Abstract

Background

Covered self-expandable metal stents (SEMS) are utilized for the management of benign and malignant esophageal conditions; however, covered SEMS are prone to migration. Endoscopic suture fixation may mitigate the migration risk of covered esophageal SEMS. Hence, we conducted a systematic review and meta-analysis to evaluate the effectiveness and safety of endoscopic suture fixation for covered esophageal SEMS.

Methods

Following PRISMA guidelines, we performed a systematic review from 2011 to 2016 to identify studies (case control/case series) reporting the technical success and migration rate of covered esophageal SEMS following endoscopic suture fixation. We searched multiple electronic databases and conference proceedings. We calculated pooled rates (and 95% confidence intervals [CI]) of technical success and stent migration using a random effects model.

Results

We identified 14 studies (212 patients) describing covered esophageal SEMS placement with endoscopic suture fixation. When reported, SEMS indications included leak/fistula (n = 75), stricture (n = 65), perforation (n = 10), and achalasia (n = 4). The pooled technical success rate was 96.7% (95% CI 92.3–98.6), without heterogeneity (I 2 = 0%). We identified 29 SEMS migrations at rate of 15.9% (95% CI 11.4–21.6), without heterogeneity (I 2 = 0%). Publication bias was observed, and using the trim-and-fill method, a more conservative estimate for stent migration was 17.0%. Suture-related adverse events were estimated to occur in 3.7% (95% CI 1.6–8.2) of cases.

Conclusions

Endoscopic suture fixation of covered esophageal SEMS appears to reduce stent migration when compared to published rates of non-anchored SEMS. However, SEMS migration still occurs in approximately 1 out of 6 cases despite excellent immediate technical success and low risk of suture-related adverse events.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

Abbreviations

SEMS:

Self-expandable metal stents

PC-SEMS:

Partially covered self-expandable metal stents

FC-SEMS:

Fully covered self-expandable metal stents

TTS:

Through-the-scope

OTS:

Over-the-scope

PRISMA:

Preferred Reporting Items for Systematic Reviews and Meta-Analyses

SEPS:

Self-expandable plastic stent

CI:

Confidence interval

References

  1. Bakken JC, Wong Kee Song LM, de Groen PC, Baron TH (2010) Use of a fully covered self-expandable metal stent for the treatment of benign esophageal diseases. Gastrointest Endosc 72:712–720

    Article  PubMed  Google Scholar 

  2. Dan DT, Gannavarapu B, Lee JG, Chang K, Muthusamy VR (2014) Removable esophageal stents have poor efficacy for the treatment of refractory benign esophageal strictures (RBES). Dis Esophagus 27:511–517

    Article  CAS  PubMed  Google Scholar 

  3. Sharma P, Kozarek R, Practice Parameters Committee of American College of Gastroenterology (2010) Role of esophageal stents in benign and malignant diseases. Am J Gastroenterol 105:258–273 quiz 74

    Article  PubMed  Google Scholar 

  4. Wang C, Lou C (2015) Randomized controlled trial to investigate the effect of metal clips on early migration during stent implantation for malignant esophageal stricture. Can J Surg 58:378–382

    Article  PubMed  PubMed Central  Google Scholar 

  5. Vanbiervliet G, Filippi J, Karimdjee BS, Venissac N, Iannelli A, Rahili A et al (2012) The role of clips in preventing migration of fully covered metallic esophageal stents: a pilot comparative study. Surg Endosc 26:53–59

    Article  PubMed  Google Scholar 

  6. Irani S, Baron TH, Gluck M, Gan I, Ross AS, Kozarek RA (2014) Preventing migration of fully covered esophageal stents with an over-the-scope clip device (with videos). Gastrointest Endosc 79:844–851

    Article  PubMed  Google Scholar 

  7. Kantsevoy SV, Bitner M (2012) Esophageal stent fixation with endoscopic suturing device (with video). Gastrointest Endosc 76:1251–1255

    Article  PubMed  Google Scholar 

  8. Fujii LL, Bonin EA, Baron TH, Gostout CJ, Wong Kee Song LM (2013) Utility of an endoscopic suturing system for prevention of covered luminal stent migration in the upper GI tract. Gastrointest Endosc 78:787–793

    Article  PubMed  Google Scholar 

  9. Sharaiha RZ, Kumta NA, DeFilippis EM, Dimaio CJ, Gonzalez S, Gonda T et al (2016) A large multicenter experience with endoscopic suturing for management of gastrointestinal defects and stent anchorage in 122 patients: a retrospective review. J Clin Gastroenterol 50:388–392

    PubMed  Google Scholar 

  10. Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gotzsche PC, Ioannidis JP et al (2009) The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. Ann Intern Med 151:W65–W94

    Article  PubMed  Google Scholar 

  11. Viswanathan M, Ansari MT, Berkman ND, Chang S, Hartling L, McPheeters M et al (2012) Assessing the risk of bias of individual studies in systematic reviews of health care interventions. In: Methods guide for effectiveness and comparative effectiveness reviews. US Department of Health and Human Services, Rockville

  12. Guyatt GH, Oxman AD, Kunz R, Brozek J, Alonso-Coello P, Rind D et al (2011) GRADE guidelines 6. Rating the quality of evidence–imprecision. J Clin Epidemiol 64:1283–1293

    Article  PubMed  Google Scholar 

  13. Egger M, Davey Smith G, Schneider M, Minder C (1997) Bias in meta-analysis detected by a simple, graphical test. BMJ 315:629–634

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  14. Easterbrook PJ, Berlin JA, Gopalan R, Matthews DR (1991) Publication bias in clinical research. Lancet 337:867–872

    Article  CAS  PubMed  Google Scholar 

  15. Duval S, Tweedie R (2000) Trim and fill: a simple funnel-plot-based method of testing and adjusting for publication bias in meta-analysis. Biometrics 56:455–463

    Article  CAS  PubMed  Google Scholar 

  16. Bick BL, Imperiale TF, Calley C, DeWitt JM (2016) Endoscopic suturing of fully covered SEMS for benign esophageal disease decreases migration rates compared to unsecured stents. Gastrointest Endosc 83:AB579–AB580

    Article  Google Scholar 

  17. Catalano MF, Sorser SA, Henderson JB, Ali S, Atia A (2014) Esophageal stent fixation using a novel endoscopic suturing device. Gastrointest Endosc 79:AB518

    Article  Google Scholar 

  18. Fakhreddine A, Datta A, Hunt D, Chavalitdhamrong D, Eysselein VE, Reicher S (2015) Endoscopic suturing system for fully covered metal stent fixation: initial experience in a safety net population. Gastrointest Endosc 81:AB457

    Article  Google Scholar 

  19. Gonzalez-Panizo-Tamargo F, Rojas-Sanchez A, Miranda-Garcia P, Juzgado-Lucas D, Vazquez-Sequeiros E (2014) Endoscopic management of common patologies with a new suturing system. Rev Esp Enferm Dig 106:413–417

    PubMed  Google Scholar 

  20. Rieder E, Dunst CM, Martinec DV, Cassera MA, Swanstrom LL (2012) Endoscopic suture fixation of gastrointestinal stents: proof of biomechanical principles and early clinical experience. Endoscopy 44:1121–1126

    Article  CAS  PubMed  Google Scholar 

  21. Sendino O, Kane SD, Binmoeller K, Shah JN, Hamerski CM, Bhat YM (2015) Comparison of migration rates of FCSEMS after stent anchoring using a novel endoscopic suturing system (Overstitch) with standard stent placement in benign upper GI disease. Gastrointest Endosc 81:AB449–AB450

    Google Scholar 

  22. Stavropoulos SN, Modayil RJ, Friedel D (2015) Endoscopic suturing. 2 Year experience of a high volume center. Gastrointest Endosc 81:AB394

    Article  Google Scholar 

  23. Wright A, Chang A, Bedi AO, Wamsteker EJ, Elta G, Kwon RS et al (2016) Endoscopic suture fixation is associated with reduced migration of esophageal fully covered self-expandable metal stents (FCSEMS). Surg Endosc. doi:10.1007/s00464-016-5374-z

    Google Scholar 

  24. Yang J, Siddiqui AA, Kowalski TE, Loren DE, Khalid A, Soomro A et al (2017) Esophageal stent fixation with endoscopic suturing device improves clinical outcomes and reduces complications in patients with locally advanced esophageal cancer prior to neoadjuvant therapy: a large multicenter experience. Surg Endosc 31:1414–1419

    Article  PubMed  Google Scholar 

  25. Johnson BJ, Gitelis ME, Linn JG, Denham W, Carbray J, Ujiki MB (2015) Multipurpose use of an endoscopic suturing device. Gastrointest Endosc 148:S-1139

    Google Scholar 

  26. DeFelice KM, Wong Kee Song LM, Buttar N (2015) Fixation of esophageal self-expanding metal stents with endoscopic suturing reduces stent migration. Gastrointest Endosc 81:AB528

    Article  Google Scholar 

  27. Rieder E, Asari R, Lenglinger J, Schoppmann SF (2014) Endolumenal stent suture-fixation compared to endoscopic clips for prevention of esophageal stent migration during prolonged dilatation in recurrent achalasia. Gastroenterology 146:S-1041

    Article  Google Scholar 

  28. Law R, Baron TH (2016) Choosing the appropriate esophageal stent for your patient. Am J Gastroenterol 111:1669–1671

    Article  PubMed  Google Scholar 

  29. Suzuki T, Siddiqui A, Taylor LJ, Cox K, Hasan RA, Laique SN et al (2016) Clinical outcomes, efficacy, and adverse events in patients undergoing esophageal stent placement for benign indications: a large multicenter study. J Clin Gastroenterol 50:373–378

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Contributions

RL: Conception of article, drafting of the manuscript, final manuscript approval. AP, LF-L: Critical review, literature review, final manuscript approval. CS: Literature search, drafting of the manuscript. SS: Critical review, statistical analysis.

Corresponding author

Correspondence to Ryan Law.

Ethics declarations

Disclosures

Ryan Law, Anoop Prabhu, Larissa Fujii-Lau, Carol Shannon, and Siddharth Singh have no conflicts of interest or financial ties to disclose.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary material 1 (DOCX 28 kb)

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Law, R., Prabhu, A., Fujii-Lau, L. et al. Stent migration following endoscopic suture fixation of esophageal self-expandable metal stents: a systematic review and meta-analysis. Surg Endosc 32, 675–681 (2018). https://doi.org/10.1007/s00464-017-5720-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00464-017-5720-9

Keywords

Navigation