Skip to main content
Log in

VATS-guided epicardial pacemaker implantation

Hand-sutured fixation of atrioventricular leads in an experimental setting

  • Published:
Surgical Endoscopy Aims and scope Submit manuscript

Abstract

Background: In neonates and infants epicardial stimulation may be preferred to endocardial stimulation because of growth-associated lead problems and the risk of vascular complications associated with transvenous electrodes. This study analyzes the feasibility of atrioventricular implantation of a new epicardial lead using the video-assisted thoracic surgical (VATS) technique in an animal model.

Methods: Bipolar steroid-eluting epicardial leads were implanted in seven young white pigs. In five animals bipolar atrial and ventricular pacing leads (n= 10) were inserted and fixed by the VATS technique, while two animals served as controls and underwent implantation through anterolateral thoracotomy. Surgical feasibility, pacing, and sensing thresholds of the leads as well as hemodynamic parameters during pacing were studied. Histological changes beneath the electrodes were evaluated 1 week after the implantation.

Results: All animals survived the pacemaker lead implantation. One animal which underwent thoracotomy died because of irreversible ventricular fibrillation induced by rapid ventricular pacing. One animal in the VATS group exhibited intraoperative herniation of the heart through the pericardial window. All animals with left-sided VATS implantations demonstrated good individual pacing and sensing threshold values. The mean cardiac output was 1.6 times higher during AAI-mode pacing as compared to VVI-mode pacing at a heart rate of 140/min. One animal died postoperatively due to respiratory failure. No displacements of the pacemaker leads were observed in the survivors.

Conclusion: While VATS-guided implantation of epicardial, atrial, and ventricular leads is feasible, technical improvements of the system are mandatory for safe clinical application.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 14 October 1996/Accepted: 14 April 1997

Rights and permissions

Reprints and permissions

About this article

Cite this article

Furrer, M., Fuhrer, J., Altermatt, H. et al. VATS-guided epicardial pacemaker implantation. Surg Endosc 11, 1167–1170 (1997). https://doi.org/10.1007/s004649900562

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s004649900562

Navigation