Abstract
Purpose
To review the existing technology for self-tonometry and evaluate methods for continuous monitoring of IOP currently undergoing development and clinical trials in humans.
Background
Glaucoma is one of the leading causes of blindness worldwide. Current glaucoma therapy is focused primarily on lowering intraocular pressures (IOP). Decisions to modify treatment regiments are primarily based on office IOP. Since IOP fluctuates throughout the day, values obtained in the office may be a poor representation of the patient’s disease. IOP measurements outside of the physician’s office environment would provide better knowledge of the disease state and allow for better-informed medical decision making.
Methods and results
We performed a literature search using Medline and IEEE database for studies investigating technologies that have been developed for continuous 24-hour IOP monitoring.
Conclusion
There is currently no technology that has been approved for use to allow for continuous monitoring of IOP fluctuations. New experimental technology being developed and currently undergoing clinical trials has demonstrated potential for changing the diagnosis and management of glaucoma.
Similar content being viewed by others
References
Resnikoff S, Pascolini D, Etya’ale D et al. (2004) Global data on visual impairment in the year 2002. Bull World Health Organ 82(11):844–851
Quigley HA, Broman AT (2006) The number of people with glaucoma worldwide in 2010 and 2020. Br J Ophthalmol 90(3):262–267
Goldmann H, Schmidt T (1957) Applanation Tonometry. Ophthalmologica 134(4):221–242
Wessels IF, Oh Y (1990) Tonometer utilization, accuracy, and calibration under field conditions. Arch Ophthalmol 108(12):1709–112
Tonnu PA, Ho T, Newson T et al. (2005) The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry. Br J Ophthalmol 89(7):851–854
Collaer N, Zeyen T, Caprioli J (2005) Sequential office pressure measurements in the management of glaucoma. J Glaucoma 14(3):196–200
Asrani S, Zeimer R, Wilensky J et al. (2000) Large diurnal fluctuations in intraocular pressures are an independent risk factor in patients with glaucoma. J Glaucoma 9:134–142
Caprioli J, Coleman AL (2008) Intraocular Pressure Fluctuation: A risk factor for visual field progression at low intraocular pressures in the Advanced Glaucoma Intervention Study. Ophthalmology 115(7):1123–1129
McLaren JW, Brubaker RF, FitzSimon JS (1996) Continuous measurement of intraocular pressure in rabbits by telemetry. Investig Ophthalmol Vis Sci 37(6):966–975
Carlson KH, McLaren JW, Topper JE et al. (1987) Effect of body position on intraocular pressure and aqueous flow. Invest Ophthalmol Vis Sci 28(8):1346–1352
Yamabayashi S, Aguilar RN, Hosoda M et al. (1991) Postural change of intraocular and blood pressures in ocular hypertension and low tension glaucoma. Br J Ophthalmol 75(11):652–655
Kothe AC (1994) The effect of posture on intraocular pressure and pulsatile ocular blood flow in normal and glaucomatous eyes. Surv Ophthalmol 38:S191–S197
Smedt SD, Mermound A, Schnyder C (2012) 24-Hour intraocular pressure fluctuation monitoring using an ocular telemetry sensor: Tolerability and functionality in healthy subjects. J Glaucoma 21(8):539–544
Hara T, Hara T, Tsuru T (2006) Increase of peak intraocular pressure during sleep in reproduced diurnal changes by posture. Arch Ophthalmol 124(2):165–168
Liu JKH, Kripke DF, Hoffman RE et al. (1998) Nocturnal elevation of intraocular pressure in young adults. Invest Ophthalmol Vis Sci 39:2707–2712
Liu JKH, Bouligny RP, Kripke DF et al. (2003) Nocturnal elevation of intraocular pressure is detectable in the sitting position. Invest Ophthalmol Vis Sci 44(10):4439–4442
Brown B, Morris P, Muller C et al. (1988) Fluctuations in intra-ocular pressure with sleep: I. Time course of IOP increase after the onset of sleep. Ophthalmic Physiol Opt 8(3):246–248
Brown B, Burton P, Mann S et al. (1988) Fluctuations in intra-ocular pressure with sleep: II. Time course of IOP decrease after waking from sleep. Ophthalmic Physiol Opt 8(3):249–252
Orzalesi N, Rossetti L, Invernizzi T et al. (2000) Effect of timolol, latanoprost, and dorzolamide on circadian IOP in glaucoma or ocular hypertension. Invest Ophthal VisSci 41(9):2566–2573
Orzalesi N, Rossetti L, Bottoli A et al. (2003) The effect of latanoprost, brimonidine, and a fixed combination of timolol and dorzolamide on circadian intraocular pressure in patients with glaucoma or ocular hypertension. Arch Ophthalmol 121(4):453–457
De Alva M, Jervell J (1997) Self monitoring improves quality of life and prognosis of people with diabetes. Br Med J 315:184–185
Hughes E, Spry P, Diamond J (2003) 24-Hour monitoring of intraocular pressure in glaucoma management: a retrospective review. J Glaucoma 12:232–236
Draeger J, Schwartz R, Groenhoff et al. (1993) Self-tonometry under microgravity conditions. Clin Investigator 71:700–703
Theofylaktopoulos I, Diestelhorst M, Krieglstein GK (1999) Self-tonometry with the Ocuton S versus Goldmann tonometry. Graefes Arch Clin Exp Ophthalmol 237:720–724
Sacu S, Vass C, Schemper M et al. (2004) Self-tonometry with the Ocuton S: evaluation of accuracy in glaucoma patients. Acta Ophthalmol Scand 82:405–409
Marchini G, Babighian S, Specchia L et al. (2002) Evaluation of the new Ocuton S tonometer. Acta Ophthalmol Scand 80:167–171
Kothy P, Vargha P, Hollo G (2001) Ocuton S self-tonometry vs Goldmann tonometry; a diurnal comparison study. Acta Ophthalmologica Scandanavica 79:294–297
Lanfermann E, Jurgens C, Grossjohann R et al. (2009) Intraocular pressure measurements with the newly Ocuton S*TT-MV self-tonometer in comparison to Goldmann applanation tonometry in glaucoma patients. Med Sci Monit 15(11):CR556–CR562
Rosenwasser GO, Holland S, Pflugfelder SC et al. (1990) Topical anesthetic abuse. Ophthalmology 97(8):967–972
McGee HT, Fraunfelder FW (2007) Toxicities of topical ophthalmic anesthetics. Expert Opin Drug Saf 6(6):637–640
Yeniad B, Canturk S, Esin Ozdemir F et al. (2010) Toxic keratopathy due to abuse of topical anesthetic drugs. Cutan Ocul Toxicol 29(2):105–109
Lam DSC, Leung DYL, Chiu TYH et al. (2004) Pressure phosphene self-tonometry: A comparison with Goldmann Tonometry in Glaucoma Patients. Investig Ophthalmol Vis Sci 45(9):3131–3136
Tai MC, Chen PL, Wu JN et al. (2005) Clinical evaluation of the intraocular pressure in patients with glaucoma or ocular hypertension by a self-assessable tonometer. J Ocul Pharmacol Ther 21(1):55–61
Danesh-Meyer HV, Niederer R, Gaskin BJ, Gamble (2004) Comparison of the Proview pressure phosphene tonometer performed by the patient and examiner with the Goldmann applanation tonometer. Clin Exp Ophthalmol 32(1):29–32
Rai S, Moster MR, Kesen M et al. (2005) Level of disagreement between Proview Phosphene Tonometer and Goldmann Applanation Tonometer Intraocular Pressure Readings. J Glaucoma 14:120–123
Li J, Herndon LW, Asrani SG et al. (2004) Clinical comparison of the Proview Eye Pressure Monitor with Goldmann Applanation Tonometer and the TonoPen. Arch Ophthalmol 122:1117–1121
Rosentreter A, Jablonski KS, Mellein AC et al. (2011) A new rebound tonometer for home monitoring of intraocular pressure. Graefes Arch Clin Exp Ophthalmol 249(11):1713–1719
Asrani S, Chatterjee A, Wallace DK et al. (2011) Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device. J Glaucoma 20:74–79
Muttuvelu DV, Baggesen K, Ehlers N (2012) Precision and accuracy of the ICare tonometer – Peripheral and central IOP measurements by rebound tonometry. Acta Ophthalmol (Copenh) 90(4):322–326
Chiu W, Lam A, Chen D et al. (2008) The influence of corneal properties on rebound tonometry. Ophthalmology 115:80–84
Kupin TH, Shin DH, Juzych MS et al. (1993) Use of a Tono-Pen for long-term home tonometry. Am J Ophthalmol 116(5):643–644
Ruokonen PC, Schwenteck T, Draeger J (2007) Evaluation of the impedance tonometer TGDc-01 and iCare according to international ocular tonometer standards ISO 8612. Graefes Arch Clin Exp Ophthalmol 245:1259–1265
Boles Carenini B, Brogliatti B, Tonetto C et al. (1992) The Pulsair-Keeler non-contact tonometer in self-tonometry: preliminary results. Int Ophthalmol 16:295–297
Hjortdal J, Jensen PK (1995) In vitro measurement of corneal strain, thickness, and curvature using digital image processing. Acta Ophthalmol Scand 73:5–11
Van der Jagt LH, Jansonius NM (2005) Three portable tonometers, the TGDc-01, the ICARE, and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt 25:429–435
Resua CG, Fernandez MJG, Exposito AC et al. (2005) Clinical evaluation of the new TGDc-01 “PRA” palpebral tonometer: comparison with contact and non-contact tonometry. Optom Vis Sci 82(2):143–150
Cook JA, Botello AP, Elders A et al. (2012) Systematic review of the agreement of tonometers with Goldmann Applanation Tonometry. Ophthalmology 119:1552–1557
Lam AKC, Douthwaite WA (1997) The effect of an artificially elevated intraocular pressure on the central corneal curvature. Ophthalmic Physiol Opt 17:18–24
Leonardi M, Leuenberger P, Bertrand D et al. (2004) First steps towards noninvasive intraocular pressure monitoring with a sensing contact lens. Invest Ophthalmol Vis Sci 45(9):3113–3117
Leonardi M, Pitchon EM, Bertsch A et al. (2009) Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes. Acta Ophthalmol (Copenh) 87(4):433–437
Faschinger C, Mossbock G (2010) Kontinuierliche 24-h-Aufzeichnung von Augendruckschwankungen mittels drahtlosem Kontaktlensensor Triggerfish. Ophthalmologe 107(10):918–922
Mansouri K, Shaarawy T (2011) Continuous intraocular pressure monitoring with a wireless ocular telemetry sensor: initial clinical experience in patients with open angle glaucoma. Br J Ophthalmol 95:627–629
Mansouri K, Medeiros FA, Tafresh et al. (2012) Continuous 24-hour monitoring of intraocular pressure patterns with a contact lens sensor: Safety, tolerability, and reproducibility in patients with glaucoma. Arch Ophthalmol 130(12):1534–1539
Lorenz K, Korb C, Herzog N et al. (2013) Tolerability of 24-hour intraocular pressure monitoring of a pressure-sensitive contact lens. J Glaucoma 22(4):311–316
Matsumoto T, Nagata TR, Slashin M et al. (1978) Measurement by holographic interferometry of the deformation of the eye accompanying changes in intraocular pressure. Appl Opt 17(22):3538–3539
Sanchez I, Laukhin V, Moya A et al. (2011) Prototype of a nanostructured sensing contact lens for noninvasive intraocular pressure monitoring. Investig Ophthalmol Vis Sci 52(11):8310–8315
Collins CC (1967) Miniature passive pressure transensor for implanting in the eye. IEEE Trans Biomed Eng 14:74–83
Collins CC (1967) Passive telemetry with glass transensors. Proc Nat Telemetring Conf 146–151
Wolbarsht ML, Wortman J, Schwartz B et al. (1980) A sclera buckle pressure gauge for continuous monitoring of intraocular pressure. Int Ophthalmol 3:11–17
Rizq RN, Choi WH, Eilers D et al. (2001) Intraocular pressure measurement at choroid surface: a feasibility study with implications for implantable microsystems. Br J Ophthalmol 85:868–871
Svedbergh B, Backlund Y, Hok B, Rosengren L (1992) The IOP-IOL. A prob into the eye. Acta Ophthalmol (Copenhagen) 70(2):266–268
Walter P, Schnakenberg U, vom Bogel G et al. (2000) Development of a completely encapsulated intraocular pressure sensor. Ophthalmic Res 32:278–284
Hille K, Draeger J, Eggers T et al. (2001) Technischer aufbau, kalibrierung und ergebnisse mit einem neuen intraokularen drucksensor mit telemetrischer Ubertragung. Klin Monatsbl Augenheilkd 218(5):376–380
Chen PJ, Rodger DC, Humayun MS (2006) Unpowered spiral-tube parylene pressure sensor for intraocular pressure sensing. Sensors and Actuators A:Physical 127(2):276–282
Chen PJ, Rodger DC, Meng E, et al. (2006) Implantable unpowered parylene MEMS intraocular pressure sensor. Technical Digest, IEEE Engineering in Medicine and Biology Society Special Topic Conference on Microtechnologies in Medicine and Biology. 256-259
Data provided courtesy of Implandata Ophthalmic Products GmbH
Todani A, Behlau I, Fava MA et al. (2011) Intraocular pressure measurement by radio wave telemetry. Invest Ophthalmol Vis Sci 52(13):9573–9580
Chen G, Ghead H, Haque R, et al. (2011) A cubic-millimeter energy-autonomous wireless intraocular pressure monitor. ISSCC Digest of Technical Papers, pp 310-312
Lee Y, Kim G, Bang S, et al. (2012) A modular 1 mm3 die-stacked sensing platform with optical communication and multi-modal energy harvesting. ISSCC Digest of Technical Papers, pp 402-404
Conflicts of interest
There are no conflicts of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Yung, E., Trubnik, V. & Katz, L.J. An overview of home tonometry and telemetry for intraocular pressure monitoring in humans. Graefes Arch Clin Exp Ophthalmol 252, 1179–1188 (2014). https://doi.org/10.1007/s00417-014-2668-5
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00417-014-2668-5