Resorbable osteosynthetic devices in pediatric traumatology: a prospective series of 24 cases

  • Original Article • PAEDIATRICS - TRAUMA
  • Published:
European Journal of Orthopaedic Surgery & Traumatology Aims and scope Submit manuscript

An Erratum to this article was published on 17 July 2015

Abstract

Bioresorbable devices are commonly used in traumatology. The biomechanical stability of these materials has improved in the past decade, and they have proven to be biologically non-hazardous, while their main advantage is that their use avoids reintervention for removal of the device. A prospective monocentric study was conducted: 24 patients presenting with a fracture that was amenable to osteosynthesis by small-diameter screws were included. These comprised ten tibial spine fractures, four osteochondritis dissecans of the distal femur, eight fractures of the medial epicondyle of the distal humerus, and two distal tibial apophyseal fractures. One or more screws were used that were made of a copolymer of poly-l-lactide-poly-d-lactide acid and trimethylene carbonate with a diameter of 2.8 mm. All patients were immobilized with a cast. Clinical and radiographic monitoring was conducted every month. The entire follow-up protocol had a duration of 24 months. One patient with osteochondritis dissecans presented with joint effusion. Joint stiffness at the time of cast removal resolved completely after 4 months, except for with three children (one epicondyle fracture, two tibial spine fractures). No subjective or objective instability could be detected by clinical examination. Radiographic follow-up revealed no secondary displacement, and all of the fractures had healed. No osteolysis was seen around the screws. No growth disturbances were noticed. Bioresorbable materials thus appear to be a suitable alternative approach for certain pediatric fractures. Their use resulted in outcomes similar to traditional techniques in terms of functional properties and bone healing. Although initial costs are presumably slightly higher, by avoiding a removal operation the total financial burden is most likely reduced.

Level of evidence

III.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Gogolewski S (2000) Bioresorbable polymers in trauma and bone surgery. Injury 31(Suppl 4):28–32

    Article  PubMed  Google Scholar 

  2. Ashammakhi N, Renier D, Arnaud E, Marchac D, Ninkovic M, Donaway D et al (2004) Successful use of biosorb osteofixation devices in 165 cranial and maxillofacial cases: a multicenter report. J Craniofac Surg 15:692–701

    Article  PubMed  Google Scholar 

  3. Cheung LK, Chow LK, Chiu WK (2004) A randomized controlled trial of resorbable versus titanium fixation for orthognathic surgery. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 98:386–397

    Article  PubMed  Google Scholar 

  4. Eppley BL (2007) Bioabsorbable plate and screw fixation in orthognathic surgery. J Craniofac Surg 18:818–825

    Article  PubMed  Google Scholar 

  5. Eppley BL (2005) Use of resorbable plates and screws in pediatric facial fractures. J Oral Maxillofac Surg 63:385–391

    Article  PubMed  Google Scholar 

  6. Eppley BL, Morales L, Wood R, Pensler J, Goldstein J, Havlik RJ et al (2004) Resorbable PLLA-PGA plate and screw fixation in pediatric craniofacial surgery: clinical experience in 1883 patients. Plast Reconstr Surg 114:850–856

    Article  PubMed  Google Scholar 

  7. Bucholz RW, Henry S, Henley MB (1994) Fixation with bioabsorbable screws for the treatment of fractures of the ankle. J Bone Joint Surg Am 76:319–324

    CAS  PubMed  Google Scholar 

  8. Hovis WD, Kaiser BW, Watson JT, Bucholz RW (2002) Treatment of syndesmotic disruptions of the ankle with bioabsorbable screw fixation. J Bone Joint Surg Am 84-A:26–31

    PubMed  Google Scholar 

  9. Kankare J, Partio EK, Hirvensalo E, Böstman O, Rokkanen P (1996) Biodegradable self-reinforced polyglycolide screws and rods in the fixation of displaced malleolar fractures in the elderly. A comparison with metallic implants. Ann Chir Gynaecol 85:263–270

    CAS  PubMed  Google Scholar 

  10. Sinisaari IP, Lüthje PMJ, Mikkonen RHM (2002) Ruptured tibio-fibular syndesmosis: comparison study of metallic to bioabsorbable fixation. Foot Ankle Int 23:744–748

    PubMed  Google Scholar 

  11. Thordarson DB, Samuelson M, Shepherd LE, Merkle PF, Lee J (2001) Bioabsorbable versus stainless steel screw fixation of the syndesmosis in pronation-lateral rotation ankle fractures: a prospective randomized trial. Foot Ankle Int 22:335–338

    CAS  PubMed  Google Scholar 

  12. Nieminen T, Rantala I, Hiidenheimo I, Keränen J, Kainulainen H, Wuolijoki E et al (2008) Degradative and mechanical properties of a novel resorbable plating system during a 3-year follow-up in vivo and in vitro. J Mater Sci Mater Med 19:1155–1163

    Article  CAS  PubMed  Google Scholar 

  13. Meyers MH, McKeever FM (1970) Fracture of the intercondylar eminence of the tibia. J Bone Joint Surg Am 52:1677–1684

    CAS  PubMed  Google Scholar 

  14. Meyers MH, McKeever FM (1959) Fracture of the intercondylar eminence of the tibia. J Bone Joint Surg Am 41-A:209–222

    CAS  PubMed  Google Scholar 

  15. Mavrogenis AF, Kanellopoulos AD, Nomikos GN, Papagelopoulos PJ, Soucacos PN (2009) Early experience with biodegradable implants in pediatric patients. Clin Orthop Relat Res 467:1591–1598

    Article  PubMed Central  PubMed  Google Scholar 

  16. Kukk A, Nurmi JT (2009) A retrospective follow-up of ankle fracture patients treated with a biodegradable plate and screws. Foot Ankle Surg 15:192–197

    Article  PubMed  Google Scholar 

  17. Casteleyn PP, Handelberg F, Haentjens P (1992) Biodegradable rods versus Kirschner wire fixation of wrist fractures. A randomised trial. J Bone Joint Surg Br 74:858–861

    CAS  PubMed  Google Scholar 

  18. Fraser RK, Cole WG (1992) Osteolysis after biodegradable pin fixation of fractures in children. J Bone Joint Surg Br 74:929–930

    CAS  PubMed  Google Scholar 

  19. Rokkanen P, Böstman O, Vainionpää S, Makela EA, Hirvensalo E, Partio EK et al (1996) Absorbable devices in the fixation of fractures. J Trauma 40:S123–S127

    Article  CAS  PubMed  Google Scholar 

  20. Eitenmüller J, David A, Pommer A, Muhr G (1996) Surgical treatment of ankle joint fractures with biodegradable screws and plates of poly-l-lactide. Chirurg 67:413–418

    PubMed  Google Scholar 

  21. Böstman OM, Pihlajamäki HK (2000) Adverse tissue reactions to bioabsorbable fixation devices. Clin Orthop Relat Res 371:216–227

    Article  PubMed  Google Scholar 

  22. Raney EM, Freccero DM, Dolan LA, Lighter DE, Fillman RR, Chambers HG (2008) Evidence-based analysis of removal of orthopaedic implants in the pediatric population. J Pediatr Orthop 28:701–704

    Article  PubMed  Google Scholar 

  23. Böstman OM (1996) Metallic or absorbable fracture fixation devices. A cost minimization analysis. Clin Orthop Relat Res 329:233–239

    Article  PubMed  Google Scholar 

  24. Louahem DM, Bourelle S, Buscayret F, Mazeau P, Kelly P, Dimeglio A et al (2010) Displaced medial epicondyle fractures of the humerus: surgical treatment and results. A report of 139 cases. Arch Orthop Trauma Surg 130:649–655

    Article  PubMed  Google Scholar 

  25. Casalonga A, Bourelle S, Chalencon F, De Oliviera L, Gautheron V, Cottalorda J (2010) Tibial intercondylar eminence fractures in children: the long-term perspective. Orthop Traumatol Surg Res 96:525–530

    Article  CAS  PubMed  Google Scholar 

  26. Tuompo P, Arvela V, Partio EK, Rokkanen P (1997) Osteochondritis dissecans of the knee fixed with biodegradable self-reinforced polyglycolide and polylactide rods in 24 patients. Int Orthop 21:355–360

    Article  CAS  PubMed  Google Scholar 

  27. Larsen MW, Pietrzak WS, DeLee JC (2005) Fixation of osteochondritis dissecans lesions using poly(l-lactic acid)/poly(glycolic acid) copolymer bioabsorbable screws. Am J Sports Med 33:68–76

    Article  PubMed  Google Scholar 

  28. Barfod G, Svendsen RN (1992) Synovitis of the knee after intraarticular fracture fixation with Biofix. Report of two cases. Acta Orthop Scand 63:680–681

    CAS  PubMed  Google Scholar 

  29. Fridén T, Rydholm U (1992) Severe aseptic synovitis of the knee after biodegradable internal fixation. A case report. Acta Orthop Scand 63:94–97

    Article  PubMed  Google Scholar 

  30. Lefort G, Moyen B, Beaufils P, de Billy B, Breda R, Cadilhac C et al (2006) Osteochondritis dissecans of the femoral condyles: report of 892 cases. Rev Chir Orthop Reparatrice Appar Mot 92:2S97–92S141

    CAS  PubMed  Google Scholar 

Download references

Conflict of interest

None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to P. Journeau.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Poircuitte, J.M., Popkov, P., Huber, D.H. et al. Resorbable osteosynthetic devices in pediatric traumatology: a prospective series of 24 cases. Eur J Orthop Surg Traumatol 25, 997–1004 (2015). https://doi.org/10.1007/s00590-015-1656-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00590-015-1656-8

Keywords

Navigation