A PC program for estimating organ dose and effective dose values in computed tomography

  • Physics radiology
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract.

Dose values in CT are specified by the manufacturers for all CT systems and operating conditions in phantoms. It is not trivial, however, to derive dose values in patients from this information. Therefore, we have developed a PC-based program which calculates organ dose and effective dose values for arbitrary scan parameters and anatomical ranges. Values for primary radiation are derived from measurements or manufacturer specifications; values for scattered radiation are derived from Monte Carlo calculations tabulated for standard anthropomorphic phantoms. Based on these values, organ doses can be computed by the program for arbitrary scan protocols in conventional and in spiral CT. Effective dose values are also provided, both with ICRP 26 and ICRP 60 tissue-weighting coefficients. Results for several standard CT protocols are presented in tabular form in this paper. In addition, potential for dose reduction is demonstrated, for example, in spiral CT and in quantitative CT. Providing realistic patient dose estimates for arbitrary CT protocols is relevant both for the physician and the patient, and it is particularly useful for educational and training purposes. The program, called WinDose, is now in use at the Erlangen University hospitals (Germany) as an information tool for radiologists and patients. Further extensions are planned.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 9 March 1998; Revision received: 4 June 1998; Accepted: 4 November 1998

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kalender, W., Schmidt, B., Zankl, M. et al. A PC program for estimating organ dose and effective dose values in computed tomography. Eur Radiol 9, 555–562 (1999). https://doi.org/10.1007/s003300050709

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s003300050709

Navigation