. J. Magnuson, M. W. Pagnano, and K. D. Zhao. Characterizing fluoroscopy based kinematic accuracy as a function of pulse width and velocity. J. Biomech. 49:3741–3745, 2016.
Escamilla, R. F., T. D. Macleod, K. E. Wilk, L. Paulos, and J. R. Andrews. Anterior cruciate ligament strain and tensile forces for weight-bearing and non-weight-bearing exercises: a guide to exercise selection. J. Orthop. Sports Phys. Ther. 42:208–220, 2012.
Fabiszewska, E., K. Pasicz, I. Grabska, W. Skrzynski, W. Slusarczyk-Kacprzyk, and W. Bulski. Evaluation of imaging parameters of ultrasound scanners: baseline for future testing. Polish J. Radiol. 82:773–782, 2017.
Farris, D. J., E. Buckeridge, G. Trewartha, and M. P. McGuigan. The effects of orthotic heel lifts on Achilles tendon force and strain during running. J Appl Biomech 28:511–519, 2012.
Farris, D. J., G. Trewartha, and M. P. McGuigan. Could intra-tendinous hyperthermia during running explain chronic injury of the human Achilles tendon? J. Biomech. 44:822–826, 2011.
Farris, D. J., G. Trewartha, and M. P. McGuigan. The effects of a 30-min run on the mechanics of the human Achilles tendon. Eur. J. Appl. Physiol. 112:653–660, 2012.
Farris, D. J., G. Trewartha, M. P. McGuigan, and G. A. Lichtwark. Differential strain patterns of the human Achilles tendon determined in vivo with freehand three-dimensional ultrasound imaging. J. Exp. Biol. 216:594–600, 2013.
Finni, T., P. V. Komi, and J. Lukkariniemi. Achilles tendon loading during walking: application of a novel optic fiber technique. Eur. J. Appl. Physiol. Occup. Physiol. 77:289–291, 1998.
Fleming, B. C., and B. D. Beynnon. In vivo measurement of ligament/tendon strains and forces: a review. Ann. Biomed. Eng. 32:318–328, 2004.
Fleming, B. C., B. D. Beynnon, C. E. Nichols, R. J. Johnson, and M. H. Pope. An in vivo comparison of anterior tibial translation and strain in the anteromedial band of the anterior cruciate ligament. J. Biomech. 26:51–58, 1993.
Fleming, B. C., B. D. Beynnon, P. A. Renstrom, G. D. Peura, C. E. Nichols, and R. J. Johnson. The strain behavior of the anterior cruciate ligament during bicycling. An in vivo study. Am. J. Sports Med. 26:109–118, 1998.
Fleming, B. C., B. D. Beynnon, H. Tohyama, R. J. Johnson, C. E. Nichols, P. Renstrom, and M. H. Pope. Determination of a zero strain reference for the anteromedial band of the anterior cruciate ligament. J. Orthop. Res. 12:789–795, 1994.
Fleming, B. C., P. A. Renstrom, G. Ohlen, R. J. Johnson, G. D. Peura, B. D. Beynnon, and G. J. Badger. The gastrocnemius muscle is an antagonist of the anterior cruciate ligament. J. Orthop. Res. 19:1178–1184, 2001.
Foldhazy, Z., A. Arndt, C. Milgrom, A. Finestone, and I. Ekenman. Exercise-induced strain and strain rate in the distal radius. J. Bone Jt. Surg. Br. 87:261–266, 2005.
Franz, J. R., L. C. Slane, K. Rasske, and D. G. Thelen. Non-uniform in vivo deformations of the human Achilles tendon during walking. Gait Posture 41:192–197, 2015.
Friden, J., and R. L. Lieber. Structural and mechanical basis of exercise-induced muscle injury. Med. Sci. Sports Exerc. 24:521–530, 1992.
Fukashiro, S., P. V. Komi, M. Jarvinen, and M. Miyashita. In vivo Achilles tendon loading during jumping in humans. Eur. J. Appl. Physiol. Occup. Physiol. 71:453–458, 1995.
Fung, Y.-C. Biomechanics: Mechanical Properties of Living Tissues. New York: Springer, 2013.
Gellhorn, A. C., and M. J. Carlson. Inter-rater, intra-rater, and inter-machine reliability of quantitative ultrasound measurements of the patellar tendon. Ultrasound Med. Biol. 39:791–796, 2013.
Goto, A., T. Murase, H. Moritomo, K. Oka, K. Sugamoto, and H. Yoshikawa. Three-dimensional in vivo kinematics during elbow flexion in patients with lateral humeral condyle nonunion by an image-matching technique. J. Shoulder Elbow Surg. 23:318–326, 2014.
Graves, N. C., D. D. Rettedal, J. J. Marshall, K. Frush, and V. Vardaxis. Ultrasound assessment of dorsal lisfranc ligament strain under clinically relevant loads. J. Am. Podiatr. Med. Assoc. 104:11–18, 2014.
Gray, H. A., S. Guan, and M. G. Pandy. Accuracy of mobile biplane X-ray imaging in measuring 6-degree -of-freedom patellofemoral kinematics during overground gait. J. Biomech. 57:152–156, 2017.
Griffin, L. Y., J. Agel, M. J. Albohm, E. A. Arendt, R. W. Dick, W. E. Garrett, J. G. Garrick, T. E. Hewett, L. Huston, M. L. Ireland, R. J. Johnson, W. B. Kibler, S. Lephart, J. L. Lewis, T. N. Lindenfeld, B. R. Mandelbaum, P. Marchak, C. C. Teitz, and E. M. Wojtys. Noncontact anterior cruciate ligament injuries: risk factors and prevention strategies. J. Am. Acad. Orthop. Surg. 8:141–150, 2000.
Guan, S., H. A. Gray, F. Keynejad, and M. G. Pandy. Mobile biplane X-ray imaging system for measuring 3D dynamic joint motion during overground gait. IEEE Trans. Med. Imaging 35:326–336, 2016.
Halilaj, E., M. J. Rainbow, D. C. Moore, D. H. Laidlaw, A. P. Weiss, A. L. Ladd, and J. J. Crisco. In vivo recruitment patterns in the anterior oblique and dorsoradial ligaments of the first carpometacarpal joint. J. Biomech. 48:1893–1898, 2015.
Haut, R. C. Age-dependent influence of strain rate on the tensile failure of rat-tail tendon. J. Biomech. Eng. 105:296–299, 1983.
Heijne, A., B. C. Fleming, P. A. Renstrom, G. D. Peura, B. D. Beynnon, and S. Werner. Strain on the anterior cruciate ligament during closed kinetic chain exercises. Med. Sci. Sports Exerc. 36:935–941, 2004.
Herzog, W. Muscle properties and coordination during voluntary movement. J. Sports Sci. 18:141–152, 2000.
Higuchi, T., Y. Arai, H. Takamiya, T. Miyamoto, D. Tokunaga, and T. Kubo. An analysis of the medial patellofemoral ligament length change pattern using open-MRI. Knee Surg. Sports Traumatol. Arthrosc. 18:1470–1475, 2010.
Hirata, K., H. Kanehisa, E. Miyamoto-Mikami, and N. Miyamoto. Evidence for intermuscle difference in slack angle in human triceps surae. J. Biomech. 48:1210–1213, 2015.
Hitz, M., P. Schutz, M. Angst, W. R. Taylor, and R. List. Influence of the moving fluoroscope on gait patterns. PLoS ONE 13:e0200608, 2018.
Hosseini Nasab, S. H., R. List, K. Oberhofer, S. F. Fucentese, J. G. Snedeker, and W. R. Taylor. Loading patterns of the posterior cruciate ligament in the healthy knee: a systematic review. PLoS ONE 11:e0167106, 2016.
Hosseini Nasab, S. H., C. R. Smith, P. Schutz, P. Damm, A. Trepczynski, R. List, and W. R. Taylor. Length-change patterns of the collateral ligaments during functional activities after total knee arthroplasty. Ann. Biomed. Eng. 48:1396–1406, 2020.
Hosseini Nasab, S. H., C. Smith, P. Schutz, B. Postolka, S. Ferguson, W. R. Taylor, and R. List. Elongation patterns of the posterior cruciate ligament after total knee arthroplasty. J. Clin. Med. 9:2078, 2020.
Hosseini Nasab, S. H., C. R. Smith, P. Schutz, B. Postolka, R. List, and W. R. Taylor. Elongation patterns of the collateral ligaments after total knee arthroplasty are dominated by the knee flexion angle. Front. Bioeng. Biotechnol. 7:323, 2019.
Hosseini, A., W. Qi, T. Y. Tsai, Y. J. Liu, H. Rubash, and G. A. Li. In vivo length change patterns of the medial and lateral collateral ligaments along the flexion path of the knee. Knee Surg. Sports Traumatol. Arthrosc. 23:3055–3061, 2015.
Howe, J. G., C. Wertheimer, R. J. Johnson, C. E. Nichols, M. H. Pope, and B. Beynnon. Arthroscopic strain gauge measurement of the normal anterior cruciate ligament. Arthroscopy 6:198–204, 1990.
Huang, Q. A., L. Dong, and L. F. Wang. LC passive wireless sensors toward a wireless sensing platform: status, prospects, and challenges. J. Microelectromech. Syst. 25:822–841, 2016.
Huber, C., Q. Zhang, W. R. Taylor, A. A. Amis, C. Smith, and S. H. Hosseini Nasab. Properties and function of the medial patellofemoral ligament: a systematic review. Am. J. Sports Med. 48:754–766, 2020.
Hug, F., L. Lacourpaille, O. Maisetti, and A. Nordez. Slack length of gastrocnemius medialis and Achilles tendon occurs at different ankle angles. J. Biomech. 46:2534–2538, 2013.
Huston, L. J., M. L. Greenfield, and E. M. Wojtys. Anterior cruciate ligament injuries in the female athlete. Potential risk factors. Clin. Orthop. Relat. Res. 372:50–63, 2000.
Hwang, B. U., J. H. Lee, T. Q. Trung, E. Roh, D. I. Kim, S. W. Kim, and N. E. Lee. Transparent stretchable self-powered patchable sensor platform with ultrasensitive recognition of human activities. ACS Nano 9:8801–8810, 2015.
Intziegianni, K., M. Cassel, N. Konig, S. Muller, K. Frohlich, and F. Mayer. Ultrasonography for the assessment of the structural properties of the Achilles tendon in asymptomatic individuals: An intra-rater reproducibility study. Isokinet. Exerc. Sci. 23:263–270, 2015.
Ishikawa, M., and P. V. Komi. The role of the stretch reflex in the gastrocnemius muscle during human locomotion at various speeds. J. Appl. Physiol. 103:1030–1036, 2007.
Ishikawa, M., P. V. Komi, M. J. Grey, V. Lepola, and G. P. Bruggemann. Muscle-tendon interaction and elastic energy usage in human walking. J. Appl. Physiol. 99:603–608, 2005.
Ishikawa, M., J. Pakaslahti, and P. V. Komi. Medial gastrocnemius muscle behavior during human running and walking. Gait Posture 25:380–384, 2007.
Iwahashi, T., K. Shino, K. Nakata, N. Nakamura, Y. Yamada, H. Yoshikawa, and K. Sugamoto. Assessment of the “functional length” of the three bundles of the anterior cruciate ligament. Knee Surg. Sports Traumatol. Arthrosc. 16:167–174, 2008.
Izadpanah, K., E. Weitzel, M. Honal, J. Winterer, M. Vicari, D. Maier, M. Jaeger, E. Kotter, J. Hennig, M. Weigel, and N. P. Sudkamp. In vivo analysis of coracoclavicular ligament kinematics during shoulder abduction. Am. J. Sports Med. 40:185–192, 2012.
Jansen, M. O., H. C. Schamhardt, A. J. van den Bogert, and W. Hartman. Mechanical properties of the tendinous equine interosseus muscle are affected by in vivo transducer implantation. J. Biomech. 31:485–490, 1998.
Jeong, W. S., Y. S. Yoo, D. Y. Kim, N. S. Shetty, P. Smolinski, K. Logishetty, and A. Ranawat. An analysis of the posterior cruciate ligament isometric position using an in vivo 3-dimensional computed tomography-based knee joint model. Arthroscopy 26:1333–1339, 2010.
Jonsson, H., L. G. Elmqvist, J. Karrholm, and A. Fugl-Meyer. Lengthening of anterior cruciate ligament graft. Roentgen stereophotogrammetry of 32 cases 2 years after repair. Acta Orthop. Scand. 63:587–592, 1992.
Jordan, S. S., L. E. DeFrate, K. W. Nha, R. Papannagari, T. J. Gill, and G. Li. The in vivo kinematics of the anteromedial and posterolateral bundles of the anterior cruciate ligament during weightbearing knee flexion. Am. J. Sports Med. 35:547–554, 2007.
Joseph, C. W., E. J. Bradshaw, T. P. Furness, J. Kemp, and R. A. Clark. Early changes in Achilles tendon behaviour in vivo following downhill backwards walking. J. Sports Sci. 34:1215–1221, 2016.
Karnovsky, S. C., and M. C. Drakos. Revision Achilles reconstruction with hamstring autograft and FHL tendon transfer in an athlete. HSS J. 13:302–306, 2017.
Kernkamp, W. A., C. Wang, C. Z. Li, H. Hu, E. R. A. van Arkel, R. G. H. H. Nelissen, R. F. LaPrade, S. K. van de Velde, and T. Y. Tsai. The medial patellofemoral ligament is a dynamic and anisometric structure: an in vivo study on length changes and isometry. Am. J. Sports Med. 47:1645–1653, 2019.
King, A. J., Q. Deng, R. Tyson, J. C. Sharp, J. Matwiy, B. Tomanek, and J. F. Dunn. In vivo open-bore MRI reveals region- and sub-arc-specific lengthening of the unloaded human posterior cruciate ligament. PLoS ONE 7:e48714, 2012.
Kittl, C., H. El-Daou, K. K. Athwal, C. M. Gupte, A. Weiler, A. Williams, and A. A. Amis. The Role of the Anterolateral Structures and the ACL in Controlling Laxity of the Intact and ACL-Deficient Knee. Am. J. Sports Med. 44:345–354, 2016.
Klimstra, M., J. Dowling, J. L. Durkin, and M. MacDonald. The effect of ultrasound probe orientation on muscle architecture measurement. J. Electromyogr. Kinesiol. 17:504–514, 2007.
Kongsgaard, M., C. H. Nielsen, S. Hegnsvad, P. Aagaard, and S. P. Magnusson. Mechanical properties of the human Achilles tendon, in vivo. Clin Biomech (Bristol, Avon) 26:772–777, 2011.
Korstanje, J. W., R. W. Selles, H. J. Stam, S. E. Hovius, and J. G. Bosch. Development and validation of ultrasound speckle tracking to quantify tendon displacement. J. Biomech. 43:1373–1379, 2010.
Kremkau, F. W. Seeing is believing? Sonographic artifacts. Phys. Today 60:84–85, 2007.
Kwah, L. K., R. Z. Pinto, J. Diong, and R. D. Herbert. Reliability and validity of ultrasound measurements of muscle fascicle length and pennation in humans: a systematic review. J. Appl. Physiol. 114(761–769):2013, 1985.
LaPrade, R. F., A. H. Engebretsen, T. V. Ly, S. Johansen, F. A. Wentorf, and L. Engebretsen. The anatomy of the medial part of the knee. J. Bone Jt. Surg. Am. 89:2000–2010, 2007.
Lee, S. J., K. J. Aadalen, P. Malaviya, E. P. Lorenz, J. K. Hayden, J. Farr, R. W. Kang, and B. J. Cole. Tibiofemoral contact mechanics after serial medial meniscectomies in the human cadaveric knee. Am. J. Sports Med. 34:1334–1344, 2006.
Lee, S. H., and S. J. Yun. Efficiency of knee ultrasound for diagnosing anterior cruciate ligament and posterior cruciate ligament injuries: a systematic review and meta-analysis. Skel. Radiol. 49:1599–1610, 2019.
Leitner, C., P. A. Hager, H. Penasso, M. Tilp, L. Benini, C. Peham, and C. Baumgartner. Ultrasound as a tool to study muscle-tendon functions during locomotion: a systematic review of applications. Sensors (Basel) 19:4316, 2019.
Lichtwark, G. A., K. Bougoulias, and A. M. Wilson. Muscle fascicle and series elastic element length changes along the length of the human gastrocnemius during walking and running. J. Biomech. 40:157–164, 2007.
Lichtwark, G. A., A. G. Cresswell, and R. J. Newsham-West. Effects of running on human Achilles tendon length-tension properties in the free and gastrocnemius components. J. Exp. Biol. 216:4388–4394, 2013.
Lichtwark, G. A., and A. M. Wilson. In vivo mechanical properties of the human Achilles tendon during one-legged hopping. J. Exp. Biol. 208:4715–4725, 2005.
Lichtwark, G. A., and A. M. Wilson. Interactions between the human gastrocnemius muscle and the Achilles tendon during incline, level and decline locomotion. J. Exp. Biol. 209:4379–4388, 2006.
Lippens, E., N. De Smet, S. Schauvliege, A. Martens, F. Gasthuys, E. Schacht, and R. Cornelissen. Biocompatibility properties of surface-modified poly(dimethylsiloxane) for urinary applications. J. Biomater. Appl. 27:651–660, 2013.
List, R., B. Postolka, P. Schutz, M. Hitz, P. Schwilch, H. Gerber, S. J. Ferguson, and W. R. Taylor. A moving fluoroscope to capture tibiofemoral kinematics during complete cycles of free level and downhill walking as well as stair descent. PLoS ONE 12:e0185952, 2017.
Liu, F., H. R. Gadikota, M. Kozanek, A. Hosseini, B. Yue, T. J. Gill, H. E. Rubash, and G. Li. In vivo length patterns of the medial collateral ligament during the stance phase of gait. Knee Surg. Sports Traumatol. Arthrosc. 19:719–727, 2011.
Lyman, S., P. Koulouvaris, S. Sherman, H. Do, L. A. Mandl, and R. G. Marx. Epidemiology of anterior cruciate ligament reconstruction: trends, readmissions, and subsequent knee surgery. J. Bone Jt. Surg. Am. 91:2321–2328, 2009.
Marshall, J. J., N. C. Graves, D. D. Rettedal, K. Frush, and V. Vardaxis. Ultrasound assessment of bilateral symmetry in dorsal Lisfranc ligament. J. Foot Ankle Surg. 52:319–323, 2013.
Martin, J. A., S. C. E. Brandon, E. M. Keuler, J. R. Hermus, A. C. Ehlers, D. J. Segalman, M. S. Allen, and D. G. Thelen. Gauging force by tapping tendons. Nat. Commun. 9:1592, 2018.
Martin, J. A., S. C. E. Brandon, E. M. Keuler, J. R. Hermus, A. C. Ehlers, D. J. Segalman, M. S. Allen, and D. G. Thelen. Gauging force by tapping tendons. Nat. Commun. 9:1–9, 2018.
Massimini, D. F., P. J. Boyer, R. Papannagari, T. J. Gill, J. P. Warner, and G. Li. In-vivo glenohumeral translation and ligament elongation during abduction and abduction with internal and external rotation. J. Orthop. Surg. Res. 7:29, 2012.
McAuliffe, S., K. McCreesh, H. Purtill, and K. O’Sullivan. A systematic review of the reliability of diagnostic ultrasound imaging in measuring tendon size: Is the error clinically acceptable? Phys. Ther. Sport 26:52–63, 2017.
Mian, O. S., J. M. Thom, L. P. Ardigo, A. E. Minetti, and M. V. Narici. Gastrocnemius muscle-tendon behaviour during walking in young and older adults. Acta Physiol. (Oxf) 189:57–65, 2007.
Mian, O. S., J. M. Thom, L. P. Ardigo, A. E. Minetti, and M. V. Narici. Gastrocnemius muscle-tendon behaviour during walking in young and older adults. Acta Physiol. 189:57–65, 2007.
Miller, R. M., J. Thunes, S. Maiti, V. Musahl, and R. E. Debski. Effects of tendon degeneration on predictions of supraspinatus tear propagation. Ann. Biomed. Eng. 47:154–161, 2019.
Moo, E. K., D. R. Peterson, T. R. Leonard, M. Kaya, and W. Herzog. In vivo muscle force and muscle power during near-maximal frog jumps. PLoS ONE 12:e0173415, 2017.
Moraux, A., L. Vandenbussche, X. Demondion, R. Gheno, V. Pansini, and A. Cotten. Anatomical study of the pisotriquetral joint ligaments using ultrasonography. Skelet. Radiol. 41:321–328, 2012.
Nadeau, M. J., A. Desrochers, M. Lamontagne, C. Lariviere, and D. H. Gagnon. Quantitative ultrasound imaging of Achilles tendon integrity in symptomatic and asymptomatic individuals: reliability and minimal detectable change. J. Foot Ankle Res. 9:30, 2016.
Nagai, K., T. Gale, D. Chiba, F. Su, F. Fu, and W. Anderst. The complex relationship between in vivo ACL elongation and knee kinematics during walking and running. J. Orthop. Res. 37:1920–1928, 2019.
Nagaraj, R., and M. N. Kumar. Revision anterior cruciate ligament reconstruction in the nonathlete population. Indian J. Orthop. 53:154–159, 2019.
Nagda, S. H., G. G. Altobelli, K. A. Bowdry, C. E. Brewster, and S. J. Lombardo. Cost analysis of outpatient anterior cruciate ligament reconstruction: autograft versus allograft. Clin. Orthop. Relat. Res. 468:1418–1422, 2010.
Noyes, F. R., J. L. DeLucas, and P. J. Torvik. Biomechanics of anterior cruciate ligament failure: an analysis of strain-rate sensitivity and mechanisms of failure in primates. J. Bone Jt. Surg. Am. 56:236–253, 1974.
Nuri, L., S. J. Obst, R. Newsham-West, and R. S. Barrett. Regional three-dimensional deformation of human Achilles tendon during conditioning. Scand. J. Med. Sci. Sports 27:1263–1272, 2016.
Nuri, L., S. J. Obst, R. Newsham-West, and R. S. Barrett. The tendinopathic Achilles tendon does not remain iso-volumetric upon repeated loading: insights from 3D ultrasound. J. Exp. Biol. 220:3053–3061, 2017.
Obst, S. J., L. Barber, A. Miller, and R. S. Barrett. Reliability of Achilles tendon moment arm measured in vivo using freehand three-dimensional ultrasound. J. Appl. Biomech. 33:300–304, 2017.
Obst, S. J., R. Newsham-West, and R. S. Barrett. In vivo measurement of human Achilles tendon morphology using freehand 3-D ultrasound. Ultrasound Med. Biol. 40:62–70, 2014.
Obst, S. J., R. Newsham-West, and R. S. Barrett. Three-dimensional morphology and strain of the human Achilles free tendon immediately following eccentric heel drop exercise. J. Exp. Biol. 218:3894–3900, 2015.
Ochen, Y., R. B. Beks, M. van Heijl, F. Hietbrink, L. P. H. Leenen, D. van der Velde, M. Heng, O. van der Meijden, R. H. H. Groenwold, and R. M. Houwert. Operative treatment versus nonoperative treatment of Achilles tendon ruptures: systematic review and meta-analysis. BMJ 364:k5120, 2019.
Papannagari, R., L. E. DeFrate, K. W. Nha, J. M. Moses, M. Moussa, T. J. Gill, and G. Li. Function of posterior cruciate ligament bundles during in vivo knee flexion. Am. J. Sports Med. 35:1507–1512, 2007.
Park, S. E., L. E. DeFrate, J. F. Suggs, T. J. Gill, H. E. Rubash, and G. Li. Erratum to “The change in length of the medial and lateral collateral ligaments during in vivo knee flexion”. Knee 13:77–82, 2006.
Park, K. K., A. Hosseini, T. Y. Tsai, Y. M. Kwon, and G. Li. Elongation of the collateral ligaments after cruciate retaining total knee arthroplasty and the maximum flexion of the knee. J. Biomech. 48:418–424, 2015.
Park, I., H. J. Lee, S. E. Kim, S. H. Bae, C. H. Byun, and Y. S. Kim. Which shoulder motions cause subacromial impingement? Evaluating the vertical displacement and peak strain of the coracoacromial ligament by ultrasound speckle tracking imaging. J. Shoulder Elbow Surg. 24:1801–1808, 2015.
Peltonen, J., N. J. Cronin, L. Stenroth, T. Finni, and J. Avela. Achilles tendon stiffness is unchanged one hour after a marathon. J. Exp. Biol. 215:3665–3671, 2012.
Pizzolato, C., D. G. Lloyd, M. H. Zheng, T. F. Besier, V. B. Shim, S. J. Obst, R. Newsham-West, D. J. Saxby, and R. S. Barrett. Finding the sweet spot via personalised Achilles tendon training: the future is within reach. Br. J. Sports Med. 53:11–12, 2019.
Prilutsky, B. I., and V. M. Zatsiorsky. Optimization-based models of muscle coordination. Exerc. Sport Sci. Rev. 30:32–38, 2002.
Rainbow, M. J., J. J. Crisco, D. C. Moore, R. N. Kamal, D. H. Laidlaw, E. Akelman, and S. W. Wolfe. Elongation of the dorsal carpal ligaments: a computational study of in vivo carpal kinematics. J. Hand Surg. Am. 37:1393–1399, 2012.
Rantanen, N. W., and R. L. Ewing. Principles of ultrasound application in animals. Vet. Radiol. 22:196–203, 1981.
Ravary, B., P. Pourcelot, C. Bortolussi, S. Konieczka, and N. Crevier-Denoix. Strain and force transducers used in human and veterinary tendon and ligament biomechanical studies. Clin. Biomech. (Bristol, Avon) 19:433–447, 2004.
Rayan, F., S. K. Nanjayan, C. Quah, D. Ramoutar, S. Konan, and F. S. Haddad. Review of evolution of tunnel position in anterior cruciate ligament reconstruction. World J. Orthop. 6:252–262, 2015.
Roberts, T. J., and E. Azizi. Flexible mechanisms: the diverse roles of biological springs in vertebrate movement. J. Exp. Biol. 214:353–361, 2011.
Roberts, T. J., R. L. Marsh, P. G. Weyand, and C. R. Taylor. Muscular force in running turkeys: the economy of minimizing work. Science 275:1113–1115, 1997.
Roriz, P., L. Carvalho, O. Frazao, J. L. Santos, and J. A. Simoes. From conventional sensors to fibre optic sensors for strain and force measurements in biomechanics applications: a review. J. Biomech. 47:1251–1261, 2014.
Ryba, D., N. Ibrahim, J. Choi, and V. Vardaxis. Evaluation of dorsal Lisfranc ligament deformation with load using ultrasound imaging. Foot (Edinb) 26:30–35, 2016.
Sandrin, L., M. Tanter, J. L. Gennisson, S. Catheline, and M. Fink. Shear elasticity probe for soft tissues with 1-D transient elastography. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 49:436–446, 2002.
Schepull, T., and P. Aspenberg. Early controlled tension improves the material properties of healing human achilles tendons after ruptures: a randomized trial. Am. J. Sports Med. 41:2550–2557, 2013.
Schepull, T., J. Kvist, C. Andersson, and P. Aspenberg. Mechanical properties during healing of Achilles tendon ruptures to predict final outcome: a pilot Roentgen stereophotogrammetric analysis in 10 patients. BMC Musculoskelet. Disord. 8:116, 2007.
Schepull, T., J. Kvist, and P. Aspenberg. Early E-modulus of healing Achilles tendons correlates with late function: similar results with or without surgery. Scand. J. Med. Sci. Sports 22:18–23, 2012.
Schulze, F., F. Mersmann, S. Bohm, and A. Arampatzis. A wide number of trials is required to achieve acceptable reliability for measurement patellar tendon elongation in vivo. Gait Posture 35:334–338, 2012.
Schweitzer, M. E., D. G. Mitchell, and S. M. Ehrlich. The patellar tendon—thickening, internal signal buckling, and other Mr variants. Skeletal Radiol. 22:411–416, 1993.
Serpell, B. G., J. M. Scarvell, M. R. Pickering, N. B. Ball, D. Perriman, J. Warmenhoven, and P. N. Smith. Vertical stiffness is not related to anterior cruciate ligament elongation in professional rugby union players. BMJ Open Sport Exerc. Med. 2:e000150, 2016.
SkovgaardLarsen, L. P., and O. S. Rasmussen. Diagnosis of acute rupture of the anterior cruciate ligament of the knee by sonography. Eur. J. Ultrasound 12:163–167, 2000.
Slane, L. C., S. Bogaerts, I. Mihejeva, and L. Scheys. Evidence of patellar tendon buckling during passive knee extension. Knee 23:801–806, 2016.
Slane, L. C., J. A. Slane, J. D’hooge, and L. Scheys. The challenges of measuring in vivo knee collateral ligament strains using ultrasound. J. Biomech. 61:258–262, 2017.
Slane, L. C., and D. G. Thelen. The use of 2D ultrasound elastography for measuring tendon motion and strain. J. Biomech. 47:750–754, 2014.
Spanjaard, M., N. D. Reeves, J. H. van Dieen, V. Baltzopoulos, and C. N. Maganaris. Influence of step-height and body mass on gastrocnemius muscle fascicle behavior during stair ascent. J. Biomech. 41:937–944, 2008.
Stauffer, F., Q. Zhang, K. Tybrandt, B. L. Zambrano, J. Hengsteler, A. Stoll, C. Trueb, M. Hagander, J. M. Sujata, F. Hoffmann, J. S. Stekhoven, J. Quack, H. Zilly, J. Goedejohann, M. P. Schneider, T. M. Kessler, W. R. Taylor, R. Kung, and J. Voros. Soft electronic strain sensor with chipless wireless readout: toward real-time monitoring of bladder volume. Adv. Mater. Technol. 3:1800031, 2018.
Stenroth, L., D. Thelen, and J. Franz. Biplanar ultrasound investigation of in vivo Achilles tendon displacement non-uniformity. Transl. Sports Med. 2:73–81, 2019.
Suzuki, T., R. Ogane, K. Yaeshima, and R. Kinugasa. Forefoot running requires shorter gastrocnemius fascicle length than rearfoot running. J. Sports Sci. 37:1972–1980, 2019.
Tang, J., E. Thorhauer, K. Bowman, F. H. Fu, and S. Tashman. In vivo posterior cruciate ligament elongation in running activity after anatomic and non-anatomic anterior cruciate ligament reconstruction. Knee Surg. Sports Traumatol. Arthrosc. 25:1177–1183, 2016.
Tashiro, Y., T. Gale, V. Sundaram, K. Nagai, J. J. Irrgang, W. Anderst, Y. Nakashima, S. Tashman, and F. H. Fu. The graft bending angle can affect early graft healing after anterior cruciate ligament reconstruction: in vivo analysis with 2 years’ follow-up. Am. J. Sports Med. 45:1829–1836, 2017.
Tashiro, Y., V. Sundaram, E. Thorhauer, T. Gale, W. Anderst, J. J. Irrgang, F. H. Fu, and S. Tashman. In vivo analysis of dynamic graft bending angle in anterior cruciate ligament-reconstructed knees during downward running and level walking: comparison of flexible and rigid drills for transportal technique. Arthroscopy 33:1393–1402, 2017.
Tashman, S., and W. Anderst. In-vivo measurement of dynamic joint motion using high speed biplane radiography and CT: Application to canine ACL deficiency. J. Biomech. Eng. Trans. Asme 125:238–245, 2003.
Tashman, S., P. Kolowich, D. Collon, K. Anderson, and W. Anderst. Dynamic function of the ACL-reconstructed knee during running. Clin. Orthop. Relat. Res. 454:66–73, 2007.
Taylor, K. A., H. C. Cutcliffe, R. M. Queen, G. M. Utturkar, C. E. Spritzer, W. E. Garrett, and L. E. DeFrate. In vivo measurement of ACL length and relative strain during walking. J. Biomech. 46:478–483, 2013.
Taylor, W. R., P. Schutz, G. Bergmann, R. List, B. Postolka, M. Hitz, J. Dymke, P. Damm, G. Duda, H. Gerber, V. Schwachmeyer, S. H. HosseiniNasab, A. Trepczynski, and I. Kutzner. A comprehensive assessment of the musculoskeletal system: the CAMS-Knee data set. J. Biomech. 65:32–39, 2017.
Taylor, K. A., M. E. Terry, G. M. Utturkar, C. E. Spritzer, R. M. Queen, L. A. Irribarra, W. E. Garrett, and L. E. DeFrate. Measurement of in vivo anterior cruciate ligament strain during dynamic jump landing. J. Biomech. 44:365–371, 2011.
Thoirs, K. A., and J. Childs. Are ultrasound measurements of achilles tendon size reliable? A systematic review of rater reliability. Ultrasound Med. Biol. 44:2476–2491, 2018.
Torniainen, J., A. Ristaniemi, J. K. Sarin, S. Mikkonen, I. O. Afara, L. Stenroth, R. K. Korhonen, and J. Toyras. Near Infrared Spectroscopic Evaluation of Ligament and Tendon Biomechanical Properties. Ann. Biomed. Eng. 47:213–222, 2019.
Utturkar, G. M., L. A. Irribarra, K. A. Taylor, C. E. Spritzer, D. C. Taylor, W. E. Garrett, and L. E. Defrate. The effects of a valgus collapse knee position on in vivo ACL elongation. Ann. Biomed. Eng. 41:123–130, 2013.
Van de Velde, S. K., L. E. DeFrate, T. J. Gill, J. M. Moses, R. Papannagari, and G. Li. The effect of anterior cruciate ligament deficiency on the in vivo elongation of the medial and lateral collateral ligaments. Am. J. Sports Med. 35:294–300, 2007.
Van de Velde, S. K., W. A. Kernkamp, A. Hosseini, R. F. LaPrade, E. R. van Arkel, and G. Li. In vivo length changes of the anterolateral ligament and related extra-articular reconstructions. Am. J. Sports Med. 44:2557–2562, 2016.
Verma, N. K., J. Conroy, P. E. Lyons, J. Coleman, M. P. O’Sullivan, H. Kornfeld, D. Kelleher, and Y. Volkov. Autophagy induction by silver nanowires: A new aspect in the biocompatibility assessment of nanocomposite thin films. Toxicol. Appl. Pharmacol. 264:451–461, 2012.
Wang, J. H., H. Y. Wu, F. J. Dong, B. B. Li, Z. H. Wei, Q. Z. Peng, D. Dong, M. Li, and J. F. Xu. The role of ultrasonography in the diagnosis of anterior cruciate ligament injury: a systematic review and meta-analysis. Eur. J. Sport Sci. 18:579–586, 2018.
Webster, K. E., J. A. Feller, A. J. Kimp, and T. S. Whitehead. Revision Anterior Cruciate Ligament Reconstruction Outcomes in Younger Patients: Medial Meniscal Pathology and High Rates of Return to Sport Are Associated With Third ACL Injuries. Am. J. Sports Med. 46:1137–1142, 2018.
Wertheim, M. G. Memoire sur l’elasticité et la cohésion des principaux tissus du corps humain. Ann. Chim. Phys. Paris (Ser. 3) 21:385–414, 1847.
Westin, O., M. Svensson, K. Nilsson Helander, K. Samuelsson, K. Gravare Silbernagel, N. Olsson, J. Karlsson, and E. Hansson Olofsson. Cost-effectiveness analysis of surgical versus non-surgical management of acute Achilles tendon ruptures. Knee Surg. Sports Traumatol. Arthrosc. 26:3074–3082, 2018.
Woo, S. L., C. Wu, O. Dede, F. Vercillo, and S. Noorani. Biomechanics and anterior cruciate ligament reconstruction. J. Orthop. Surg. Res. 1:2, 2006.
Wood, N., J. G. Konin, and C. Nofsinger. Diagnosis of an ulnar collateral ligament tear using musculoskeletal ultrasound in a collegiate baseball pitcher: a case report. N. Am. J. Sports Phys. Ther. 5:227–233, 2010.
Wu, J. L., A. Hosseini, M. Kozanek, H. R. Gadikota, T. J. T. Gill, and G. Li. Kinematics of the anterior cruciate ligament during gait. Am. J. Sports Med. 38:1475–1482, 2010.
Yan C. Y., J. X. Wang, W. B. Kang, M. Q. Cui, X. Wang, C. Y. Foo, K. J. Chee, and P. S. Lee. Highly stretchable piezoresistive graphene-nanocellulose nanopaper for strain sensors. Adv. Mater. 26:2022–2027, 2014.
Yang, C., A. Goto, W. Sahara, H. Yoshikawa, and K. Sugamoto. In vivo three-dimensional evaluation of the functional length of glenohumeral ligaments. Clin. Biomech. 25:137–141, 2010.
Yang, X. F., N. Wu, G. H. Cheng, Z. Y. Zhou, D. S. Yu, J. J. Beitler, W. J. Curran, and T. Liu. Automated segmentation of the parotid gland based on atlas registration and machine learning: a longitudinal mri study in head-and-neck radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 90:1225–1233, 2014.
Yang X. F. and B. W. Fei. 3D Prostate Segmentation of Ultrasound Images Combining Longitudinal Image Registration and Machine Learning. Medical Imaging 2012: Image-Guided Procedures, Robotic Interventions, and Modeling 8316: 2012.
Yao, S. S., and Y. Zhu. Wearable multifunctional sensors using printed stretchable conductors made of silver nanowires. Nanoscale 6:2345–2352, 2014.
Yasuda, K., C. F. van Eck, Y. Hoshino, F. H. Fu, and S. Tashman. Anatomic single- and double-bundle anterior cruciate ligament reconstruction, part 1: Basic science. Am. J. Sports Med. 39:1789–1799, 2011.
Yoo, Y. S., W. S. Jeong, N. S. Shetty, S. J. Ingham, P. Smolinski, and F. Fu. Changes in ACL length at different knee flexion angles: an in vivo biomechanical study. Knee Surg. Sports Traumatol. Arthrosc. 18:292–297, 2010.
Yue, B., K. M. Varadarajan, H. E. Rubash, and G. Li. In vivo function of posterior cruciate ligament before and after posterior cruciate ligament-retaining total knee arthroplasty. Int. Orthop. 36:1387–1392, 2012.
Zavras, T. D., A. Race, A. M. J. Bull, and A. A. Amis. A comparative study of ‘isometric’ points for anterior cruciate ligament graft attachment. Knee Surg. Sports Traumatol. Arthrosc. 9:28–33, 2001.
Zelik, K. E., and J. R. Franz. It’s positive to be negative: Achilles tendon work loops during human locomotion. PLoS ONE 12:e0179976, 2017.
Zens, M., J. Ruhhammer, F. Goldschmidtboeing, P. Woias, M. J. Feucht, H. O. Mayr, and P. Niemeyer. A New Approach to Determine Ligament Strain Using Polydimethylsiloxane Strain Gauges: Exemplary Measurements of the Anterolateral Ligament. J. Biomech. Eng. Trans. Asme 136:124504, 2014.
Zielinska, B., and T. L. Donahue. 3D finite element model of meniscectomy: changes in joint contact behavior. J. Biomech. Eng. 128:115–123, 2006.
Funding
Open access funding provided by Swiss Federal Institute of Technology Zurich. This study was partially funding by the Swiss Research Foundation under the Project Number 182241 and the Robert Mathys Foundation, Bettlach, Switzerland (https://www.rmsfoundation.ch/home.html) under the Project Number E17_0006.
Conflicts of interest
All authors declare that they have no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Additional information
Associate Editor Michael R. Torry oversaw the review of this article.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zhang, Q., Adam, N.C., Hosseini Nasab, S.H. et al. Techniques for In Vivo Measurement of Ligament and Tendon Strain: A Review. Ann Biomed Eng 49, 7–28 (2021). https://doi.org/10.1007/s10439-020-02635-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10439-020-02635-5