Abstract
Introduction
The aim of this study was to describe early results of a new internal locking system, PHILOS, used for the treatment of proximal humeral fractures.
Materials and methods
A chart and radiographic review of 25 cases that had proximal humeral internal locking system (PHILOS) plate for the treatment of proximal humeral fractures was performed.
Results
Of the 25 cases, 20 went to union with a mean neck/shaft angle of 127.2°. Five cases required or were considered for revision surgery for non-union or implant failure. Of the 25 implants, 4 had screw protrusion into the gleno-humeral joint, 4 had screw loosening and backing out, and 1 plate broke without further trauma.
Conclusions
Our results suggest that PHILOS is an effective system for providing fracture stabilisation to bony union but awareness of potential hardware complications is essential.
Similar content being viewed by others
References
Agel J, Jones CB, Sanzone AG, Camuso M, Henley MB (2004) Treatment of proximal humeral fractures with polarus nail fixation. J Shoulder Elbow Surg 13:191–195
Bernard J, Charalambides C, Aderinto J, Mok D (2000) Early failure of intramedullary nailing for proximal humeral fractures. Injury 31:789–792
Burton D, Wells G, Watters A, Schilders E, Venkateswaran B (2005) Early experience with the PlantTan fixator plate for 2 and 3 part fractures of the proximal humerus. Injury 36:1190–1196
Haidukewych GJ (2004) Innovations in locking plate technology. J Am Acad Orthop Surg 12:205–212
Hintermann B, Trouillier H, Schafer D (2000) Rigid internal fixation of fractures of the proximal humerus in older patients. J Bone Joint Surg Br 8:1107–1112
Iannotti JP, Gabriel JP, Schneck SL, Evans BG, Misra S (1992) The normal glenohumeral relationships. An anatomical study of one hundred and forty shoulders. J Bone Joint Surg Am 74:491–500
Kristiansen B, Christensen SW (1986) Plate fixation of proximal humeral fractures. Acta Orthop Scand 57:320–323
Machani B, Sinopidis C, Brownson P, Papadopulos P, Gibson J, Frostick SP (2006) Mid term results of PlantTan plate in the treatment of proximal humeral fractures. Injury 37:269–276
Neer CS II (1970) Displaced proximal humeral fractures. Classification and evaluation. J Bone Joint Surg Am 52:1077–1089
Perren S (2003) Backgrounds of the technology of internal fixators. Injury 34:S-B1–S-B3
Rajasekhar C, Ray P, Bhamra M (2001) Fixation of proximal humeral fractures with the polarus nail. J Shoulder Elbow Surg 10:7–10
Robinson CM, Christie J (1993) The two part-proximal humeral fracture: a review of operative treatment using two techniques. Injury 24:123–125
Robinson CM, Page R (2003) Severely impacted valgus proximal humeral fractures. J Bone Joint Surg Am 9:1647–1655
Sadowski C, Riand N, Stern R, Hoffmeyer P (2003) Fixation of fractures of the proximal humerus with the PlantTan humerus fixator plate: early experience with a new implant. J Shoulder Elbow Surg 12:148–151
Wijgman A, Roolker W, Patt T, Raaymakers E, Marti R (2002) Open reduction and internal fixation of three and four part fractures of the proximal part of the humerus. J Bone Joint Surg Am 84:1919–1924
Zyto K, Ahrengart L, Sperber A, Tornkvist H (1997) Treatment of displaced proximal humeral fractures in elderly patients. J Bone Joint Surg Br 3:412–417
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Charalambous, C.P., Siddique, I., Valluripalli, K. et al. Proximal humeral internal locking system (PHILOS) for the treatment of proximal humeral fractures. Arch Orthop Trauma Surg 127, 205–210 (2007). https://doi.org/10.1007/s00402-006-0256-9
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-006-0256-9