Abstract

Background and purpose

We describe our initial experience with a new metallic ureteric stent which has been designed to provide long-term urinary drainage in patients with malignant ureteric strictures. The aim is to achieve longer primary patency rates than conventional polyurethane ureteric stents, where encrustation and compression by malignant masses limit primary patency. The Resonance metallic double-pigtail ureteric stent (Cook, Ireland) is constructed from coiled wire spirals of a corrosion-resistant alloy designed to minimize tissue in-growth and resist encrustation, and the manufacturer recommends interval stent change at 12 months.

Methods

Seventeen Resonance stents were inserted via an antegrade approach into 15 patients between December 2004 and March 2006. The causes of ureteric obstruction were malignancies of the bladder (n = 4), colon (n = 3), gynecologic (n = 5), and others (n = 3).

Results

One patient had the stent changed after 12 months, and 3 patients had their stents changed at 6 months. These stents were draining adequately with minimal encrustation. Four patients are still alive with functioning stents in situ for 2–10 months. Seven patients died with functioning stents in place (follow-up periods of 1 week to 8 months). Three stents failed from the outset due to bulky pelvic malignancy resulting in high intravesical pressure, as occurs with conventional plastic stents.

Conclusion

Our initial experience with the Resonance metallic ureteric stent indicates that it may provide adequate long-term urinary drainage (up to 12 months) in patients with malignant ureteric obstruction but without significantly bulky pelvic disease. This obviates the need for regular stent changes and would offer significant benefit for these patients with limited life expectancy.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Docimo SG, Dewolf WC (1989) High failure rate of indwelling ureteral stents inpatients with extrinsic obstruction: Experience at 2 institutions. J Urol 142:277–279

    PubMed  CAS  Google Scholar 

  2. Chitale SV, Scott-Barrett S, Ho ET, et al. (2002) The management of ureteric obstruction secondary to malignant pelvic disease. Clin Radiol 57:1118–1121

    Article  PubMed  CAS  Google Scholar 

  3. Barbalias GA, Siablis D, Liatsikos EN, et al. (1997) Metal stents: A new treatment of malignant ureteral obstruction. J Urol 158:54–58

    Article  PubMed  CAS  Google Scholar 

  4. Lang EK, Irwin RJ, Lopez-Martinez RA, et al. (1998) Placement of metallic stents in ureters obstructed by carcinoma of the cervix to maintain renal function in patients undergoing long term chemotherapy. AJR Am J Roentgenol 171:1595–1599

    PubMed  CAS  Google Scholar 

  5. vanSonnenberg E, D’Agostino HB, O’Laoide R, Donaldson J, Sanchez RB, Hoyt A, Pittman CC (1994) Malignant ureteral obstruction: Treatment with metal stents. Technique, results, and observations with percutaneous intraluminal US. Radiology 191:765–768

    PubMed  CAS  Google Scholar 

Download references

Acknowledgments

The Resonance metallic ureteric stents and payment for the color illustrations were provided by Cook, Ireland.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Tze M. Wah.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Wah, T.M., Irving, H.C. & Cartledge, J. Initial Experience with the Resonance Metallic Stent for Antegrade Ureteric Stenting. Cardiovasc Intervent Radiol 30, 705–710 (2007). https://doi.org/10.1007/s00270-007-9043-4

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00270-007-9043-4

Keywords

Navigation