Abstract
Purpose
The purpose of this study was to compare the ultrasound image quality at three different transducer positions for ultrasound-guided lumbar plexus block (LPB).
Methods
This prospective comparative study included 30 patients who underwent total hip arthroplasty under general anesthesia in combination with LPB. Using the same ultrasound machine settings for each patient, a transverse view of the lumbar plexus (LP) at the L3–4 vertebral level was obtained with a convex transducer placed at three different positions: immediately lateral to the dorsal midline (medial position), almost 5 cm lateral to the dorsal midline (paravertebral position), and at the abdominal transverse flank (shamrock position). Ultrasound-guided LPB with catheter insertion was performed via in-plane needle insertion with the transducer randomly assigned to one of the three positions. The echo intensity (EI) ratio of the LP to psoas major muscle (PMM), the EI of the LP and PMM, and the ultrasound visibility score of the needle, local anesthetic, and catheter were recorded.
Results
The LP/PMM EI ratio was significantly higher at paravertebral position (1.4 ± 0.2) than at medial position (1.2 ± 0.2; p = 0.003) and shamrock position (1.3 ± 0.2; p = 0.040). The EI of the LP and PMM was highest at shamrock position (p < 0.001). During the block procedure, the ultrasound visibility score of the needle and local anesthetic was significantly higher at paravertebral position than at medial position.
Conclusion
Under the conditions of this study, the contrast between LP and PMM is significantly higher at paravertebral position than at medial position and at the abdominal transverse flank (shamrock position). LP and PMM at the shamrock position appear significantly brighter among the three probe positions in sonograms.
Similar content being viewed by others
References
Doi K, Sakura S, Hara K. A modified posterior approach to lumbar plexus block using a transverse ultrasound image and an approach from the lateral border of the transducer. Anaesth Intensive Care. 2010;38:213–4.
Karmakar MK, Ho AH, Li X, Kwok WH, Tsang K, Ngan Kee WD. Ultrasound-guided lumbar plexus block through the acoustic window of the lumbar ultrasound trident. Br J Anaesth. 2008;100:533–7.
Karmakar MK, Li JW, Kwok WH, Kwok WH, Hadzic A. Ultrasound-guided lumbar plexus block using a transverse scan through the lumbar intertransverse space. Reg Anesth Pain Med. 2015;40:75–81.
Sauter AR, Ullensvang K, Bendtsen TF, Børglum J. The “Shamrock Method”—a new and promising technique for ultrasound guided lumbar plexus blocks. Br J Anaesth. 2013. https://doi.org/10.1093/bja/el_9814.
de Luise C, Brimacombe M, Pedersen L, Sørensen HT. Comorbidity and mortality following hip fracture: a population-based cohort study. Aging Clin Exp Res. 2008;20:412–8.
Chidambaram R, Cobb AG. Change in the age distribution of patients undergoing primary hip and knee replacements over 13 years—an increase in the number of younger men having hip surgery. J Bone Jt Surg Br. 2009;91-B(SUPP I):152.
Chan VWS. Ultrasound imaging for regional anesthesia. 2nd ed. Toronto: Toronto Printing Company; 2009. pp. 19–41.
Li X, Karmakar MK, Lee A, Kwok WH, Critchley LAH, Gin T. Quantitative evaluation of the echo intensity of the median nerve and flexor muscles of the forearm in the young and the elderly. Br J Radiol. 2012;85:e140–5.
Dogan Z, Bakan M, Idin K, Esen A, Uslu FB, Ozturk E. Total spinal block after lumbar plexus block: a case report. Braz J Anesthesiol. 2014;64:121–3.
Vadi MG, Patel N, Stiegler MP. Local anesthetic systemic toxicity after combined psoas compartment-sciatic nerve block: analysis of decision factors and diagnostic delay. Anesthesiology. 2014;120:987–96.
Maurits NM, Bollen AE, Windhausen A, De Jager AEJ, Van Der Hoeven JH. Muscle ultrasound analysis: normal values and differentiation between myopathies and neuropathies. Ultrasound Med Biol. 2003;29:215–25.
Sipilä S, Suominen H. Muscle ultrasonography and computed tomography in elderly trained and untrained women. Muscle Nerve. 1993;16:294–300.
Karmakar MK, Li JW, Kwok WH, Kwok WH, Soh E, Hadzic A. Sonoanatomy relevant for lumbar plexus block in volunteers correlated with cross-sectional anatomic and magnetic resonance images. Reg Anesth Pain Med. 2013;38:391–7.
Strid JMC, Sauter AR, Ullensvang K, Andersen MN, Daugaard M, Bendtsen MAF, Søballe K, Pedersen EM, Børglum J, Bendtsen TF. Ultrasound-guided lumbar plexus block in volunteers; a randomized controlled trial. Br J Anaesth. 2017;118:430–8.
Acknowledgements
We thank the nursing staff and the technical staff at the Department of Device Technologies of Asahikawa Medical University Hospital.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
Makoto Sato has no conflict of interest; Tomoki Sasakawa has no conflict of interest; Yuki Izumi has no conflict of interest; Yoshiko Onodera has no conflict of interest; Takayuki Kunisawa has no conflict of interest.
About this article
Cite this article
Sato, M., Sasakawa, T., Izumi, Y. et al. Ultrasound-guided lumbar plexus block using three different techniques: a comparison of ultrasound image quality. J Anesth 32, 694–701 (2018). https://doi.org/10.1007/s00540-018-2539-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00540-018-2539-z