Abstract.
We developed the Hy-Flex II total knee and ligament balancing system (Hy-Flex II total knee) which provides (1) a flexion angle of at least 120° or more, (2) a range of motion (ROM) comparable to or greater than that before surgery, and (3) occasional full flexion. The system design has several features: small posterior femoral condyle radius, a 4° posterior tilting of the tibial joint surface, and equal tension of the bilateral soft tissues obtained by using a ligament tensor. From September 1997 to June 1999, Hy-Flex II total knee arthroplasty was carried out on 114 joints of 84 patients with rheumatoid arthritis. One year after surgery, the average flexion in our series was 122.1°±15.0°. The number of knees operated on which had a flexion angle of 120° or greater was 82 (71.9%) 1 year after surgery among the total of 114 knees. In total, 94 (82.5%) joints obtained the same level or better than the preoperative flexion; 26 (22.8%) attained full flexion 4 weeks after the operation and 15 (13.1%), 1 year after the operation. These results suggest that this system will be able to achieve our aims in almost all rheumatoid knees.
Similar content being viewed by others
Author information
Authors and Affiliations
Additional information
Electronic Publication
Rights and permissions
About this article
Cite this article
Yamazaki, J., Ishigami, S., Nagashima, M. et al. Hy-Flex II total knee system and range of motion. Arch Orthop Trauma Surg 122, 156–160 (2002). https://doi.org/10.1007/s00402-001-0354-7
Received:
Issue Date: