Introduction

Ventricular shunts have long formed the mainstay of hydrocephalus treatment with ventriculo-peritoneal shunts (VPS) proving the most popular, utilised in an estimated 3500 procedures annually in the UK [1]. Ventricular shunting is generally accepted to be associated with more complications than any other neurosurgical intervention, including valve or catheter blockage, breakage, infection [1, 2] and increasingly reported over-drainage [3,4,5,6]. The combined effects of hydrostatic forces within CSF and gravity in the upright position result in a siphon effect on CSF, leading to excess drainage of fluid from the ventricles [7, 8]. In the developing brain, such over-drainage has been demonstrated to affect cerebral morphology, leading to premature synostosis and microcephaly at the detriment of normal neurodevelopment [4, 5]. In order to mitigate this phenomenon and better mimic physiological CSF drainage, the use of differential pressure valves has become a standard practice when fitting VPS [8].

The Miethke proGAV®2.0 comprises an adjustable differential pressure unit (DPU) composed of a solid titanium body with a ball-in-cone valve. A pre-tensioned bow spring, externally adjustable via magnets, defines the opening pressure of the DPU. The pre-tensioning of the spring and thus the opening pressure can be adjusted by turning a rotor, with the valve implanted under the patient’s skin. Distal to the DPU lies a gravitational unit of fixed resistance, governed by the weight of a tantalum ball. When the patient is lying down, the gravitational unit remains open, and therefore CSF outflow from the skull meets little resistance. Hence, the opening pressure of the proGAV®2.0 is defined by the setting of the adjustable DPU. When the patient moves into an upright position, the gravitational unit closes. In addition to the opening pressure of the adjustable DPU, the weight of the tantalum ball has to be exceeded; thus, the opening pressure becomes the sum of both DPU and gravitational unit. Only when the sum of the intraventricular pressure and the hydrostatic pressure exceeds the opening pressure of the combined components is drainage permitted [9].

Typically, failure of a fixed-pressure valve system requires valve replacement which is accompanied by the risk of complication. Limiting the need for repeat surgical intervention is a crucial advantage to programmable valves; however, they are accompanied by their own complications [6, 10, 11], and there is evidence to suggest that shunt survival is not significantly different from that of fixed-pressure valves [12, 13]. Given their complex mechanisms, these valves are typically more expensive than their fixed-pressure counterparts, and the debate over the cost-benefit is ongoing [10].

In complex patients for whom multiple valve revisions are required, the need to minimise the number of operations is crucial, and recent data from our centre has demonstrated the positive impact of using fixed-pressure valves in treating hydrocephalus [14]. This study aims to describe the experience of a tertiary paediatric neurosurgical centre in using the proGAV®2.0 programmable valve.

Methods

Clinical notes of all patients receiving proGAV®2.0 valves at either initial insertion or revision of shunts at Alder Hey Children’s Hospital NHS Foundation Trust between January 2014 and September 2018 were collected. Data was collected according to the following headings: (i) patient demographics, (ii) symptomatology (pre and postoperatively), (iii) indication for shunt/valve insertion, (iv) frequency of surgical intervention, (v) frequency of valve setting alteration and (vi) ventricular linear metrics.

For comparison, Miethke fixed differential pressure valve survival data for those inserted between January 2014 and September 2018 was extracted from our centre’s database [15]. Miethke paediGAV, GAV and DualSwitch valves were assessed. Patient demographics, ventricular metrics and shunt revision rate pre- and post-insertion were not available for fixed-pressure valves.

Shunt failure was defined as a surgical intervention wherein a valve or entire shunt system was replaced. Systems in situ for <1 month were excluded from survival analysis to account for immediate postoperative or inpatient-related complications. Kaplan-Meier curves were produced for (i) overall shunt system survival, (ii) shunt system survival (excluding infection and distal failure), (iii) de novo vs replacement valve survival and (iv) ≤12 months vs >12 months on valve insertion.

Over-drainage was defined according to a combination of ventricular morphology, symptomatic presentation of the patient and, if performed, ICP monitor values. The rationale for selecting appropriate replacement valves in the cases where revision was necessary was decided on a case-by-case basis. Gravitational unit choice was made according to Miethke guidance: 20cmH2O for those under 5 years of age and 25cmH2O for over 5 years. DPU pressure setting on insertion was decided on a case-by-case basis, guided by Miethke manufacturer recommendations ranging from 5cmH2O to 10cmH2O and above in those with particularly wide ventricles. Adjustments to the DPU pressure setting were made according to either inadequate response in ventricular metrics, the patient remaining symptomatic or ICP monitor values.

For all patients, preoperative and postoperative imaging was assessed to determine changes in ventricular linear metrics. T2-weighted MRI scans were used to measure frontal and occipital horn ratio (FOHR) and frontal and occipital horn width ratio (FOHWR). In the absence of T2-weighted MRI, CT scans were used. All measurements were taken at the level of the intraventricular foramen of Monro, as identified on axial images. The FOHR was defined as the mean value of the frontal and occipital horn width divided by twice the widest biparietal diameter (BPD) [16, 17]. FOHWR was then defined as the average of the maximum width of the individual frontal and occipital horns, divided by twice the widest BPD [15]. The closest available scans prior to and following shunt insertion/revision were used to assess any changes.

The latest versions of Microsoft Excel and SPSS statistical software were used for data collection and statistical analysis respectively. Kaplan-Meier curves and log-rank tests were used to assess survival outcomes. Chi-squared test was used to compare categorical variables. A threshold of p = 0.05 was used to determine significance.

Results

Patient population

Seventy-seven patients were identified as having received a proGAV®2.0 valve, of which n = 45 (58%) were male. The overall mean age of the cohort was 5.1 years (IQR: 0.4–11.0 years). Median age for de novo proGAV®2.0 insertion was 11 months and for revision was 60 months. The valve population was dichotomised into groups ≤12 months of age (n = 35, 39.8%) and >12 months of age (n = 53, 60.2%).

Valve data

Median follow-up was 17.5 months (1.0–47.3 months). Eighty-eight proGAV®2.0 valves were inserted over the time period analysed; therefore, n = 88 cases were available for assessment. Of these, n = 42 (48%) were as part of de novo shunt insertions and the remainder as revisions (n = 46, 52%). Sixty-six (86%) patients received n = 1 proGAV® 2.0, and 11 (14%) patients received n = 2. Causes of hydrocephalus that indicated initial shunt insertion included intraventricular haemorrhage (n = 22, 28.6%), spina bifida (n = 8, 10.4%), tumours (n = 5, 6.5%) and aqueductal stenosis (n = 4, 5.2%) (Table 1).

Table 1 The indications for insertion of a proGAV2.0 valve as part of a de novo shunt system

Indications for replacement of other valves with proGAV®2.0s included valve obstruction (n = 12, 35%), infection (n = 4, 12%) and over-drainage (n = 9, 26%) (Table 2). No patients who received a proGAV®2.0 for over-drainage had recurrence of their symptoms after insertion during follow-up.

Table 2 The indications for revision of different valve subtypes in our cohort

N = 20 (23%) proGAV®2.0 valves required revision. Obstruction was the leading indication for proGAV®2.0 revision (n = 11, 55%), followed by valve mechanism failure (n = 3, 15%) and catheter failure (n = 2, 10%) (Table 2). ProGAV®2.0s were exchanged ‘like for like’ in n = 9 (20%) cases. Of these, indications for revision were valve blockage (n = 6, 67%), catheter/distal tubing problem (n = 2, 22%) and infection (n = 1, 11%). N = 5 (56%) of proGAV®2.0 ‘like for like’ revisions remained in situ at the latest data collection.

A total of 102 patients received a variety of Miethke fixed-pressure valves during the same time period: 4/24 (n = 14, 13.7%), 9/24 (n = 2, 19.6%), 9/29 (n = 82, 80.4%), 10/40 (n = 3, 2.9%) and DualSwitch (n = 1, 1%). Over-drainage led to revision in n = 2 (1.9%) fixed-pressure valves. N = 42 required revision, due to valve obstruction (n = 14, 33%), infection (n = 10, 24%) and over-drainage (n = 2, 5%) (Table 2).

The only significant difference in aetiology of failure between valve types was over-drainage, occurring significantly more often in the series of valves replaced by a proGAV®2.0 (n = 9, 26%) than in either proGAV®2.0 (n = 1, 1.1%) or Miethke fixed-pressure valves (n = 2, 1.9%) (p < 0.01).

Symptomatology

Each proGAV®2.0 inserted (n = 88) was treated as an individual case when analysing symptomatology before and after valve insertion, of which n = 78 (89%) had available data (n = 33 ≤12 months; n = 45 >12 months). In those ≤12 months, increasing head circumference, vomiting, lethargy, behavioural changes and sunsetting occurred most frequently. In those >12 months of age, headache, vomiting, nausea, lethargy and behavioural changes were most common. All patients found to be over-draining presented with low-pressure headaches (Table 3).

Table 3. Clinical presentation data of those receiving proGAV2.0 valves, grouped according to age

ICP monitoring

N = 11 of 88 (12.5%) had available ICP monitoring data pre-insertion of proGAV®2.0, and n = 5 (5.7%) had available post-insertion data. Pre-insertion, n = 6 (54.5%) had raised ICP, and n = 5 (45.5%) had low ICP. Post-insertion, n = 1 (20%) was normal, n = 1 (20%) had high pressure and n = 3 (60%) low pressure (1 of which resulted in further valve revision).

Pressure settings

N = 44 (57.1%) patients were under 5 years and therefore received 20cmH20 gravitational units and n = 33 (42.9%) over 5 years and therefore received 25cmH20 gravitational units. Initial pressure setting data for the DPU of newly inserted proGAV®2.0s was available for n = 69 valves, of which the median was 10cmH2O (range: 5–14). Over the time period analysed, ‘pressure alteration’ data was available for n = 71 valves; n = 41 (57.7%) of which required no alteration. Of the 30 valves (42.3%) requiring adjustment, n = 16 (22.5%) required 1 adjustment, n = 8 (11.2%) required 2 adjustments and n = 6 patients (8.5%) required ≥3 adjustments.

Valve survival

N = 3 (3.4%) proGAV®2.0 and n = 13 (12.7%) fixed-pressure valves were replaced within 1 month of insertion and so were excluded from further analysis. Overall mean time to valve revision in the proGAV®2.0 group was 37.1 months (95%CI: 32.8–41.3), and overall survival (OS) was 80.4% at 12 months compared to 31.0 months (95%CI: 25.7–36.3) and 58.3% at 12 months in the fixed valve group (Fig. 1) (p = 0.001). After excluding infection and distal failure as cause of failure, mean time to revision rose to 38.0 months (95%CI: 33.8–42.2) and cumulative survival to 82.7% with a proGAV®2.0, compared to 35.1 months (95%CI: 29.6–40.6) and 66.6% at 12 months with fixed-pressure valves (p = 0.02) (Fig. 2).

Fig. 1
figure 1

Comparative cumulative survival of the proGAV®2.0 valve (red) and Miethke fixed-pressure valves (blue)

Fig. 2
figure 2

Comparative cumulative survival of the proGAV®2.0 valve (red) and Miethke fixed-pressure valves (blue) with exclusion of all failures due to infection, distal catheter or cause not otherwise specified

Within the proGAV®2.0 cohort alone, de novo valves had a mean time to revision of 31.9 months and 12-month cumulative survival of 73%, compared to a mean time to revision of 40.3 months and 12-month survival of 88.6% in revised valves (p = 0.10) (Fig. 3). Of the proGAV®2.0 valves exchanged ‘like for like’, mean survival of the first valve was shorter than the second (7.3 vs 11.9 months), though not significantly so (p > 0.1). ProGAV®2.0s in patients ≤12 months had a mean OS of 35.3 months and cumulative survival at 12 months of 71.2%, compared to a mean OS of 37.3 months and 86.3% 12-month cumulative survival in those >12 months (p = 0.25) (Fig. 4).

Fig. 3
figure 3

Cumulative survival of de novo (blue) vs revised (red) proGAV®2.0 valves

Fig. 4
figure 4

Cumulative survival of proGAV®2.0 valves in patients ≤12 months of age on insertion (blue) vs those >12 months (red)

Ventricular metrics

Across the entire cohort of proGAV®2.0 insertions, a statistically significant decrease was seen in both mean FOHR and FOHWR. The mean decrease in FOHR was 0.014 (95%CI: 0.006–0.023, p = 0.002), whilst the mean decrease in FOHWR was 0.037 (95%CI: 0.005–0.069, p = 0.024) (Table 4).

Table 4 Change in ventricular size following insertion of proGAV®2.0 differential pressure valves as defined by fronto-occipital horn ratio and fronto-occipital horn width ratio

Across the cohort, n = 5 (6.5%) had slit ventricles before and after treatment (all of which were valve revisions). N = 4 of these of these had measurable ventricular metrics, of which none changed significantly following insertion; however, all of these patients had improvement in their symptoms.

Discussion

This study aims to describe a tertiary paediatric neurosurgical centre’s experience with the proGAV®2.0 programmable valve.

Across the proGAV®2.0 cohort, a marked improvement in symptoms was observed after valve insertion, accompanied by a significant decrease in ventricular metrics (p = 0.002, p = 0.024). In patients with slit ventricles and symptoms of over-drainage, despite clinically improving, little real change was seen in their ventricular metrics. Slit-ventricle syndrome is relatively poorly understood, and the persistence of slit ventricles despite intervention and clinical improvement may speak to the nature of ventricular compliance in the syndrome itself or perhaps suggest that FOHR and FOHWR are not useful metrics in this scenario. Regardless, further investigation is needed into slit-ventricle syndrome both independently and in relation to different valve/shunt systems.

Valve revision forms a significant part of the paediatric neurosurgical workload, and numerous sources have reviewed the risks associated with it [18,19,20,21,22]. Theoretically mitigating the need for shunt revision is therefore an attractive feature of programmable valves, and in this cohort the insertion of a proGAV®2.0 valve resulted in a decrease in mean revision rate from 1.77 to 0.25 per patient per year, demonstrating a significant reduction in the number of revisions per patient (p < 0.01). In the future, comparing this impact on the revision rate with that of fixed-pressure valves is crucial and may better inform valve selection for patients [10].

The predecessor to the proGAV®2.0 was extensively investigated, with overall survival rates of 81–88.7% [6, 11, 23] in keeping with those reported here. Whilst similar revision rates have been reported in both programmable and fixed-pressure valves [1013], in this study, both 12-month survival and mean time to revision were significantly greater in the proGAV®2.0 cohort (Figs. 1 and 2). This supports the wider use of programmable valves, not just in those with CSF over-drainage, but as a safe alternative for hydrocephalus in general.

In the proGAV®2.0 group alone, de novo and revised valves did not differ significantly in their mean survival (31.9 vs 40.3 months, p = 0.10) [21, 22]. The shorter survival seen in de novo valves may be explained by the fact that de novo valves were more frequently placed in a younger cohort in this study (0.9 vs 5 years), a feature that has independently been shown to increase risk of shunt failure [23]. In the proGAV®2.0 group, patients ≤12 months of age at insertion did have shorter mean valve survival times and cumulative survival at 12 months compared to those >12 months of age (35.3 months and 71.2% vs 37.3 months and 86.3% respectively), though this discrepancy was not significant (p = 0.25).

Over-drainage was the indication for revision in 11% of all valve types requiring revision (Table 2). The overall prevalence of over-drainage amongst valves that were replaced by proGAV®2.0s was significantly greater than either proGAV®2.0 or Miethke fixed-pressure valves (p < 0.01) (Table 2), though selection bias is likely responsible for this high prevalence. Both proGAV®2.0s and fixed-pressure valves had low rates of over-drainage (1.1% vs 1.9% respectively) in this study, though these were relatively small cohorts and further follow-up is required. In those that had a proGAV®2.0 inserted for over-drainage, no cases recurred following insertion, demonstrating the valve’s efficacy in tackling the condition. Infection as a cause of shunt revision constituted 10% of proGAV®2.0 revisions but occurred in only 2.2% of all proGAV®2.0 valves inserted. This is markedly lower than infection rates identified in the literature for other valves [8, 11, 24], as well as other valves in both this series. Other complications rates were similar in prevalence to the literature (6,21,22).

Whilst ICP monitoring may be utilised to determine whether valve revision is necessary prior to surgery, it played a relatively small role in our centre, with only 12.5% having ICP monitoring. Often if clinical suspicion of over-drainage persists despite unremarkable ICP values, surgery may be opted for regardless. As is evidenced by the frequent adjustments made to the valves in this cohort, programmable valves such as the proGAV®2.0 provide an opportunity to delay the need for surgery, with minimal effort and risk. Anecdotally, practitioners at the centre found the means by which to adjust the proGAV®2.0, simple to understand and straightforward to perform.

Limitations to our study include the retrospective nature of the data and the fact that accurately determining aetiology of shunt failure is subject to observer bias. Our study also does not assess the impact of fixed-pressure valves on ventricular metrics or revision rate before and after insertion. This was not examined in the original study either [15]; therefore, we hope to investigate these outcome measures in the near future. The financial cost associated with programmable valves cannot be overlooked, and whilst this study provides encouraging evidence for their wider use, we did not perform a cost-analysis, and this variable should be considered in the future.

Conclusion

The proGAV®2.0 programmable valve reduced the need for valve revision in patients with ventricular over-drainage, such that no cases re-occurred during the follow-up period. Both 12-month survival and mean time to revision were significantly greater in proGAV®2.0 programmable valves compared to Miethke fixed-pressure valves. This study provides evidence to support the wider use of programmable valves, but the impact of cost was not assessed, and further investigation is therefore required.