Abstract
Objective
Peripherally inserted central catheter (PICC) use continues to increase, leading to the development of a blind bedside technique (BST) for placement. The aim of our study was to compare the BST with the fluoroscopically guided technique (FGT), with specific regard to catheter tip position (CTP).
Materials and methods
One hundred eighty patients were randomized to either the BST or the FGT. All procedures were done by the same interventional team and included postprocedural chest X-ray to assess CTP. Depending on the international guidelines for optimal CTP, patients were classified in three types: optimal, suboptimal not needing repositioning, and nonoptimal requiring additional repositioning procedures. Fisher’s test was used for comparisons.
Results
One hundred seventy-one PICCs were successful inserted. In the BST groups, 23.3% of placements were suboptimal and 30% nonoptimal, requiring repositioning. In the FGT group, 5.6% were suboptimal and 1.1% nonoptimal. Thus, suboptimal and nonoptimal CTP were significantly lower in the FGT group (p < 0.001).
Conclusion
Tip malposition rates are high when using blind BST, exposing the patient to an increased risk of deep venous thrombosis and catheter malfunction. Using the FGT or emerging technologies that could help tip positioning are recommended, especially for long-term indications.
Key points
• Bedside and fluoroscopy guided techniques are commonly used for PICC placement.
• Catheter malposition is the major technical issue with the bedside technique.
• Catheter malposition occurred in 53% of patients with the bedside technique.
Similar content being viewed by others
Abbreviations
- PICC:
-
Peripherally inserted central catheters
- BST:
-
Blind bedside technique
- FGT:
-
Fluoroscopically guided technique
- CTP:
-
Catheter tip position
- CICV:
-
Centrally inserted venous catheters
- SVC:
-
Superior vena cava
- CAJ:
-
Cavoatrial junction
- ICU:
-
Intensive care unit
References
Amerasekera SS, Jones CM, Patel R, Cleasby MJ (2009) Imaging of the complications of peripherally inserted central venous catheters. Clin Radiol 64:832–840
Maki DG, Kluger DM, Crnich CJ (2006) The risk of bloodstream infection in adults with different intravascular devices: a systematic review of 200 published prospective studies. Mayo Clin Proc 81:1159–1171
Caers J, Fontaine C, Vinh-Hung V et al (2005) Catheter tip position as a risk factor for thrombosis associated with the use of subcutaneous infusion ports. Support Care Cancer 13:325–331
Verdino RJ (2006) The evolution of atrial fibrillation ablation from triggers to substrate. J Electrocardiol 39:S184–S187
Bivins MH, Callahan MJ (2000) Position-dependent ventricular tachycardia related to a peripherally inserted central catheter. Mayo Clin Proc 75:414–416
Orme RM, McSwiney MM, Chamberlain-Webber RF (2007) Fatal cardiac tamponade as a result of a peripherally inserted central venous catheter: a case report and review of the literature. Br J Anaesth 99:384–388
Johnston AJ, Holder A, Bishop SM, See TC, Streater CT (2014) Evaluation of the Sherlock 3CG Tip Confirmation System on peripherally inserted central catheter malposition rates. Anaesthesia 69:1322–1330
Baldinelli F, Capozzoli G, Pedrazzoli R, Marzano N (2015) Evaluation of the correct position of peripherally inserted central catheters: anatomical landmark vs. electrocardiographic technique. J Vasc Access 0:0
Ragasa J, Shah N, Watson RC (1989) Where antecubital catheters go: a study under fluoroscopic control. Anesthesiology 71:378–380
LaRue GD (1995) Improving central placement rates of peripherally inserted catheters. J Intraven Nurs 18:24–27
Burgess GE 3rd, Marino RJ, Peuler MJ (1977) Effect of head position on the location of venous catheters inserted via basilic veins. Anesthesiology 46:212–213
Schweickert WD, Herlitz J, Pohlman AS, Gehlbach BK, Hall JB, Kress JP (2009) A randomized, controlled trial evaluating postinsertion neck ultrasound in peripherally inserted central catheter procedures. Crit Care Med 37:1217–1221
Pittiruti M, Hamilton H, Biffi R, MacFie J, Pertkiewicz M (2009) ESPEN Guidelines on Parenteral Nutrition: central venous catheters (access, care, diagnosis and therapy of complications). Clin Nutr 28:365–377
Vesely TM (2003) Central venous catheter tip position: a continuing controversy. J Vasc Interv Radiol 14:527–534
Ridge CA, Litmanovich D, Molinari F, Bankier AA, Eisenberg RL (2013) Radiographic evaluation of central venous catheter position: anatomic correlation using gated coronary computed tomographic angiography. J Thorac Imaging 28:129–133
Fricke BL, Racadio JM, Duckworth T, Donnelly LF, Tamer RM, Johnson ND (2005) Placement of peripherally inserted central catheters without fluoroscopy in children: initial catheter tip position. Radiology 234:887–892
Trerotola SO, Thompson S, Chittams J, Vierregger KS (2007) Analysis of tip malposition and correction in peripherally inserted central catheters placed at bedside by a dedicated nursing team. J Vasc Interv Radiol 18:513–518
Minkovich L, Djaiani G, McCluskey SA, Mitsakakis N, Gilbert RW, Beattie WS (2011) Frequent malpositions of peripherally inserted central venous catheters in patients undergoing head and neck surgery. Can J Anaesth 58:709–713
Johnston AJ, Bishop SM, Martin L, See TC, Streater CT (2013) Defining peripherally inserted central catheter tip position and an evaluation of insertions in one unit. Anaesthesia 68:484–491
Pittiruti M, Bertollo D, Briglia E et al (2012) The intracavitary ECG method for positioning the tip of central venous catheters: results of an Italian multicenter study. J Vasc Access 13:357–365
Venkatesan T, Sen N, Korula PJ et al (2007) Blind placements of peripherally inserted antecubital central catheters: initial catheter tip position in relation to carina. Br J Anaesth 98:83–88
Ng PK, Ault MJ, Maldonado LS (1996) Peripherally inserted central catheters in the intensive care unit. J Intensive Care Med 11:49–54
Pinsky MR (1990) The effects of mechanical ventilation on the cardiovascular system. Crit Care Clin 6:663–678
(2006) Infusion Nursing Standards of Practice. J Infus Nurs 29:S1–92
Tian G, Chen B, Qi L, Zhu Y (2011) Modified insertion of a peripherally inserted central catheter: taking the chest radiograph earlier. Crit Care Nurse 31:64–69
Funaki B (2002) Central venous access: a primer for the diagnostic radiologist. AJR Am J Roentgenol 179:309–318
Acknowledgements
The scientific guarantor of this publication is Salah Dine Qanadli. The authors of this manuscript declare relationships with the following companies: Salah Dine Qanadli was a consultant for C. R Bard Inc. during the last 3 years. The authors state that this work has not received any funding. No complex statistical methods were necessary for this paper.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Glauser, F., Breault, S., Rigamonti, F. et al. Tip malposition of peripherally inserted central catheters: a prospective randomized controlled trial to compare bedside insertion to fluoroscopically guided placement. Eur Radiol 27, 2843–2849 (2017). https://doi.org/10.1007/s00330-016-4666-y
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-016-4666-y