Abstract
The uncemented glenoid implants in total anatomical shoulder arthroplasty are likely to be accused of problems like dissociations, secondary rotator cuff tear, and wear of polyethylene (PE). This work is a clinical and radiological prospective review of 143 cases of anatomical total shoulder arthroplasty using a new metal back uncemented glenoid implant (MB) in order to see if this new implant induces those complications. A total of 143 cases were operated between 2003 and 2011. In a first part, the whole series of 143 cases was radiologically studied in order to quantify the lateralisation induced by the MB implant. In a second study, 37 cases had a mean follow-up of 38 months (24–75, mean 32) and served for the clinical and radiological final study. Pre- and postoperative clinical evaluation was done using the Constant–Murley score and the simple shoulder test from Matsen. The final X-rays served to detect an eventual secondary narrowing of the joint space and to analyse the frequency of radio lucent lines (RLL) and loosenings. Despite a small radiological lateralisation in comparison with the normal contralateral side (0.36 cm, p = 0.02), the clinical results after 2 years were similar to the published cemented glenoid implants series but without any RLL, glenoid loosening or joint narrowing. Some dissociations occured in the beginning and definitely eliminated by a design modification of the PE tray. The discussion tried to show that, despite a still short follow-up, this series is encouraging to continue to use this new MB implant. Different applications of the concept of universality and conversion are discussed, this tray been also the support of a glenosphere in reverse arthroplasty.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The gold standard technique for glenoid replacement in total anatomical shoulder prosthesis is still the use of full polyethylene cemented implant like Neer concept 40 years ago [1].
The main problem is glenoid loosening. Despite good and predictable clinical results, the frequency of radio lucent lines (RLL) is high, 70 % in the most recent literature with an increasing number of glenoid loosenings, 40 % at 10 years [2–4]. However, the number of revisions is low, around 5 % [5, 6].
These worrisome findings have spurred development of new ideas. Among them cementless glenoid devices have been tried, as much as the success of reverse prosthesis has obliged to develop metal back screwed glenoid trays, able to resist to the shear forces induced by a glenosphere [7].
These new uncemented implants in anatomical shoulder replacement have been greatly criticised, accused to be responsible of loosening, dissociation, and early PE wear. Furthermore, it has been advocated that the increased thickness induced by the metal tray [8, 9] could be a risk for the rotator cuff [10].
The excellent primary fixation of these screwed implants in reverse arthroplasty has led us to extend our indications to anatomical replacements, in order to see whether with this kind of design we are able to decrease the frequency of RLL observed with the cemented glenoids.
Materials and methods
The metal back glenoid implant (MB) of the Universal Shoulder Arthroplastic System ARROW (FH orthopedics, 3 rue de la Forêt 68990-Heimsbrunn-France) is 6.5 mm thick, 3.5 for the PE and 3 for the metal tray. The deep convexe surface and the keel are covered with hydroxyapatite.
Four sizes are available 44, 46, 48 and 50. Whatever the size of the humeral head, there is a systematic mismatch between the radius of curvature of the glenoid and of the humeral head with an average of 4 mm (between 1 and 6).
The ancillary system allows a precise preparation of the glenoid with a reaming of the bone surface and a press fit preparation of the keel grove, in order to insure a perfect contact between hydroxyapatite and bone.
The primary fixation is insured by 2 axial screws and can be enhanced by a third sagittal screw. This third screw goes through an anterior plate and the keel. It can be useful in case of osteoporotic patient and glenoid bone loss, allowing an easy bone graft fixation.
On the humeral side, noncemented press fit stems were preferentially used, with grafting of the metaphysis using some cancellous bone from the humeral head. In case of osteoporotic bone, a classical cemented technique was recommended.
The clinical analysis included a pre- and postoperative evaluation of the Score of Constant and Murley [11], of the active and passive range of motion and of the simple shoulder test from Matsen [12].
Radiographic preoperative assessment consisted of plain anteroposterior radiographs with medial, neutral and lateral rotation, axillary and outlet view under fluoroscopic guidance. A systematic CT scan completed the preoperative radiographic analysis to evaluate the status of the cuff and the glenoid bone stock according to Walch classification [13].
Postoperative radiological study included an AP view with a standardised fluoroscopic technique and the X-ray beam perpendicular to the plane of the joint space. This allowed to detect an eventual narrowing, witness of a progressive polyethylene wear.
The study included 2 parts:
-
1.
A radiological study done on the first postoperative X-rays in order to check if the increased thickness of the MB component induced a measurable lateralisation. The lateral offset was measured between the centre of the glenoid bone and the lateral border of the great tuberosity (Fig. 1). This measurement was compared with the normal contralateral side, if not involved.
-
2.
The secund part consisted of analysing the final clinical and radiological results on the patients with more than 24 months of follow-up.
The Student’s t test was used for statistical analysis when two groups had to be compared. When the comparison involved more than two groups, a variance analysis was applied. The chosen level of significance (p) was set at 0.05.
Results
From November 2003 to December 2011, 143 total anatomical shoulder arthroplasties have been performed. Aetiology is summarised in Table 1, dominated by primary osteoarthritis with 90 % of normal cuff.
Delto pectoral approach was performed except in 2 cases.
The common size for the MB was 44 in 3/4 of the cases, and 46 for the remaining. The humeral stem size was mainly 10 or 12. 25 % of the humeral heads were 44, 25 % for 46 and 48. 65 % were excentric heads with a height of 16 for half of them.
Most of the biceps tendons were tenodesed (76 %).
Radiological analysis:
1. Radiological results
The results of the first part of this work are summarised in the Table 2. Seventy-eight patients had no bilateral involvement. The postoperative immediate radiological study showed a lateralisation of 0.36 cm between the operated and the contralateral normal side (p = 0.02).
With more than 2 years of follow-up 2 cases presented a well-tolerated superior migration. One, with a follow-up of 24 months, had no pain, an active flexion of 90° and a Constant score of 53 (68 %). The other, at 44 months, presented a slight pain, an active elevation of 100° and a score of 54 (74 %).
In the whole series of 143 MB prosthesis, we noted 4 cases of superior migration (2.8 %). Two were reoperated. All were anterior to 2008.
In 5 cases, the inferior screw was under the scapula with no clinical or radiological consequences.
Part of the cohort allowed a study of the joint narrowing. At 3 months, the difference between operated and contralateral normal side was 0.47 cm (n = 30, 0–1), at 1 year, we found a 0.44 of average (−1 to 1), at 2 years 0.42 (0–1) and finally 0.39 (−0.1 to 1) with no statistical significance. No lucent lines were recorded.
2. Clinical results
Among the 143 cases, 11 complications were recorded (7.7 %):
-
Three dissociations occured in the beginning of the experiment. A first design did not allow to precisely centre the PE tray before impaction. In 2007, the addition of a PE central peg allowed to insure a good alignment. This modification erased definitely this complication. Two of these cases were revised, one with a cemented glenoid with an excellent final Constant score of 85 (121 %) and an active flexion of 160°. The other one is 60 months of follow-up of a simple reimpaction of a new PE tray with also an excellent result, a Constant score of 90 (103 %) and an active elevation of 170°. The third case of dissociation did not accept any revision.
-
Three dislocations occured, all with b2 and c glenoid types. Two were revised with a conversion in reverse. In one case with a type b2 glenoid, the conversion consisted of simply changing the PE tray to a glenosphere and the humeral head to a cup. Neither the MB tray nor the humeral stem was modified. This patient with 18 months of follow-up had an active flexion of 120°, no pain and a Constant score of 54 (76 %) (Fig. 2). For the 2 cases with a glenoid type c, because of the necessity of grafting the glenoid it was necessary to take out the metal tray despite a good integration. But like the type b2 case the humeral stem was untouched.
-
Two secondary rotator cuff tears were converted to a reverse. The first one, at 17 months of the revision, had no pain but a fair active and passive mobility with a Constant score of 44 (68 %). The other one was a b2 glenoid and had also a contralateral MB glenoid with an excellent result. During revision performed at 2 years a posterosuperior, PE wear was found with a contact between the metal tray and the humeral head. At 36 months of the revision, the result was excellent with a Constant score of 64 (100 %), an active elevation of 140° (Fig. 3).
-
One case sustained a superior migration due to a bad initial choice of the size of the glenoid with a too low implantation. This allowed a quick superior migration of the humeral head above the glenoid implant but without cuff tear. The revision at 1 month consisted on a higher implantation of a bigger metal tray. The MB 44 was converted to 48. This patient was 60 months of follow-up, and this case of revision was included in the final review.
-
One patient had a postoperative painful stiffness because of a complex regional pain syndrome.
-
Another case of painful shoulder was reoperated at 6 months. No aetiology was found. The MB glenoid was converted to a cemented glenoid. At 13 months, there was a superior migration with a bad result.
Globally, the revision rate was 8/143 = 5.59 %. But if we do not take account of the dissociation cases, the percentage decreases to 4.19 % (6/143).
3. Results after 24 months or more
Thirty-seven cases in 36 patients were 2 years of follow-up or more, 38.3 months in average (24–75, mean 32), 28 female (one bilateral) 8 men, 21 right side, and 32 right handed. The average age was 69 (35–83, mean 72). Aetiology was mainly primary osteoarthritis without cuff tear (Table 3). Four patients were already operated on, 2 subacromial decompression and 2 previous arthroplasties. The first one was this already mentioned patient with a too small MB glenoid implanted too low and revised at 3 months. The second one sustained previously a total anatomical arthroplasty with a cemented glenoid, revised with a MB glenoid and a grafting because of a glenoid loosening.
4. Three revisions were recorded
-
Two presented a superior migration due to a rotator cuff tear. One was already ruptured preoperatively, and the other was checked as pathologic. As already mentioned, these 2 cases were converted to a reverse, simple and rapid procedure thanks to the universality of the ARROW system, consisting of a simple change of the intermediate devices. These 2 cases were excluded from the final review.
-
One patient presented an early dissociation revised at 2 months. A simple change of the PE tray allowed an excellent result at 39 months with an active flexion of 170° and a Constant score of 85 (102 %).
-
One patient died from a medical cause in January 2012 after the final review in November 2011. Her result was excellent, 150° of flexion and a score of 72 (101 %).
Finally, 35 cases were available for the final results which are summarised in Table 4. Pain increased from 1.6 to 13.4, flexion from 92° to 146° and Constant score from 27 (36 %) to 70 (95 %). The statistical difference between pre- and postoperative values was greatly significant.
We analysed the results according to the preoperative type of glenoid:
-
Between types A1 and A2 the preoperative clinical values were systematically inferior for the most used glenoids with no statistical significance. The same differences were noted postoperatively excepted for the activities of daily living.
-
We found the same differences between A1 and B1 and between type A and type B.
-
if we study the types A1 and B1 versus A2, B2 and C, the values were systematically lower pre- and postoperatively for the most used glenoids.
Discussion
In total shoulder anatomical arthroplasty using a cemented glenoid, the percentage of radio lucent lines is high despite a good and stable clinical results [14] Whatever the model, keeled or pegged, at 10 years, 76 % presented a RLL and 40 % a glenoid loosening. The modular second or third generation tried to give a better adaptation to the patient’s anatomy but did not solve the problem [15, 16]. One of the most recent publication, with more than 5 years of follow-up, showed 18.9 % of loosenings, among them 23 % were progressive with a functional repercussion [17].
Many modifications of the glenoid devices tried to solve this problem. Parallely, the feasibility of the metal back trays, able to sustain without failure the shear forces induced by a glenosphere in reverse shoulder arthroplasty, led to experiment these devices in anatomical arthroplasty [18]. With this kind of design, the literature evoqued early polyethylene wear, number of dissociation between PE and metal tray, loosenings and superior migration due to rotator cuff tears. The first referred cause of those failures was the thickness of the implant [4, 19, 20]. However, most of the articles referred to old conceptions [21–23] or designs which did not insure a good primary fixation [24]. However, some more recent publications presented some encouraging results [25, 26].
In 1992, we started to work on a new implant which was available for clinical use in human since 2003. The convexity was preferred to a flat back tray, being widely recognised as insuring a good bone-implant contact and transforming the shear forces in compressive forces [1, 23, 27–30]. Iannotti [29], like Neer previously [1], insisted in the better easiness of well positioning the convexe implant, which parallely decreases the frequency of lucent lines [31]. The principle of a mismatch was also adopted [15, 22, 24, 25].
In our experience, the fit induced by a precise ancillary system and the frontal and the eventual saggital screwing avoided any primary fixation problems and enhanced the indications to the cases necessitating a glenoid bone graft. The hydroxyapatite coverage on all the parts in contact with the glenoid bone insured a good secondary fixation with no migrations, loosenings or even RLL.
We had on the beginning some cases of dissociations between PE and MB. A modification of the design, consisting of adding a small central peg on the PE allowed to well centre the PE before impaction on the metal tray. No more dissociation happened after this modification.
The total thickness of the implant is 6.5 mm. Our results showed an increasing in the lateral offset in comparison with a normal contralateral side. Another work (accepted as a free paper in the SOFCOT 2012 meeting) comparing the results between MB and cemented glenoid showed that this induced lateralisation did not influence the clinical results. Radiologically, no loosening, no lucent lines, no narrowing of the joint space, witness of a polyethylene wear were found. However, our tendency is actually to increase the PE thickness [26].
The rigidity of the metal back device was suspected to induce stress shielding and osteolysis under the metal [24, 32]. These experimental publications [33], in our knowledge, were never confirmed by clinical studies [34] and not confirmed in our work.
Some rotator cuff tears happened in our series, but with the same frequency than the most recent publications on cemented glenoid designs [17].
Despite the fact that, in the contrary than the literature [35, 36], the results were not statistically different between types a and b, the type of preoperative glenoid bone wear influenced our choice, as the only cases of dislocations were on posteriorly used glenoid bones, type b2 or c. In those cases where the posterior wear is not too important, we recommend to increase the anterior reaming, to add a posterior bone graft and eventually to utilise the long keeled implant, designed for revisions.
In old patients with a thin cuff, a cemented glenoid can be preferred to avoid any tension on the tendons. However, we prefer the noncemented glenoid, as, in case of secondary superior migration, the conversion in a reverse is facilitated, the same metal back tray been the support of the glenosphere. Moreover, this new design, thanks to the anterior plate and the 2 available directions of screwing, allows to extend the indications to big glenoid bone loss necessiting a bone graft [37].
However, this study presented some limitations:
-
This is a medium-term study with a mean follow-up of 3 years.
-
The measurement of the radiological lateral offset is technician dependent, and no scoring of the RLL has been used.
-
This is a multicentric not randomised study, and the choice of either a noncemented or a cemented prosthesis was often subjective and surgeon dependent .
However, in the view of our good and predictable results, our indications of using a MB glenoid in total anatomical shoulder arthroplasty are increasing (Fig. 4).
In conclusion, we did not find in this medium-term clinical and radiological review of a noncemented metal back glenoid implant, the classical complications pointed out in the literature for those uncemented glenoid implants.
With this new design, despite a radiological increase in the lateral offset, there is no proved risk for the cuff, no early polyethylene wear, no dissociation and the clinical results seem to be similar to the cemented glenoids but avoiding their frequent troubles such as evolutive lucent lines and loosenings.
References
Neer CS (1974) Replacement arthroplasty for glenohumeral osteoarthritis. J Bone Joint Surg Am 56:1–13
Bohsali KJ, Wirth MA, Rockwood CA (2006) Complications of total shoulder arthroplasty. J Bone Joint Surg Am 88:2279–2292
Brems J (1993) The glenoid component in total shoulder arthroplasty. J Shoulder Elbow Surg 2:47–54
Cheung EV, Sperling JW, Cofield RH (2008) Revision shoulder arthroplasty for glenoid component loosening. J Shoulder Elbow Surg 17:371–375
Sperling JW, Rowland CM, Neer CS (1998) Hemiarthroplasty and Neer total shoulder arthroplasty in patients fifty years old or less. Long-term results. J Bone Joint Surg Am 80:464–473
Torchia ME, Cofield RH, Settergren CR (1997) Total shoulder arthroplasty with the Neer prosthesis: long-term results. J Shoulder Elbow Surg 6:495–505
Katz D, O’Toole G, Cogswell L, Valenti P, Sauzières P (2007) A history of the reverse shoulder prosthesis. Int J Shoulder Surg 1:108–113
Orr TE, Cater DR, Shurman DJ (1988) Stress analyses of glenoid component designs. Clin Orthop 232:217–224
Cheung EV, Sperling JW, Cofield RH (2007) Polyethylene insert exchange for wear after total shoulder arthroplasty. J Shoulder Elbow Surg 16:574–578
Tammachote N, Sperling JW, Vathana T, Cofield RH, Harmsen WS, Schleck CD (2009) Long-term results of cemented metal-backed glenoid components for osteoarthritis of the shoulder. J Bone Joint Surg Am 91:160–166
Constant CR, Murley AH (1987) A clinical method of functional assessment of the shoulder. Clin Orthop Relat Res 204:160–164
Lippitt SB, Harryman DT, Matsen FA (1993) A practical tool for evaluating function: the simple shoulder test. In: Matsen FA, Fu FH, Hawkins RJ (eds) The Shoulder: a balance of mobility and stability. American Academy of Orthopedic Surgeons, Rosemont, pp 501–530
Walch G, Boileau P (1999) Prosthetic adaptability: a new concept for shoulder arthroplasty. J Shoulder Elbow Surg 8:443–451
LazarusMD Jensen KL, Southworth C, Matsen FA (2002) The radiographic evaluation of keeled and pegged glenoid component insertion. J Bone Joint Surg Am 84:1174–1182
Norris TR, Iannotti JP (2002) Functional outcome after shoulder arthroplasty for primary osteoarthritis. A multicenter study. J Shoulder Elbow Surg 11:130–135
Walch G, Edwards TB, Boulahia A, Boileau P, Mole D, Adeleine P (2002) The influence of glenohumeral prosthetic mismatch on glenoid lucent lines: results of a multicenter study. J Bone Joint Surg Am 84:2186–2191
Walch G, Young AA, Melis B, Gazielly D, Loew M, Boileau P (2011) Results of a convex-back cemented keeled glenoid component in primary osteoarthritis: multicenter study with a follow-up greater than 5 years. J Shoulder Elbow Surg 20:385–394
Katz D, Sauzières P, Valenti P, Kany J (2012) The case for the metal-backed glenoid design in total anatomical shoulder arthroplasty. Eur J Orthop Surg Traumatol 22:9–16. doi:10.1007/s00590-011-0796-8
Cofield RH (1994) Uncemented total shoulder arthroplasty: a review. Clin Orthop 307:86–93
Cofield RH, Daly P (1992) Total shoulder arthroplasty with a tissue ingrowth glenoid component. J Shoulder Elbow Surg 1:77–85
Martin SD, Zurakowski D, Thornhill T (2005) Uncemented glenoid component in Total Shoulder Arthroplasty. J Bone Joint Surg 87:1284–1289
Taunton MJ, McIntosh AL, Sperling JW, Cofield RH (2008) Total Shoulder Arthroplasty with a metal-backed bone-ingrowth glenoid component. J Bone Joint Surg Am 90:2180–2188
Churchill RS (2011) Trends in glenoid component design in unconstrained shoulder arthroplasty. J Shoulder Elbow Surg 20:S41–S46
Boileau P, Avidor C, Krishnan SG, Walch G, Kempf JF, Mole D (2002) Cemented polyethylene versus uncemented metal-back glenoid component in total shoulder arthroplasty:a prospective, double blind, randomized study. J Shoulder Elbow Surg 11:351–359
Castagna A, Randelli M, Garofalo R, Maradei L, Giardella A, Borroni M (2010) Mid-term results of a metal-backed glenoid component in total shoulder replacement. J Bone Joint Surg (Br) 92-B:1410–1415
Clement ND, Mathur K, Collins R, Stirrat AN (2010) The metal-back glenoid component in rheumatoid disease:eight to fourteen year follow-up. J Shoulder Elbow Surg 19:749–756
Anglin C, Wyss UP, Pichora DR (2000) Mechanical testing of shoulder prostheses and recommendations for glenoid design. J Shoulder Elbow Surg 9:323–331
Szabo I, Buscayret F, Edwards TB, Nemoz C, Boileau P, Walch G (2005) Radiographic comparison of flat-back and convex-back glenoid components in total shoulder arthoplasty. J Shoulder Elbow Surg 14:636–642
Iannotti JP, Spencer EE, Winter U, Deffenbaugh D, Williams G (2005) Prosthetic positioning in total shoulder arthroplasty. J Shoulder Elbow Surg 14(1 suppl S):111S–121S
Williams GR, Abboud JA (2005) Total shoulder arthroplasty: glenoid component design. J Shoulder Elbow Surg 14:122S–128S
Collin P, Tay AKL, Melis B, Boileau P, Walch G (2011) A ten-year radiologic comparison of two-all polyethylene glenoid component designs: a prospective trial. J Shoulder Elbow Surg 20:1217–1233
Stone KD, Grabowski JJ, Cofield RH, Morrey BF, An KN (1999) Stress analysis of glenoid components in total shoulder arthroplasty. J Shoulder Elbow Surg 8:151–158
Pelletier MH, Langdown A, Gilles RM, Sonnabend DH, Walsh WR (2008) Photoelastic comparison of strains in the underlying glenoid metal-back and all-polyethylene implants. J Shoulder Elbow Surg 127:779–783
Gupta S, van Dre Helm FC, Van Keulen F (2004) The possibilities of uncemented glenoid component. Clin Biomech (Bristol, Avon) 19:292–302
Barteld R, Sperling JW, Schleck CD, Cofield RH (2011) Arthroplasty in patients aged fifty-five years or younger with osteoarthritis. J Shoulder Elbow Surg 20:123–130. doi:10.1016/jse.2010.05.006
Farron A, Terrier A, Buchler P (2006) Risks of loosening of a prosthetic glenoid implanted in retroversion. J Shoulder Elbow Surg 15:521–526
Katz D, Valenti P, El Hadi A (2004) 14 cas de luxations invétérées de l’épaule traitées par arthroplastie. Réflexions pronostiques et thérapeutiques. Rev Chir Orthop 90(Sup 6):2 S 145
Conflict of interest
The authors have received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this manuscript.
Open Access
This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Katz, D., Kany, J., Valenti, P. et al. New design of a cementless glenoid component in unconstrained shoulder arthroplasty: a prospective medium-term analysis of 143 cases. Eur J Orthop Surg Traumatol 23, 27–34 (2013). https://doi.org/10.1007/s00590-012-1109-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00590-012-1109-6