Thin-section CT of the lung: influence of 0.5-s gantry rotation and ECG triggering on image quality

  • Chest
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract.

The aim of this study was to determine if ECG triggering and a shorter acquisition time of 0.5-s rotation decrease cardiac motion artifacts of thin-section CT of the lung. In 25 patients referred for thin-section thoracic CT, 1-mm thin-section slices were performed with a scanning time of 0.5 s with ECG gating, 0.5 s and 1 s during the diastolic phase of the heart at five identical anatomical levels from the aortic arch to lung basis. At each anatomical level and for each lung, cardiac motion artifacts were graded independently on a four-point scale by three readers. Patients were divided into two groups according to their heart rate. A four-way analysis of variance was used to assess differences between the three modalities. Mean cardiac motion artifacts scores were rated 1.23±0.02, 1.47±0.02, and 1.79±0.02, at 0.5 s with ECG gating, 0.5 s without ECG gating, and 1 s, respectively (F=139, p<0.0001). At the four anatomical levels below the aortic arch, the left lung scores were greater than the right lung score for the three modalities. For the modality 0.5 s with ECG gating no difference of scores was found between patients grouped according to their cardiac frequency. The 0.5-s gantry rotation with or without ECG gating scans reduces cardiac motion artifacts on pulmonary thin-section CT images and is mainly beneficial for the lower part of the left lung.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Electronic Publication

Rights and permissions

Reprints and permissions

About this article

Cite this article

Montaudon, M., Berger, P., Blachère, H. et al. Thin-section CT of the lung: influence of 0.5-s gantry rotation and ECG triggering on image quality. Eur Radiol 11, 1681–1687 (2001). https://doi.org/10.1007/s003300000810

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s003300000810

Navigation