Prospective randomized comparison of radiation exposure from full spine radiographs obtained in three different techniques

  • Original Article
  • Published:
European Spine Journal Aims and scope Submit manuscript

Abstract

Problem: The purpose of the investigation was a comparison of two different digital X-ray techniques with conventional standing full spine films. Evaluation of dose area product, image quality and inter-observer error of Cobb-angle measurement in patients with scoliosis and kyphoscoliosis were studied.

Methods: A consecutive series of 150 patients were prospectively randomized into three groups. Patients in group 1 (n=53) received a conventional standing postero-anterior full spine radiograph. All films were evaluated on the light box. Patients in group 2 (n=48) received a X-ray using the digital storage phosphor plate system (CR). For group 3 (n=49) digital pulsed fluoroscopy was used. In groups 2 and 3 images were exported to a picture archiving and communicating system (PACS) workstation and viewed on a monitor (Siemens SMM 21140P, Germany). Dose area product measurements were performed in all three groups (Diamentor-M, PTW, Freiburg). Three experienced investigators independently reviewed all pictures. Pedicles and endplates were counted. Cobb-angles of the main curves were measured.

Results: The mean dose area product was 97.0 cGy cm² (37.0–380.0 cGy cm²) for conventional films, 31.5 cGy cm² (6.0–66.0 cGy cm²) for CR imaging and 5.0 cGy cm² (1.0–29.0 cGy cm²) for digital fluoroscopy. The differences of Cobb-angle measurements were not significantly different for the three methods. Differences in the count of pedicles and endplates between the investigators were significantly lower for the conventional film as an indicator for the best detail presentation.

Conclusion: A significant reduction in dose area product is possible with modern digital X-ray methods. The inter-observer error of Cobb-angle measurement is not significantly altered. The detail information is decreased in comparison to conventional films.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Carman DL, Browne RH, Birch JG (1990) Measurements of scoliosis and kyphosis radiographs: intraobserver and interoberserver variation. J Bone Joint Surg 72:328–333

    PubMed  Google Scholar 

  2. Geijer H, Beckmann KW, Jonsson B, Andersson T, Persliden J (2001) Digital radiography of scoliosis with a scanning method: initial evaluation. Radiology 218:402–410

    PubMed  Google Scholar 

  3. Geijer H, Verdonck B, Beckman KW, Andersson T, Persliden J (2003) Digital radiography of scoliosis with a scanning method: radiation dose optimization. Eur Radiol 13:543–551

    PubMed  Google Scholar 

  4. Gross C, Gross M, Kuschner S (1983) Error analysis of scoliosis curvature measurement. Bull Hosp Joint Dis Orthop Inst 43:171–177

    Google Scholar 

  5. Hellström G, Irstam L, Nachemson A (1983) Reduction of radiation dose in radiologic examination of patients with scoliosis. Spine 8:28–30

    PubMed  Google Scholar 

  6. Levy AR, Goldberg MS, Mayo NE, Hanley JA, Poitras B (1996) Reducing the lifetime risk of cancer from spinal radiographs among people with adolescent idiopathic scoliosis. Spine 21:1540–1547

    Article  PubMed  Google Scholar 

  7. Lonstein JE, Carlson JM (1984) The prediction of curve progression in untreated idiopathic scoliosis during groth. J Bone Joint Surg 66:1061–1071

    PubMed  Google Scholar 

  8. Nash CL Jr, Gregg EC, Brown RH, Pillai K (1979) Risks of exposure to x-rays in patients undergoing long term treatment for scoliosis. J Bone Joint Surg 61:371–374

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Torsten Kluba.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kluba, T., Schäfer, J., Hahnfeldt, T. et al. Prospective randomized comparison of radiation exposure from full spine radiographs obtained in three different techniques. Eur Spine J 15, 752–756 (2006). https://doi.org/10.1007/s00586-005-1005-4

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00586-005-1005-4

Keywords

Navigation