Cementless femoral stems with lower canal fill ratio have similar mid-term to long-term outcomes to those with adequate fill ratio in Dorr type C femurs

  • Hip Arthroplasty
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Background

Lower canal fill ratio was reported to correlate with aseptic loosening in many studies. However, the most widely used standard of fill ratio seemed inapplicable to Dorr type C femurs. We aimed to adapt the method of measuring the fill ratio in Dorr type C femurs and compare the outcomes among patients with different fill ratios.

Methods

Twenty patients with Corail stems implanted in their Dorr type C femurs received spectrum CT to evaluate the whole-stem's fill ratio. Pearson Correlation Coefficient was calculated to assess the correlation between the fill ratio in X-ray film and spectrum CT. Then 87 THAs were involved in this study, divided into the fill ratio ≤ 80% group and the fill ratio > 80% group. Clinical and radiological outcomes were evaluated with a mean follow-up of 8.2 years.

Results

Fill ratio at 2 cm below the lesser trochanter in anterior–posterior X-ray film correlated with the whole-stem's fill ratio (r = 0.50, P = 0.02). Survival rate of stem, function scores, and radiological outcomes between the two groups showed no significant difference. In the fill ratio > 80% group, intraoperative fracture was significantly higher (19% VS 5%, P < 0.05).

Conclusion

Patients with lower fill ratios at 2 cm below the lesser trochanter did not have poorer functional scores or more subsidence, but had a lower intraoperative fracture rate. The revision rates of the two groups presented no significant difference, but this result need to be confirmed in larger cohort in the future. In Dorr type C femurs, risk of fracture and the special morphology of the femur should be noted, and high fill ratio is not the most decisive factor for stem size selecting.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

Availability of data and materials

The authors confirm that the data supporting the findings of this study are available within the article.

References

  1. Achten J, Parsons NR, Edlin RP, Griffin DR, Costa ML (2010) A randomised controlled trial of total hip arthroplasty versus resurfacing arthroplasty in the treatment of young patients with arthritis of the hip joint. BMC Musculoskelet Disord 11:8

    Article  Google Scholar 

  2. Al-Najjim M, Khattak U, Sim J, Chambers I (2016) Differences in subsidence rate between alternative designs of a commonly used uncemented femoral stem. J Orthop 13(4):322–326

    Article  Google Scholar 

  3. Al-Dirini RM, Huff D, Zhang J, Besier T, Clement JG, Taylor M (2018) Influence of collars on the primary stability of cementless femoral stems: a finite element study using a diverse patient cohort. J Orthop Res 36(4):1185–1195

    Article  CAS  Google Scholar 

  4. Aldinger PR, Jung AW, Breusch SJ, Ewerbeck V, Parsch D (2009) Survival of the cementless Spotorno® stem in the second decade. Clin Orthop Relat Res 467(9):2297–2304

    Article  Google Scholar 

  5. Biedermann R, Krismer M, Stöckl B, Mayrhofer P, Ornstein E, Franzén H (1999) Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement Einzel-Bild-Röntgen-Analyse-femoral component analysis. J Bone Jt Surg Br Vol 81(2):266–272

    Article  CAS  Google Scholar 

  6. Camine VM, Rüdiger HA, Pioletti DP, Terrier A (2018) Effect of a collar on subsidence and local micromotion of cementless femoral stems: in vitro comparative study based on micro-computerised tomography. Int Orthop 42(1):49–57

    Article  Google Scholar 

  7. Dorr LD, Faugere MC, Mackel AM, Gruen TA, Bognar B, Malluche HH (1993) Structural and cellular assessment of bone quality of proximal femur. Bone 14(3):231–242

    Article  CAS  Google Scholar 

  8. Engh CA, Bobyn JD, Glassman AH (1987) Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Jt Surg Br Vol 69(1):45–55

    Article  CAS  Google Scholar 

  9. Engh CA, Massin P, Suthers KE (1990) Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res 257:107–128

    Article  Google Scholar 

  10. Evola F, Evola G, Graceffa A et al (2014) Performance of the CLS Spotorno uncemented stem in the third decade after implantation. Bone Jt J 96(4):455–461

    Article  Google Scholar 

  11. Fottner A, Woiczinski M, Kistler M et al (2017) Influence of undersized cementless hip stems on primary stability and strain distribution. Arch Orthop Trauma Surg 137(10):1435–1441

    Article  Google Scholar 

  12. Gromov K, Bersang A, Nielsen CS, Kallemose T, Husted H, Troelsen A (2017) Risk factors for post-operative periprosthetic fractures following primary total hip arthroplasty with a proximally coated double-tapered cementless femoral component. Bone Jt J 99-b(4):451–457

    Article  CAS  Google Scholar 

  13. Hulley SBCS, Browner WS, Grady D, Newman TB (2013) Designing clinical research: an epidemiologic approach, 4th edn. Lippincott Williams & Wilkins, Philadelphia, p 75 (Appendix 6B)

    Google Scholar 

  14. Hwang K-T, Kim Y-H, Kim Y-S, Choi I-Y (2012) Total hip arthroplasty using cementless grit-blasted femoral component: a minimum 10-year follow-up study. J Arthroplasty 27(8):1554–1561

    Article  Google Scholar 

  15. Ishii S, Homma Y, Baba T, Ozaki Y, Matsumoto M, Kaneko K (2016) Does the canal fill ratio and femoral morphology of Asian females influence early radiographic outcomes of total hip arthroplasty with an uncemented proximally coated, tapered-wedge stem? J Arthroplasty 31(7):1524–1528

    Article  Google Scholar 

  16. Khalily C, Lester DK (2002) Results of a tapered cementless femoral stem implanted in varus. J Arthroplasty 17(4):463–466

    Article  Google Scholar 

  17. Krismer M, Biedermann R, Stöckl B, Fischer M, Bauer R, Haid C (1999) The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA. Einzel-Bild-Roentgen-Analyse-femoral component analysis. J Bone Jt Surg Br Vol 81(2):273–280

    Article  CAS  Google Scholar 

  18. McLaughlin JR, Lee KR (2016) Long-term results of uncemented total hip arthroplasty with the Taperloc femoral component in patients with Dorr type C proximal femoral morphology. Bone Jt J 98-b(5):595–600

    Article  CAS  Google Scholar 

  19. Nakaya R, Takao M, Hamada H, Sakai T, Sugano N (2019) Reproducibility of the Dorr classification and its quantitative indices on plain radiographs. Orthop Traumatol Surg Res OTSR 105(1):17–21

    Article  Google Scholar 

  20. Nash W, Harris A (2014) The Dorr type and cortical thickness index of the proximal femur for predicting peri-operative complications during hemiarthroplasty. J Orthop Surg 22(1):92–95

    Article  Google Scholar 

  21. Noble PC, Alexander JW, Lindahl LJ, Yew DT, Granberry WM, Tullos HS (1988) The anatomic basis of femoral component design. Clin Orthop Relat Res 235:148–165

    Article  Google Scholar 

  22. Panichkul P, Bavonratanavech S, Arirachakaran A, Kongtharvonskul J (2019) Comparative outcomes between collared versus collarless and short versus long stem of direct anterior approach total hip arthroplasty: a systematic review and indirect meta-analysis. Eur J Orthop Surg Traumatol 29(8):1693–1704

    Article  Google Scholar 

  23. Rattanaprichavej P, Laoruengthana A, Chotanaphuti T, Khuangsirikul S, Phreethanutt C, Pongpirul K (2019) Subsidence of hydroxyapatite-coated femoral stem in Dorr type C proximal femoral morphology. J Arthroplasty 34(9):2011–2015

    Article  Google Scholar 

  24. Ries C, Boese CK, Dietrich F, Miehlke W, Heisel C (2019) Femoral stem subsidence in cementless total hip arthroplasty: a retrospective single-centre study. Int Orthop 43(2):307–314

    Article  Google Scholar 

  25. Streit MR, Innmann MM, Merle C, Bruckner T, Aldinger PR, Gotterbarm T (2013) Long-term (20- to 25-year) results of an uncemented tapered titanium femoral component and factors affecting survivorship. Clin Orthop Relat Res 471(10):3262–3269

    Article  Google Scholar 

  26. Syed F, Hussein A, Katam K, Saunders P, Young SK, Faisal M (2018) Risk of subsidence and peri-prosthetic fractures using collared hydroxyapatite-coated stem for hip arthroplasty in the elderly. HIP Int 28(6):663–667

    Article  Google Scholar 

  27. Vidalain J-P, Selmi TAS, Beverland D et al (2011) The CORAIL® Hip System: a practical approach based on 25 years of experience. Springer, pp 23–29

  28. Young S, Michel MC, Selmi TAS et al (2011) The Corail® hip stemthe Corail® hip system. Springer, Berlin, pp 53–88

    Book  Google Scholar 

Download references

Funding

This study was funded by the National Natural Science Foundation of China Program (No. 81974347) and National Clinical Research Center for Geriatrics, West China Hospital, Sichuan University (NO. Z20191008).

Author information

Authors and Affiliations

Authors

Contributions

Mingyang Li, Bin Shen conceived and designed the analysis. Yi Zeng, Yuangang Wu collected the data. Yuan Liu, Limin Wu performed the analysis. Mingyang Li, Bin Shen wrote the paper.

Corresponding author

Correspondence to Bin Shen.

Ethics declarations

Conflict of interest

All authors declared that there is no conflict of interest.

Ethical approval

Ethical approval was obtained from the institutional review board of West China Hospital (Chengdu, Sichuan Province, China).

Consent to participate

All included patients provided informed consent.

Consent to publish

All authors provided publish consent.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, M., Zeng, Y., Wu, Y. et al. Cementless femoral stems with lower canal fill ratio have similar mid-term to long-term outcomes to those with adequate fill ratio in Dorr type C femurs. Arch Orthop Trauma Surg 142, 1265–1273 (2022). https://doi.org/10.1007/s00402-021-03916-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-021-03916-x

Keywords

Navigation