Core herniation after implantation of a cervical artificial disc: case report

  • Case Report
  • Published:
European Spine Journal Aims and scope Submit manuscript

Abstract

Introduction

Cervical artificial discs (CADs) represent an established surgical option in selected patients with cervical spinal disc degeneration. Though CADs have been available for many years, there is a lack of information concerning long-term safety, durability and implant-related failure rates.

Materials and methods

The authors describe the failure of a M6-C CAD (Spinal Kinetics, Sunnyvale, CA, USA).

Results

Eight years after implantation of a CAD of the M6 type, a 39-year-old female presented with new clinical signs of cervical myelopathy. Radiologically, medullar compression due to posterior core herniation was the suspected cause. The damaged CAD was removed and the segment fused. During revision surgery, rupture of the posterior structures could be detected. Possible mechanisms leading to implant failure are discussed.

Conclusion

As there is no standard regarding clinical and radiological follow-up for patients with CADs, radiological long-term follow-up investigations seem to be justified for exclusion of implant failure.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Anderson PA, Hashimoto R (2012) Total disc replacement in the cervical spine: a systematic review evaluating long-term safety. Evid Based Spine Care J 3:9–18. doi:10.1055/s-0031-1298604

    PubMed Central  PubMed  Google Scholar 

  2. Lauryssen C, Coric D, Dimmig T, Musante D, Ohnmeiss DD, Stubbs HA (2012) Cervical total disc replacement using a novel compressible prothesis: results from a prospective Food and Drug Administration—regulated feasibility study with 24-month follow-up. Int J Spine Surg 6:71–77

    Article  PubMed Central  PubMed  Google Scholar 

  3. Patwardhan AG, Tzermiadianos MN, Tsitsopoulos PP, Voronov LI, Renner SM, Reo ML, Carandang G, Ritter-Lang K, Havey RM (2012) Primary and coupled motions after cervical total disc replacement using a compressible six-degree-of-freedom prosthesis. Eur Spine J 21(Suppl 5):S618–S629. doi:10.1007/s00586-010-1575-7

    Article  PubMed  Google Scholar 

  4. Reyes-Sanchez A, Miramontes V, Olivarez LM, Aquirre AA, Quiroz AO, Zarate-Kalfopulos B (2010) Initial clinical experience with a next-generation artificial disc for the treatment of symptomatic degenerative cervical radiculopathy. SAS J 4:9–15

    Article  PubMed Central  PubMed  Google Scholar 

  5. Thomas S, Willems K, Van den Daelen L, Linden P, Ciocci MC, Bocher P (2013) The M6-C cervical disc prosthesis: first clinical experience in 33 patients. J Spinal Disord Tech. doi:10.1097/BSD.0000000000000025

    PubMed  Google Scholar 

  6. Fan H, Wu S, Wu Z, Wang Z, Guo Z (2012) Implant failure of Bryan cervical disc due to broken polyurethane sheath: a case report. Spine (Phila Pa 1976) 37:E814–E816. doi:10.1097/BRS.0b013e3182477d85

  7. Anderson PA, Sasso RC, Rouleau JP, Carlson CS, Goffin J (2004) The Bryan Cervical Disc: wear properties and early clinical results. Spine J 4:303S–309S. doi:10.1016/j.spinee.2004.07.026

    Article  PubMed  Google Scholar 

Download references

Conflict of interest

The authors have no disclosures to make.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Christopher Brenke.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Brenke, C., Schmieder, K. & Barth, M. Core herniation after implantation of a cervical artificial disc: case report. Eur Spine J 24 (Suppl 4), 536–539 (2015). https://doi.org/10.1007/s00586-014-3677-0

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00586-014-3677-0

Keywords

Navigation