Skip to main content
Log in

Optically sliced micro-PIV using confocal laser scanning microscopy (CLSM)

  • Original
  • Published:
Experiments in Fluids Aims and scope Submit manuscript

Abstract

Optically sliced microscopic-particle image velocimetry (micro-PIV) is developed using confocal laser scanning microscopy (CLSM). The developed PIV system shows a unique optical slicing capability allowing true depth-wise resolved micro-PIV vector field mapping. A comparative study between CLSM micro-PIV and conventional epi-fluorescence micro-PIV is presented. Both techniques have been applied to the creeping Poiseuille flows in two different microtubes of 99-μm (Re=0.00275) and 516-μm ID diameters (Re=0.021), which are respectively imaged by a 40×-0.75NA objective with an estimated 2.8-μm optical slice thickness, and by a 10×-0.30NA objective with a 26.7-μm slicing. Compared to conventional micro-PIV, CLSM micro-PIV consistently shows significantly improved particle image contrasts, definitions, and measured flow vector fields agreeing more accurately with predictions based on the Poiseuille flow fields. The data improvement due to the optical slicing of CLSM micro-PIV is more pronounced with higher magnification imaging with higher NA objectives for a smaller microtube.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1a, b
Fig. 2
Fig. 3a, b
Fig. 4
Fig. 5a, b
Fig. 6
Fig. 7a, b
Fig. 8
Fig. 9
Fig. 10
Fig. 11

Similar content being viewed by others

Notes

  1. This is also called a “far-field” diffraction condition, which is defined as R>a 2/λ, where R is the smaller of the two distances from the particle to the objective lens and the objective lens to the imaging detector, a is the particle radius, and λ is the wavelength in the medium. For typical conditions for micro-PIV, R~1 mm, a~200 nm, and λ~500 nm, the inequality is well satisfied by a ratio of greater than 12,000.

  2. Numerical aperture, NA, is defined as \(NA \equiv n_{i} \sin \theta _{{\max }} \), where n i is the refractive index of the immersing medium (air, water, oil, etc.) adjacent to the objective lens, and θ max is the half-angle of the maximum cone of the light apertured by the lens.

  3. Modified pinhole diameter, PD, is defined as pinhole diameter/magnification, with the pinhole diameter measured in μm.

  4. Airy unit, \({\text{AU}} \equiv {1.22\lambda _{{{\text{ex}}}} } \mathord{\left/ {\vphantom {{1.22\lambda _{{{\text{ex}}}} } {NA}}} \right. \kern-\nulldelimiterspace} {NA} \), with λ ex being the fluorescent excitation wavelength.

  5. The mean wavelength is defined as, \(\ifmmode\expandafter\bar\else\expandafter\=\fi{\lambda } = {\sqrt 2 }\frac{{\lambda _{{{\text{ex}}}} \cdot \lambda _{{{\text{em}}}} }}{{{\sqrt {\lambda _{{{\text{ex}}}} ^{2} + \lambda _{{{\text{em}}}} ^{2} } }}} \).

  6. The reduced apparent depth, h a, is derived as h/n, based on Snell’s law of refraction (Hecht 2002). Strictly speaking, the analysis assumes a planar interface and zero ray-incident angle, thus, for the case of a circular microtube, it is only valid along the centerline (refer to Fig. 9).

  7. The spatial uncertainty of the imaging planes is estimated as an rms of the one-half of the micro-stage reading resolution, 0.5 μm, and the uncertainty level for identifying the top-end point is estimated to be approximately identical to the image depth-of-field (DOF), 0.92 µm for 40× and 5.72 µm for 10× magnification, for the conventional microscope. Note that the values of DOF for CLSM are 0.63 µm and 4.66 µm for 40× and 10× magnification, respectively.

  8. The background noise from the off-focus particle images can be reduced to an acceptable level by limiting the PIV measurement depth to a base-cut level where the field-wide-averaged image intensity reaches one-tenth of the maximum in-focus image intensity (Meinhart et al. 2000).

  9. The actual recorded image of a seed particle on the CCD is a convolution of the geometric particle image, Md p, with the FPS, d s, of the recording optics. Approximating both of the geometric and diffraction-limited images as Gaussian functions, the image diameter, d e, can be expressed as (Born and Wolf 1999) \(d_{{\text{e}}} = {\left[ {M^{2} d_{{\text{p}}} ^{2} + d_{{\text{s}}} ^{2} } \right]}^{{1/2}} \) where d e is the effective particle diameter in the CCD, M is the magnification of the microscope, d p is the particle diameter, and d s is the characteristic diameter of the PSF. For magnifications much larger than unity, the diameter of the diffraction-limited PSF, in the image plane, is given by \(d_{{\text{s}}} = 2.44M\frac{\lambda }{{2NA}} \) where NA is the numerical aperture and λ is the wavelength of light.

  10. Optical path length (OPL) is defined as the local medium thickness multiplied by the local refractive index.

References

  • Born M, Wolf E (1999) Principles of optics, 7th edn. Cambridge University Press, Cambridge, UK, pp 491

  • Conchello JA, Lichtman JW (1994) Theoretical analysis of a rotating-disk partially confocal scanning microscope. Appl Opt 33:585–596

    Google Scholar 

  • Diaspro A (ed) (2002) Confocal and two-photon microscopy: Foundations, application, and advances. Wiley-Liss, New York, pp 101–125, 245–247

  • Egner A, Andersen V, Hell SW (2002) Comparison of the axial resolution of practical Nipkow-disk confocal fluorescence microscopy with that of multifocal multiphoton microscopy: Theory and experiment. J Microsc 206:24–32

    Article  CAS  PubMed  Google Scholar 

  • Hecht E (2002) Optics, 4th edn. Addison Wesley, Reading, MA, pp 467–471

  • Hell S, Reiner G, Cremer C, Stelzer EHK (1993) Aberrations in confocal fluorescence microscopy induced by mismatches in refractive index. J Microsc 169:391–405

    Google Scholar 

  • Ichihara A, Tanaami T, Isozaki K, Sugiyama Y, Kosugi Y, Mikuriya K, Abe M, Uemura I (1996) High-speed confocal fluorescence microscopy using a Nipkow scanner with microlenses for 3-D imaging of single fluorescent molecule in real time. Bioimaging 4:57–62

    Google Scholar 

  • Kimura S, Munakata C (1990) Depth resolution of the fluorescent confocal scanning optical microscope. Appl Opt 29:489–494

    CAS  Google Scholar 

  • Lew HS, Fung YC (1970) Entry flow into blood vessels at arbitrary Reynolds number. J Biomech 3:23–28

    CAS  PubMed  Google Scholar 

  • Meinhart CD, Wereley ST, Gray MHB (2000) Volume illumination for two-dimensional particle image velocimetry. Meas Sci Technol 11:809–814

    CAS  Google Scholar 

  • Merzkirch W (1987) Flow visualization, 2nd edn. Academic Press, Orlando

  • Minsky M (1998) Memoir on inventing the confocal scanning microscope. Scanning 10:128–138

    Google Scholar 

  • Olsen MG, Adrian RJ (2000) Out-of-focus effects on particle image visibility and correlation in microscopic particle image Velocimetry. Exp Fluids 29:S166-S174

    Article  Google Scholar 

  • Prenel JP, Bailly Y (1998) Theoretical determination of light distributions in various laser light sheets for flow visualization. J Flow Vis Image Proc 5:211–224

    Google Scholar 

  • Qian H, Elson EL (1991) Analysis of confocal laser-microscope optics for 3-D fluorescence correlation spectroscopy. Appl Opt 30:1185–1195

    CAS  Google Scholar 

  • Raffel M, Willert CE, Kompenhans J (1998) Particle image velocimetry: A practical guide. Springer, Berlin Heidelberg New York

    Google Scholar 

  • Sandison DR, Webb WW (1994) Background rejection and signal-to-noise optimization in confocal and alternative fluorescence microscopes. Appl Opt 33:603–615

    Google Scholar 

  • Santiago JG, Wereley ST, Meinhart CD, Beebe DJ, Adrian RJ (1998) A particle image velocimetry system for microfluidics. Exp Fluids 25:316–319

    CAS  Google Scholar 

  • Sheppard CJR, Gu M (1991) Aberration compensation in confocal microscopy. Appl Opt 30:3563–3568

    Google Scholar 

  • Sugii Y, Nishio S, Okamoto K (2002) In vivo PIV measurement of red blood cell velocity field in microvessels considering mesentery motion. Physiol Meas 23:403–416

    Article  PubMed  Google Scholar 

  • Thiery L, Prenel JP, Porcar R (1996) Theoretical and experimental intensity analysis of laser sheets for flow visualization. Opt Commun 123:801–809

    Article  CAS  Google Scholar 

  • Tiziani HJ, Uhde HM (1994) Three-dimensional analysis by a microlens-array confocal arrangement. Appl Opt 33:567–572

    Google Scholar 

  • Webb RH (1996) Confocal optical microscopy. Rep Prog Phys 59:427–471

    Article  Google Scholar 

  • Wilhelm S, Grobler B, Gluch M, Heinz H (2003) Confocal laser scanning microscopy: Principles. Carl Zeiss, pp 7–16

    Google Scholar 

Download references

Acknowledgement

The confocal laser scanning microscope (CLSM) system was purchased by the Texas A&M Permanent University Facility (PUF) Award granted to Dr. Kihm’s Micro/nano-scale Fluidics and Heat Transport Laboratory http://go.to/microlab. The authors acknowledge that the current research has been partially sponsored by the NASA-Fluid Physics Research Program, grant no. NAG 3–2712, and partially by a subcontract from the R4D Program at the National Center for Microgravity Research (NCMR). The presented technical contents are not necessarily the representative views of NASA or NCMR.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Kenneth D. Kihm.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Park, J.S., Choi, C.K. & D. Kihm, K. Optically sliced micro-PIV using confocal laser scanning microscopy (CLSM). Exp Fluids 37, 105–119 (2004). https://doi.org/10.1007/s00348-004-0790-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00348-004-0790-6

Keywords

Navigation