Radiation exposure in the treatment of pediatric supracondylar humerus fractures

  • Orthopaedic Surgery
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Purpose

To determine the factors that influence radiation exposure during repair of supracondylar humerus fractures.

Methods

Medical records of almost 200 children with supracondylar fractures were retrospectively analyzed for variables correlated with fluoroscopy time and radiation dose as measures of radiation exposure.

Results

There was no statistically significant difference in fluoroscopy time (27 vs. 22 s p = 0.345) or direct radiation dose (0.394 vs. 0.318 mSv p = 0.290) between uniplanar and biplanar C-arm use. No statistically significant differences in fluoroscopy time or radiation dose were found for surgical technique, comorbid ipsilateral fractures, preoperative neurovascular compromise, or resident participation. There was a significant 8.3 s increase in fluoroscopy time (p = 0.022) and 0.249 mSv increase in radiation dose (p = 0.020) as the fracture type increased from II to III. An increase in one pin during CRPP resulted in a statistically significant 10.4 s increase in fluoroscopy time and a 0.205 mSv increase in radiation dose. There were significant differences between the physician with the lowest fluoroscopy time and radiation dose compared with the physicians with the two highest values for both fluoroscopy time and radiation dose (p < 0.01).

Conclusions

We found no significant difference in direct radiation exposure or fluoroscopy time when comparing biplanar to uniplanar C-arm use, resident participation, preoperative neurovascular compromise, or for comorbid ipsilateral fractures. Both outcomes increased significantly as fracture type increased from II to III and as the number of pins used during CRPP increased. Both outcomes were significantly different between the surgeons performing CRPP.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Niera JZSD, Prada-Cañizares A, Marti-Circuelos R, Pretell-Mazzini J (2015) Supracondylar humeral fractures in children:current concepts for the management and prognosis. Int Orthop 29(11):22872296

    Google Scholar 

  2. Giachino AA, Cheng M (1980) Irradiation of the surgeon during pinning of femoral fractures. J Bone Jt Surg 62-B(2):227–229

    Article  CAS  Google Scholar 

  3. Martus JE, Hilmes MA, Grice JV et al (2018) Radiation exposure during operative fixation of pediatric supracondylar humerus fractures. J Pediatr Orthop 38(5):249–253

    Article  Google Scholar 

  4. National Research Council of the National Academies (2006) Health risks from exposure to low levels of ionizing radiation: BEIR VII Phase 2. The National Academies Press, Washington, DC

    Google Scholar 

  5. ICRP (2007) The 2007 recommendations of the international commission on radiological protection. ICRP publication 103. Ann ICRP 37(2–4):1–332

    Google Scholar 

  6. Badman BL, Rill L, Butkovich B, Arreola M, Griend RAV (2005) Radiation exposure with use of the mini-C-arm for routine orthopaedic imaging procedures. J Bone Jt Surg Am. 87(1):13–17

    Article  Google Scholar 

  7. Bar-On R, Weigl DM, Becker T, Katz K, Konen O (2010) Intraoperative C-arm radiation affecting factors and reduction by an intervention program. J Pediatr Orthop 30(4):320–323

    Article  Google Scholar 

  8. Eismann EA, Qall EJ, Thomas EC, Little MA (2014) Direct beam radiation exposure to surgeons during pinning of supracondylar humerus fractures. J Pediatr Orthop 34(2):166–171

    Article  Google Scholar 

  9. Esen E, Dogramaci Y, Gultekin S et al (2009) Comparison of radiation exposure times in the treatment of pediatric supracondylar humeral fractures with open-closed reduction and internal fixation. Acta Orthop Traumatol Turc 43:400–405

    Article  Google Scholar 

  10. Hsu RY, Lareau CR et al (2014) The effect of C-arm position on radiation exposure during fixation of pediatric supracondylar fractures of the humerus. J Bone Jt Surg Am 96(15):e129

    Article  Google Scholar 

  11. Kalenderer Ö, Turgut A, Bilgin E, Erkus S, İpci FB, Edizsoy TP (2018) The influence of resident level of training on fluoroscopy time in pediatric supracondylar humeral fractures treated with closed reduction and percutaneous pinning. Cureus 10(2):e2245

    PubMed  PubMed Central  Google Scholar 

  12. Na Y, Bai R, Zhao Z et al (2018) Comparison of lateral entry with crossed entry pinning for pediatric supracondylar humeral fractures: a meta-analysis. J Orthop Surg Res 13(1):68 (Published 2018 Apr 3)

    Article  Google Scholar 

  13. Riley SA (1989) Radiation exposure from fluoroscopy during orthopedic surgical procedures. Clin Orthop Relat Res (248):257–260

  14. Singer G, Herron B, Herron D (2011) Exposure from the large C-arm versus the mini C-arm using hand/wrist and elbow phantoms. J Hand Surg 36(4):628–631

    Article  Google Scholar 

  15. Tremains MR, Georgiadis GM, Dennis MJ (2001) Radiation exposure with use of the inverted-c-arm technique in upper-extremity surgery. J Bone Jt Surg Am 83-A(5):674–678

    Article  Google Scholar 

  16. Dekker AE, Krijnen P, Schipper IB (2016) Results of crossed versus lateral entry K-wire fixation of displaced pediatric supracondylar humeral fractures: a systematic review and meta-analysis. Injury 47(11):2391–2398

    Article  CAS  Google Scholar 

  17. Goldstone KE, Wright IH, Cohen B (1993) Radiation exposure to the hands of orthopaedic surgeons during procedures under fluoroscopic X-ray control. Br J Radiol 66(790):899–901

    Article  CAS  Google Scholar 

  18. Müller LP, Suffner J, Wenda K, Mohr W, Rommens PM (1998) Radiation exposure to the hands and the thyroid of the surgeon during intramedullary nailing. Injury 29(6):461–468

    Article  Google Scholar 

  19. Bushberg JT (2011) The essential physics of medical imaging. Wolters Kluwer, Philadelphia

    Google Scholar 

  20. Schueler BA, Vrieze TJ, Bjarnason H, Stanson AW (2006) An investigation of operator exposure in interventional radiology. Radiographics 26(5):1533–1541

    Article  Google Scholar 

  21. Stahl CM, Meisinger QC, Andre MP, Kinney TB, Newton IG (2016) Radiation risk to the fluoroscopy operator and staff. Am J Roentgenol 207(4):737–744

    Article  Google Scholar 

  22. Lee K, Lee KM, Park MS, Lee B, Kwon DG, Chung CY (2012) Measurements of surgeons’ exposure to ionizing radiation dose during intraoperative use of C-arm fluoroscopy. Spine 37(14):1240–1244

    Article  Google Scholar 

  23. Giordano BD, Grauer JN, Miller CP et al (2011) Radiation exposure issues in orthopaedics. J Bone Jt Surg Am 93(e69):1–10

    Google Scholar 

  24. Khong P-L, Ringertz H, Donoghue V et al (2013) ICRP publication 121: radiological protection in paediatric diagnostic and interventional radiology. Ann ICRP. 42(2):1–63

    Article  Google Scholar 

  25. Mastrangelo G, Fedeli U, Fadda E, Giovanazzi A, Scoizzato L, Saia B (2005) Increased cancer risk among surgeons in an orthopaedic hospital. Occup Med 55(6):498–500

    Article  Google Scholar 

  26. Singer G (2005) Occupational radiation exposure to the surgeon. J Am Acad Orthop Surg 13(1):69–76

    Article  Google Scholar 

  27. Rampersaud YR, Foley KT, Shen AC, Williams S, Solomito M (2000) Radiation exposure to the spine surgeon during fluoroscopically assisted pedicle screw insertion. Spine 25(20):2637–2645

    Article  CAS  Google Scholar 

  28. Hayda RA, Hsu RY, DePasse JM, Gil JA (2018) Radiation exposure and health risks for orthopaedic surgeons. J Am Acad Orthop Surg 26(8):268–277

    Article  Google Scholar 

  29. Sung KH, Min E, Chung CY, Jo BC, Park MS, Lee K (2016) Measurements of surgeons’ exposure to ionizing radiation dose: comparison of conventional and mini C-arm fluoroscopy. J Hand Surg Eur Vol 41(3):340–345

    Article  CAS  Google Scholar 

Download references

Funding

This study received no outside funding.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Michael Albert.

Ethics declarations

Conflict of interest

Author Alex Schmucker, Author Roy Chen, Author Shobhan Vachhrajani, Author Melissa Martinek, and Author Michael Albert declare that they have no conflict of interest.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors. The study IRB approval as a retrospective chart review. This study was reviewed and approved by the Institutional Review Board of the academic healthcare facility where patients were seen.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Schmucker, A., Chen, R., Vachhrajani, S. et al. Radiation exposure in the treatment of pediatric supracondylar humerus fractures. Arch Orthop Trauma Surg 140, 449–455 (2020). https://doi.org/10.1007/s00402-019-03251-2

Download citation

  • Received:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-019-03251-2

Keywords

Navigation