Abstract
Background
Angular deformities can be treated with corrective osteotomies and application of internal or external fixation. In children, this major intervention can be avoided with temporary hemiepiphysiodesis (i.e., guided growth). Recently, a new device called the eight-Plate Guided Growth System, consisting of a two-hole plate and two screws, was presented as an alternative to the widely used Blount staple to perform temporary hemiepiphysiodesis in children.
Methods
Forty-three patients (54 physes, 51 limbs) underwent treatment between August 2004 and December 2005 with average follow-up after plate insertion of 2 years 2 months (range, 1 year 6 months to 2 years 6 months). Rate of correction and reversibility of this intervention were calculated.
Results
Average age at eight-Plate implantation was 9 years 7 months (age range, 4 years 0 months to 14 years 3 months). eight-Plates were inserted for an average 14.2 months (range, 5.0–27.4 months). No growth disturbance was observed. Mechanical lateral distal femoral angle changed an average 10.00 degrees (range, 1–18 degrees) or 0.65 degrees/month (range, 0.05–1.22 degrees/month). Medial proximal tibial angle changed an average 7.78 degrees (range, 0–14 degrees) or 0.58 degrees/month (range, 0.13–1.67 degrees/month). In the two distal tibial cases, lateral distal tibial angle improved 6 degrees and 10 degrees (average change, 0.44 degrees/month). Mechanical axis deviation improved an average 25.4 mm (range, 0–74 mm) or 1.73 mm/month (range, 0–6.4 mm/month). Ten patients (13 limbs) had more than 10 months of radiographic follow-up after plate removal; ten limbs showed average rebound of 15.7 mm or 1.0 mm/month, indicating the reversibility of this procedure. Four cases failed to achieve correction.
Conclusions
The eight-Plate effectively treats angular deformities in growing children and is less likely to extrude spontaneously than the Blount staple. We have not observed growth disturbance or other complications related to this device.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Blount WP, Clarke GR. Control of bone growth by epiphyseal stapling: a preliminary report. J Bone Joint Surg Am 1949;31:464–478.
Eidelman M, D’Agostino P. Hemiepiphysiodesis around the knee by percutaneously guided and grooved staple. J Pediatr Orthop B 2005;14:434–435.
Haas SL. Retardation of bone growth by a wire loop. J Bone Joint Surg [Am] 1945;27:25–36.
Horton GA, Olney BW. Epiphysiodesis of the lower extremity: results of the percutaneous technique. J Pediatr Orthop 1996;16:180–182.
Khoury JG, Tavares JO, McConnell S, Zeiders G, Sanders JO. Results of screw epiphysiodesis for the treatment of limb length discrepancy and angular deformity. J Pediatr Orthop 2007; 27:623–628.
Métaizeau JP, Wong-Chung J, Bertrand H, Pasquier P. Percutaneous epiphysiodesis using transphyseal screws (PETS). J Pediatr Orthop 1998;18:363–369.
Nouth F, Kuo LA. Percutaneous epiphysiodesis using transphyseal screws (PETS): prospective case study and review. J Pediatr Orthop 2004;24:721–725.
Phemister DB. Operative arrestment of longitudinal growth of bones in the treatment of deformities. J Bone Joint Surg Am 1933;15:1–15.
Blount WP. A mature look at epiphyseal stapling. Clin Orthop Relat Res 1971;77:158–163.
Stevens PM, Maguire M, Dales MD, Robins AJ. Physeal stapling for idiopathic genu valgum. J Pediatr Orthop 1999;19:645–649.
Brockway A, Craig WA, Cockreli BR Jr. End-result study of 62 stapling operations. J Bone Joint Surg Am 1954;36:1063–1070.
Frantz CH. Epiphyseal stapling: a comprehensive review. Clin Orthop Relat Res 1971;77:149–157.
Mielke CH, Stevens PM. Hemiepiphyseal stapling for knee deformities in children younger than 10 years: a preliminary report. J Pediatr Orthop 1996;16:423–429.
Stevens PM. Guided growth for angular correction: a preliminary series using a tension band plate. J Pediatr Orthop 2007;27:253–259.
Stevens PM, Klatt JB. Guided growth for pathological physis: radiographic improvement during realignment. J Pediatr Orthop 2008;28:632–639.
Burghardt RD, Herzenberg JE, Standard SC, Paley D. Temporary hemiepiphyseal arrest with a screw and plate device to treat knee and ankle deformities in children: a preliminary report. J Child Orthop 2008;2:187–197.
Paley D. Frontal plane mechanical and anatomic axis planning. In: Paley D. Principles of deformity correction. 1st ed, corrected 3rd printing. Berlin Tokyo Heidelberg New York: Springer-Verlag; 2005:61–97.
Bowen JR, Leahey JL, Zhang ZH, MacEwen GD. Partial epiphysiodesis at the knee to correct angular deformity. Clin Orthop Relat Res 1985;198:184–190.
Author information
Authors and Affiliations
Additional information
Study conducted at the International Center for Limb Lengthening at the Rubin Institute for Advanced Orthopedics, Sinai Hospital of Baltimore, Baltimore, Maryland, USA
Rights and permissions
This article is published under an open access license. Please check the 'Copyright Information' section either on this page or in the PDF for details of this license and what re-use is permitted. If your intended use exceeds what is permitted by the license or if you are unable to locate the licence and re-use information, please contact the Rights and Permissions team.
About this article
Cite this article
Burghardt, R.D., Herzenberg, J.E. Temporary hemiepiphysiodesis with the eight-Plate for angular deformities: mid-term results. J Orthop Sci 15, 699–704 (2010). https://doi.org/10.1007/s00776-010-1514-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00776-010-1514-9