In Vivo Evaluation of Wearable Head Impact Sensors

  • Published:
Annals of Biomedical Engineering Aims and scope Submit manuscript

Abstract

Inertial sensors are commonly used to measure human head motion. Some sensors have been tested with dummy or cadaver experiments with mixed results, and methods to evaluate sensors in vivo are lacking. Here we present an in vivo method using high speed video to test teeth-mounted (mouthguard), soft tissue-mounted (skin patch), and headgear-mounted (skull cap) sensors during 6–13 g sagittal soccer head impacts. Sensor coupling to the skull was quantified by displacement from an ear-canal reference. Mouthguard displacements were within video measurement error (<1 mm), while the skin patch and skull cap displaced up to 4 and 13 mm from the ear-canal reference, respectively. We used the mouthguard, which had the least displacement from skull, as the reference to assess 6-degree-of-freedom skin patch and skull cap measurements. Linear and rotational acceleration magnitudes were over-predicted by both the skin patch (with 120% NRMS error for \(a_{\rm mag}\), 290% for \(\alpha _{\rm mag}\)) and the skull cap (320% NRMS error for \(a_{\rm mag}\), 500% for \(\alpha _{\rm mag}\)). Such over-predictions were largely due to out-of-plane motion. To model sensor error, we found that in-plane skin patch linear acceleration in the anterior–posterior direction could be modeled by an underdamped viscoelastic system. In summary, the mouthguard showed tighter skull coupling than the other sensor mounting approaches. Furthermore, the in vivo methods presented are valuable for investigating skull acceleration sensor technologies.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Figure 1
Figure 2
Figure 3
Figure 4
Figure 5
Figure 6
Figure 7
Figure 8

Similar content being viewed by others

References

  1. Bartsch, A., S. Samorezov, E. Benzel, V. Miele, and D. Brett. Validation of an “intelligent mouthguard” single event head impact dosimeter. Stapp Car Crash J. 58:1–27, 2014.

    PubMed  Google Scholar 

  2. Beckwith, J. G., R. M. Greenwald, and J. J. Chu. Measuring head kinematics in football: correlation between the head impact telemetry system and hybrid III headform. Ann. Biomed. Eng. 40(1):237–248, 2012.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Bouguet, J. Y. Camera Calibration Toolbox for Matlab. http://www.vision.caltech.edu/bouguetj/calib_doc/, 2013.

  4. Camarillo, D. B, P. B. Shull, J. Mattson, R. Shultz, and D. Garza. An instrumented mouthguard for measuring linear and angular head impact kinematics in American football. Ann. Biomed. Eng. 41 (9): 1939–1949, 2013.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Christopher, J. J., M. R. Sochor, J. Pellettiere, and R. S. Salzar. Assessment of Ear- and Tooth-Mounted Accelerometers as Representative of Human Head Response. SAE Technical Paper No. 2013-01-0805, 2013.

  6. Duma, S. M., S. J. Manoogian, W. R. Bussone, P. G. Brolinson, M. W. Goforth, J. J. Donnenwerth, R. M. Greenwald, J. J. Chu, and J. J. Crisco. Analysis of real-time head accelerations in collegiate football players. Clin. J. Sport Med. 15(1):3–8, 2005.

    Article  PubMed  Google Scholar 

  7. Ewins, D. J. Modal Testing: Theory, Practice and Application, 2nd edn. Hertfordshire: Research Studies Press Ltd., 2000.

    Google Scholar 

  8. Federal Motor Vehicle Safety Standards (FMVSS). 571.202a. Section 571, Standard 202a-Head restraints, 2014.

  9. Funk, J. R., J. M. Cormier, C. E. Bain, H. Guzman, and E. Bonugli. Validation and Application of a Methodology to Calculate Head Accelerations and Neck Loading in Soccer Ball Impacts. SAE Technical Paper No. 2009-01-0251, 2009.

  10. Heikkila, J. and O. Silvén. A Four-step Camera Calibration Procedure with Implicit Image Correction. In: Proceedings of IEEE Computer Society Conference on Computer Vision and Pattern Recognition, 1997, pp. 1106–1112.

  11. Hernandez, F., P. B. Shull, and D. B. Camarillo. Evaluation of a laboratory model of human head impact biomechanics. J. Biomech., 2015 (in press).

  12. Hernandez, F., L. C. Wu, M. C. Yip, K. Laksari, A. R. Hoffman, J. R. Lopez, G. A. Grant, S. Kleiven, and D. B. Camarillo. Six degree-of-freedom measurements of human mild traumatic brain injury. Ann. Biomed. Eng. 43(8):1918–1934, 2014.

    Article  PubMed  Google Scholar 

  13. Higgins, M., P. D. Halstead, L. Snyder-Mackler, and D. Barlow. Measurement of impact acceleration: mouthpiece accelerometer versus helmet accelerometer. J. Athl. Train. 42(1):5–10, 2007.

    PubMed  PubMed Central  Google Scholar 

  14. Jadischke, R., D. C. Viano, N. Dau, A. I. King, and J. McCarthy. On the accuracy of the Head Impact Telemetry (HIT) System used in football helmets. J. Biomech. 46(13):2310–2315, 2013.

    Article  PubMed  Google Scholar 

  15. Kim, W., A. S. Voloshin, S. H. Johnson, and A. Simkin. Measurement of the impulsive bone motion by skin-mounted accelerometers. J. Biomech. Eng. 115(1):47–52, 1993.

    Article  CAS  PubMed  Google Scholar 

  16. Knox, T. Validation of Earplug Accelerometers as a Means of Measuring Head Motion. SAE Technical Paper No. 2004-01-3538, 2004.

  17. Lucchetti, L., A. Cappozzo, A. Cappello, and U. D. Croce. Skin movement artefact assessment and compensation in the estimation of knee-joint kinematics. J. Biomech. 31(11):977–984, 1998.

    Article  CAS  PubMed  Google Scholar 

  18. Pellman, E. J., D. C. Viano, A. M. Tucker, and I. R. Casson. Concussion in professional football, part 1: reconstruction of game impacts and injuries. Neurosurgery 53(4):796, 2003.

    Article  Google Scholar 

  19. Reinschmidt, C., B. M. Nigg, A. Lundberg, A. J. van den Bogert, and N. Murphy. Effect of skin movement on the analysis of skeletal knee joint motion during running. J. Biomech. 30(1):729–732, 1997.

    Article  CAS  PubMed  Google Scholar 

  20. Rowson, S. and S. M. Duma. Development of the STAR evaluation system for football helmets: integrating player head impact exposure and risk of concussion. Ann. Biomed. Eng. 39(8): 2130–2140, 2011.

    Article  PubMed  Google Scholar 

  21. Rowson, S., J. G. Beckwith, J. J. Chu, D. S. Leonard, R. M. Greenwald, and S. M. Duma. A six degree of freedom head acceleration measurement device for use in football. J. Appl. Biomech. 27(1):8–14, 2011.

    PubMed  Google Scholar 

  22. Salzar, R. S., C. R. Dale Bass, and J. A. Pellettiere. Improving earpiece accelerometer coupling to the head. SAE Technical Paper No. 2008-01-2978, 2014.

  23. Shewchenko, N., C. Withnall, M. Keown, R. Gittens, and J. Dvorak. Heading in football. Part 1: development of biomechanical methods to investigate head response. Br. J. Sports Med. 39(Suppl 1):10–25, 2005.

    Article  Google Scholar 

  24. Shultz, R., A. E. Kedgley, and T. R. Jenkyn. Quantifying skin motion artifact error of the hindfoot and forefoot marker clusters with the optical tracking of a multi-segment foot model using single-plane fluoroscopy. Gait Posture 34(1):44–48, 2011.

    Article  CAS  PubMed  Google Scholar 

  25. Takhounts, E. G., M. J. Craig, K. Moorhouse, J. McFadden, and V. Hasija. Development of brain injury criteria (BrIC). Stapp Car Crash J. 57:243–266, 2013.

    Google Scholar 

  26. Trujillo, D. M. and H. R. Busby. A mathematical method for the measurement of bone motion with skin-mounted accelerometers. J. Biomech. Eng. 112:229–231, 1990.

    Article  CAS  PubMed  Google Scholar 

  27. Wu, L. C., L. Zarnescu, V. Nangia, B. Cam, and D. B. Camarillo. A head impact detection system using SVM classification and proximity sensing in an instrumented mouthguard. IEEE Trans. Biomed. Eng. 61(11):2659–2668, 2014.

    Article  PubMed  Google Scholar 

  28. Zhang, Z. Flexible camera calibration by viewing a plane from unknown orientations. In: Proceedings of the Seventh IEEE International Conference on Computer Vision, vol. 1, pp. 666–673. 1999.

Download references

Acknowledgements

We thank X2 Biosystems Inc. for supplying skin patch sensors. The study was supported by the National Institutes of Health (NIH) National Institute of Biomedical Imaging and Bioengineering (NIBIB) 3R21EB01761101S1, David and Lucile Packard Foundation 38454, and the Stanford Child Health Research Institute Transdisciplinary Initiatives Program.

Conflict of interest

The authors have no personal or financial conflicts of interest related to this study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to David B. Camarillo.

Additional information

Associate Editor Joel D. Stitzel oversaw the review of this article.

Electronic supplementary material

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wu, L.C., Nangia, V., Bui, K. et al. In Vivo Evaluation of Wearable Head Impact Sensors. Ann Biomed Eng 44, 1234–1245 (2016). https://doi.org/10.1007/s10439-015-1423-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10439-015-1423-3

Keywords

Navigation