Abstract
The flow field of the newly developed inflow cannula designed for a bridge-to-decision circulatory support was numerically analyzed by computational fluid dynamics. This new cannula has elastic struts at the tip that enable minimal invasive insertion into the left ventricle while maintaining a wide inflow area by its lantern-like tip. The cannula’s hydrodynamic loss, including change in pressure loss due to deformation, and its thrombus potential were numerically examined. Hydraulic resistance of the cannula with blood analog fluid was 31 mmHg at the flow rate of 5.0 L/min. There were regions on the inner surface of the struts where the shear rate was <100 s−1, and these regions can be a potential for thrombus formation, especially at low flow rates or under limited anticoagulant therapy.
References
Reynolds HR, Hochman JS. Cardiogenic shock: current concepts and improving outcomes. Circulation. 2008;117:686–97.
Leshnower BG, Gleason TG, O’Hare ML, Pochettino A, Woo YJ, Morris JR, Gardner TJ, Acker MA, et al. Ann Thorac Surg. 2006;81:1365–71.
Yamazaki K, Saito S, Kihara S, Tagusari O, Kurosawa H. Completely pulsatile high flow circulatory support with a constant-speed centrifugal blood pump. Gen Rhorac Cardiovasc Surg. 2007;55:158–62.
Kurihara C, Ono M, Nishimura T, Saito A, Taketani T, Hisagi M, Nawata K, Kinoshita O, Morota T, Motomura N, Kyo S. Use of DuraHeart support for more than 1 year as the first successful bridge to transplantation in Japan. J Artif Organs. 2011;14(1):67–69
Kirklin JK, Naftel DC, Kormos RL, Stevenson LW, Pagani FD, Miller MA, Ullsney KL, Baldwin JT, Young JB. Third INTERMACS Annual Report: the evolution of destination therapy in the United States. J Heart Lung Transplant. 2011;30:115–23.
Slaughter MS, Rogers JG, Milano CA, Russell SD, Conte JV, Feldman D, Sun B, Tatooles AJ, Delgado RM 3rd, Long JW, Wozniak TC, Ghumman W, Farrar DJ, Frazier OH, et al. Advanced heart failure treated with continuous-flow left ventricular assist device. Engl J Med. 2009;361:2241–51.
Ziemba EA, John R. Mechanical circulatory support for bridge to decision: which device and when to decide. J Card Surg. 2010;25:425–33.
Crumpstone T, Martin TD, Yang JJ, Peng YG, et al. Misplacement of LVAD inflow cannula leads to insufficient output, tissue hypoperfusion. J. Artif Organs. 2010;13:225–7.
Komoda T, Weng Y, Nojiri C, Hetzer R. Implantation technique for the Duraheart left ventricular assist system. J Artif Organs. 2007;10(2):124–127
Fraser KH, Zhang T, Taskin ME, Griffith BP, Wu ZJ, et al. Computational fluid dynamics analysis of thrombosis potential in left ventricular assist device drainage Cannulae. ASAIO J. 2010;56:157–63.
Sumikura H, Tsukiya T, Toda K, Takewa Y, Watanabe F, Taenaka Y, Tatsumi E. Development and hydrodynamic evaluation of a novel inflow Cannula in a mechanical circulatory support for bridge to decision. Artif Organs. 2011 (in press).
Yamanaka H, Rosenberg G, Weiss WJ, Snyder AJ, Zapanta CM, Siedlecki CA. Multiscale analysis of surface thrombosis in vivo in a left ventricular assist system. ASAIO J. 2005;51:567–77.
Hochareon P, Manning KB, Fontaine AA, Tarbell JM, Deutsch S. Correlation of in vivo clot deposition with the flow characteristics in the 50 cc Penn State Artificial Heart: a preliminary study. ASAIO J. 2004;50:537–42.
Author information
Authors and Affiliations
Corresponding authors
Rights and permissions
About this article
Cite this article
Tsukiya, T., Toda, K., Sumikura, H. et al. Computational fluid dynamic analysis of the flow field in the newly developed inflow cannula for a bridge-to-decision mechanical circulatory support. J Artif Organs 14, 381–384 (2011). https://doi.org/10.1007/s10047-011-0599-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10047-011-0599-z