Abstract
Background
The main objective is related to the capability of integrating into minimally invasive and ultra-thin disposable micro-endoscopic tool, a modality of realizing high-resolution imaging through scattering medium such as blood while performing medical procedure. In this research we aim for the first time to present a time-multiplexing super-resolving approach exhibiting enhanced focus sensitivity, generated by 3D spatial filtering, for significant contrast increase in images collected through scattering medium.
Method
Our innovative method of imaging through scattering media provides imaging of only one specific object plane in scattering medium’s volume while suppressing the noise coming from all other planes. The method should be assisted with axial scanning to perform imaging of the entire 3D object’s volume. In our developed optical system noise suppression is achieved by 3D spatial filtering approach while more than an order of magnitude of suppression is experimentally demonstrated. The sensitivity to defocus and noise suppression is dramatically enhanced by placing an array of micro-lenses combined with pinholes raster positioned between two modules of telecentric lenses.
Results
We present our novel conceptual designs for the enhanced signal-to-noise ratio (SNR) when imaging through scattering medium and present preliminary experimental results demonstrating both quality imaging performed on resolution bars target as well as SNR quantified results in which SNR enhancement of more than one order of magnitude was obtained.
Conclusions
In this paper, to the best of our knowledge, we present the first ever design of time-multiplexing-based approach for super-resolved imaging through scattering medium. The approach includes a time-multiplexing optical design significantly increasing the depth of focus sensitivity and after performing axial scanning yielding a significant enhancement of the SNR of the 3D object that is being imaged through the scattering medium. Right after the contrast (the SNR) enhancement we scan the object with the projected array of spots (raster) and map it continuously and with high imaging resolution.
Similar content being viewed by others
References
Matz G, Messerschmidt B, Göbel W et al (2017) “Chip-on-the-tip compact flexible endoscopic epifluorescence video-microscope for in-vivo imaging in medicine and biomedical research. Biomed Opt Express. https://doi.org/10.1364/BOE.8.003329
Shimoji K, Fujioka H, Onodera M et al (1991) Observation of spinal canal and cisternae with the newly developed small-diameter, flexible fiberscopes. Anesthesiology 75(2):341–344. https://doi.org/10.1097/00000542-199108000-00024 (PMID: 1859021)
Shahmoon A, Aharon S, Kruchik O, Hohmann M, Slovin H, Douplik A, Zalevsky Z (2013) In vivo minimally invasive interstitial multi-functional microendoscopy. Sci Rep. https://doi.org/10.1038/srep01805
Douplik A, Hohmann M, Shahmoon A, Zam A, Zalevsky Z, Schmidt M, and Schaaf H (2013) “Microendoscopy of small ducts as a potential tool for guiding and monitoring intruductual biopsy and therapy,” SPIE BIOS conf. of Advanced Biomedical and Clinical Diagnostic Systems XI, Conference 8572
Shahmoon A, Zalevsky Z (2017) “Multi Core Needle Endoscopy”, Society of American Gastrointestinal and endoscopic surgeons SAGES. Houston, USA
Wagner O., Pandya A., Chemla Y., Pinhas H., Schelkanova I., Shahmoon A., Mandel Y., Douplik A. and Zalevsky Z., 2018 “Lens-less Micro-endoscopy through highly scattering media,” CLEO (Conference on Lasers and Electro-Optics), San Jose, USA
Lengenfelder B, Mehari F, Tang Y, Klämpfl F, Zalevsky Z, Schmidt M (2017) Towards non-contact photo-acoustic endoscopy using speckle pattern analysis. Photons Plus Ultrasound Imaging and Sensing 10064:442–447
Brezinski ME, Tearney GJ, Bouma BE, Boppart SA, Hee MR, Swanson EA, Southern JF, Fujimoto JG (1996) Imaging of coronary artery microstructure (in vitro) with optical coherence tomography. Am J Cardiol 77(1):92–93. https://doi.org/10.1016/S0002-9149(97)89143-6
Kawasaki M (2013) Optical coherence tomography for coronary artery plaques – a comparison with intravascular ultrasound. IntechOpen book publisher. https://doi.org/10.5772/56293
Katz O, Heidmann P, Fink M et al (2014) Non-invasive single-shot imaging through scattering layers and around corners via speckle correlations. Nature Photon 8:784–790. https://doi.org/10.1038/nphoton.2014.189
Sanjeev A, Kapellner Y, Shabairou N, Gur E, Sinvani M, Zalevsky Z (2019) Non-invasive imaging through scattering medium by using a reverse response wavefront shaping technique. Sci Rep 9(1):1–11
Sanjeev A, Trivedi V, Zalevsky Z (2022) Optical reciprocity induced wavefront shaping for axial and lateral shifting of focus through a scattering medium. Sci Rep 12(1):1–15
Wagner O, Shahmoon A, Zalevsky Z (2019) “Imaging Through Blood Super-Resolution Based Flexible Microendoscope”, Scientific session of the Society of American Gastrointestinal and endoscopic surgeons SAGES. Baltimore, USA
Zalevsky Z., Shahmoon A., Meiri A., Herman O., Elkabetz S. and Rudnitsky A.,2022 “System and method for imaging through scattering medium,” 17/696935 PCT application.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Disclosure
Zeev Zalevsky, Shimon Elkabetz, Arkady Rudnitsky, Oran Herman, Amihai Meiri and Asaf Shahmoon all have financial ties and equity interest in Zsquare.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Springer Nature or its licensor holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Zalevsky, Z., Elkabetz, S., Rudnitsky, A. et al. S194–Imaging through scattering media by 3D spatial filtering embedded into micro-endoscope. Surg Endosc 37, 3162–3172 (2023). https://doi.org/10.1007/s00464-022-09511-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-022-09511-4