An overview of home tonometry and telemetry for intraocular pressure monitoring in humans

  • Review Article
  • Published:
Graefe's Archive for Clinical and Experimental Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To review the existing technology for self-tonometry and evaluate methods for continuous monitoring of IOP currently undergoing development and clinical trials in humans.

Background

Glaucoma is one of the leading causes of blindness worldwide. Current glaucoma therapy is focused primarily on lowering intraocular pressures (IOP). Decisions to modify treatment regiments are primarily based on office IOP. Since IOP fluctuates throughout the day, values obtained in the office may be a poor representation of the patient’s disease. IOP measurements outside of the physician’s office environment would provide better knowledge of the disease state and allow for better-informed medical decision making.

Methods and results

We performed a literature search using Medline and IEEE database for studies investigating technologies that have been developed for continuous 24-hour IOP monitoring.

Conclusion

There is currently no technology that has been approved for use to allow for continuous monitoring of IOP fluctuations. New experimental technology being developed and currently undergoing clinical trials has demonstrated potential for changing the diagnosis and management of glaucoma.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

References

  1. Resnikoff S, Pascolini D, Etya’ale D et al. (2004) Global data on visual impairment in the year 2002. Bull World Health Organ 82(11):844–851

    PubMed Central  PubMed  Google Scholar 

  2. Quigley HA, Broman AT (2006) The number of people with glaucoma worldwide in 2010 and 2020. Br J Ophthalmol 90(3):262–267

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  3. Goldmann H, Schmidt T (1957) Applanation Tonometry. Ophthalmologica 134(4):221–242

    Article  CAS  PubMed  Google Scholar 

  4. Wessels IF, Oh Y (1990) Tonometer utilization, accuracy, and calibration under field conditions. Arch Ophthalmol 108(12):1709–112

    Article  CAS  PubMed  Google Scholar 

  5. Tonnu PA, Ho T, Newson T et al. (2005) The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry. Br J Ophthalmol 89(7):851–854

    Article  PubMed Central  PubMed  Google Scholar 

  6. Collaer N, Zeyen T, Caprioli J (2005) Sequential office pressure measurements in the management of glaucoma. J Glaucoma 14(3):196–200

    Article  PubMed  Google Scholar 

  7. Asrani S, Zeimer R, Wilensky J et al. (2000) Large diurnal fluctuations in intraocular pressures are an independent risk factor in patients with glaucoma. J Glaucoma 9:134–142

    Article  CAS  PubMed  Google Scholar 

  8. Caprioli J, Coleman AL (2008) Intraocular Pressure Fluctuation: A risk factor for visual field progression at low intraocular pressures in the Advanced Glaucoma Intervention Study. Ophthalmology 115(7):1123–1129

    Article  PubMed  Google Scholar 

  9. McLaren JW, Brubaker RF, FitzSimon JS (1996) Continuous measurement of intraocular pressure in rabbits by telemetry. Investig Ophthalmol Vis Sci 37(6):966–975

    CAS  Google Scholar 

  10. Carlson KH, McLaren JW, Topper JE et al. (1987) Effect of body position on intraocular pressure and aqueous flow. Invest Ophthalmol Vis Sci 28(8):1346–1352

    CAS  PubMed  Google Scholar 

  11. Yamabayashi S, Aguilar RN, Hosoda M et al. (1991) Postural change of intraocular and blood pressures in ocular hypertension and low tension glaucoma. Br J Ophthalmol 75(11):652–655

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  12. Kothe AC (1994) The effect of posture on intraocular pressure and pulsatile ocular blood flow in normal and glaucomatous eyes. Surv Ophthalmol 38:S191–S197

    Article  PubMed  Google Scholar 

  13. Smedt SD, Mermound A, Schnyder C (2012) 24-Hour intraocular pressure fluctuation monitoring using an ocular telemetry sensor: Tolerability and functionality in healthy subjects. J Glaucoma 21(8):539–544

    Article  PubMed  Google Scholar 

  14. Hara T, Hara T, Tsuru T (2006) Increase of peak intraocular pressure during sleep in reproduced diurnal changes by posture. Arch Ophthalmol 124(2):165–168

    Article  PubMed  Google Scholar 

  15. Liu JKH, Kripke DF, Hoffman RE et al. (1998) Nocturnal elevation of intraocular pressure in young adults. Invest Ophthalmol Vis Sci 39:2707–2712

    CAS  PubMed  Google Scholar 

  16. Liu JKH, Bouligny RP, Kripke DF et al. (2003) Nocturnal elevation of intraocular pressure is detectable in the sitting position. Invest Ophthalmol Vis Sci 44(10):4439–4442

    Article  PubMed  Google Scholar 

  17. Brown B, Morris P, Muller C et al. (1988) Fluctuations in intra-ocular pressure with sleep: I. Time course of IOP increase after the onset of sleep. Ophthalmic Physiol Opt 8(3):246–248

    CAS  PubMed  Google Scholar 

  18. Brown B, Burton P, Mann S et al. (1988) Fluctuations in intra-ocular pressure with sleep: II. Time course of IOP decrease after waking from sleep. Ophthalmic Physiol Opt 8(3):249–252

    CAS  PubMed  Google Scholar 

  19. Orzalesi N, Rossetti L, Invernizzi T et al. (2000) Effect of timolol, latanoprost, and dorzolamide on circadian IOP in glaucoma or ocular hypertension. Invest Ophthal VisSci 41(9):2566–2573

    CAS  Google Scholar 

  20. Orzalesi N, Rossetti L, Bottoli A et al. (2003) The effect of latanoprost, brimonidine, and a fixed combination of timolol and dorzolamide on circadian intraocular pressure in patients with glaucoma or ocular hypertension. Arch Ophthalmol 121(4):453–457

    Article  CAS  PubMed  Google Scholar 

  21. De Alva M, Jervell J (1997) Self monitoring improves quality of life and prognosis of people with diabetes. Br Med J 315:184–185

    Article  Google Scholar 

  22. Hughes E, Spry P, Diamond J (2003) 24-Hour monitoring of intraocular pressure in glaucoma management: a retrospective review. J Glaucoma 12:232–236

    Article  PubMed  Google Scholar 

  23. Draeger J, Schwartz R, Groenhoff et al. (1993) Self-tonometry under microgravity conditions. Clin Investigator 71:700–703

    Article  CAS  Google Scholar 

  24. Theofylaktopoulos I, Diestelhorst M, Krieglstein GK (1999) Self-tonometry with the Ocuton S versus Goldmann tonometry. Graefes Arch Clin Exp Ophthalmol 237:720–724

    Article  CAS  PubMed  Google Scholar 

  25. Sacu S, Vass C, Schemper M et al. (2004) Self-tonometry with the Ocuton S: evaluation of accuracy in glaucoma patients. Acta Ophthalmol Scand 82:405–409

    Article  PubMed  Google Scholar 

  26. Marchini G, Babighian S, Specchia L et al. (2002) Evaluation of the new Ocuton S tonometer. Acta Ophthalmol Scand 80:167–171

    Article  PubMed  Google Scholar 

  27. Kothy P, Vargha P, Hollo G (2001) Ocuton S self-tonometry vs Goldmann tonometry; a diurnal comparison study. Acta Ophthalmologica Scandanavica 79:294–297

    Article  CAS  Google Scholar 

  28. Lanfermann E, Jurgens C, Grossjohann R et al. (2009) Intraocular pressure measurements with the newly Ocuton S*TT-MV self-tonometer in comparison to Goldmann applanation tonometry in glaucoma patients. Med Sci Monit 15(11):CR556–CR562

    PubMed  Google Scholar 

  29. Rosenwasser GO, Holland S, Pflugfelder SC et al. (1990) Topical anesthetic abuse. Ophthalmology 97(8):967–972

    Article  CAS  PubMed  Google Scholar 

  30. McGee HT, Fraunfelder FW (2007) Toxicities of topical ophthalmic anesthetics. Expert Opin Drug Saf 6(6):637–640

    Article  CAS  PubMed  Google Scholar 

  31. Yeniad B, Canturk S, Esin Ozdemir F et al. (2010) Toxic keratopathy due to abuse of topical anesthetic drugs. Cutan Ocul Toxicol 29(2):105–109

    Article  PubMed  Google Scholar 

  32. Lam DSC, Leung DYL, Chiu TYH et al. (2004) Pressure phosphene self-tonometry: A comparison with Goldmann Tonometry in Glaucoma Patients. Investig Ophthalmol Vis Sci 45(9):3131–3136

    Article  Google Scholar 

  33. Tai MC, Chen PL, Wu JN et al. (2005) Clinical evaluation of the intraocular pressure in patients with glaucoma or ocular hypertension by a self-assessable tonometer. J Ocul Pharmacol Ther 21(1):55–61

    Article  CAS  PubMed  Google Scholar 

  34. Danesh-Meyer HV, Niederer R, Gaskin BJ, Gamble (2004) Comparison of the Proview pressure phosphene tonometer performed by the patient and examiner with the Goldmann applanation tonometer. Clin Exp Ophthalmol 32(1):29–32

    Article  Google Scholar 

  35. Rai S, Moster MR, Kesen M et al. (2005) Level of disagreement between Proview Phosphene Tonometer and Goldmann Applanation Tonometer Intraocular Pressure Readings. J Glaucoma 14:120–123

    Article  PubMed  Google Scholar 

  36. Li J, Herndon LW, Asrani SG et al. (2004) Clinical comparison of the Proview Eye Pressure Monitor with Goldmann Applanation Tonometer and the TonoPen. Arch Ophthalmol 122:1117–1121

    Article  PubMed  Google Scholar 

  37. Rosentreter A, Jablonski KS, Mellein AC et al. (2011) A new rebound tonometer for home monitoring of intraocular pressure. Graefes Arch Clin Exp Ophthalmol 249(11):1713–1719

    Article  PubMed  Google Scholar 

  38. Asrani S, Chatterjee A, Wallace DK et al. (2011) Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device. J Glaucoma 20:74–79

    Article  PubMed  Google Scholar 

  39. Muttuvelu DV, Baggesen K, Ehlers N (2012) Precision and accuracy of the ICare tonometer – Peripheral and central IOP measurements by rebound tonometry. Acta Ophthalmol (Copenh) 90(4):322–326

    Article  Google Scholar 

  40. Chiu W, Lam A, Chen D et al. (2008) The influence of corneal properties on rebound tonometry. Ophthalmology 115:80–84

    Article  PubMed  Google Scholar 

  41. Kupin TH, Shin DH, Juzych MS et al. (1993) Use of a Tono-Pen for long-term home tonometry. Am J Ophthalmol 116(5):643–644

    CAS  PubMed  Google Scholar 

  42. Ruokonen PC, Schwenteck T, Draeger J (2007) Evaluation of the impedance tonometer TGDc-01 and iCare according to international ocular tonometer standards ISO 8612. Graefes Arch Clin Exp Ophthalmol 245:1259–1265

    Article  PubMed  Google Scholar 

  43. Boles Carenini B, Brogliatti B, Tonetto C et al. (1992) The Pulsair-Keeler non-contact tonometer in self-tonometry: preliminary results. Int Ophthalmol 16:295–297

    Article  CAS  PubMed  Google Scholar 

  44. Hjortdal J, Jensen PK (1995) In vitro measurement of corneal strain, thickness, and curvature using digital image processing. Acta Ophthalmol Scand 73:5–11

    Article  CAS  PubMed  Google Scholar 

  45. Van der Jagt LH, Jansonius NM (2005) Three portable tonometers, the TGDc-01, the ICARE, and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt 25:429–435

    Article  PubMed  Google Scholar 

  46. Resua CG, Fernandez MJG, Exposito AC et al. (2005) Clinical evaluation of the new TGDc-01 “PRA” palpebral tonometer: comparison with contact and non-contact tonometry. Optom Vis Sci 82(2):143–150

    Article  Google Scholar 

  47. Cook JA, Botello AP, Elders A et al. (2012) Systematic review of the agreement of tonometers with Goldmann Applanation Tonometry. Ophthalmology 119:1552–1557

    Article  PubMed  Google Scholar 

  48. Lam AKC, Douthwaite WA (1997) The effect of an artificially elevated intraocular pressure on the central corneal curvature. Ophthalmic Physiol Opt 17:18–24

    Article  CAS  PubMed  Google Scholar 

  49. Leonardi M, Leuenberger P, Bertrand D et al. (2004) First steps towards noninvasive intraocular pressure monitoring with a sensing contact lens. Invest Ophthalmol Vis Sci 45(9):3113–3117

    Article  PubMed  Google Scholar 

  50. Leonardi M, Pitchon EM, Bertsch A et al. (2009) Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes. Acta Ophthalmol (Copenh) 87(4):433–437

    Article  Google Scholar 

  51. Faschinger C, Mossbock G (2010) Kontinuierliche 24-h-Aufzeichnung von Augendruckschwankungen mittels drahtlosem Kontaktlensensor Triggerfish. Ophthalmologe 107(10):918–922

    Article  CAS  PubMed  Google Scholar 

  52. Mansouri K, Shaarawy T (2011) Continuous intraocular pressure monitoring with a wireless ocular telemetry sensor: initial clinical experience in patients with open angle glaucoma. Br J Ophthalmol 95:627–629

    Article  PubMed  Google Scholar 

  53. Mansouri K, Medeiros FA, Tafresh et al. (2012) Continuous 24-hour monitoring of intraocular pressure patterns with a contact lens sensor: Safety, tolerability, and reproducibility in patients with glaucoma. Arch Ophthalmol 130(12):1534–1539

    Article  Google Scholar 

  54. Lorenz K, Korb C, Herzog N et al. (2013) Tolerability of 24-hour intraocular pressure monitoring of a pressure-sensitive contact lens. J Glaucoma 22(4):311–316

    Article  PubMed  Google Scholar 

  55. Matsumoto T, Nagata TR, Slashin M et al. (1978) Measurement by holographic interferometry of the deformation of the eye accompanying changes in intraocular pressure. Appl Opt 17(22):3538–3539

    Article  CAS  PubMed  Google Scholar 

  56. Sanchez I, Laukhin V, Moya A et al. (2011) Prototype of a nanostructured sensing contact lens for noninvasive intraocular pressure monitoring. Investig Ophthalmol Vis Sci 52(11):8310–8315

    Article  Google Scholar 

  57. Collins CC (1967) Miniature passive pressure transensor for implanting in the eye. IEEE Trans Biomed Eng 14:74–83

    Article  CAS  PubMed  Google Scholar 

  58. Collins CC (1967) Passive telemetry with glass transensors. Proc Nat Telemetring Conf 146–151

  59. Wolbarsht ML, Wortman J, Schwartz B et al. (1980) A sclera buckle pressure gauge for continuous monitoring of intraocular pressure. Int Ophthalmol 3:11–17

    Article  CAS  PubMed  Google Scholar 

  60. Rizq RN, Choi WH, Eilers D et al. (2001) Intraocular pressure measurement at choroid surface: a feasibility study with implications for implantable microsystems. Br J Ophthalmol 85:868–871

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  61. Svedbergh B, Backlund Y, Hok B, Rosengren L (1992) The IOP-IOL. A prob into the eye. Acta Ophthalmol (Copenhagen) 70(2):266–268

    Article  CAS  Google Scholar 

  62. Walter P, Schnakenberg U, vom Bogel G et al. (2000) Development of a completely encapsulated intraocular pressure sensor. Ophthalmic Res 32:278–284

    Article  CAS  PubMed  Google Scholar 

  63. Hille K, Draeger J, Eggers T et al. (2001) Technischer aufbau, kalibrierung und ergebnisse mit einem neuen intraokularen drucksensor mit telemetrischer Ubertragung. Klin Monatsbl Augenheilkd 218(5):376–380

    Article  CAS  PubMed  Google Scholar 

  64. Chen PJ, Rodger DC, Humayun MS (2006) Unpowered spiral-tube parylene pressure sensor for intraocular pressure sensing. Sensors and Actuators A:Physical 127(2):276–282

    Article  CAS  Google Scholar 

  65. Chen PJ, Rodger DC, Meng E, et al. (2006) Implantable unpowered parylene MEMS intraocular pressure sensor. Technical Digest, IEEE Engineering in Medicine and Biology Society Special Topic Conference on Microtechnologies in Medicine and Biology. 256-259

  66. Data provided courtesy of Implandata Ophthalmic Products GmbH

  67. Todani A, Behlau I, Fava MA et al. (2011) Intraocular pressure measurement by radio wave telemetry. Invest Ophthalmol Vis Sci 52(13):9573–9580

    Article  PubMed  Google Scholar 

  68. Chen G, Ghead H, Haque R, et al. (2011) A cubic-millimeter energy-autonomous wireless intraocular pressure monitor. ISSCC Digest of Technical Papers, pp 310-312

  69. Lee Y, Kim G, Bang S, et al. (2012) A modular 1 mm3 die-stacked sensing platform with optical communication and multi-modal energy harvesting. ISSCC Digest of Technical Papers, pp 402-404

Download references

Conflicts of interest

There are no conflicts of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Edward Yung.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Yung, E., Trubnik, V. & Katz, L.J. An overview of home tonometry and telemetry for intraocular pressure monitoring in humans. Graefes Arch Clin Exp Ophthalmol 252, 1179–1188 (2014). https://doi.org/10.1007/s00417-014-2668-5

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00417-014-2668-5

Keywords

Navigation