Abstract

Introduction

Although the Nuss procedure for pectus excavatum is widely employed, a variety of complications have been reported with relatively high frequency; those that involve cardiac and pericardial injuries can be life threatening. To reduce such dangers, we present here a newly developed sternal elevator.

Materials and methods

The elevator is horseshoe shaped. Its elevator side has the same curvature as a Nuss introducer, so that interference between devices is minimal and no extra skin incision is needed for the elevator insertion. The elevator holds the sternum forward and enlarges the retrosternal space for safer passage of thoracoscopically guided introducer.

Results

The authors have used the elevator for 61 pectus excavatum cases between March 2004 and December 2009 without any major complications. The entire process of substernal tunneling was endoscopically observed, which eliminated any blunt and blind dissection, even in a significantly depressed funnel chest case. With the device, the sternum was effectively elevated again for the placement of the second plate in 30 cases.

Conclusion

Our newly developed sternum elevator makes the Nuss procedure safer and more affordable without introducing any extra scarring.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Nuss D, Kelly RE Jr, Croitoru DP, Katz ME (1998) A 10-year review of a minimally invasive technique for the correction of pectus excavatum. J Pediatr Surg 33(4):545–552 (pii:S0022-3468(98)90314-1)

    Article  PubMed  CAS  Google Scholar 

  2. Shamberger RC (2000) Cardiopulmonary effects of anterior chest wall deformities. Chest Surg Clin N Am 10(2):245–252

    PubMed  CAS  Google Scholar 

  3. Castellani C, Schalamon J, Saxena AK, Hoellwarth ME (2008) Early complications of the Nuss procedure for pectus excavatum: a prospective study. Pediatr Surg Int 24(6):659–666. doi:10.1007/s00383-008-2106-z

    Article  PubMed  Google Scholar 

  4. Kelly RE Jr, Shamberger RC, Mellins RB, Mitchell KK, Lawson ML, Oldham K, Azizkhan RG, Hebra AV, Nuss D, Goretsky MJ, Sharp RJ, Holcomb GW 3rd, Shim WK, Megison SM, Moss RL, Fecteau AH, Colombani PM, Bagley TC, Moskowitz AB (2007) Prospective multicenter study of surgical correction of pectus excavatum: design, perioperative complications, pain, and baseline pulmonary function facilitated by internet-based data collection. J Am Coll Surg 205(2):205–216. doi:10.1016/j.jamcollsurg.2007.03.027

    Article  PubMed  Google Scholar 

  5. Park HJ, Lee SY, Lee CS (2004) Complications associated with the Nuss procedure: analysis of risk factors and suggested measures for prevention of complications. J Pediatr Surg 39(3):391–395 (pii:S0022346803008704)

    Article  PubMed  Google Scholar 

  6. Boehm RA, Muensterer OJ, Till H (2004) Comparing minimally invasive funnel chest repair versus the conventional technique: an outcome analysis in children. Plast Reconstr Surg 114(3):668–673 (discussion 674–665. pii:00006534-200409010-00007)

    Article  PubMed  Google Scholar 

  7. Uemura S, Nakagawa Y, Yoshida A, Choda Y (2003) Experience in 100 cases with the Nuss procedure using a technique for stabilization of the pectus bar. Pediatr Surg Int 19(3):186–189. doi:10.1007/s00383-002-0915-z

    PubMed  Google Scholar 

  8. Nuss D, Croitoru DP, Kelly RE Jr, Goretsky MJ, Nuss KJ, Gustin TS (2002) Review and discussion of the complications of minimally invasive pectus excavatum repair. Eur J Pediatr Surg 12(4):230–234. doi:10.1055/s-2002-34485

    Article  PubMed  CAS  Google Scholar 

  9. Croitoru DP, Kelly RE Jr, Goretsky MJ, Lawson ML, Swoveland B, Nuss D (2002) Experience and modification update for the minimally invasive Nuss technique for pectus excavatum repair in 303 patients. J Pediatr Surg 37(3):437–445 (pii:S0022346802303683)

    Article  PubMed  Google Scholar 

  10. Miller KA, Woods RK, Sharp RJ, Gittes GK, Wade K, Ashcraft KW, Snyder CL, Andrews WM, Murphy JP, Holcomb GW 3rd (2001) Minimally invasive repair of pectus excavatum: a single institution’s experience. Surgery 130(4):652–657. doi:10.1067/msy.2001.116917 (discussion 657-659)

    Article  PubMed  CAS  Google Scholar 

  11. Hebra A, Swoveland B, Egbert M, Tagge EP, Georgeson K, Othersen HB Jr, Nuss D (2000) Outcome analysis of minimally invasive repair of pectus excavatum: review of 251 cases. J Pediatr Surg 35(2):252–257 (discussion 257–258. pii:S0022-3468(00)90019-8)

    Article  PubMed  CAS  Google Scholar 

  12. Belcher E, Arora S, Samancilar O, Goldstraw P (2008) Reducing cardiac injury during minimally invasive repair of pectus excavatum. Eur J Cardiothorac Surg 33(5):931–933. doi:10.1016/j.ejcts.2008.01.030

    Article  PubMed  Google Scholar 

  13. Gips H, Zaitsev K, Hiss J (2008) Cardiac perforation by a pectus bar after surgical correction of pectus excavatum: case report and review of the literature. Pediatr Surg Int 24(5):617–620. doi:10.1007/s00383-007-2097-1

    Article  PubMed  Google Scholar 

  14. Moss RL, Albanese CT, Reynolds M (2001) Major complications after minimally invasive repair of pectus excavatum: case reports. J Pediatr Surg 36(1):155–158 (pii:S0022346801420410)

    Article  PubMed  CAS  Google Scholar 

  15. Willekes CL, Backer CL, Mavroudis C (1999) A 26-year review of pectus deformity repairs, including simultaneous intracardiac repair. Ann Thorac Surg 67(2):511–518 (pii:S0003-4975(99)00015-6)

    Article  PubMed  CAS  Google Scholar 

  16. Hamajima A, Suzuki N, Kuroiwa M, Suzuki M, Sakamoto J, Tsuchida Y (2003) A technique for the Nuss procedure using a U-shaped elevator. Jpn J Pediatr Surg 35:722–726

    Google Scholar 

Download references

Conflict of interest

The authors declare that they have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Satoshi Takagi.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Takagi, S., Oyama, T., Tomokazu, N. et al. A new sternum elevator reduces severe complications during minimally invasive repair of the pectus excavatum. Pediatr Surg Int 28, 623–626 (2012). https://doi.org/10.1007/s00383-012-3087-5

Download citation

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00383-012-3087-5

Keywords

Navigation