Atrial and ventricular lead insulation defects with resulting inappropriate shocks and end-of-life of the ICD pulse generator in a young bodybuilder with congenital long-QT-syndrome

  • ORIGINAL PAPER
  • Published:
Clinical Research in Cardiology Aims and scope Submit manuscript

Abstract

We report on a 19-year-old male with a congenital Long-QT syndrome who was admitted to our hospital because of insulation defects of both atrial and ventricular ICD leads resulting in inappropriate delivered shocks and a non-responding pulse generator during telemetrical evaluation. The insulation defects led to multiple arc marks within the ICD pocket and there was a short circuit between denuded leads and the electrically active pulse generator. The high current flow generated sufficient heat to damage several circuits of the generator. A new ICD system was implanted successfully and the postoperative course has been remained uneventful for almost 6 years.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Bracke FA, Meijer A, Van Gelder LM (2002) Malfunction of endocardial defibrillator leads and lead extraction: where do they meet? Europace 4:19–24

    Article  PubMed  CAS  Google Scholar 

  2. Ellenbogen KA, Wood MA, Shepard RK, Clemo HF, Vaughn T, Holloman K, Dow M, Leffler J, Abeyratne A, Verness D (2003) Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications. J Am Coll Cardiol 41:73–80

    Article  PubMed  Google Scholar 

  3. Gradaus R, Breithardt G, Bocker D (2003) ICD leads: design and chronic dysfunctions. Pacing Clin Electrophysiol 26:649–657

    Article  PubMed  Google Scholar 

  4. Heidbuchel H (2007) Implantable cardioverter defibrillator therapy in athletes. Cardiol Clin 25:467–482

    Article  PubMed  Google Scholar 

  5. Kitamura S, Satomi K, Kurita T, Shimizu W, Suyama K, Aihara N, Niwaya K, Kobayashi J, Kamakura S (2006) Long-term follow-up of transvenous defibrillation leads: high incidence of fracture in coaxial polyurethane lead. Circ J 70:273–277

    Article  PubMed  Google Scholar 

  6. Kleemann T, Becker T, Doenges K, Vater M, Senges J, Schneider S, Saggau W, Weisse U, Seidl K (2007) Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation 115:2474–2480

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Claudius Diez MD.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Diez, C., Hofmann, HS. & Börgermann, J. Atrial and ventricular lead insulation defects with resulting inappropriate shocks and end-of-life of the ICD pulse generator in a young bodybuilder with congenital long-QT-syndrome. Clin Res Cardiol 98, 16–18 (2009). https://doi.org/10.1007/s00392-008-0711-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00392-008-0711-4

Keywords

Navigation