Skip to main content
Log in

High-pressure liquid chromatographic determination of toluenein urine as a marker of occupational exposure to toluene

  • ORIGINAL ARTICLE
  • Published:
International Archives of Occupational and Environmental Health Aims and scope Submit manuscript

Abstract

Objective: To establish a convenient method by high-pressure liquid chromatography (HPLC) to measure toluene in urine as a marker of occupational exposure to toluene. Methods: As soon after sampling as possible, 1 ml of urine was mixed with an equal volume of acetonitrile in a 2.2-ml HPLC glass bottle, and the bottle was tightly sealed and stored at 4 °C. Immediately before HPLC determination, 100 μl methanol was added to the mixture to prevent confounding effects of glycosuria, and the bottle was spun to remove any suspended matter. An aliquot of the supernate was introduced into the HPLC system and analyzed on a PRODIGY column, with an acetonitrile – perchloric acid – phosphoric acid – water mixture serving as the mobile phase. The effluent was monitored at 191 nm. Results: The method can measure toluene in urine every 20 min; the detection limit was 2 μg/l, the coefficient of variation was less than 5%, and the recovery rate was 100%. No significant reduction in toluene concentration was observed for 1 week after storage at 4 °C. When the method was applied to end-of-shift urine samples from 13 male workers exposed to toluene at 18–140 ppm and also to urine samples from 10 nonexposed male controls, toluene in urine was linearly related to toluene exposure concentration, with a regression line passing close to the origin. The correlation coefficient was as high as 0.97 (n = 23). No toluene was detected in control urine samples. Calculations suggest that urinary toluene accounts for as little as less than 0.01% of the toluene absorbed via inhalation and that the absorbed toluene is converted almost quantitatively to hippuric acid and, by less than 0.1%, to o-cresol.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 25 August 1997 / Accepted: 13 February 1998

Rights and permissions

Reprints and permissions

About this article

Cite this article

Inoue, O., Kanno, E., Kudo, S. et al. High-pressure liquid chromatographic determination of toluenein urine as a marker of occupational exposure to toluene. Int Arch Occup Environ Health 71, 302–308 (1998). https://doi.org/10.1007/s004200050285

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s004200050285

Navigation