This study presents an arthroscopic removal technique for proud metallic suture anchor after Bankart repair and analyzes the cause of anchor failures.
Patients
Six male patients with an average age 23 years who had proud anchor on the glenoid surface were included. The diagnosis of six patients at the time of the primary surgery was traumatic anterior shoulder instability. Four patients had arthroscopic repair and two had open Bankart repair previously. Four patients complained of pain accompanying a metallic clicking sound during shoulder motion which increased with abduction and external rotation. One patient had pain with apprehension of dislocation and another patient suffered from only pain. Most symptoms had been revealed during the rehabilitation period (average 8.3 months) and confused with postoperative pain. The protruded anchors were detected through plain radiographs in four patients and during arthroscopic examination in two patients.
Method
To extract the proud anchor arthroscopically, a screw driver of a larger diameter than that of the proud suture anchor was used to retrieve the anchor. A larger screw driver was striked with a hammer along the previous suture anchor hole to make a room between the suture anchor and the adjacent glenoid bone so that the hole of the suture anchor became larger. After the hole was widened, the suture anchor had enough room to move freely and it could be removed with a grasper or a mosquito easily. The location of the proud anchor was 2, 3 and 5 o’clock in three patients and 4 o’clock in three patients. In two patients, the suture anchor was malpositioned about 5 mm medial from the anterior glenoid rim. All patients had chondral damage on the humeral head.
Results
Following the procedure none had shoulder instability in 3 years of follow-up. Preoperative visual analog scale score for pain was an average of 3.5. The visual analog scale score for pain was decreased to 1.2 after surgery. All patients had a slight limitation of external rotation preoperatively, and they showed a normal range of motion postoperatively. Constant score improved from 65 to 89, and similarly, American Shoulder and Elbow Society score increased from 67 to 88 after the operation.
Conclusion
Despite small numbers of patients, a successful removal of proud metal suture anchors was achieved using a large empty suture anchor screw driver, which is a simple and reproducible method to remove a proud metallic suture anchor arthroscopically.
Similar content being viewed by others
References
Cole BJ, Romeo AA (2001) Arthroscopic shoulder stabilization with suture anchors: technique, technology, and pitfalls. Clin Orthop Relat Res 390:17–30
Kim SH, Ha KI (2002) Bankart repair in traumatic anterior shoulder instability: open versus arthroscopic technique. Arthroscopy 18:755–763
Cole BJ, L’Insalata J, Irrgang J, Warner JJ (2000) Comparison of arthroscopic and open anterior shoulder stabilization. A two to six-year follow-up study. J Bone Joint Surg Am 82:1108–1114
Mohtadi NG, Bitar IJ, Sasyniuk TM, Hollinshead RM, Harper WP (2005) Arthroscopic versus open repair for traumatic anterior shoulder instability: a meta-analysis. Arthroscopy 21:652–658
Detrisac DA, Johnson LL (1993) Arthroscopic shoulder capsulorrhaphy using metal staples. Orthop Clin North Am 24:71–88
Lane JG, Sachs RA, Riehl B (1993) Arthroscopic staple capsulorrhaphy: a long-term follow-up. Arthroscopy 9:190–194
Rhee YG, Lee DH, Chun IH, Bae SC (2004) Glenohumeral arthropathy after arthroscopic anterior shoulder stabilization. Arthroscopy 20:402–406
Shaffer BS, Tibone JE (1999) Arthroscopic shoulder instability surgery. Complications. Clin Sports Med 18:737–767
Weber SC, Abrams JS, Nottage WM (2002) Complications associated with arthroscopic shoulder surgery. Arthroscopy 18:88–95
Barber FA, Herbert MA, Click JN (1995) The ultimate strength of suture anchors. Arthroscopy 11:21–28
Kaar TK, Schenck RC Jr, Wirth MA, Rockwood CA Jr (2001) Complications of metallic suture anchors in shoulder surgery: a report of 8 cases. Arthroscopy 17:31–37
Bynum CK, Lee S, Mahar A, Tasto J, Pedowitz R (2005) Failure mode of suture anchors as a function of insertion depth. Am J Sports Med 33:1030–1034
Roth CA, Bartolozzi AR, Ciccotti MG et al (1998) Failure properties of suture anchors in the glenoid and the effects of cortical thickness. Arthroscopy 14:186–191
Zuckerman JD, Matsen FA 3rd (1984) Complications about the glenohumeral joint related to the use of screws and staples. J Bone Joint Surg Am 66:175–180
Ekelund A (1998) Cartilage injuries in the shoulder joint caused by migration of suture anchors or mini screw. J Shoulder Elbow Surg 7:537–539
Silver MD, Daigneault JP (2000) Symptomatic interarticular migration of glenoid suture anchors. Arthroscopy 16:102–105
Agneskirchner JD, Hosseini H, Troger M, Lobenhoffer P (2006) Komplikationen bei rekonstruktiven arthroskopischen achulteroperationen. Arthroskopie 19:157–171
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Jeong, JH., Shin, SJ. Arthroscopic removal of proud metallic suture anchors after Bankart repair. Arch Orthop Trauma Surg 129, 1109–1115 (2009). https://doi.org/10.1007/s00402-009-0847-3
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-009-0847-3