Stability of L-shaped and inverted triangle fixation assemblies in treating Pauwels type II femoral neck fracture: a comparative mechanical study

  • Original Article • HIP - FRACTURES
  • Published:
European Journal of Orthopaedic Surgery & Traumatology Aims and scope Submit manuscript

Abstract

Purpose

The aim of our study is to compare the mechanical resistance of two screw configurations in fixating type II Pauwels femoral neck fractures.

Methods

Fifteen synthetic models of femur bones in young adults were divided into three equal groups: intact (G1), models with fixation of a 5.0-mm failure zone created in the posterior cortex of the femoral neck using an L-shaped screw arrangement (G2, n = 5), and models with an identical failure zone fixated using an inverted triangle assembly (G3, n = 5). Model strength (axial loading) and rotational deviation of the fragments were load-tested until a 5.0-mm displacement was reached (step 1) and then until failure, here considered as 10.0 mm displacement in G2 and G3 or femoral neck fracture in G1 (step 2).

Results

In step 1, the mean resistance in G1 was 1593 N (standard deviation [SD] of 62 N); this value in G2 was 1261 N (SD 49 N) and in G3 was 1074 N (SD 153 N). During step 2, the value for G1 was 2247 N (SD 84 N), for G2 was 1895 N (SD 69 N), and for G3 was 1523 N (SD 280 N). G3 (the inverted triangle assembly) showed a significantly lower maximum load than the group using the L-shaped assembly (G2) and the control group (G1), which was significant using Kruskal–Wallis analysis of variance (p = 0.002).

Conclusion

Under test conditions in synthetic bone, fixation using a L-shaped screw assembly provides greater mechanical resistance than an inverted triangle assembly.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
Fig. 11

Similar content being viewed by others

References

  1. Asnis SE (1985) The guided screw system in intracapsular fractures of the hip. Contemp Orthop 10(6):33–42

    Google Scholar 

  2. Ye Y, Hao J, Mauffrey C, Hammerberg EM, Stahel PF, Hak DJ (2015) Optimizing stability in femoral neck fracture fixation. Orthopedics 38(10):625–631

    Article  PubMed  Google Scholar 

  3. Kauffman JI, Simon JA, Kummer FJ, Pearlman CJ, Zuckerman JD, Koval KJ (1999) Internal fixation of femoral neck fractures with posterior comminution: a biomechanical study. J Orthop Trauma 13:155–159

    Article  PubMed  CAS  Google Scholar 

  4. Shen M, Wang C, Chen H, Rui YF, Zhao S (2016) An update on the Pauwels classification. J Orthop Surg Res 11(1):161

    Article  PubMed  PubMed Central  Google Scholar 

  5. Freitas A, Lula WF, de Oliveira JS, Maciel RA, Souto DR, Godinho PF (2014) Analysis of mechanical strength to fixing the femoral neck fracture in synthetic bone type Asnis. Acta Ortop Bras 22(4):206–209

    Article  PubMed  PubMed Central  Google Scholar 

  6. Hollander M, Wolfe DA (1999) Nonparametric statistical methods, 2nd edn. Wiley, New York, p 787

    Google Scholar 

  7. Dunn OJ (1964) Multiple comparisons using rank sums. Technometrics 6(3):241–252

    Article  Google Scholar 

  8. Caviglia HA, Osorio PQ, Comando D (2002) Classification and diagnosis of intracapsular fractures of the proximal femur. Clin Orthop Relat Res 399:17–27

    Article  Google Scholar 

  9. Scheck M (1980) The significance of posterior comminution in femoral neck fractures. Clin Orthop Relat Res 152:138–142

    Google Scholar 

  10. Zahid M, Bin Sabir A, Asif N, Julfigar M, Khan AQ, Ahmad S, Siddiqui YS (2012) Fixation using cannulated screws and fibular strut grafts for fresh femoral neck fractures with posterior comminution. J Orthop Surg (Hong Kong) 20(2):191–195

    Article  Google Scholar 

  11. Rawall S, Bali K, Upendra B, Garg B, Yadav CS, Jayaswal A (2012) Displaced femoral neck fractures in the young: significance of posterior comminution and raised intracapsular pressure. Arch Orthop Trauma Surg 132:73–79

    Article  PubMed  Google Scholar 

  12. Upadhyay A, Jain P, Mishra P, Maini L, Gautum VK, Dhaon BK (2004) Delayed internal fixation of fractures of the neck of the femur in young adults. A prospective, randomized study comparing closed and open reduction. J Bone Jt Surg Br 86:1035–1040

    Article  CAS  Google Scholar 

  13. Christie J, Howie CR, Armour PC (1988) Fixation of displaced subcapital femoral fractures. Compression screw fixation versus double divergent pins. J Bone Jt Surg Br 70:199–201

    Article  CAS  Google Scholar 

  14. Holmes CA, Edwards WT, Myers ER, Lewallen DG, White AA 3rd, Hayes WC (1993) Biomechanics of pin and screw fixation of femoral neck fractures. J Orthop Trauma 7:242–247

    Article  PubMed  CAS  Google Scholar 

  15. Walker E, Mukherjee DP, Ogden AL, Sadasivan KK, Albright JA (2007) A biomechanical study of simulated femoral neck fracture fixation by cannulated screws: effects of placement angle and number of screws. Am J Orthop (Belle Mead NJ) 36:680–684

    Google Scholar 

  16. Zdero R, Keast-Butler O, Schemitsch EH (2010) A biomechanical comparison of two triple-screw methods for femoral neck fracture fixation in a synthetic bone model. J Trauma 69:1537–1544

    Article  PubMed  Google Scholar 

  17. Zlowodzki M, Brink O, Switzer J, Wingerter S, Woodall J Jr, Petrisor BA, Kregor PJ, Bruinsma DR, Bhandari M (2008) The effect of shortening and varus collapse of the femoral neck on function after fixation of intracapsular fracture of the hip: a multi-centre cohort study. J Bone Jt Surg Br 90(11):1487–1494

    Article  CAS  Google Scholar 

  18. Lindequist S, Wredmark T, Eriksson SA, Samnegård E (1993) Screw positions in femoral neck fractures. Comparison of two different screw positions in cadavers. Acta Orthop Scand 64(1):67–70

    Article  PubMed  CAS  Google Scholar 

  19. Panteli M, Rodham P, Giannoudis PV (2015) Biomechanical rationale for implant choices in femoral neck fracture fixation in the non-elderly. Injury 46(3):445–452

    Article  PubMed  Google Scholar 

  20. Aminian A, Gao F, Fedoriw WW, Zhang LQ, Kalainov DM, Merk BR (2007) Vertically oriented femoral neck fractures: mechanical analysis of four fixation techniques. J Orthop Trauma 21(8):544–548

    Article  PubMed  Google Scholar 

  21. Cristofolini L, Viceconti M, Cappello A, Toni A (1996) Mechanical validation of whole bone composite femur models. J Biomech 29(4):525–535

    Article  PubMed  CAS  Google Scholar 

  22. Giordano V, Luz EM, Lima-Filho JCC, Serafim MAM, Amaral NP, Scappini W Jr (2006) Estudo eletromiográfico das forças atuantes no quadril em apoio monopodal: uma nova visão sobre o modelo de Koch. Rev Bras Ortop 41(7):278–282

    Google Scholar 

  23. Gok K, Inal S (2015) Biomechanical comparison using finite element analysis of different screw configurations in the fixation of femoral neck fractures. Mech Sci 6:173–179

    Article  Google Scholar 

  24. Bishop JA, Behn AW, Castillo TN (2014) The biomechanical significance of washer use with screw fixation. J Orthop Trauma 28(2):114–117

    Article  PubMed  Google Scholar 

  25. Zlowodzki M, Wijdicks CA, Armitage BM, Cole PA (2015) The value of washers in internal fixation of femoral neck fractures with cancellous screws: a biomechanical evaluation. J Orthop Trauma 29(2):e69–e72

    Article  PubMed  Google Scholar 

  26. Zlowodzki M, Weening B, Petrisor B, Bhandari M (2005) The value of washers in cannulated screw fixation of femoral neck fractures. J Trauma 59(4):969–975

    Article  PubMed  Google Scholar 

  27. Ma JX, Kuang MJ, Xing F, Zhao YL, Chen HT, Zhang LK, Fan ZR, Han C, Ma XL (2018) Sliding hip screw versus cannulated cancellous screws for fixation of femoral neck fracture in adults: a systematic review. Int J Surg 52:89–97

    Article  PubMed  Google Scholar 

  28. Gupta M, Arya RK, Kumar S, Jain VK, Sinha S, Naik AK (2016) Comparative study of multiple cancellous screws versus sliding hip screws in femoral neck fractures of Young adults. Chin J Traumatol 19(4):209–212

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Contributions

VG helped in protocol creation, manuscript creation, manuscript editing, data collection, and data analysis. RPP, DDA, and ABA contributed to protocol creation and data collection. WB, MG, and AF helped in manuscript editing and critical review. Senior author HAK contributed to manuscript editing and critical review.

Corresponding author

Correspondence to Vincenzo Giordano.

Ethics declarations

Conflict of interest

On behalf of all authors, the corresponding author states that there is no conflict of interest.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Giordano, V., Paes, R.P., Alves, D.D. et al. Stability of L-shaped and inverted triangle fixation assemblies in treating Pauwels type II femoral neck fracture: a comparative mechanical study. Eur J Orthop Surg Traumatol 28, 1359–1367 (2018). https://doi.org/10.1007/s00590-018-2207-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00590-018-2207-x

Keywords

Navigation