Abstract
Purpose
To determine the reliability of measurements obtained by the Good Strength dynamometer, determining isometric abdominal wall and back muscle strength in patients with ventral incisional hernia (VIH) and healthy volunteers with an intact abdominal wall.
Methods
Ten patients with VIH and ten healthy volunteers with an intact abdominal wall were each examined twice with a 1 week interval. Examination included the assessment of truncal flexion and extension as measured with the Good Strength dynamometer, the completion of the International Physical Activity Questionnaire (IPAQ) and the self-assessment of truncal strength on a visual analogue scale (SATS). The test–retest reliability of truncal flexion and extension was assessed by interclass correlation coefficient (ICC), and Bland and Altman graphs. Finally, correlations between truncal strength, and IPAQ and SATS were examined.
Results
Truncal flexion and extension showed excellent test–retest reliability for both patients with VIH (ICC 0.91 and 0.99) and healthy controls (ICC 0.97 and 0.96). Bland and Altman plots showed that no systematic bias was present for neither truncal flexion nor extension when assessing reliability. For patients with VIH, no significant correlations between objective measures of truncal strength and IPAQ or SATS were found. For healthy controls, both truncal flexion (τ 0.58, p = 0.025) and extension (τ 0.58, p = 0.025) correlated significantly with SATS, while no other significant correlation between truncal strength measures and IPAQ was found.
Conclusions
The Good Strength dynamometer provided a reliable, low-cost measure of truncal flexion and extension in patients with VIH.
Similar content being viewed by others
References
Fink C, Baumann P, Wente MN, Knebel P, Bruckner T, Ulrich A, Werner J, Buchler MW, Diener MK (2014) Incisional hernia rate 3 years after midline laparotomy. Br J Surg 101:51–54
Sorensen LT, Hemmingsen UB, Kirkeby LT, Kallehave F, Jorgensen LN (2005) Smoking is a risk factor for incisional hernia. Arch Surg 140:119–123
van Ramshorst GH, Eker HH, Hop WC, Jeekel J, Lange JF (2012) Impact of incisional hernia on health-related quality of life and body image: a prospective cohort study. Am J Surg 204:144–150
Bellows CF, Robinson C, Fitzgibbons RJ, Webber LS, Berger DH (2014) Watchful waiting for ventral hernias: a longitudinal study. Am Surg 80:245–252
Jensen KK, Henriksen NA, Harling H (2014) Standardized measurement of quality of life after incisional hernia repair: a systematic review. Am J Surg 208:485–493
Criss CN, Petro CC, Krpata DM, Seafler CM, Lai N, Fiutem J, Novitsky YW, Rosen MJ (2014) Functional abdominal wall reconstruction improves core physiology and quality-of-life. Surgery 156:176–182
Gunnarsson U, Johansson M, Strigard K (2011) Assessment of abdominal muscle function using the Biodex System-4. Validity and reliability in healthy volunteers and patients with giant ventral hernia. Hernia 15:417–421
Stark B, Emanuelsson P, Gunnarsson U, Strigard K (2012) Validation of Biodex system 4 for measuring the strength of muscles in patients with rectus diastasis. J Plast Surg Hand Surg 46:102–105
Fraser A, Vallow J, Preston A, Cooper RG (1999) Predicting ‘normal’ grip strength for rheumatoid arthritis patients. Rheumatology (Oxford, England) 38:521–528
Craig CL, Marshall AL, Sjostrom M, Bauman AE, Booth ML, Ainsworth BE, Pratt M, Ekelund U, Yngve A, Sallis JF, Oja P (2003) International physical activity questionnaire: 12-country reliability and validity. Med Sci Sports Exerc 35:1381–1395
Rankin G, Stokes M (1998) Reliability of assessment tools in rehabilitation: an illustration of appropriate statistical analyses. Clin Rehabil 12:187–199
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310
Hopkins WG (2000) Measures of reliability in sports medicine and science. Sports Med 30:1–15
Giavarina D (2015) Understanding Bland Altman analysis. Biochem Med (Zagreb) 25:141–151
Chastin SF, Culhane B, Dall PM (2014) Comparison of self-reported measure of sitting time (IPAQ) with objective measurement (activPAL). Physiol Meas 35:2319–2328
Hallal PC, Gomez LF, Parra DC, Lobelo F, Mosquera J, Florindo AA, Reis RS, Pratt M, Sarmiento OL (2010) Lessons learned after 10 years of IPAQ use in Brazil and Colombia. J Phys Act Health 7(Suppl 2):S259–S264
Lachin JM (2004) The role of measurement reliability in clinical trials. Clin Trials 1:553–566
Hopman-Rock M, Kraaimaat FW, Bijlsma JW (1997) Quality of life in elderly subjects with pain in the hip or knee. Qual Life Res 6:67–76
Badia X, Monserrat S, Roset M, Herdman M (1999) Feasibility, validity and test-retest reliability of scaling methods for health states: the visual analogue scale and the time trade-off. Qual Life Res 8:303–310
Stroyer J, Essendrop M, Jensen LD, Warming S, Avlund K, Schibye B (2007) Validity and reliability of self-assessed physical fitness using visual analogue scales. Percept Mot Skills 104:519–533
Janevic MR, McLaughlin SJ, Connell CM (2012) Overestimation of physical activity among a nationally representative sample of underactive individuals with diabetes. Med Care 50:441–445
Jensen KK, Kjaer M, Jorgensen LN (2014) Abdominal muscle function and incisional hernia: a systematic review. Hernia 18:481–486
Parker M, Goldberg RF, Dinkins MM, Asbun HJ, Daniel Smith C, Preissler S, Bowers SP (2011) Pilot study on objective measurement of abdominal wall strength in patients with ventral incisional hernia. Surg Endosc 25:3503–3508
DuBay DA, Choi W, Urbanchek MG, Wang X, Adamson B, Dennis RG, Kuzon WM Jr, Franz MG (2007) Incisional herniation induces decreased abdominal wall compliance via oblique muscle atrophy and fibrosis. Ann Surg 245:140–146
Culbertson EJ, Xing L, Wen Y, Franz MG (2013) Reversibility of abdominal wall atrophy and fibrosis after primary or mesh herniorrhaphy. Ann Surg 257:142–149
Colavita PD, Tsirline VB, Belyansky I, Walters AL, Lincourt AE, Sing RF, Heniford BT (2012) Prospective, long-term comparison of quality of life in laparoscopic versus open ventral hernia repair. Ann Surg 256:714–722
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
KJ declares no conflicts of interest. MK declares no conflicts of interest. LJ declares no conflicts of interest.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Rights and permissions
About this article
Cite this article
Jensen, K.K., Kjaer, M. & Jorgensen, L.N. Isometric abdominal wall muscle strength assessment in individuals with incisional hernia: a prospective reliability study. Hernia 20, 831–837 (2016). https://doi.org/10.1007/s10029-016-1497-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10029-016-1497-3