Abstract
Endomedullary femur nails with a head–neck implant offer an up to now unused point of reference for the determination of rotation during osteosynthesis. The lateral femur nail (LFN) or the long Proximal Femur Nail (long PFNA), for example, have recon screws or a blade placed at a 10° angle to the distal locking screw in the transversal plane. If the head–neck implant is inserted properly, the rotation of the femur can be measured with the C-arm taking the hole of the distal locking screw and the posterior condylar tangential line as reference lines. If the posterior condylar tangential line is parallel to the axes of the hole of the distal locking screw, then the rotation or anteversion of the femur is equal to the angle between the head–neck implant and the distal locking screw which is 10° with LFN or long PFNA. Differing rotation angles can be made visible by rotating the C-arm from the axial projection of the distal locking screw up to the posterior condylar tangential line. Three exemplary cases are presented. The recommended method was effective. The rotation of the femur can be measured intraoperatively with this technique and many revisions can be avoided.
References
Braten M, Terjesen T, Rossvoll I (1993) Torsional deformity after intramedullary nailing of femoral shaft fractures. Measurement of anteversion angles in 110 patients. J Bone Jt Surg 75:799–803
Buddenbrock B, Wissing H, Mller RD, John V (1997) Radiological determination of femoral rotation deformity—computerized tomography, optimized measurement accuracy and exposure dosage. Zeitschrift für Orthopädie und ihre Grenzgebiete 135:9–16
Dunn DM (1952) Anteversion of the neck of the femur; a method of measurement. J Bone Jt Surg B 34:181–186
Jaarsma RL, Pakvis DFM, Verdonschot N, Biert J, van Kampen A (2004) A rotational malalignment after intramedullary nailing of femoral fractures. J Orthop Trauma 18:403–409
Jend HH (1986) Computed tomographic determination of the anteversion angle. Premises and possibilities. RöFo 144:447–452
Jerosch J, Finken N, Filler T, Peuker E, Rahgozar M, Lahmer A, Witzel U (1999) Robotics-assisted implantation of femoral components in hip endoprosthetics-an experimental study. Zeitschrift für Orthopädie und ihre Grenzgebiete 137:Oa7–O12
Jiang X, Li X, Wang M, Gu X, Zhang B, Sun L, Zhang L, Liu Y, Liu D, Rong G (2002) Measurement of fracture malrotation after interlocking intramedullary nailing of femoral shaft fracture. zhong hua wai ke za zhi 40:55–58
Kingsley PC, Olmsted KL (1948) A study to determine the angle of anteversion of the neck of the femur. J Bone Jt Surg A 30:745–751
Krettek C, Gler S, Schandelmaier P, Tscherne H (1996) Intramedullary nailing of open fractures. Der Orthopäde 25:223–233
Starker M, Hanusek S, Rittmeister M, Thoma W (1998) Validation of computerized tomography antetorsion angle measurement of the femur. Zeitschrift für Orthopädie und ihre Grenzgebiete 136(5):420–427
Strecker W, Suger G, Kinzl L (1996) Local complications of intramedullary nailing. Der Orthopäde 25:274–291
Sugano N, Noble PC, Kamaric E (1998) A comparison of alternative methods of measuring femoral anteversion. J Comput Assist Tomogr 22(4):610–614
Tönnis D, Heinecke A (1999) Acetabular and femoral anteversion: relationship with osteoarthritis of the hip. J Bone Jt Surg 81:1747–1770
Winquist RA, Hansen ST, Clawson DK (1984) Closed intramedullary nailing of femoral fractures. A report of five hundred and twenty cases. J Bone Jt Surg 66:529–539
Conflict of interest
We do not have any competing interests.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Hoigné, D., Hauck, R. & Babst, R. Technique for intraoperative determination of femoral rotation with a lateral femur nail (LFN, Synthes, Oberdorf, Switzerland). Arch Orthop Trauma Surg 131, 1649–1654 (2011). https://doi.org/10.1007/s00402-011-1361-y
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-011-1361-y